← Product Code [NRY](/productcode/NRY) · K132641

# MODIFIED TREVO PROVUE RETRIEVER (K132641)

_Concentric Medical, Inc. · NRY · Jan 13, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132641

## Device Facts

- **Applicant:** Concentric Medical, Inc.
- **Product Code:** [NRY](/productcode/NRY.md)
- **Decision Date:** Jan 13, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

## Device Story

Modified Trevo ProVue Retriever is a neurovascular thrombus retrieval device; consists of flexible tapered core with shaped distal section; features radiopaque platinum wires and new truncated distal end for fluoroscopic visualization; hydrophilic coating reduces friction. Used in neurovasculature to restore blood flow by removing thrombus. Operated by physicians in clinical settings; device is navigated to thrombus site; shaped section engages clot for removal. Proximal end compatible with guide wire extensions for catheter exchange. Benefits patients by restoring cerebral perfusion in acute ischemic stroke. Device is supplied with torque device and insertion tool.

## Clinical Evidence

Bench testing only. Testing included dimensional verification, tensile strength (proximal, mid, tip, platinum wire), radial force, tip flexibility, torque tensile durability, radiopacity, deliverability, and simulated use in a 37°C silicone neurovascular model with simulated thrombus. Biocompatibility testing (cytotoxicity, hemolysis, physicochemical) confirmed compliance with ISO 10993-1.

## Technological Characteristics

Flexible tapered core with shaped distal section; radiopaque platinum wires; hydrophilic coating. Dimensions vary by size. Sterilization method same as predicate. Materials identical to predicate (ISO 10993-1 compliant). Standalone device; mechanical operation via manual manipulation under fluoroscopy.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Concentric Trevo ProVue Retriever ([K122478](/device/K122478.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 5. 510(k) Summary

| Trade Name:          | Modified Trevo ProVue Retriever                                                                                                                       |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Catheter, Thrombus Retriever                                                                                                                          |
| Classification Name: | Thrombus Retriever, 21CFR 870.1250 Class II                                                                                                           |
| Product Code:        | NRY                                                                                                                                                   |
| Submitter:           | Concentric Medical, Inc.<br>301 E. Evelyn Avenue<br>Mountain View, CA 94041<br>Tel 510-413-2681<br>Fax 510-413-2558<br>Facility Registration #2954917 |
| Contact:             | Sarah Meyer<br>Senior Regulatory Affairs Specialist                                                                                                   |
| Date Prepared:       | February 25, 2014                                                                                                                                     |
| Predicate Device:    | Concentric Trevo ProVue Retriever (K122478)                                                                                                           |

#### Device Description

The Modified Trevo ProVue Retriever consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. Radiopaque platinum wires in the shaped section and new radiopaque markers on the new truncated distal end allow fluoroscopic visualization. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel. The Modified Trevo ProVue Retriever is offered in two sizes.

#### Indications for Use

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

{1}------------------------------------------------

# Technological Characteristics and Product Feature Comparison

The Modified Trevo ProVue Retriever is substantially equivalent to the predicate device in terms of basic design, materials used, and function. A comparison of the subject device with predicate device is summarized in the below table.

| Feature                                  | Results                                                                             |
|------------------------------------------|-------------------------------------------------------------------------------------|
| Indications for Use                      | Same                                                                                |
| Device Description                       | Same except new truncated distal end with new distal<br>radiopaque markers.         |
| Target Population                        | Same                                                                                |
| Accessory Devices Provided               | Same                                                                                |
| Materials                                | Same                                                                                |
| Overall Length                           | Same except 3x20 size is longer.                                                    |
| Total Shaped Section Length              | Shorter than predicate device.                                                      |
| Active Shaped Section<br>Length          | Same                                                                                |
| Distal Tip and Distal Taper<br>Length    | Not applicable to subject device, modified device distal<br>end has been truncated. |
| Proximal Core Wire Diameter              | Same except 3x20 size has a smaller diameter                                        |
| Shaped Section Diameter                  | Same except 3x20 size has a smaller diameter                                        |
| Packaging Materials and<br>Configuration | Same                                                                                |
| Sterilization Method                     | Same                                                                                |
| How Supplied                             | Same                                                                                |

#### Product Feature Comparison of Subject Device with Predicate Device

### Risk Assessment

Risk assessment of the modifications has been conducted in accordance with EN ISO 14971:2012. Concentric Medical, Inc. has determined the modifications to the predicate device raise no new questions of safety or effectiveness.

