K132691 · Radiometer Medical Aps · MQM · Nov 13, 2014 · Clinical Chemistry
Device Facts
Record ID
K132691
Device Name
ABL90 FLEX
Applicant
Radiometer Medical Aps
Product Code
MQM · Clinical Chemistry
Decision Date
Nov 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1113
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The ABL90 FLEX analyzer is an in vitro diagnostic, portable, automated analyzer that quantitatively measures neonatal bilirubin in heparinised capillary whole blood. The ABL90 FLEX analyzer is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. Bilirubin measurements on the ABL90 FLEX analyzer are intended to aid in assessing the risk of kernicterus in neonates.
Device Story
The ABL90 FLEX is a portable, automated in vitro diagnostic analyzer for whole blood analysis. It utilizes a sensor cassette containing electrochemical sensors and an optical oxygen sensor, alongside a 256-pixel array spectrophotometer for co-oximetry and bilirubin measurements. Operated by trained clinicians in laboratory or point-of-care settings, the device processes heparinized capillary whole blood samples. The spectrophotometric module performs multi-component analysis on hemolyzed sample portions to quantify neonatal bilirubin. Results are provided to healthcare providers to aid in assessing kernicterus risk in neonates, facilitating timely clinical decision-making. The system uses a two-point liquid calibration method.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP05-A2 across three POC sites and laboratory settings using aqueous samples and spiked whole blood (N=240 per aqueous sample; N=75 per spiked blood sample). Method comparison per CLSI EP09-A2 (N=224 capillary, N=210 syringe) showed good correlation with the predicate (R²=0.99). Linearity confirmed per CLSI EP6-A. Interference testing per CLSI EP07-A2 identified specific substances (e.g., Fluorescein, Methylene Blue) causing significant interference. LoB/LoD/LoQ determined per CLSI EP17-A2 as 1.1 mg/dL, 1.60 mg/dL, and 1.60 mg/dL, respectively.
Technological Characteristics
Portable, automated analyzer. Features miniaturized electrochemical sensors (film technology) and an optical oxygen sensor integrated into a replaceable sensor cassette. Uses a 256-pixel array spectrophotometer for optical measurements. Connectivity includes laboratory/POC integration. Calibration via two-point liquid method. Software-based analysis of spectrophotometric data.
Indications for Use
Indicated for neonates requiring quantitative measurement of bilirubin in heparinised capillary whole blood to aid in assessing the risk of kernicterus. Intended for use by trained healthcare professionals in laboratory or point-of-care settings under physician order.
Regulatory Classification
Identification
A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K132691
B. Purpose for Submission:
Addition of neonatal bilirubin assay to the ABL90 FLEX Analyzer
C. Measurand:
Neonatal Bilirubin
D. Type of Test:
Quantitative, Spectrophotometric
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
ABL90 FLEX
G. Regulatory Information:
| Classification name | Regulation | Device Class | Product Code | Panel |
| --- | --- | --- | --- | --- |
| Bilirubin in the neonate test system | 862.1113 | I, reserved | MQM | Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See Indication(s) for use.
{1}
2. Indication(s) for use:
The ABL90 FLEX analyzer is an in vitro diagnostic, portable, automated analyzer that quantitatively measures neonatal bilirubin in heparinised capillary whole blood. The ABL90 FLEX analyzer is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. Bilirubin measurements on the ABL90 FLEX analyzer are intended to aid in assessing the risk of kernicterus in neonates.
3. Special conditions for use statement(s):
For Prescription Use.
This test system is intended for Point-of-Care (POC) or clinical laboratory settings.
For neonatal use.
