The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy. It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors: • Uterine activity by using an external, pressure-sensitive TOCO transducer, and • Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer. Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment. This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
Device Story
Wireless fetal monitoring system accessory for Huntleigh FM800E monitors; replaces wired transducers. System comprises receiver unit and two wireless, water-tight transducers: pulsed Doppler ultrasound transducer for fetal heart rate (FHR) and pressure-sensitive TOCO transducer for uterine activity. Receiver docks transducers for battery charging and connects to FM800E monitor via RF/WMTS frequency band. Used in clinical/hospital settings by physicians or trained practitioners. Enables patient mobility, including use in bath or shower. Provides continuous monitoring data to FM800E monitor for clinical decision-making regarding fetal well-being and labor progression.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including ultrasound acoustic output (IEC 60601-2-37), environmental testing (ISO 60068-2), electrical safety (EN 60601-1), EMC (EN 60601-1-2), software validation (EN 62304), and biocompatibility (ISO 10993).
Technological Characteristics
Wireless transducer system using WMTS frequency band. Components: Receiver with docking stations, pulsed Doppler ultrasound transducer, pressure-sensitive TOCO transducer. Powered by internal rechargeable lithium batteries. Water-tight construction. Connectivity: RF transmission to receiver, wired interface to FM800E monitor. Standards: IEC 60601-2-37 (acoustic), ISO 60068-2 (environmental), EN 60601-1 (safety), EN 60601-1-2 (EMC), EN 62304 (software), ISO 10993 (biocompatibility).
Indications for Use
Indicated for monitoring fetal heart rate and maternal uterine contractions in pregnant women during antepartum and intrapartum periods. Suitable for use in clinical and hospital settings. Requires supervision by a licensed physician or trained health practitioner.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Avalon CTS Cordless Fetal Transducer System (K023931)
Submission Summary (Full Text)
{0}------------------------------------------------
K132650
Huntleigh Healthcare Ltd
FEB - 7 2014
# HUNTLEIGH DIAGNOSTIC PRODUCTS DIVISION
## Section 5: 510(k) Summary as required by Section 807.92(c)
#### APPLICANT INFORMATION 5.1
| Huntleigh Healthcare Limited, |
|------------------------------------------|
| Diagnostic Products Division |
| 35 Portmanmoor Road |
| CARDIFF, CF24 5HN |
| United Kingdom |
| Laura Garcia Deacon |
| Regulatory Affairs Engineer |
| +44 2920 447703 |
| +(44) 2920 492520 |
| Laura.deacon@huntleigh-diagnostics.co.uk |
| January 9, 2014 |
| |
## 5.2 DEVICE INFORMATION
| Proprietary Name: | Sonicaid™ Freedom<br>Wireless Fetal Monitoring System | |
|-----------------------------|-------------------------------------------------------|--|
| Common/Classification Name: | Perinatal Monitoring Device & Accessories | |
| Classification | Class II | |
| Product Code | HGM | |
| Classification Regulation | 884.2740 | |
#### 5.3 PREDICATE DEVICE
| Predicate Device: | Avalon CTS Cordless Fetal Transducer System<br>(K023931) |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Owned by: | Philips Medical Systems,<br>Nederland B.V. Post bus 10.000,<br>5680 DA Best,<br>Netherlands |
| Manufactured by: | Philips Medizin Systeme Boeblingen GmbH,<br>Cardiac and Monitoring Systems,<br>Hewlett-Packard Str.2,<br>71034 Boeblingen,<br>Germany |
{1}------------------------------------------------
#### 5.4 INDICATIONS FOR USE
#### Sonicaid™ Freedom Wireless Fetal Monitoring System
The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy.
It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors:
• Uterine activity by using an external, pressure-sensitive TOCO transducer, and
• Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer.
Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment.
This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
#### DEVICE DESCRIPTION 5.5
The Sonicaid™ Freedom is intended to monitor fetal heart movement and uterine activity by using wireless, non-invasive transducers on pregnant women during antepartum and intrapartum phases of pregnancy. This system is an optional accessory that must be connected to, and used with, a Sonicaid™ FM800E Monitor when a woman wishes to be mobile or in a water environment when being monitored or giving birth.
The Freedom has three major components; the Receiver and Ultrasound ('US') and Tocograph ('TOCO') transducers. The wireless transducers are powered by internal batteries that are recharged when docked on the Receiver.
The Sonicaid ™Freedom operates in the WMTS frequency bandwidth. There are 100 selectable channels available for use. If there is interference when installing the system, then the channel may be changed to achieve clear transmission.
#### Technology Overview
#### Receiver
The Receiver must be connected to a Sonicaid™ FM800E Monitor to operate the wireless transducers. The front panel has two docking stations for charging the TOCO and US transducers and for storing them and three sets of visual indicators, which inform an operator about the status of the transducers when docked or in use, include:
- . The charging indicators ensure that transducers are recognised by the receiver, docked properly and recharging until the battery is fully recharged.
- . The transmission indicators confirm that the transducers are in range and there is no interference during use.
- The battery indicator provides the operator with the relative level of battery capacity. ●
These indicators also inform the operator when there is interference, low battery capacity, when a patient is out of transmission range or if there is a fault.
