K132618 · Lifescan, Inc. · MRZ · Dec 16, 2013 · General Hospital
Device Facts
Record ID
K132618
Device Name
ONETOUCH REVEAL
Applicant
Lifescan, Inc.
Product Code
MRZ · General Hospital
Decision Date
Dec 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
OneTouch® Reveal™ is indicated for use by individuals or health care professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server database to support diabetes management. The device is indicated for professional use and over-the-counter sales.
Device Story
OneTouch® Reveal™ is a web-based diabetes management system; facilitates data transmission from home monitoring devices (glucose meters, insulin pumps) to a server database. Used by patients and healthcare professionals in homes or clinical facilities. System identifies glycemic patterns; generates reports; enables remote data viewing. Assists in tracking blood glucose levels and insulin doses to support glycemic control. Healthcare providers use output to inform clinical decision-making; patients benefit from improved diabetes management through pattern recognition and data tracking.
Clinical Evidence
Bench testing only. Evidence consists of software verification and validation testing and Human Factors Usability Studies demonstrating the device meets required specifications and performs as intended.
Technological Characteristics
Web-based software application; functions as a data management system for glucose meters and insulin pumps. Connectivity via server database. No hardware transmitter required. Software-based pattern recognition and reporting.
Indications for Use
Indicated for individuals with diabetes or health care professionals managing diabetes; used in home or clinical settings to transmit data from glucose meters and insulin pumps to a server for management.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
·KV32618
Image /page/0/Picture/1 description: The image shows the LifeScan logo. The logo features the word "LIFESCAN" in bold, sans-serif font. Below the word is a tagline in a smaller font. Above the word is a curved line that goes over the word. On the left side of the word is a circular design.
DEC 1 6 2013
.....
## 4. 510(k) Summary
| Name of Firm: | LifeScan Inc.<br>A Johnson & Johnson Company<br>965 Chesterbrook Bld.<br>Wayne, PA 19087 |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary 510(k) Contact: | Gabrielle Logan<br>Regulatory Affairs Analyst<br>LifeScan Inc.<br>A Johnson & Johnson Diabetes Care Franchise<br>Phone: 484-328-6159<br>Fax: 610-651-7271<br>Email: GLogan2@its.jnj.com |
| Secondary 510(k) Contact: | Amy Smith<br>Director, Regulatory Affairs<br>Advanced Sterilization Products<br>A Johnson & Johnson Company<br>Phone: 949-789-3803<br>Email: ASmith21@its.jnj.com |
| Date Prepared: | August 19, 2013 |
| Device Trade Name: | OneTouch® Reveal™ |
| Device Generic Name: | Accessories, Pump, Infusion System<br>Test, Blood Glucose, Over The Counter |
| Device Class: | II |
| Review Panel: | General Hospital |
| Product Code: | Classification Product Code: MRZ<br>Subsequent Product Code: NBW |
| Regulation Number: | Infusion Pump 21 CFR § 880.5725 |
| Predicate Device: | K101806 -- Aidera Diasend System (Aidera AB) |
| Device Description: | OneTouch® Reveal™ is a Web-based Diabetes management<br>system. The application is designed to assist health care<br>professionals and people with diabetes to track blood glucose<br>levels and insulin doses. The application identifies patterns to<br>help patients manage glycemic control. OneTouch® Reveal™<br>includes pattern recognition messages, reports, and the ability to<br>view patient data remotely. |
| Traditional 510(k) | Confidential |
| Intended Use /<br>Indications for Use: | OneTouch® Reveal™ is indicated for use by individuals or<br>health care professionals in the home or health care facilities for<br>transmitting data from home monitoring devices such as<br>glucose meters and insulin pumps to a server database to<br>support diabetes management. The device is indicated for<br>professional use and over-the-counter sales. |
| Comparison of the<br>technological<br>characteristics of the<br>device to the predicate<br>device: | The LifeScan Inc.'s OneTouch® Reveal™ is substantially<br>equivalent to the legally marketed predicate, Aidera Diasend<br>System and incorporates the following similar technology and<br>functionality including:<br>OTC and Rx Web-based software application Manage patient accounts, filter/sort/search patients, print reports Unlike the predicate, the LifeScan Inc.'s OneTouch® Reveal™<br>does not use a transmitter to transfer data. |
| Performance Data<br>(Nonclinical and/or<br>Clinical): | Performance testing consisted of software verification and<br>validation testing and Human Factors Usability Studies that<br>demonstrated OneTouch® Reveal™ meets its required<br>specifications and performs as intended. |
| Substantial Equivalence to<br>Predicate Device: | Based on the information presented in this submission,<br>OneTouch® Reveal™ does not raise new questions of safety<br>and effectiveness and is substantially equivalent in its intended<br>use, performance, safety, effectiveness and underlying scientific<br>and operating principles compared to the predicate. |
: : : :
LifeScan OneTouch® Reveal™
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for LifeScan, a Johnson & Johnson company. The logo features the word "LIFESCAN" in bold, black letters. Below the company name is the text "a Johnson & Johnson company" in a smaller font. To the left of the company name is a graphic element that appears to be a stylized representation of a globe or sphere.
","text","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","title","titl
·
,
.
:
. '
: · ·
Traditional 510(k) LifeScan OneTouch® Reveal™ Confidential
.
・・
.
、
.
...
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three parallel lines above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G605 Silver Spring, MD 20993-0002
### December 16, 2013
LifeScan Incorporated A Johnson & Johnson Company Ms. Gabrielle Logan Regulatory Affairs Analyst 965 Chesterbrook Bld Wayne, PA 19087
Re: K132618
Trade/Device Name: OneTouch® Reveal™ Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ, NBW Dated: November 13, 2013 Received: November 14, 2013
### Dear Ms. Logan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Logan
..............................................................................................................................................................................
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
- S
Sincerely yours,
# Kwame O. Ulmer
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K132618
### Device Name
OneTouch® Reveal™
### Indications for Use (Describe)
OneTouch® Reveal™ is indicated for use by individuals or health care professionals in the home or health care facilities for transmitting data from home monitoring devices such as glucose meters and insulin pumps to a server databes management. The device is indicated for professional use and over-the-counter sales.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
.. FOR FDA USE ONLY . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Digitally signed by Richard C. Chapman Date: 2013.12.16 12:15:19 -05'00' PSC Publishing Services (301) 443-6740 FORM FDA 3881 (9/13) EI
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.