ULTRACONGRUENT INSERT

K132752 · United Orthopedic Corporation · JWH · Mar 13, 2014 · Orthopedic

Device Facts

Record IDK132752
Device NameULTRACONGRUENT INSERT
ApplicantUnited Orthopedic Corporation
Product CodeJWH · Orthopedic
Decision DateMar 13, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device system is intended for cemented use only in the U.S.A.

Device Story

Ultracongruent Insert is a semi-constrained tibial bearing component for total knee arthroplasty; used with cruciate-retained (CR) femoral components. Designed for patients with PCL deficiency (incomplete, absent, nonfunctional, or released). Features a prominent anterior lip to increase articulating surface area and stabilize the femur during flexion. Available in seven sizes and multiple thicknesses. Implanted by orthopedic surgeons in clinical settings; intended for cemented fixation. Provides mechanical stability to the knee joint; replaces articulating surfaces to reduce pain and restore function.

Clinical Evidence

Bench testing only. Performance evaluation included constraint testing, contact area and contact pressure analysis, range of motion assessment, and interlocking strength testing. No clinical data provided.

Technological Characteristics

Material: Irradiated UHMWPE (ASTM F2565, ASTM F648, ISO 5834). Form factor: Tibial insert with prominent anterior lip for femoral stabilization. Sizes: 7 proportional sizes; 5 thicknesses (9mm-18mm). Minimum bearing surface thickness: 6mm. Fixation: Cemented. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring knee arthroplasty for pain relief or improved function due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, pseudogout, or posttraumatic joint loss. Applicable for patients with patellofemoral erosion, dysfunction, prior patellectomy, or moderate valgus/varus/flexion deformities. Also indicated for salvage of failed surgical attempts if the knee can be balanced and stabilized.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 1 3 2014 ## ( Ultracongruent Insert 510(k) Summary ## 510(k) Summary of Safety and Effectiveness | Submitted by: | United Orthopedic Corporation | |-----------------------|----------------------------------------------------------| | Address: | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan | | Phone Number: | +886-3-5773351 ext. 2212 | | Fax Number: | +886-3-577156 | | Date of Summary: | August 16, 2013 | | Contact Person | Fang-Yuan Ho | | | Regulation and Document Management | | Proprietary Name: | Ultracongruent Insert | | Common Name: | Semi-constrained total knee prostheses | | Device Classification | Knee joint patellofemorotibial polymer/metal/polymer | | Name and Reference: | semi-constrained cemented prosthesis per 21CFR 888.3560. | | | This falls under the Orthopedics panel. | | Device Class | Class II | | Panel Code | Orthopaedics Device | | Device Product Code: | JWH | | Predicate Device: | 1. "United" U2 Total Knee System (K051640) | | | 2. "United" U2 XPE Total Knee System (K103733) | | | 3. "Ortho Development" Balanced Knee System (K090705) | | | 4. "Biomet" Vanguard™ Complete Knee System (K050222) | #### Device Description: The Ultracongruent Insert is designed for the replacement of the bearing and/or articulating surfaces of the proximal tibia composed of an articulating bearing surface fixed in a metal tibial baseplate, and is used for patients with PCL incomplete, absent, nonfunctional or required release. It is used with the cruciate retained (CR) type Femoral Component. The anterior lip of Ultracongruent Insert is more prominent when comparing with the U2 CR Insert (K051640, K103733), which increases the articulating surface area and expands the {1}------------------------------------------------ ## (9) Ultracongruent Insert 510(k) Summary circumference to accommodate and stabilize the femur during knee flexion. The Ultracongruent Insert is available in seven proportional sizes (#1~ #7) and five thicknesses (thicknesses of insert + tibial baseplate: 9mm, 11mm, 15mm, and 18mm). The minimum thickness of Ultracongruent Insert is 6 mm on the bearing surface. It is manufactured from irradiated UHMWPE which conform to ASTM F2565, while the UHMWPE raw material is in accordance with ASTM F648 and ISO 5834. ### Indications: The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device system is intended for cemented use only in the U.S.A. #### Basis for Substantial Equivalence: The Ultracongruent Insert has the same indications, size distribution, materials and the same manufacturing and sterilization method as the XPE tibial insert of the "United" U2 Total Knee System. Besides, the basic design, prominent anterior lip height and intended use of the subjected device are similar with the tibial insert of "Ortho Development" Balanced Knee System® (K090705) and "Biomet" Vanguard"" Complete Knee System (K050222). #### Performance Data: Tests as follows were conducted to evaluate the safety and effectiveness of the subjected device, and the test results demonstrated that this device is safe and effective. {2}------------------------------------------------ # ( Ultracongruent Insert 510(k) Summary - Constraint Test a. - Contact Area and Contact Pressure b. - Range of Motion c. - d. Interlocking Strength Test {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with outstretched arms, representing care and protection. The figure is positioned within a circle that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 13, 2014 United Orthopedic Corporation Ms. Fang-Yuan Ho Regulatory Affairs Manager No. 57, Park Avenue 2, Science Park Hsinchu, 300 TAIWAN Re: K132752 Trade/Device Name: Ultracongruent Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: January 28, 2014 Received: January 29, 2014 Dear Ms. Ho: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {4}------------------------------------------------ Page 2 - Ms. Fang-Yuan Ho (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indication for Use K132752 510 (k) Number (if known): _ Device Name: Ultracongruent Insert #### Indications for Use: The U2 Total Knee system is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device system is intended for cemented use only in the U.S.A. Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/5/Picture/7 description: The image shows the words "Casey L. Hanley, Ph.D." in bold black font. There is a black line underneath the name. Below the line, the words "Division of Orthopedic Devices" are written in bold black font. Page 1 of 1 Page: 6-1/1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...