K132782 · Biosense Webster, Inc. · DQK · Jan 24, 2014 · Cardiovascular
Device Facts
Record ID
K132782
Device Name
CARTO3 EP NAVIGATION SYSTEM, VERSION 3.2
Applicant
Biosense Webster, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Jan 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.
Device Story
Catheter-based cardiac mapping system; acquires/analyzes intracardiac electrograms (IECG) and body surface ECG signals; utilizes magnetic sensor and Advanced Catheter Location (ACL) technology to track catheter position. Transforms input signals into real-time 3D anatomical and electroanatomical maps of the heart. Used in clinical settings by physicians during EP procedures to visualize cardiac electrical activity and catheter location. Includes SMARTTOUCH™ module for contact force monitoring, Advanced FAM (Fast Anatomical Mapping) for visualization, Remote Magnetic Technology (RMT) support, and PaSo™ (Pace Mapping) module. Output displayed on screen to assist clinical decision-making during ablation or diagnostic procedures. Benefits include enhanced navigation and mapping precision for cardiac interventions.
Clinical Evidence
Evidence includes non-clinical bench testing, animal testing, and an IDE clinical investigation. Testing verified modified features and confirmed that software updates did not negatively impact existing functionality. Results demonstrated the system is as safe and effective as the predicate and reference devices.
Technological Characteristics
Programmable diagnostic computer; utilizes magnetic sensor and Advanced Catheter Location (ACL) technology. Hardware platform is identical to predicate Version 3.0. Software-based system for 3D cardiac mapping and electrogram display. Connectivity includes video export to recording systems. No specific materials or sterilization methods listed for the console/software components.
Indications for Use
Indicated for patients eligible for conventional catheter-based cardiac electrophysiological (EP) procedures. No specific age or gender restrictions; no special contraindications.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
CARTO® 3 EP Navigation System, Version 3.0 and Accessories (K120550)
Reference Devices
CARTO® 3 XP EP Navigation System, Version 1.0 (K093566)
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 4 2014
### 14 510(K) SUMMARY
-" " ::
i
### 14.1 General Information
| Applicant: | Biosense Webster, Inc.<br>3333 Diamond Canyon Road<br>Diamond Bar, CA 91765 USA<br>Phone: 909-839-8597<br>Fax: 909-839-8804 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | September 3, 2013 |
| Contact Person: | Wayne R. Hohman<br>Project Manager Regulatory Affairs |
| Authored By: | Moshe Hochmitz<br>Quality and Regulatory Manager<br>Biosense Webster (ISRAEL), Ltd. |
| Trade/Proprietary Device Name: | CARTO® 3 EP Navigation System, Version 3.2 and<br>Accessories |
| Manufacturing Part Number: | FG-5400-00 |
| Common Device Name: | Cardiac mapping system |
| Classification Name: | Programmable diagnostic computer<br>Class II, 21 CFR 870.1425, Product Code DQK |
| Predicate Device: | CARTO® 3 EP Navigation System, Version 3.0 and<br>Accessories<br>510(k) K120550, May 7, 2012 |
| | Reference Device:<br>CARTO® 3 XP EP Navigation, System Version 1.0<br>510(k) K093566, June 18, 2010 |
| Manufacturing Facilities: | Biosense Webster (Israel), Ltd.<br>a Johnson & Johnson Company<br>4 Hatnufa Street, POB 275<br>Yokneam 20692<br>ISRAEL<br><br>Biosense Webster, Inc.<br>a Johnson & Johnson Company<br>3333 Diamond Canyon Road<br>Diamond Bar, California 91765 USA |
Biosense Webster, Inc.
