K132747 · Stryker Communications · FSY · Dec 30, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132747
Device Name
VISUM BLADE LED SURGICAL LIGHTING SYSTEM
Applicant
Stryker Communications
Product Code
FSY · General, Plastic Surgery
Decision Date
Dec 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Visum Blade LED Surgical Light is intended to illuminate the operative site during surgical procedures with high intensity light.
Device Story
Visum Blade LED Surgical Light System provides high-intensity illumination (up to 160,000 Lux) for surgical procedures. System comprises 1-4 light heads on a pivoting ceiling-mounted suspension system; includes power supply, drop tube, and optional monitor mounts. Operated by surgeons via light handle for intensity control; optional wall control panel available for non-sterile staff. Features LED light source with Fresnel lenses for cool light and reduced shadow resolution. Optional in-light camera available. Device used in ORs, exam rooms, and emergency rooms to provide clear visualization of operative sites, aiding clinical decision-making during surgery.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Ceiling-mounted surgical light; LED light source with Fresnel lenses; variable intensity up to 160,000 Lux; sterilizable light handle; pivoting suspension system; optional wall control panel; optional in-light camera. No mechanical focus mechanism.
Indications for Use
Indicated for use in operating rooms, examination rooms, emergency rooms, and other health care facilities requiring additional illumination for surgical procedures, using various mounting configurations.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
Burton's AIM 200 OR Major Surgical Light (K101537)
Stryker's Visum LED Surgical Light System (K060802)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the first letter, which is capitalized. There is a registered trademark symbol to the upper right of the word. The word is likely a company logo or brand name.
DEC 3 0 2013
1410 Lakeside Parkway, #100 Flower Mound, TX 75028 t: 972 410 7100 f: 972 410 7001 www.stryker.com
#### Communications
# Section 5 - 510(k) Summary for the Visum Blade LED Surgical Light System
PER 21 CFR 807.92
| Date: | December 26, 2013 |
|------------------------|---------------------------------------------------------------------------------------------------------|
| 510(k) Owner/Sponsor: | Stryker Communications |
| Address: | 1410 Lakeside Parkway, #100 |
| | Flower Mound, Texas 75028 |
| Establishment Number: | 2031963 |
| Telephone Number: | (972) 410-7181 |
| Fax Number: | (972) 410-7201 |
| Contact Person: | Trey Thorsen; Senior Regulatory Compliance Analyst |
| Email Address: | trey.thorsen@stryker.com |
| Proposed Device: | Visum Blade LED Surgical Light System (Blade) |
| Device Common Name: | Surgical Light, Surgical Lamp |
| Product Code: | FSY |
| FDA Regulation Number: | 21 CFR 878.4580 |
| Device Classification: | Class II |
| Predicate Devices: | Burton's AIM 200 OR Major Surgical Light (K101537). Stryker's Visum LED Surgical Light System (K060802) |
| Device Common Name: | Surgical Light, Surgical Lamp |
| Product Code: | FSY |
| FDA Regulation Number: | 21 CFR 878.4580 |
| Device Classification: | Class II |
| | |
#### Device Description
The Visum Blade LED Surgical Light System (Blade) is suitable for all major and minor surgical procedures in operating rooms and all other healthcare facilities where the need for illumination exists. The light intensity is variable up to 160,000 Lux. The light quality is based upon high quality LED's to provide cool light with reduced shadow resolution.
A system consists of 1 - 4 light heads supported by a pivoting suspension system that is attached to the heath care facility's super structure by a mounting plate. The system requires at least one of the following: a power supply box, ceiling cover, drop tube, central axis and spring arm. The system may include optional end effectors such as but not limited to monitor mounts. The mounting plates may be configured to accommodate a single or tandem mount suspension system.
{1}------------------------------------------------
Independent, control of each light's intensity is via the light handle. An optional electronic wall control panel may be installed in the operating room for control by non-sterile users. Blade allows for an in-light camera option that may be added to a light head. This option requires a wall control for camera operation.
#### Intended Use
The Visum Blade LED Surgical Light is intended to illuminate the operative site during surgical procedures with high intensity light.
#### Indications for Use
The Visum Blade LED Surgical Light System is indicated to be used with various mounting configurations in operating rooms, examination rooms, emergency rooms, and all other health care facilities where the need for additional illumination exists.
#### Technological Comparison
Like its predicates, Blade is a ceiling mounted surgical light that is available in single or tandem light head configurations. Both of the predicates and Blade provide illumination with reduced shadow resolution. All three devices have a removable, sterilizable light handle.
Blade uses an LED light source with Fresnel lenses to direct the light. Predicate devices use LED or halogen light sources with reflectors to direct the light.
Blade has two options for control of light intensity. Control is on the light head that allows the surgeon to control the light intensity from the sterile field. This is different than the predicates. Blade also allows light intensity to be controlled through a wall control, which is an optional feature.
The predicates have spot size adjustment (mechanical focus mechanism). Blade does not have this feature. The absence of this feature reduces cost, size, and complexity of the luminaire allowing Blade be marketable to a broader range of healthcare facilities.
The differences in the technological characteristics do not impact safety or effectiveness. For a complete comparison refer to Section 13 Substantial Equivalence of this premarket notification.
#### Performance Testing
See Section 18, Performance, Bench.
{2}------------------------------------------------
### Conclusion
The submitted information in this premarket notification shows that Blade raises no new questions of safety and effectiveness and the performance can be considered equivalent to the predicate devices. This conclusion is based on the comparison of Blade to indications/intended use, design, energy used/delivered, materials, performance, safety, effectiveness, and labeling of the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of a human figure embracing or protecting another figure, representing the department's mission of protecting the health of all Americans and providing essential human services.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 30, 2013
Stryker Communications Mr. Trev Thorsen Senior Regulatory Compliance Analyst 1410 Lakeside Parkway, Suite 100 Flower Mound, Texas 75028
Re: K132747
Trade/Device Name: Visum Blade LED Surgical Light System Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: November 8, 2013 Received: November 12, 2013
Dear Mr. Thorsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Trey Thorsen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Binita S. Ashar -- S 2013.12.30 75:56:53 -05'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
- * * * * * * *
1410 Lakeside Parkway, #100 Flower Mound, TX 75028 t: 972 410 7100 f: 972 410 7001 www.stryker.com
Communications
## Indications for Use
510(k) Number (if known): __K132747
Device Name: Visum Blade LED Surgical Light System
Indications for Use:
The Visum Blade LED Surgical Light System (Blade) is indicated to be used with various mounting configurations in operating rooms, examination rooms, emergency rooms, and all other health care facilities where the need for additional illumination exists.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen -A 요 11월 11일 : 1월 11일 : 1월 11일 : 1월 11일 : 1월 11일 : 1월 11일 : 1월 11일 : 1월 11일 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월 : 1월
for BSA
(Division Sign-off) Division of Surgical Devices 510(k) Number: K132747
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.