Stereotactic spatial positioning and orientation of an instrument holder or tool guide to be used by a surgeon to manually guide standard neurosurgical instruments.
Device Story
Neuromate Frameless Gen II is a computer-controlled, electromechanical, multi-jointed robotic arm used in operating rooms by neurosurgeons. It provides spatial positioning and orientation for neurosurgical instrument holders. Input consists of preoperative 3D imaging data processed via VoXim software; registration is performed using an ultrasound system. The surgeon defines the tool, trajectory, point of access, and orientation on the workstation; the robot then positions the tool holder accordingly. The device ensures stability during manual insertion of instruments by the surgeon. It supports both frame-based (mechanical registration) and frameless (ultrasound registration) modes. Benefits include precise, stable guidance for standard neurosurgical instruments, potentially improving procedural accuracy.
Clinical Evidence
No clinical data submitted. Bench and performance testing only. Device found compliant with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (electromagnetic compatibility).
Technological Characteristics
Electromechanical multi-jointed arm with 5 degrees of freedom. Components: motors, planetary/gear reducers, worm-wheel adjustment system. Absolute digital encoders. Connectivity: workstation-based. Software: VoXim neuromate 6.4. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 62366, ISO 14971, IEC 62304, DICOM 3, ISO 17665-1.
Indications for Use
Indicated for neurosurgeons performing stereotactic spatial positioning and orientation of an instrument holder or tool guide to manually guide standard neurosurgical instruments during neurological surgery on the head.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
# RENISHAW mayfie
## 510(k) SUMMARY
APR 2 9 2014
- 1. SUBMITTER: . Renishaw Mayfield sarl 31. rue Ampère 69680 Chassieu FRANCE 011-33 4 37 23 89 00 Phone: Fax: 011-33 4 78 90 75 22
Contact person:
Stéphane Vinot Quality Assurance & Regulatory affairs Manager stephane.vinot@renishawmayfield.com
Preparation date:
July 11, 2013
#### 2. DEVICE NAME:
| Trade name: | neuromate Frameless Gen II |
|----------------------|----------------------------------|
| Common name: | Stereotactic neurological system |
| Classification name: | Stereotaxic Instrument (HAW) |
#### 3. PREDICATE DEVICE(s):
K991081 - Frameless Neuromate - Integrated Surgical Systems Inc
#### DEVICE DESCRIPTION: 4.
A stereotactic system with an electromechanical, multi jointed arm for spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a preoperative plan developed with three-dimensional imaging software and utilizes ultrasonic registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments.
#### 5. INTENDED USE:
Stereotactic spatial positioning and orientation of an instrument holder or tool guide to be used by a surgeon to manually guide standard neurosurgical instruments.
Renishaw Mayfield sarl 31 rue ampère F-69680 Chassieu - France Téléphone +33 (0) 4 37 23 89 00 +33 (0) 4 78 90 75 22 Fax
RCS Lyon 480 820 307 SIRET 48082030700027 APE 3250A Nº identification TVA FR26 480 820 307 Capital : 250 000 €
Société Générale Lyon République Code Banque 30003 Code quichet 01200 N° compte 00020839167 Clé 57 IBAN FR76 30003 01200 00020839167 BIC SOGEFRPP
{1}------------------------------------------------
#### 6. COMPARISON TO PREDICATE DEVICE:
## 6.1 Similarities
The neuromate system is a technical update of the cleared Frameless NeuroMate. It has the same intend and is similar in all the following respects:
- the patient benefits .
- the architecture of the product .
- the materials and components used .
- . the construction and method of manufacture
- . the embedded software
- . the planning and imaging software
- the operational modes offered ●
- . the attachments used
## 6.2 Differences
The only areas of difference between the neuromate and the cleared Frameless NeuroMate are:
- a. The new neuromate system is type BF applied part while the cleared Frameless NeuroMate is type B applied part.
- Note: The terms "Type BF applied part" and "Type B applied part" are used in the IEC60601-1 standard (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance) to define the level of electrical isolation of an electrical medical equipment that has a part(s) in contact with a patient under normal use. The electrical isolation is between the patient and an external electrical source. In the IEC60601-1 standard, a "Type BF applied part" has a higher degree of electrical isolation compared to a "Type B applied part".