CARTO® 3 EP Navigation System, V3.2
Traditional 510(k)
Page 5 of 268
{1}------------------------------------------------
### 14.2 Substantial Equivalence
ﺪ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
The CARTO® 3 EP Navigation System, Version 3.2 and Accessories are substantially equivalent to the legally marketed CARTO® 3 EP Navigation System, Version 3.0 and Accessories and the CARTO® XP EP Navigation System. Version 10 as shown in Table 7:
| Table 7<br>Predicate Devices for CARTO® 3 EP Navigation System, Version 3.2 | | |
|-----------------------------------------------------------------------------|------------------|-----------------------------------------|
| Device Name | 510(k)<br>Number | Equivalence Criteria |
| CARTO® 3 EP Navigation System,<br>Version 3.0 and Accessories | K120550 | Legacy Functionality |
| CARTO XP EP Navigation<br>System, Version 10 | K093566 | Pace Mapping Software (PaSo™)<br>Module |
### 14.3 Description of Device
The CARTO®3 EP Navigation 3 System, Version 3.2 and Accessories is a catheterbased atrial and ventricular mapping system designed to acquire and analyze data points, and use this information to display 3D anatomical and electroanatomical maps of the human heart in real-time. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheter and reference device. The system allows real-time display of electrograms and cardiac maps based on the received intracardiac signals from the catheters in a number of different formats. The acquired patient signals, including body surface ECG and intracardiac electrograms (IECG) may also be displayed on the display screen. The CARTO® 3 System V3.2 uses two distinct types of location technology magnetic sensor technology and Advanced Catheter Location (ACL) technology.
#### 14.4 Indications for Use
The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.
# 14.4a Technological Characteristics
The Proposed CARTO® 3 Electrophysiology Navigation System. Version 3.2, has the same technological characteristics (i. e., design, material, chemical composition, energy source) as the predicate CARTO® 3 Electrophysiology Navigation System, Version 3.0. A summary of the technological characteristics of the new device compared to the predicate device is as follows:
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2 Traditional 510(k) Page 88 of 2538
Traditional 510(k) Page 6 of 268
{2}------------------------------------------------
- . The hardware platform of the new Version 3.2 is identical to the predicate Version 3.0.
- . The differences between Versions 3.2 and 3.0 involve only software as follows:
- The main change in this submission is to activate an installed but previously o blocked SMARTTOUCHTM Module.
- o The software of three other features of this system is enhanced (i. e., updated) in this submission: 1) Advanced FAM - Visual improvement of the V2 FAM Feature, 2) Remote Magnetic Technology (RMT) Improvements, and 3) Video Export to Recording System.
- o Finally, Biosense Webster is updating FDA on a feature that was incorporated via 510(k) Letter to File since the last submission: Improved Pace Mapping Software (PaSoTM) Module. The PaSoTM Module was originally cleared via in an earlier version of this device and this device is included in this Application as a Reference Device for the PaSo™ Module feature: CARTO® XP EP Navigation System, Version 10 (510(k) K093566, cleared June 18, 2010).
### 14.5 Summary of Non-Clinical Performance Testing
The CARTO® 3 EP Navigation System, Version 3.2 underwent extensive Bench and Animal Testing to verify the modified features and to demonstrate with regression testing that the new features did not negatively affect existing features. The CARTO® 3 EP Navigation System, Version 3.2 passed all tests in accordance with appropriate test criteria and standards, and the CARTO® 3 EP Navigation System, Version 3.2 and Accessories did not raise new questions of safety or effectiveness.
#### 14.6 Conclusions
. ...
Extensive Non-Clinical (Bench) Testing, Animal Testing, and an IDE Clinical Investigation all demonstrated that the CARTO® 3 EP Navigation System, Version 3.2 and Accessories as used with the activated SMARTOUCH® Module are as safe, as effective, and performs as well as or better than the Predicate Device and the Reference Device. This testing program supports the determination of substantial equivalence of the CARTO® 3 EP Navigation System, Version 3.2 and Accessories to existing predicate devices.
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2 Traditional 510(k) Page 89 of 2538
Traditional 510(k) Page 7 of 268
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 24, 2014
Biosense Webster, Inc. Wayne Hohman 3333 Diamond Canyon Rd Diamond Bar, CA 91765 US
K132782 Re:
> Trade/Device Name: CARTO 3 EP Navigation System, version 3.2 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: November 5, 2013 Received: November 6, 2013
Dear Mr. Hohman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) No (if known):
Device Name: CARTO® 3 EP Navigation System, Version 3.2
Indications for Use:
The intended use of the CARTO® 3 System is catheter-based cardiac electrophysiological (EP) procedures. The CARTO® 3 System provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure. The system has no special contraindications.
Prescription Use ___ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by.
Owen P. Faris -S
Date: 2014.01.24
16:20:52 -05'00'
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2
Biosense Webster, Inc. CARTO® 3 EP Navigation System, V3.2 Traditional 510(k) Page 42 of 2538
Traditional 510(k) Page 9 of 268
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.