The Heat and Moisture Exchanger and Filter is a disposable single-use device indicated for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The products mentioned above are designed for disposable use and should be changed at least every 24 hours.
Device Story
Disposable, single-use heat and moisture exchanger and filter (HMEF) for respiratory circuits; incorporates electrostatic or mechanical filter pad within plastic housing. Connects to endotracheal tubes or face masks via ISO 5356 standard 15mm/22mm connectors; includes luer lock port for gas monitoring. Used in clinical settings (ICU, anesthesia, transport) by qualified medical personnel to provide humidification and bacterial/viral filtration for ventilator gases. Device reduces cross-contamination risk; requires replacement every 24 hours. Performance relies on internal filter media to balance moisture retention, filtration efficiency, and pressure drop.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including Bacterial Filtration Efficiency (BFE), Viral Filtration Efficiency (VFE), ISO 23328-1 salt testing, pressure drop, internal volume, moisture loss (ISO 9360), and ISO 5356 connector performance. Biocompatibility verified per ISO 10993-1.
Technological Characteristics
Housing: K-Resin (Styrene-Butadiene Copolymer). Filter: Polyolefin (electrostatic) or paper glass fiber (mechanical). Foam: Polyurethane impregnated with calcium chloride. Connectors: ISO 5356 compliant. Non-sterile. Standalone device.
Indications for Use
Indicated for adult patients requiring humidification during ventilator gas delivery and filtration to reduce cross-contamination between patient and equipment. For use by qualified medical personnel in hospitals, ICUs, anesthesia, respiratory therapy, transport, and with resuscitators.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
Drager Medical Breathing Circuit Bacterial Filter (K072002)
Engineering Medical Systems HEPA Filter / HME (K013089)
Submission Summary (Full Text)
{0}------------------------------------------------
K132709
VR Medical Heat and Moisture Exchanger and Filter Premarket Notification [510(k)] Application
#### 510(k) SUMMARY SECTION 5
Date of Submission: August 16, 2013
Owner/Submitter:
90 Gao Xin Road, ZhouZhuang Kunshan 215325, China Phone: 86 512 5720 0500
Contact Person:
Lee Pan, Ph.D. CEO, VR Medical Technology Co. 90 Gao Xin Road, ZhouZhuang Kunshan 215325, China T: 86 512 5720 0500 C: 137 8897 5958 I (US) 6262275727 lpan@vrmedical.net
Proprietary Name: VR Medical Heat and Moisture Exchanger and Filter
VR Medica! Technology Co.
Common/ Usual Name: Breathing Circuit Bacterial Filter
Classification Reference: 21 CFR 868.5260
Product Code: САН
Datex - Ohmeda HMEF 750 (K031653) Predicate Devices: Datex- Ohmeda HMEF Mini (K023641) Drager Medical Breathing Circuit Bacterial Filter (K072002) Engineering Medical Systems HEPA Filter / HME (K013089)
Indication for Use:
The Heat and Moisture Exchanger and Filter is a disposable single-use device indicated for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The products mentioned above are designed for disposable use and should be changed at least every 24 hours.
Patient Population/ Environment of Use:
The Heat and Moisture Exchanger and Filter is for use in hospital, ICU, anesthesia, respiratory therapy, during transport and with resuscitators. The device can be used on adult patients. The device is indicated for use by qualified medical personnel only.
{1}------------------------------------------------
### 9 of77 VR Medical Heat and Moisture Exchanger and Filter Premarket Notification [510(k)] Application
# Device Description:
The Heat and Moisture Exchanger and Filter (HMEF) consists of a plastic body which incorporates an electrostatic/ mechanical filter pad. 22 mm female / 15 mm male connectors in accordance with ISO 5356 (Anaesthetic and respiratory equipment-Conical connectors- Part 1: Cones and sockets) and a luer lock connector which is only used for gas monitoring.
| Item | HMEF | HMEF |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| | VR003 | VR005 |
| Filtration efficiency<br>(By Nelson Lab) | 99.999% (BFE*)<br>99.99% (VFE*) | 99.9999% (BFE*)<br>99.999% (VFE*) |
| Filtration mechanism | electrostatic | mechanical |
| Pressure Drop | 0.7 cm H2O at 30 L/min | 1.5 cm H2O at 30 L/min |
| Internal Volume | 60mL | 85mL |
| Recommended tidal<br>volume | 250 to 1500 mL | 250 to 1500 mL |
| Conical connectors<br>ISO 5356 | 15 mm Inner Diameter (ID) X22<br>mm Outer Diameter (OD) fitting to<br>connect to the endotracheal tube or<br>face mask 15 mm OD X22 mm ID<br>fitting to connect to the breathing<br>circuit Y-piece | 15 mm iD X22 mm OD<br>fitting to connect to the<br>endotracheal tube or face<br>mask 15 mm OD.fitting to<br>connect to the breathing<br>circuitY-piece |
| Moisture loss ISO 9360* | 10.5 mg/L at VT=500 mL | 7.8 mg/L at VT=500 mL |
| Gas sampling port | Yes | Yes |
| Material | | |
| Housing: | K-Resin (Styrene-Butadiene<br>Copolymer Plastic) | K-Resin (Styrene-Butadiene<br>Copolymer Plastic) |
| Filter medium | Synthetic fibers (Polyolefin) | Paper glass fiber |
| Foam | Polyurethane (PU) impregnated<br>with calcium chloride (CaCl2) | Polyurethane (PU)<br>impregnated with calcium<br>chloride (CaCl2) |
# * Notes
BFE: Bacterial filtration efficiency
VFE: Virus filtration efficiency
ISO 9360: Anaesthetic and respiratory equipment-Heat and moisture exchangers (HMEs) for humidifying respired gases in humans-Part 1:HMEs for use with minimum tidal volumes of 250 ml
Design verification tests (as well as an evaluation under ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process) were conducted on the HMEF as a result of the resulting risk analysis and product requirements.