4. Special instrument requirements:
ABL90 FLEX Analyzer
I. Device Description:
Neonate Bilirubin (nBili) is a new test offered on the ABL90 FLEX Analyzer. The ABL90 FLEX Analyzer is a point of care and clinical laboratory instrument and has been previously cleared (k092686) for pH, pO2, pCO2, Sodium, Potassium, Calcium, Chloride, Glucose, Lactate, ctHB, sO2, fO₂Hb, fCOHb, fMetHb, fHHb and fHbF. Enabling the nBili measurement is accomplished through software design changes introduced in software version 2.6 by adding the bilirubin calculation algorithm. No hardware or mechanical changes were needed.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ABL800 FLEX
2. Predicate 510(k) number(s):
K050869
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: nBili on ABL 90 FLEX | Predicate Device: nBili on ABL800 FLEX (k050869) |
| Intended Use | The ABL90 FLEX analyzer is an in vitro diagnostic, portable, automated analyzer that quantitatively measures neonatal bilirubin in heparinised capillary whole blood. The ABL90 FLEX analyzer is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order. Bilirubin measurements on the ABL90 FLEX analyzer are intended to aid in assessing the risk of kernicterus in neonates. | Same |
| Principle of Operation | Using spectrophotometric multi-component analysis through the instrument's existing CO-Oximetry module on a hemolyzed part of the sample. | Same |
| Test Principle | Optical | Same |
| Calibrations | 2-point | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: nBili on ABL 90 FLEX | Predicate Device: nBili on ABL800 FLEX (k050869) |
| Intended use site | Laboratory and point-of-care. | Laboratory |
| Neonatal bilirubin measuring range | μmol/L: 218 - 6438 mg/dL: 1.26 – 37.93 mg/L: 126 – 3793 | μmol/L: 1 - 1000 mg/dL: 0.0 – 58.5 mg/L: 0 - 585 |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition.
CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
{3}
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition.
CLSI EP9-A2, Method Comparison and Bias Estimation Using Patient Samples, 2002
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
ISO 14971:2007, Second edition 2007-03-01, Medical devices - Application of risk management to medical devices.
IEC 62304 First edition 2006-05, Medical device software - Software life cycle processes.
## L. Test Principle:
Neonatal bilirubin measured in heparinized whole blood on the ABL90 FLEX analyzer using spectrophotometric multi-component analysis through the instrument’s existing CO-Oximetry module on a hemolyzed part of the sample. This optical system of the ABL90 FLEX analyzer is designed to measure total hemoglobin and hemoglobin fractions as well as oxygen saturation.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
##### Internal Testing
The internal precision study was performed as follows:
- A 20 day internal precision study was performed as specified in the CLSI Guideline EP5-A2 using one analyzer and three operators. Three levels of aqueous control solutions covering the device measuring range were analyzed in 2 runs per day with 2 replicates for each level. This generated within-run and total imprecision data for aqueous solutions of bilirubin. The results for both capillary and syringe mode are summarized below.
- An additional internal precision study was performed using freshly drawn whole blood, and freshly drawn cord blood, both spiked with bilirubin to obtain low, medium, and high bilirubin levels covering the device measuring range. All samples were adjusted to total hemoglobin (tHb) ~18g/dL and tonometered to 95 – 100% saturation to mimic neonate whole blood. All six levels of the prepared samples were analyzed in 5 runs with 5 replicates for each level. The results for both capillary and syringe mode are summarized below.
4
{4}
Bilirubin Syringe:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.05 | 2.4 | 0.07 | 3.2 |
| Mid Control | 80 | 7.4 | 0.11 | 1.5 | 0.14 | 1.9 |
| High Control | 80 | 30.9 | 0.45 | 1.4 | 0.57 | 1.8 |
| Sample 1 Adult Blood | 25 | 2.6 | 0.13 | 5.0 | 0.21 | 8.3 |
| Sample 2 Adult Blood | 25 | 7.4 | 0.09 | 1.2 | 0.15 | 2.1 |
| Sample 3 Adult Blood | 25 | 29.4 | 0.51 | 1.7 | 0.52 | 1.8 |
| Sample 4 Cord Blood | 25 | 2.4 | 0.12 | 4.9 | 0.18 | 7.4 |
| Sample 5 Cord Blood | 25 | 7.1 | 0.24 | 3.4 | 0.28 | 4.0 |
| Sample 6 Cord Blood | 25 | 29.9 | 0.21 | 0.7 | 0.26 | 0.9 |
Bilirubin Capillary:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.05 | 2.3 | 0.07 | 3.4 |
| Mid Control | 80 | 7.4 | 0.12 | 1.6 | 0.13 | 1.7 |
| High Control | 80 | 30.7 | 0.48 | 1.6 | 0.46 | 1.5 |
| Sample 1 Adult Blood | 25 | 2.3 | 0.15 | 6.7 | 0.18 | 7.7 |
| Sample 2 Adult Blood | 25 | 7.8 | 0.22 | 2.8 | 0.34 | 4.4 |
| Sample 3 Adult Blood | 25 | 30.2 | 0.73 | 2.4 | 0.80 | 2.6 |
| Sample 4 Cord Blood | 25 | 2.0 | 0.10 | 4.9 | 0.13 | 6.7 |
| Sample 5 Cord Blood | 25 | 7.0 | 0.19 | 2.7 | 0.24 | 3.4 |
| Sample 6 Cord Blood | 25 | 30.6 | 0.32 | 1.1 | 0.34 | 1.1 |
## External Testing
The external precision study was performed at three Point-of-Care (POC) sites as follows:
- A 20 day external precision study was performed as specified in the CLSI Guideline EP5-A2 using one analyzer and three operators per site. Three levels of aqueous control solutions covering the device measuring range were analyzed in 2 runs per day with 2 replicates for each level. This generated within-run and total imprecision data for aqueous solutions of bilirubin. The results for both capillary and syringe mode are summarized below.