- b. The neuromate system uses absolute digital encoders to control the position of the axis of the robot arm of the system while the cleared Frameless NeuroMate uses analogical ones.
- c. The image-feedback sensor has been removed from the standard tool-holder, thereby removing the image-feedback functionality of the system
## 6.3 Conclusion
## Except for the above stated changes, the proposed Neuromate is identical in design, intended use, construction and operation to the predicate K991081.
The itemised comparison between the device and its predicate is provided in paragraph 10 (Substantial Equivalence Summary) of this document.
{2}------------------------------------------------
## 7. NONCLINICAL DATA:
Bench and performance testing showed that neuromate performed as designed and intended. The neuromate was tested to and found compliant with the requirements of IEC 60601-1 and IEC 60601-1-2.
Software was developed in compliance with FDA software guidance and IEC 62304.
#### CLINICAL DATA: 8.
No clinical data are submitted.
#### CONCLUSIONS DRAWN FROM TESTING: 9.
Øi
The testing showed that the neuromate performs as designed and intended, and is equivalent in design and performance to the predicate device.
## 10. SUBSTANTIAL EQUIVALENCE SUMMARY
| Device | neuromate Frameless Gen II<br>(this submission) | Frameless Neuromate<br>(510(k) K991081) | SE | |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|--|
| Manufacturer | Renishaw Mayfield | Integrated Surgical Systems Inc | SE | |
| | Intended Use/Indications for Use | | | |
| Intended Use | Stereotactic spatial positioning and<br>orientation of an instrument holder or<br>tool quide to be used by a surgeon<br>to manually guide standard<br>neurosurgical instruments. | Stereotactic spatial positioning and<br>orientation of an instrument holder or<br>tool quide to be used by a surgeon<br>to manually guide standard<br>neurosurgical instruments. | Same | |
| Clinical<br>application | Neurological surgery | Neurological surgery | Same | |
| Sites of use | Operating rooms | Operating rooms | Same | |
| User | Neurosurgeon | Neurosurgeon | Same | |
| Anatomical site | Head | Head | Same | |
| Technical Characteristics | | | | |
| General device<br>description | Computer-controlled<br>electromechanical multi-jointed arm | Computer-controlled<br>electromechanical multi-jointed arm | Same | |
| Principle of<br>Operation | Patient imaging data is visualized<br>on the supplied workstation running<br>the VoXim software. It is the<br>surgeon who chooses the tool,<br>trajectory, point of access and<br>orientation.<br>The data is communicated to the<br>robot which locates the tool holder<br>accordingly and accurately. | Patient imaging data is visualized<br>on the supplied workstation running<br>the VoXim software. It is the<br>surgeon who chooses the tool,<br>trajectory, point of access and<br>orientation.<br>The data is communicated to the<br>robot which locates the tool holder<br>accordingly and accurately. | Same | |
| Device | neuromate Frameless Gen II<br>(this submission) | Frameless Neuromate<br>(510(k) K991081) | SE | |
| | The surgeon performs the<br>procedure by inserting tools through<br>the tool-holder, the robot ensuring<br>the continued stability of the<br>position. | The surgeon performs the<br>procedure by inserting tools through<br>the tool-holder, the robot ensuring<br>the continued stability of the<br>position. | | |
| | Patient registration uses an<br>ultrasound system. | Patient registration uses an<br>ultrasound system. | | |
| May be used in<br>frame-based<br>mode | Yes: the patient wears a stereotactic<br>frame and the frame is attached to<br>the robot (mechanical registration) | Yes: the patient wears a<br>stereotactic frame and the frame is<br>attached to the robot (mechanical<br>registration) | Same | |
| May be used in<br>frameless mode | Yes: the patient wears a light-weight<br>microphone frame and the<br>registration is made by ultrasound. | Yes: the patient wears a light-<br>weight microphone frame and the<br>registration is made by ultrasound. | Same | |
| Image feedback | No: the surgeon controls the depth<br>of insertion of a biopsy needle<br>through the rule on the cannula. | Yes: the position of the tool's<br>extremity is shown superposed to<br>the brain scan image. | | |
| | | However, the surgeon is advised<br>not to use this information to control<br>tool depth (use instead the rule on<br>the cannula) | SE | |
