K132699 · Tyrx, Inc. · FTL · Nov 26, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132699
Device Name
AIGIS RX N (MEDIUM), AIGIS RX N (LARGE)
Applicant
Tyrx, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Nov 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AIGISRx N is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body. AIGISRx N contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the gluteal region.
Device Story
AIGISRx N is a sterile, dual-component (resorbable and non-resorbable) surgical mesh envelope; constructed of knitted polypropylene filaments coated with a bioresorbable polyarylate polymer containing rifampin and minocycline. Device is implanted surgically by a physician to house a pulse generator (vagus nerve stimulator or spinal cord neuromodulator). The envelope secures the device in the body, creating a stable environment; antimicrobial agents elute to reduce infection risk. No electronic components or software; device is purely mechanical/chemical in function. Benefits include device stabilization and localized antimicrobial protection.
Clinical Evidence
No human clinical trials. Evidence includes bench testing (biocompatibility per ISO 10993, sterility per ISO 11137) and an in vivo sheep study. Sheep study (n=8) evaluated minocycline/rifampin diffusion in plasma and CSF following SCNM implantation with the envelope. Results showed no quantifiable concentrations of antimicrobial agents in plasma or CSF after 7 days, indicating no significant risk for CNS side effects.
Technological Characteristics
Knitted polypropylene mesh coated with bioresorbable polyarylate polymer containing rifampin and minocycline. Sterile, non-pyrogenic. Gamma sterilized. Same physical, chemical, and mechanical properties (suture retention, burst strength) as predicate.
Indications for Use
Indicated for patients requiring implantation of a vagus nerve stimulator (infraclavicular fossa) or spinal cord neuromodulator (lateral to body midline, superior to gluteal region) to provide a stable environment and reduce infection risk via antimicrobial agents.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
KB2699
NOV 2 6 2013
## 510(k) Summary
In accordance with 21 CFR 807.92, TYRX Inc. provides this summary of the safety and effectiveness information available for AIGISRx® N, as well as the substantial equivalence decision making process used for the AIGISRs® N subject device.
Sponsor/Applicant Name and Address:
Establishment Registration Number:
Sponsor Contact Information:
TYRX Inc. 1 Deer Park Drive Suite G Monmouth Junction, N.J. 08852
3005619263
Regina Novak Manager, Regulatory Affairs Phone: 732-964-1492 Fax: 732-964-1490 Email: rnovak@tyrx.com
Alternate Contact: Susan Olinger, JD Vice President, Regulatory Affairs Phone: 732-964-1110 Fax: 732-964-1490 Email: solinger@tyrx.com
August 28, 2013
AIGISRx® N
Surgical Mesh, Class II PROCODE: FTL
Date of Preparation of 510(k) Summary: New Device Trade/Proprietary Name:
Device Common/Classification Name:
{1}------------------------------------------------
#### Predicate Device Name and 510(k) Number:
### AIGISRS®N (K131007)
#### Device Description:
AIGISRs N is a dual component (resorbable and non-resorbable) sterile prosthesis designed to hold a vagus nerve stimulator or spinal cord neuromodulator securely to create a stable environment when implanted in the body. It is constructed of knitted filaments of polypropylene that are coated with a bioresorbable polyarylate polymer containing the antimicrobial agents rifampin and minocycline. Rifampin and minocycline have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of an implantable electronic device.
#### Device Intended Use:
AIGIS& "N is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body. AIGISRS® N contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the gluteal region.
The only difference between the subject and predicate devices is that the subject device has an expanded indication for use to include spinal cord neuromodulators (SCNM).
#### Technological Characteristics:
The physical, chemical and mechanical properties of the AIGIS N subject device, such as mesh knit characteristics, suture retention strength and burst strength and burst strength are the same as the AIGISRs N predicate device. There is no change to the design, materials of construction, or manufacturing processes for the subject device. There are no technological differences between the subject and predicate devices, and there are no design changes to the predicate device.
{2}------------------------------------------------
#### Performance Data:
AIGISRs N is designed to be a biocompatible, sterile device intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body. The AIGISRs N subject device is identical to the AIGISRs N predicate device. Information on the cleared predicate device, K131007, is included by reference in this 510(k).
AIGISRx N is sterile, biocompatible, and non-pyrogenic. Sterility conforms to ISO 11137, and bench testing shows that gamma sterilization has no detrimental effect on the chemical structure or thermal properties of the polypropylene substrate mesh. Standard ISO 10993 testing demonstrated the biocompatibility and safety of the device. An in vivo functionality study showed that AIGIS devices do not alter or interfere with an implantable pulse generator. The AIGISR: device (K063091) was the device used in the aforementioned studies and it was the predicate for AIGIS N (K131007). which was cleared on July 10, 2013. The AIGISRx N subject device is identical to the AIGIS&x N predicate device (K131007), there is no change to the design, materials of construction or manufacturing processes for the subject device. There are no technological differences or design changes. AIGIS N (K131007) is the predicate for our subject device, therefore, we believe that these studies are directly applicable to this new 510K. Information on the cleared predicate device. K131007, is included by reference in this 510(k).
To provide additional evidence on the safety of AIGIS N when implanted with SCNM, a study was conducted to address the question of minocycline diffusion from the AIGIS N device and possible CNS effects. The purpose of the study was to determine concentrations of minocycline and rifampin in the plasma and cerebrospinal fluid (CSF) of sheep implanted with a SCNM and an AIGISR: antibacterial envelope, with or without a spinal lead.
This study consisted of four (4) treatment groups: sheep in Group 1 received AIGIS or AIGIS R envelope with a SCNM: sheep in Groups 2 & 3 received AIGIS or AIGIS R and SCNM and a lead. Sheep in Group 4 were implanted with a single SCNM without the AIGIS device and received oral doses of minocycline and rifampin daily. Each group consisted of 2 sheep and the study duration was 7 days.
The study demonstrated that the AIGISRx N Antibacterial Envelope was safe in the sheep model as assessed by the absence of adverse clinical signs. There were no quantifiable concentrations of minocycline or rifampin in plasma and CSF samples collected up to 7 days after implantation of the SCNM enclosed in the AIGISRsN Antibacterial envelope. with or without leads. This study demonstrated that the amount of minocycline in CSF
{3}------------------------------------------------
and plasma, as a result of AIGISRs N implantation, are not detectable and do not pose a significant risk for CNS side effects.
## Conclusions:
Both the AIGISRs N predicate and the AIGISRx N subject devices are safe and effective for their intended uses.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three overlapping, curved lines forming its body and wings. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 26, 2013
TYRX, Inc. c/o Regina Novak Manager, Regulatory Affairs 1 Deer Park Drive, Suite G Monmouth Junction, NJ 08852
Re: K132699
Trade/Device Name: AIGISRx N Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: September 4, 2013 Received: September 5, 2013
Dear Ms. Novak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Prox (100) the the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misicading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that i Dri has intacted and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Regina Novak
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Joyce M. Whang -S
Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K132699
Device Name: AIGISRx N Antibacterial Envelope
Indications For Use:
AIGISRx N is intended to hold a vagus nerve stimulator or spinal cord neuromodulator securely in order to create a stable environment when implanted in the body. AIGISRx N contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators implanted in the infraclavicular fossa, or spinal cord neuromodulators implanted laterally to the body midline and slightly superior to the qluteal region.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of _ 1 _ _ _ _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.