- An additional external precision study was performed using freshly drawn adult whole blood spiked with bilirubin to obtain low, medium, and high bilirubin levels covering the device measuring range. All samples were adjusted to tHb ~18g/dL and tonometered to 95 – 100% saturation to mimic neonate whole blood. Three levels of bilirubin were analyzed in 5 runs with 5 replicates for each level in one day. The results for both capillary and syringe mode are summarized below.
{5}
Site 1:
Bilirubin Syringe:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.05 | 2.4 | 0.08 | 4.1 |
| Mid Control | 80 | 7.3 | 0.05 | 0.7 | 0.09 | 1.3 |
| High Control | 80 | 30.3 | 0.09 | 0.3 | 0.50 | 1.6 |
| Low Adult Blood | 25 | 2.5 | 0.12 | 4.8 | 0.12 | 4.9 |
| Mid Adult Blood | 25 | 7.2 | 0.14 | 1.9 | 0.14 | 1.9 |
| High Adult Blood | 25 | 31.1 | 0.21 | 0.7 | 0.19 | 0.6 |
Bilirubin Capillary:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.04 | 1.8 | 0.08 | 3.7 |
| Mid Control | 80 | 7.3 | 0.06 | 0.8 | 0.10 | 1.3 |
| High Control | 80 | 30.2 | 0.19 | 0.6 | 0.49 | 1.6 |
| Low Adult Blood | 25 | 2.3 | 0.14 | 6.1 | 0.31 | 13.4 |
| Mid Adult Blood | 25 | 7.1 | 0.15 | 2.1 | 0.61 | 8.5 |
| High Adult Blood | 25 | 30.4 | 0.16 | 0.5 | 0.37 | 1.2 |
Site 2:
Bilirubin Syringe:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.2 | 0.06 | 2.9 | 0.11 | 4.9 |
| Mid Control | 80 | 7.6 | 0.07 | 0.9 | 0.13 | 1.6 |
| High Control | 80 | 31.6 | 0.28 | 0.9 | 0.41 | 1.3 |
| Low Adult Blood | 25 | 2.6 | 0.11 | 4.2 | 0.12 | 4.5 |
| Mid Adult Blood | 25 | 7.5 | 0.16 | 2.1 | 0.15 | 2.0 |
| High Adult Blood | 25 | 30.9 | 0.21 | 0.7 | 0.20 | 0.6 |
Bilirubin Capillary:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.2 | 0.05 | 2.3 | 0.08 | 3.6 |
| Mid Control | 80 | 7.5 | 0.10 | 1.4 | 0.13 | 1.7 |
| High Control | 80 | 31.3 | 0.38 | 1.2 | 0.40 | 1.3 |
| Low Adult Blood | 25 | 1.8 | 0.11 | 5.7 | 0.14 | 7.3 |
| Mid Adult Blood | 25 | 7.2 | 0.19 | 2.6 | 0.20 | 2.8 |
| High Adult Blood | 25 | 29.5 | 0.22 | 0.7 | 0.41 | 1.4 |
{6}
Site 3:
Bilirubin Syringe:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.06 | 2.8 | 0.10 | 4.8 |
| Mid Control | 80 | 7.3 | 0.06 | 0.9 | 0.11 | 1.5 |
| High Control | 80 | 30.4 | 0.20 | 0.7 | 0.27 | 0.9 |
| Low Adult Blood | 25 | 2.4 | 0.15 | 6.1 | 0.34 | 13.9 |
| Mid Adult Blood | 25 | 6.8 | 0.13 | 1.9 | 0.30 | 4.5 |
| High Adult Blood | 25 | 30.5 | 0.36 | 1.2 | 0.46 | 1.5 |
Bilirubin Capillary:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 80 | 2.1 | 0.05 | 2.3 | 0.09 | 4.2 |
| Mid Control | 80 | 7.3 | 0.09 | 1.2 | 0.11 | 1.5 |
| High Control | 80 | 30.3 | 0.37 | 1.2 | 0.34 | 1.1 |
| Low Adult Blood | 25 | 2.2 | 0.11 | 5.2 | 0.39 | 17.7 |
| Mid Adult Blood | 25 | 6.9 | 0.14 | 2.1 | 0.42 | 6.2 |
| High Adult Blood | 25 | 30.0 | 0.40 | 1.3 | 0.65 | 2.2 |
All 3 Sites Combined:
Bilirubin Syringe:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 240 | 2.1 | 0.06 | 2.7 | 0.10 | 4.6 |
