← Product Code [LYZ](/productcode/LYZ) · K132637

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR (K132637)

_Zibo Intco Medical Products Co., Ltd. · LYZ · May 1, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132637

## Device Facts

- **Applicant:** Zibo Intco Medical Products Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** May 1, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination gloves; yellow color. Worn by healthcare examiners on hands or fingers to provide a barrier against contamination between patient and examiner. Manufactured from poly vinyl chloride, polyurethane, and diisononyl phthalate (DINP). Used in clinical or general medical settings. Provides physical protection; prevents cross-contamination. Evaluated for tensile strength, elongation, pinhole resistance, and biocompatibility (skin irritation/sensitization).

## Clinical Evidence

Bench testing only. No clinical data. Testing included physical dimensions, tensile strength, ultimate elongation (before and after aging), 1000 ml water fill test for pinholes (AQL 2.5), residual powder content (< 2 mg/glove), and biocompatibility (primary skin irritation and dermal sensitization).

## Technological Characteristics

Material: Poly vinyl chloride, polyurethane, diisononyl phthalate (DINP). Standards: ASTM D5250-06, ASTM D6124-06 (residual powder), ASTM D5151-06 (pinhole). Form factor: Disposable, powder-free, yellow, hand-shaped garment. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free ([K992821](/device/K992821.md))

## Reference Devices

- Shanghai Jiawang Industrial Co., Ltd. ([K031854](/device/K031854.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510 (K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.

- 1. Submitter's Identification:
Zibo Intco Medical Products, Co. Ltd. No. 18, Qingtian Road. Qilu Chemical Industry Park Zibo. Shandong Province China

Date summary prepared: March 23, 2014

- 2. Name of the Device:
Zibo Intco Medical Products, Co. Ltd. Powder-free Vinyl Patient Examination Gloves. Yellow Color

- Common name/classification name of the Device: 3.
Powder-Free Vinyl Patient Examination Gloves, Yellow Color

#### 4. Contact Person:

Jennifer Guo, Tel: 909-548-4828

#### Predicate Device Information: ಳು

Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)

#### Device Description: 6.

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80L YZ, and meets all requirement of ASTM Standard D5250-06.

- 7. Intended Use:
A patient examination ylove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner.

m

{1}------------------------------------------------

#### 8. Comparison to Predicate Devices:

Zibo Intco Medical Products, Co. Ltd. Powder-free Vinyl Patient Examination Gloves, Yellow Color are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd. (K992821) and Shanghai Jiawang Industrial Co., Ltd. (K031854)

### 9. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:

The standards used for Zibo Intco Medical Products. Co. Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2. AQL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5. Inspection Level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

#### Discussion of Clinical Tests Performed: 10.

Not Applicable - There is no hypoallergenic claim.

#### 11. Conclusions:

Zibo Intco Medical Products, Co. Ltd. Powder-free Vinyl Patient Examination Gloves. Yellow Color conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