Results of verification and validation testing have demonstrated the Modified Trevo ProVue Retriever is substantially equivalent to the predicate device. Furthermore, the modifications did not result in any new failure modes nor were there any changes to existing failure modes.

## Testing Summary

{2}------------------------------------------------

The results of verification and validation testing conducted on the Modified Trevo ProVue Retriever demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the predicate device. Specifically, the following tests were performed on the proposed device:

|      | Test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Test Method Summary                                                                                                                                                                                                                                                                                                    | Conclusions                                                                     |
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 1.   | Dimensional Verification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Verified dimensions using specified<br>measurement tool.                                                                                                                                                                                                                                                               | Dimensional verification meets<br>acceptance criteria.                          |
| 2.   | Retriever Proximal Joint<br>Tensile Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identified joint and cut sample for<br>test.  Recorded peak tensile force<br>results.                                                                                                                                                                                                                                  | Retriever Proximal Joint Tensile<br>Strength meets acceptance criteria.         |
| 3.   | Retriever Mid Joint<br>Tensile Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identified joint and cut sample for<br>test.  Recorded peak tensile force<br>results.                                                                                                                                                                                                                                  | Retriever Mid Joint Tensile Strength<br>meets acceptance criteria.              |
| 4.   | Retriever Tip Tensile<br>Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Loaded sample. Recorded peak<br>tensile force results.                                                                                                                                                                                                                                                                 | Retriever Tip Tensile Strength<br>meets acceptance criteria.                    |
| 5.   | Retriever Shaped<br>Section Radial Force                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Constrained shaped section of<br>retriever to specified diameter.<br>Recorded radial force results.                                                                                                                                                                                                                    | Retriever Shaped Section Radial<br>Force meets acceptance criteria.             |
| 6.   | Retriever/Vessel<br>Interaction (Tip<br>Flexibility)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Loaded sample so that the distal tip<br>was flexed.  Recorded peak<br>compression/flex force results.                                                                                                                                                                                                                  | Retriever/Vessel Interaction (Tip<br>Flexibility) meets acceptance<br>criteria. |
| 7.   | Retriever Torque<br>Tensile Durability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Gripped device and applied rotations<br>to torque device.  Pulled tensile<br>cycles to a max load then last cycle<br>to failure. Recorded results.                                                                                                                                                                     | Retriever Torque Tensile Durability<br>meets acceptance criteria.               |
| 8.   | Retriever Platinum Wire<br>Joint Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identified joint and cut sample for<br>test.  Recorded peak tensile force on<br>each individual platinum wire results.                                                                                                                                                                                                 | Retriever Platinum Wire Joint<br>Strength meets acceptance criteria.            |
| 9.   | Retriever Platinum Wire<br>and Joint Durability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Wrapped and unwrapped the entire<br>length of the shaped section of the<br>retriever (sheathed in insertion tool)<br>around a pin and repeat.  Performed<br>visual inspection and recorded<br>results. Performed deploy/reload<br>cycles into insertion tool.  Performed<br>visual inspection and recorded<br>results. | Retriever Platinum Wire and Joint<br>Durability meets acceptance criteria.      |
| 10.  | Radiopacity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Radiopacity was assessed based on<br>visual assessment of the device<br>being used under fluoroscopy.                                                                                                                                                                                                                  | Radiopacity meets acceptance<br>criteria.                                       |
| 11.  | Retriever/Microcatheter<br>Deliverability                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Measured the force to push the<br>device through a tortuous model.                                                                                                                                                                                                                                                     | Retriever/Microcatheter<br>Deliverability meets acceptance<br>criteria.         |
| 12.  | Proximal Curl<br>Resistance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Applied rotations to distal end of core<br>wire with device constrained.<br>Recorded the number of rotations<br>until a loop results or stop rotations<br>at specified number if no loop<br>results.                                                                                                                   | Proximal Curl Resistance meets<br>acceptance criteria.                          |
| 13.  | Simulated Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Simulated use testing used a silicone<br>neurovascular model cast from<br>actual human neurovascular arteries.<br>This bench testing model replicates<br>the tortuosity, diameter and location                                                                                                                         | Simulated Use meets acceptance<br>criteria.                                     |
| Test | Test Method Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Conclusions                                                                                                                                                                                                                                                                                                            |                                                                                 |
|      | of the arteries in the<br>neurovasculature including the<br>internal carotid artery (ICA) siphon.<br>The model ends at the mid carotid<br>arteries and proximal support is<br>provided by a guide catheter. The<br>model incorporates a re-circulating<br>water bath at 37°C pressurized<br>between 2 – 2.5 psi (100 – 126 mm<br>Hg) to simulate the human arterial<br>circulation. All testing follows the<br>procedural instructions outlined in<br>the Instructions for Use. Simulated<br>thrombus is used to assess the<br>devices ability to retrieve clot. |                                                                                                                                                                                                                                                                                                                        |                                                                                 |