| Mid Control | 240 | 7.4 | 0.06 | 0.8 | 0.11 | 1.5 |
| High Control | 240 | 30.7 | 0.20 | 0.7 | 0.40 | 1.3 |
| Low Adult Blood | 75 | 2.5 | 0.13 | 5.0 | 0.22 | 8.7 |
| Mid Adult Blood | 75 | 7.2 | 0.14 | 2.0 | 0.21 | 2.9 |
| High Adult Blood | 75 | 30.8 | 0.27 | 0.9 | 0.31 | 1.0 |
Bilirubin Capillary:
| Sample | N | Mean (mg/dL) | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Low Control | 240 | 2.1 | 0.04 | 2.1 | 0.08 | 3.8 |
| Mid Control | 240 | 7.4 | 0.09 | 1.2 | 0.11 | 1.5 |
| High Control | 240 | 30.6 | 0.32 | 1.1 | 0.42 | 1.4 |
| Low Adult Blood | 75 | 2.1 | 0.12 | 5.7 | 0.30 | 14.0 |
| Mid Adult Blood | 75 | 7.1 | 0.16 | 2.3 | 0.44 | 6.3 |
| High Adult Blood | 75 | 30.0 | 0.28 | 0.9 | 0.49 | 1.6 |
{7}
b. Linearity/assay reportable range:
A linearity study was performed using a low concentration bilirubin sample, a high concentration bilirubin sample, and nine intermediate concentration bilirubin samples made by combining proportions of the low and high samples volumetrically. The high, low, and intermediate samples were run in replicates of four in one day. The data was analyzed in accordance with CLSI EP-6 with sample range tested from 0.9 to 40.3 mg/dL. The method is linear (first order) over the entire measuring range and yields the following linear regression results: Y = 0.968X + 0.2984, R² = 0.9996. These linearity results demonstrate the linearity of the device across the claimed measuring range of 1.6 – 37.9 mg/dL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
There is no unique calibration measurement for nBili. The concentration of total hemoglobin (ctHb) calibration curve is used in nBili measurements. The ctHb calibrator was previously cleared in k092686.
The ctHb calibrators are traceable to Hemoglobin Cyanide Standard purchased commercially, which is traceable to the NIST Reference materials.
d. Detection limit:
The limit of detection was determined in accordance with CLSI EP-17A. For the blank samples, whole cord blood specimens had the plasma fraction replaced with saline. In order to calculate the limit of detection (LOD), five whole blood samples, with bilirubin concentrations in the range of 0.5 to 2.0 mg/dL were each assessed in replicates of four on two instruments over a minimum of three days. The limit of blank (LOB) was calculated from these results using the 95th percentile of the ascending rank ordered values for each instrument. The mean LOB calculated from the two instruments was 0.82 mg/dL. The LOD was calculated as LOD = LOB + C_p (Std Dev) and resulted in a mean LOD from the two instruments of 1.2 mg/dL. To determine LoQ, fourteen whole blood samples with bilirubin concentrations in the range of 0.35 to 2.45 mg/dL were each assessed in replicates of five on two instruments over a minimum of three days. LoQ was determined to be 1.2 mg/dL based on a mean %CV equal to or less than 20%.