It must be noted that the HMEF is provided non-sterile. Therefore, all corresponding sterilization tests are not necessary and were not performed.
{2}------------------------------------------------
# Substantial Equivalence:
The intended use and indications for use for the proposed device are the same as the predicate devices.
The comparison of the data demonstrates similar values for the key performance data. The VRM HMEF devices demonstrate similar values in filtration efficiency, internal volume, pressure drop and recommended tidal volumes when compared to the legally marketed devices.
It must be pointed out that there is an interdependency regarding filtration efficiency. moisture loss, internal volume and pressure drop. A higher amount of media will result in a higher efficiency in filtration or humidification but with the consequence of having more internal volume and higher pressure drop.
The main differences that may influence the technological characteristics between the proposed device and the predicate devices are the different shape and size of the housing. The shape of the proposed device is cuboid-shaped, but the predicate devices are cylindershaped. These differences can influence the performance of internal volume and related performances, but as the designed volume of the housing is substantial equivalent to predicate devices, the proposed device have the similar internal volume and related performances as the predicated devices. Thus the differences between the subject device and the predicate devices do not raise new questions of safety and effectiveness. .
The Heat and Moisture Exchanger and Filter is manufactured from K-Resin (Styrene-Butadiene Copolymer Plastic), Polyurethane (PU), Synthetic fibers (Polyolefin), Paper glass fibers. All materials have been evaluated in accordance with ISO 10993-1 Biological Evaluation of Medical Device -Part 1: Evaluation and testing within a risk management process and FDA General Program Memorandum #G95-1. Thus, VR Medical conducted two confirmatory in-vitro tests (utilize both polar and non-polar solvents) and five confirmatory tests (cytotoxicity, sensitization, irritation, genotoxicity, and implantation) which confirmed the acceptable biocompatible status of these materials,
#### Summary of Non-Clinical Tests:
The following in-vitro tests were conducted:
- Bacterial Filtration Efficiency (BFE) .
- . Viral Filtration Efficiency (VFE)
- . Salt Test (ISO 23328-1)
- . Pressure Drop
- . Internal Volume
- . Moisture Loss
- Conical Connectors Related Performance (ISO 5356) (Including the General Requirements and Special Requirement)
{3}------------------------------------------------
- . Package Integrity
- . Accelerated Aging
Package Transport Test**
- Constant Temperature and Humidity .
- . Compression Testing
- . Fixed Displacement Vibration Testing
- . Free Fall Drop Testing
- Package Transport Test Standards Utilized
- 1STA 2A ISTA Packaged Products 150 1b (68 kg) or less; ISTA 2 Series Partial Simulation Performance Test Procedure
- ASTM D4169-08 Standard Practices for Performing Testing of Shipping Containers and Systems
Biocompatibility
- · According to evaluation using ISO 10993-1 and FDA 095-1;
- · The materials are either identical to, or substantially equivalent to predicate devices;
- . Furthermore, two confirmatory in-vitro tests (utilize both polar and non-polar solvents) and five confirmatory tests (cytotoxicity, sensitization, irritation, genotoxicity, and implantation) demonstrated the materials acceptable biocompatible status
- Cytotoxicity (ISO 10993-5:2009) o
- Sensitization (ISO 10993-10:2010) 0
- Irritation (ISO 10993-10:2010) 0
- Genotoxity (ISO 10993-3:2003) 0
- 0 Implantation (ISO 10993-6:2007)
# Summary of Clinical Tests:
The subject of this premarket submission, VR Medical Heat and Moisture Exchanger and Filter did not require clinical studies to support substantial equivalence.
## Conclusion
VR Medical has demonstrated that the proposed device is as safe, as effective and the performance testing results are substantially equivalent to the described predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the branches of government. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
July 30, 2014
VR Medical Technology Co. Lee Pan, Ph.D. CEO 90 Gao Xin Road, ZhouZhuang Kun Shan, Jiangsu 215325 CHINA
Re: K132709
Trade/Device Name: VR Medical Heat and Moisture Exchanger and Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Class: Il Product Code: CAH Dated: May 21, 2014 Received: June 30, 2014
# Dear Dr. Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2 - Dr. Pan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm_ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fdagov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
7 of 77 VR Medical Heat and Moisture Exchanger and Filter Premarket Notification [510(k)] Application
.
INDICATION FOR USE STATEMENT SECTION 4
Page 1 of 1
510(k) Number (if known): _K132709
Device Name: VR Medical I-leat and Moisture Exchanger and Filter
Indication for Use:
The Heat and Moisture Exchanger and Filter is a disposable single-use device indicated for patients who require humidification during the delivery of ventilator gases and provide filtration for reducing possible cross contamination between patient and equipment. The products mentioned above are designed for disposable use and should be changed at least every 24 hours.
Patient Population/ Environment of Use:
The Heat and Moisture Exchanger and Filter is for use in hospital, ICU, anesthesia, respiratory therapy, during transport and with resuscitators. The device can be used on adult patients. The device is indicated for use by qualified medical personnel only.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
K132709
Anya C. Harry -S 2014.07.29 13:58:36 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.