|                                                                        | Proposed Device                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device<br>(K992821)                                             |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Description                                                            | Zibo Intco Medical Products, Co. Ltd<br>Powder-Free Vinyl Patient<br>Examination Gloves, Yellow Color                                                                                                                                                                                                                                                                                   | Shijiazhuang Hongxiang<br>Powder-free Vinyl Patient<br>Examination Gloves |
| Product Code                                                           | LYZ                                                                                                                                                                                                                                                                                                                                                                                     | LYZ                                                                       |
| Indications for Use                                                    | A disposable device intended for<br>medical purposes that is worn upon the<br>examiner's hands or fingers to prevent<br>contamination between patient and<br>examiner.                                                                                                                                                                                                                  | Substantially equivalent                                                  |
| Type of Use                                                            | Over-The-Counter Use                                                                                                                                                                                                                                                                                                                                                                    | Substantially equivalent                                                  |
| Labeling: Instruction<br>for use                                       | A garment covering the hand and wrist<br>area. Clovers have separate openings<br>for each finger and the thumb.                                                                                                                                                                                                                                                                         | Substantially equivalent                                                  |
| Labeling: Labels on<br>the carton                                      | Labels include: Product name; color;<br>"single use Only" size, piece count, lot<br>number, distributor name. and<br>manufacturer address.                                                                                                                                                                                                                                              | Substantially equivalent                                                  |
| Device Materials                                                       | Poly Vinyl Chloride<br>Polyurethane<br>Diisononyl Phthalate (DINP)                                                                                                                                                                                                                                                                                                                      | Substantially equivalent                                                  |
| Before Aging:<br>Tensile,<br>Strength(Mpa) and<br>Ultimate Elongations | Average Tensile Strength (Mpa): 19.2<br>Average Ultimate Elongations: 501%                                                                                                                                                                                                                                                                                                              | Substantially equivalent                                                  |
| After Aging: Tensile<br>Strength(Mpa) and<br>Ultimate Elongations      | Average Tensile Strength (Mpa): 16.8<br>Average Ultimate Elongations: 472%                                                                                                                                                                                                                                                                                                              | Substantially equivalent                                                  |
| Overall Length on<br>Medium Size                                       | Average over 249mm                                                                                                                                                                                                                                                                                                                                                                      | Substantially equivalent                                                  |
| Width of Palm on<br>Medium Size                                        | Average 95mm                                                                                                                                                                                                                                                                                                                                                                            | Substantially equivalent                                                  |
| Palm Thickness                                                         | Average 0.085 mm                                                                                                                                                                                                                                                                                                                                                                        | Substantially equivalent                                                  |
| Figure Thickness                                                       | Average 0.088 mm                                                                                                                                                                                                                                                                                                                                                                        | Substantially equivalent                                                  |
| Residual Powder                                                        | According to ASTM D6124-06<br>Standard Test Method for Residual<br>Powder on Medical gloves for the<br>determination of residual powder<br>content. Testing result indicates the<br>weight of all types of residual or<br>powder on finished powder-free<br>gloves as < 2 mg per glove and there is<br>no defect glove found according to<br>ASTM D6124-06.                             | Substantially equivalent                                                  |
| Pinhole Results                                                        | According to ASTM D5151-06,<br>Testing result indicates pinhole were<br>found less than two pieces gloves out<br>of 125 pieces gloves. AQL 2.5 is met.                                                                                                                                                                                                                                  | Substantially equivalent                                                  |
| Biocompatibility<br>Result: Primary Skin<br>Irritation                 | Not an irritant                                                                                                                                                                                                                                                                                                                                                                         | Substantially equivalent                                                  |
| Dermal Sensitization                                                   | Not a sensitizer                                                                                                                                                                                                                                                                                                                                                                        | Substantially equivalent                                                  |
| Summary of<br>comparison                                               | Zibo Intco Medical Products, Co. Ltd Powder-Free Vinyl Patient<br>Examination Gloves, Yellow Color (subject device) and Shijiazhuang<br>Hongxiang Powder-free Vinyl Patient Examination Gloves (predicate<br>device) are substantially equivalent in all technological<br>characteristics, including tensile strength, ultimate elongations,<br>thickness, residual powder and pinhole. |                                                                           |

# Table 7-2. Side-by-Side Comparison of Intended Use, Design, Material, Physical, Biocompatibility, and Performance Testing

:

. ,

.

{3}------------------------------------------------

:

. . . . .

:

۔۔۔

:

:

:

:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and centered horizontally. The background is plain white.

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its body and wings. The eagle is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66 Silver Spring, MD 20993-0002

May 1, 2014

Zibo Intco Medical Products Company, Limited C/O Frank Liu Basic Medical Industries, INC 12390 East End Avenue Chino, California 91710

Re: K132637

Trade/Device Name: Powder - free Vinyl Patient Examination Gloves, Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: March 24, 2014 Received: March 27, 2014

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Tejashri Purohit Sheth, M.D.
Director
Tejashri Purohit-Shath MD Chilean Deputy Dir
DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Orug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K132637

## Device Name

Powder-free Vinyl Patient Examination Gloves. Yellow Color

Indications for Use (Describe)

A patient examination glove is a disposable device inenced for medical purposes that is worn upon the examiner's hands of fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

図 Over-The-Counter Use (21 CFR 801 Subpan C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Elizabeth F. Clave 2014.04.30 15:39:09

FORM FDA 3881 (1/14)

---

**Source:** [https://fda-staging.innolitics.com/device/K132637](https://fda-staging.innolitics.com/device/K132637)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).