The nbili measuring range is 1.6 – 37.9 mg/dL.
e. Analytical specificity:
Interference testing was performed by spiking the test matrix with cord blood samples to tHb ~18 g/dL and sO2 close to 100% in order to mimic neonatal whole blood, then spiking these samples with interferents at the concentrations shown below. The interference has been tested at 2 levels of bilirubin, 5 mg/dL and 15 mg/dL. Sponsor defines non-significant interference as <10% bias between the spiked and unspiked samples. Sponsor defines significant interference to be ≥10% bias. The tables below summarize the interference testing results.
{8}
There was non-significant interference with Evans Blue, Intralipid, HbF, Hemolysis, and Triglyceride at the highest concentration indicated below: (Sponsor defines non-significant interference as $< \pm 10\%$ )
| Substances tested | Highest concentration tested with non-significant interference |
| --- | --- |
| Evans Blue | 5 mg/L |
| Intralipid | 1000 mg/dL |
| HbF | 82% |
| Hemolysis | 20% (equivalent to approximately 3000 mg/dL hemoglobin) |
| Triglyceride | 500 mg/dL |
Based on the initial screening interference testing, the following substances have the following interferences observed.
Fluorescein tested at $40\mathrm{mg / L}$ yielded a $-309\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-111\%$ bias at $15\mathrm{mg / dL}$ bilirubin. Patent Blue V tested at $10\mathrm{mg / L}$ yielded a $-94\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-44\%$ bias at $15\mathrm{mg / dL}$ bilirubin. Methylene Blue tested at $60\mathrm{mg / L}$ yielded a $-449\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-120\%$ bias at $15\mathrm{mg / dL}$ bilirubin. Cardio Green tested at $30\mathrm{mg / L}$ yielded a $-82\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-36\%$ bias at $15\mathrm{mg / dL}$ bilirubin. SHb tested at $10\%$ yielded a $150\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $35\%$ bias at $15\mathrm{mg / dL}$ bilirubin. Hydroxocobalamin Hydrochloride tested at $2\mathrm{mg / L}$ yielded a $-15.8\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-85\%$ bias at $15\mathrm{mg / dL}$ bilirubin. Cyanocobalamin tested at $2\mathrm{mg / L}$ yielded a $-9\%$ bias at $5\mathrm{mg / dL}$ bilirubin level and a $-73\%$ bias at $15\mathrm{mg / dL}$ bilirubin.
Since there was significant interference for Fluorescein, Patent Blue V, Methylene Blue, Cardio Green, SHb, Hydroxocobalamin Hydrochloride, and Cyanocobalamin based on an initial screening test. (Sponsor defines significant interference as $\geq \pm 10\%$ ) Sponsor performed a dose response interference test per CLSI EP7-A2 and determined the concentration at which the interfering substance does not interfere with the assay. The dose response test results are provided in the table below.
{9}
| Interferents tested | Bilirubin concentration tested (mg/dL) | Concentration of interfering substance above which interference occurs |
| --- | --- | --- |
| Fluorescein | 5 | 1.5 mg/L |
| | 15 | 4 mg/L |
| Patent Blue V | 5 | 1.5 mg/L |
| | 15 | 2.5 mg/L |
| Methylene Blue | 5 | 0.75 mg/L |
| | 15 | 2 mg/L |
| Cardio Green | 5 | 3 mg/L |
| | 15 | 10 mg/L |
| SHb | 5 | 1.1% |
| | 15 | 1.6% |
| Hydroxocobalamin Hydrochloride | 5 | 0.19 g/L |
| | 15 | 0.5 g/L |
| Cyanocobalamin | 5 | 0.2 g/L |
| | 15 | 0.7 g/L |
pH interference was tested at levels from 6.8 to 7.9 and no significant interference was found.