| System<br>Architecture | • a robot, made of a computer-<br>controlled articulated motorized<br>arm and the base that supports it<br>and contains the control<br>electronics;<br>• a workstation on which runs<br>software for scanned image<br>processing, procedure planning,<br>robot configuration, robot control<br>and system diagnostic;<br>• a series of attachments to connect<br>the robot to the patient, to perform<br>patient registration and to support<br>the tools used by the surgeon | • a robot, made of a computer-<br>controlled articulated motorized<br>arm and the base that supports it<br>and contains the control<br>electronics;<br>• a workstation on which runs<br>software for scanned image<br>processing, procedure planning,<br>robot configuration, robot control<br>and system diagnostic;<br>a series of attachments to<br>connect the robot to the patient, to<br>perform patient registration and to<br>support the tools used by the<br>surgeon | Same | |
| Attachments | • remote control handset<br>• standard tool-holder<br>• tool laser holder<br>• frame adapter<br>• head-holder<br>• ultrasound tool-holder<br>• microphone frame<br>• temporary implant insert<br>• CT image localizer<br>• MRI image localizer<br>• angiographic module localizer | • remote control handset<br>• standard tool-holder<br>• tool laser holder<br>• frame adapter<br>• head-holder<br>• ultrasound tool-holder<br>• microphone frame<br>• temporary implant insert<br>• CT image localizer<br>• MRI image localizer<br>• angiographic module localizer | Same | |
| Device | neuromate Frameless Gen II<br>(this submission) | Frameless Neuromate<br>(510(k) K991081) | SE | |
| Construction | • Mechanical arm in four parts with<br>5 joints (5 degrees of freedom)<br>• Rigid fixation between the<br>stereotactic ring attachment or<br>head holder attachment and the<br>neuromate robot<br>• Each arm joint is an electrome-<br>chanical subassembly composed<br>of a motor; planetary reducer;<br>gear reducer; and worm-wheel<br>adjustment system with variable<br>backlash, providing non-reversible<br>motion and guaranteeing rigid<br>position locking. | • Mechanical arm in four parts with<br>5 joints (5 degrees of freedom)<br>• Rigid fixation between the<br>stereotactic ring attachment or<br>head holder attachment and the<br>neuromate robot<br>• Each arm joint is an electrome-<br>chanical subassembly composed<br>of a motor; planetary reducer;<br>gear reducer; and worm-wheel<br>adjustment system with variable<br>backlash, providing non-reversible<br>motion and guaranteeing rigid<br>position locking. | Same | |
| Axis control<br>technology | Incremental digital encoders<br>Absolute digital encoders | Incremental digital encoder<br>Absolute analogical encoder | SE | |
| Software/firmware<br>/ microprocessor<br>control | Yes | Yes | Same | |
| Controller S/W | Main Controller | Main Controller | Same | |
| Planning S/W | VoXim neuromate 6.4 | VoXim 1.0 | SE | |
| Imaging S/W | VoXim neuromate 6.4 | VoXim 1.0 | SE | |
| Compliance with<br>voluntary<br>standards | Yes<br>IEC 60601-1:2005 + AC1 + AC2<br>IEC 60601-1-2:2007<br>IEC 60601-1-4:2000<br>IEC 62366: 2007<br>ISO 14971: 2007<br>IEC 62304: 2006<br>DICOM 3<br>ISO 17665-1:2006 | Yes<br>EN 55011, Group 1, Class B<br>EN 50082-1 (IEC 801-2)<br>EN 50082-1 (IEC 801-3)<br>EN 50082-1 (IEC 801-4)<br>IEC 601.1 | SE | |
{3}------------------------------------------------
・
-
.
:
{4}------------------------------------------------
:
:
:
.
.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2014
Renishaw Mayfield Sarl Mr. Stephane Vinot Quality Assurance & Regulatory Affairs Manager 31 Rue Ampere Chassieu. France 69680
Re: K132755
Trade/Device Name: Neuromate Frameless Gen II Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: January 21, 2014 Received: January 29, 2014
Dear Mr. Vinot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be fray of subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a dolor regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 – Mr. Stephane Vinot
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132755
### Device Name Neuromate Frameless Gen II
Indications for Use (Describe)
Stereotactic spatial positioning and orientation of an instrument holder or tool guide to be used by a surgeon to manually guide standard neurosurgical instruments.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY :54 : 1 . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Date: 2014.04.2 18:06:23 -04'00'
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number,"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.