{2}------------------------------------------------
The back of the Receiver contains the connecting points for the receiving antenna, RF-channel selector, interface to the Sonicaid FM800E Monitor and the power supply input.
The Receiver's internal components drive the system through the microcontroller, selectable radio frequency channels, an antenna and a near field communication ('NFC') transmitter.
#### Transducers
The US and TOCO transducers are both wireless, water tight and may be used when patients are stationary, mobile or in a water environment.
#### Ultrasound Transducer
The US transducer detects fetal heart movement by pulsed Doppier ultrasound. This is driven by a master clock that produces the pulse streams necessary for the detection of FHR. Other components include the battery power supply and an RF transmitter.
#### TOCO Transducer
The TOCO transducer is a pressure sensitive device that detects contractions or uterine activity. The key components within the TOCO transducer are the NFC transmitter, pressure sensor and the battery power supply.
#### 5.6 SUBSTANTIAL EQUIVALENCE
The Sonicaid™ Freedom is substantially equivalent to the cleared predicate device, the Avalon CTS Cordless Fetal Transducer System (K023931).
### 5.7 TECHNOLOGY COMPARISONS
The Sonicaid™ Freedom and the predicate device are equivalent in the following ways:
- . Each system has a Receiver unit with docking stations for each transducer. Both systems have a least one Ultrasound transducer and one TOCO transducer.
- · Both wireless transducer systems have been designed as accessories for use with compatible fetal monitoring systems with wired transducers.
- Neither unit can be used as a stand-alone system; both must be used only when they are . connected to specified fetal monitoring systems.
- · Both systems have TOCO and Ultrasound transducers that are wireless, powered by rechargeable lithium batteries and have the same water ingress rating; these features enable patients to be monitored when they are stationary, mobile or in a water environment.
- · Communication between the transducers and the Receiver is via a WMTS frequency band.
- . Both systems automatically recharge the transducers' batteries when transducers are docked on their respective Receivers.
- · Both systems have similar indicators to monitor the battery capacity and the operating status of the transducers.
- · The intended use, indications for use, patient population, clinical settings and operating conditions for monitoring fetal heart movement and uterine contractions are the same.
- Both systems use the same or similar methods and technologies for determining FHR and uterine activity; FHR by Doppler ultrasound and UA by pressure detection.
{3}------------------------------------------------
- • The Ultrasound transducer elements from both systems have a similar structure and their acoustic output levels are comparable and within safety limits.
#### 5.8 BENCH TESTS
Verification and validation tests were conducted to demonstrate performance, safety and substantial equivalence to the predicate device. Compliance to the following was achieved:
- · Ultrasound Transducer Acoustic Output (IEC 60601-2-37)
- · Environmental Requirements (ISO 60068-2)
- · Electrical Safety (EN 60601-1)
- · EMC (EN 60601-1-2 /FCC)
- · Software (EN 62304)
- · Material Biocompatibility (ISO 10993)
Simulated bench tests were used to demonstrate that the performance of the Sonicaid™ Freedom Ultrasound and TOCO transducers is comparable to the Sonicaid™ FM800E transducers.
#### CLINICAL TESTS 5.9
No clinical tests were required to determine substantial equivalence between the Freedom and the predicate device.
#### 5.10 CONCLUSIONS
The conclusions drawn from the data detailed within this submission demonstrate that the Sonicaid™ Freedom is as safe, as effective, and performs as well as the legally marketed Avalon CTS Cordless Fetal Transducer System (K023931).
The Sonicaid™ Freedom does not raise any new concerns regarding safety and effectiveness.
Project Number: HD 778
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## February 7, 2014
Huntleigh Healthcare Limited Laura Garcia Deacon Regulatory Affairs Engineer 35 Portmanmoor Road Cardiff, Wales CF24 5HN United Kingdom
Re: K132650
Trade/Device Name: Sonicaid™ Freedom Wireless Fetal Monitoring System Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: January 9, 2014 Received: January 16, 2014
Dear Laura Garcia Deacon,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Laura Garcia Deacon
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
# Herbert P. Lerner -S
for
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
| 510(k) Number: | K132650 |
|----------------|---------|
|----------------|---------|
Sonicaid™ Freedom Wireless Fetal Monitoring System Device Name:
Indications for Use:
The Sonicaid™ Freedom Wireless Fetal Monitoring System ('Freedom') is a wireless transducer system for monitoring fetal heart movement and maternal contractions during intrapartum and antepartum periods of pregnancy.
It is an optional accessory for use with Huntleigh Healthcare Limited's Sonicaid™ FM820 and FM830 Encore Fetal Monitors ('FM800E Monitors') as an alternative to their wired transducers. When connected to an FM800E Monitor, the system monitors:
- · Uterine activity by using an external, pressure-sensitive TOCO transducer, and
- Fetal Heart Rate (FHR) by pulsed Doppler ultrasound using an external Ultrasound transducer.
Freedom is suitable for use with pregnant women in clinical and hospital facilities. The transducers are water tight allowing pregnant women to be monitored while they are mobile, stationary or in a bath or shower environment.
This system should only be used by, or under the supervision of, a licensed physician or other health practitioner who is trained in the use of FHR monitors.
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert P. Lerner -S 2014.02.07 12:04:08 -05'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.