In addition, Cyanmethemoglobulin (HiCN) and Beta-carotene are known interferents with the assay because the spectra for HiCN and Beta-carotene overlap with the spectrum of bilirubin. Results from samples containing these substances should not be used. The sponsor has the following limitation listed in the label:
“Since the spectra for HiCN and Beta-carotene overlap with the spectrum of bilirubin, these are known interfering substances. Results from samples containing these substances should not be used.”
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison study versus the predicate device (ABL835 analyzer) was conducted according to CLSI guideline “Method Comparison and Bias Estimation Using Patient Samples”, EP09-A2. The study was conducted for both capillary and syringe mode at three point-of-care sites and included a total of 224 samples for capillary mode and 210 samples for syringe mode spanning the entire measuring range. Freshly drawn whole blood neonate samples were tested on the predicate device first and the left over samples were tested on the candidate device. In order to cover the reportable range spiked blood samples were used. Spiked samples contributed no more than 20% of the total number of samples at each site for this
{10}
study. Site 1 used 8 spiked samples in both capillary and syringe testing. Site 2 and 3 used 9 spiked samples in both capillary and syringe testing. The linear regression results are summarized below:
Syringe mode:
| Site | N | Slope (95% CI) | Intercept (95% CI) mg/dL | R² | Sample range tested mg/dL |
| --- | --- | --- | --- | --- | --- |
| Site 1 | 74 | 0.9922 (0.964 – 1.020) | 1.0207 (0.64 – 1.40) | 0.9857 | 1.8 – 35.9 |
| Site 2 | 51 | 1.0054 (0.980 – 1.031) | 0.3744 (0.00 – 0.75) | 0.9924 | 2.0 – 37.9 |
| Site 3 | 85 | 0.9917 (0.969 – 1.014) | 0.3623 (0.01 – 0.71) | 0.9895 | 2.7 – 37.1 |
| All sites combined | 210 | 0.9903 (0.975 – 1.005) | 0.6574 (0.44 – 0.88) | 0.9878 | 1.8 – 37.9 |
Capillary mode:
| Site | N | Slope (95% CI) | Intercept (95% CI) mg/dL | R² | Sample range tested mg/dL |
| --- | --- | --- | --- | --- | --- |
| Site 1 | 77 | 0.9774 (0.950 – 1.005) | 1.1199 (0.76 – 1.48) | 0.9853 | 1.8 – 35.5 |
| Site 2 | 56 | 0.9977 (0.974 – 1.021) | 0.5358 (0.19 – 0.88) | 0.9927 | 2.1 – 37.3 |
| Site 3 | 91 | 0.9737 (0.948 – 0.999) | 0.4862 (0.09 – 0.88) | 0.9845 | 3.0 – 36.7 |
| All sites combined | 224 | 0.9760 (0.961 – 0.991) | 0.7741 (0.55 – 1.00) | 0.9861 | 1.8 – 37.3 |
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
{11}
4. Clinical cut-off: Not applicable
5. Expected values/Reference range:
The reference ranges are as follows:
| Age | Bilirubin |
| --- | --- |
| ≤24 hrs, premature | 103-205 μmol/L
1.0-8.0 mg/dL
10-80 mg/L |
| ≤24 hrs, full-term | 34-103 μmol/L
2.0-6.0 mg/dL
20-60 mg/L |
| ≤48 hrs, premature | 103-205 μmol/L
6-12 mg/dL
60-120 mg/L |
| ≤48 hrs | 103-171 μmol/L
6-10 mg/dL
60-100 mg/L |
| 3-5 days, premature | 171-239 μmol/L
10-14 mg/dL
100-140 mg/L |
| 3-5 days, full-term | 68-137 μmol/L
4-8 mg/dL
40-80 mg/L |
| >1 month | 3.4-17 μmol/L
0.2-1.0 mg/dL
2-10 mg/L |
Reference: ${}^{1}$ Tietz NW, Logan NM. Reference ranges. In: Tietz NW, ed. Fundamentals of clinical chemistry. 3rd ed. Philadelphia: WB Saunders Company, 1987: 944-75.
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.