{3}------------------------------------------------

#### Biocompatibility

The modified Trevo ProVue Retriever was assessed for impact to biocompatibility. Materials used in the modified Trevo ProVue Retriever are all the same materials used in the cleared Trevo ProVue Retriever (K122478). Furthermore, screening testing was performed successfully, providing further assurance that any design or process changes did not adversely affect the biocompatibility of the modified Trevo ProVue Retriever. Screening testing included cytotoxicity, hemolysis, and physicochemical testing, summarized below. Both the modified Trevo ProVue Retriever and the cleared Trevo ProVue Retriever meet biological safety requirements per ISO 10993-1 for externally communicating medical devices with circulating blood contact for less than 24 hours.

| Test<br>Description                                                 | Result                                                                                                                     |                            |           | Conclusion                                                                                                                                                                                                                                                |                                |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Cytotoxicity<br>Using the<br>ISO Elution<br>Method                  | Dilution                                                                                                                   | Result Grade               | Pass/Fail | Non-cytotoxic.                                                                                                                                                                                                                                            |                                |
|                                                                     | Undiluted                                                                                                                  | 3 – Moderate<br>reactivity | *         | The score of “3” is not attributed to the<br>changes made to the modified Trevo<br>ProVue Retriever since this result is<br>consistent with historical cytotoxicity data<br>collected on Retriever devices (such as<br>K122478) with hydrophilic coating. |                                |
|                                                                     | 50%                                                                                                                        | 0 – No reactivity          | Pass      |                                                                                                                                                                                                                                                           |                                |
|                                                                     | 25%                                                                                                                        | 0 – No reactivity          | Pass      |                                                                                                                                                                                                                                                           |                                |
|                                                                     | 12.5%                                                                                                                      | 0 – No reactivity          | Pass      |                                                                                                                                                                                                                                                           |                                |
|                                                                     | 6.25%                                                                                                                      | 0 – No reactivity          | Pass      |                                                                                                                                                                                                                                                           |                                |
|                                                                     | 3.13%                                                                                                                      | 0 – No reactivity          | Pass      |                                                                                                                                                                                                                                                           |                                |
| ASTM<br>Hemolysis<br>(combined<br>direct<br>contact and<br>extract) | Pass.<br>Hemolytic index: 0.0% (direct contact) and 0.4%<br>(extract)                                                      |                            |           | Non-Hemolytic                                                                                                                                                                                                                                             |                                |
| Physicoche-<br>mical Tests<br>Plastics                              | Pass.<br>Non-volatile residue: 2 mg<br>Residue on ignition: ≤ 2 mg<br>Heavy metal: < 1 ppm<br>Buffering capacity: < 1.0 mL |                            |           |                                                                                                                                                                                                                                                           | No heavy metals or leachables. |

{4}------------------------------------------------

## Summary of Substantial Equivalence

The Modified Trevo ProVue Retriever is substantially equivalent to the predicate device with regard to device design, materials, intended use, and patient population. The conclusions drawn from risk assessments and the verification and validation testing conducted using the Modified Trevo ProVue Retriever demonstrate that the device performs as designed, is suitable for its intended use and is substantially equivalent to the legally marketed predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and three wavy lines extending from the top. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 13, 2014

Concentric Medical, Inc. c/o Ms. Sarah Meyer Senior Regulatory Affairs Specialist 301 E. Evelyn Ave. Mountain View, CA 94041

Re: K132641

Trade/Device Name: Modified Trevo ProVue Retriever Regulation Number: 21 CFR 870.1250 Regulation Name: Catheter, Thrombus Retriever Regulatory Class: Class II Product Code: NRY Dated: October 14, 2013 Received: October 15, 2013

Dear Ms. Meyer:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 - Ms. Sarah Meyer

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Joyce M. Whang -S

for

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Indications for Use

510(k) Number (if known): K132641

Device Name: Modified Trevo ProVue Retriever

Indications For Use:

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

# Joyce M. Whang -S

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K132641](https://fda-staging.innolitics.com/device/K132641)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).