ONTEX ROLL WADDING PLASTIC APPLICATOR COMPACT SIZE UNSCENTED TAMPONS
Applicant
Ontex Bvba
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Apr 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The Ontex Roll wadding plastic Applicator Compact size Unscented Tampons are inserted into the vagina to absorb menstrual discharge.
Device Story
Menstrual tampon consisting of viscose absorbent pledget with polymeric overwrap and polyester/cotton withdrawal cord; features telescopic plastic (polyethylene) applicator for compact storage. User inserts tampon into vagina to absorb menstrual discharge; applicator facilitates placement and is discarded after use. Available in three absorbencies (6-9g, 9-12g, 12-15g). Device is non-sterile, single-use, and intended for over-the-counter use.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and microbiological analysis. Bench tests included Syngina absorptive capacity, fiber loss (<1 mg), cord strength (<50N), and applicator expulsion force (<6N). Microbiological testing confirmed absence of pathogens (Candida albicans, gram-negative bacteria, S. aureus, P. aeruginosa) and compliance with total viable count specifications (<10 cfu/g).
Technological Characteristics
Materials: viscose pledget, polymeric overwrap, polyester/cotton cord, polyethylene applicator. Design: roll wadding form with telescopic compact applicator. Absorbency: 6-15g range. Compliance: ISO 10993-1 for biocompatibility. Single-use, non-sterile.
Indications for Use
Indicated for individuals requiring vaginal absorption of menstrual discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Ontex Digital and Plastic/Cardboard Applicator Unscented Tampons (K122603)
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary
#### for
## ONTEX
# ROLL WADDING PLASTIC APPLICATOR COMPACT SIZE UNSCENTED TAMPONS
#### 1. Submission Sponsor
'
ONTEX BVBA SPINNERIJSTRAAT 12 9240 ZELE BELGIUM Phone: + 32 9 376 77 06 Fax: +32 9 378 13 33 Contact: HENRI LESAGE, R&D Manager Strategic Projects
#### 2. Submission Correspondent
Emergo Europe Consulting Prinsessegracht 20 2514 AP The Hague Cell Phone: +33 (0)6 89 83 16 09 Office Phone: + 31 (0) 70 345 8570 Direct: + 31 (0) 70 850 8249 Fax: + 31 (0) 70 346 7299 Contact: Rachel PAUL, Senior Consultant, QA/RA Email: pmteurope@emergogroup.com
- 3. Date Prepared
16th August 2013
#### 4. Device Identification
ONTEX ROLL WADDING PLASTIC APPLICATOR COMPACT SIZE Trade/Proprietary Name: UNSCENTED TAMPONS UNSCENTED MENSTRUAL TAMPONS Common/Usual Name: UNSCENTED MENSTRUAL TAMPON Classification Name: 21 CFR 884.5470 Classification Regulation: Product Code: HEB Class II Device Class: OBGYN, Obstetrics/Gynecology Classification Panel:
{1}------------------------------------------------
#### 5. Legally Marketed Predicate Device(s)
Ontex Digital and Plastic/Cardboard Applicator Unscented Tampons K122603
#### Device Description 6.
The Roll wadding plastic Applicator Compact size Unscented Tampon is a menstrual tampon.
These tampons are unscented tampons it means without any perfume.
These tampons are made from viscose material and polymeric overwrap. The withdrawal cord is made from polyester and cotton.
These tampons have an applicator. The applicator is made from plastic (polyethylene). . The applicator is in a compact size meaning that the inner tube and the outer tube are slided into each other telescopically so the inner tube needs to be retracted before usage.
These tampons have a Roll wadding form. In a roll tampon, the fleece band is folded and then rolled (like a Swiss roll) before being radially pressed.
The Roll wadding plastic Applicator Compact size Unscented Tampon is available in 3 absorbencies: regular, super plus absorbency (regular (6-9g), super (9-12g), and super plus (12-15g)).
| Tampon Type | Applicator material | Applicator Size | Absorbencies |
|--------------|---------------------|-----------------|---------------------|
| Roll wadding | Plastic | Compact | 6-9g, 9-12g, 12-15g |
The materials used in these tampons are exactly identical to the materials used to manufacture the predicate devices tampons.
#### 7. Indication for Use Statement
The Roll wadding plastic Applicator Compact size Unscented Tampons are inserted into the vagina to absorb menstrual discharge.
#### 8. Substantial Equivalence Discussion
The following table compares the ROLL WADDING PLASTIC APPLICATOR COMPACT SIZE, UNSCENTED TAMPONS to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{2}------------------------------------------------
# K132595
Page 3 of 5
| | Table SA -General Companson of Characteristics | | |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Manufacturer | ONTEX · | ONTEX | SIGNIFICANT<br>DIFFERENCES |
| Trade Name | ROLL WADDING<br>PLASTIC APPLICATOR<br>COMPACT SIZE<br>UNSCENTED TAMPONS | DIGITAL AND<br>PLASTIC/CARDBOARD<br>UNSCENTED<br>APPLICATOR TAMPONS<br>(K122603) | Roll wadding design<br>combined with a<br>compact size applicator |
| 510(k) Number | Not yet defined | K122603 | |
| Product Code | HEB | HEB | same |
| Regulation<br>Number | 21 CFR PART 884.5470 | 21 CFR PART 884.5470 | same |
| Regulation Name | Unscented menstrual<br>tampons | Unscented menstrual<br>tampons | same |
| Indications for Use | Inserted into the vagina to<br>absorb menstrual<br>discharge | Inserted into the vagina<br>to absorb menstrual<br>discharge | same |
| Material | Absorbent pledget in<br>viscose, polymeric<br>overwrap, cotton<br>polyester cord.<br>Plastic applicators in<br>polyethylene. | Absorbent pledget in<br>viscose, polymeric<br>overwrap, cotton<br>polyester cord.<br>Plastic applicators in<br>polyethylene .<br>Cardboard applicators<br>in paper. | same |
| Tampon Type and<br>Applicator<br>(material and size) | Roll wadding tampon type<br>with a plastic applicator in<br>compact size | Roll wadding tampon<br>digital (without<br>applicator)<br>W wadding tampon<br>type with a plastic<br>applicator in full size<br>(long) and in compact<br>size | Different, roll wadding<br>form combined to<br>compact applicator size |
| Absorbencies | 6-9g, 9-12g, 12-15g | <6g, 6-9g, 9-12g, 12-15g | Different no light<br>absorbency for roll<br>wadding plastic<br>applicator compact size<br>Tampons |
| Sterile | no | no | same |
| Single-Use | yes | yes | same |
| Complies with ISO<br>10993-1 | yes | yes | same |
### てったけ
.
:
:
.
-
: .
:
{3}------------------------------------------------
#### Non-Clinical Performance Data 9.
As the intended use (same type and duration of user contact), raw materials and manufacturing process are the same, it was not necessary to test again the biocompatibility (irritation, sensitization, cytotoxicity and acute systemic toxicity) and microbiology (growth of S. oureus, TSST-1, normal vaginal microflora) of the Roll wadding plastic Applicator Compact Size Unscented Tampons. The results of tests performed on the predicate are still applicable. The modifications do not adverselv alter the safety profile of the tampons.
The Roll wadding plastic Applicator Compact size Unscented Tampons are recognized to be non-cytotoxic, non-irritant, with no terminal or gross observations in the reproductive tracts of any of the animals, with no exhibiting toxic signs, and with a negligible dermal response. They did not present a potential for dermal irritation or allergic contact sensitization.
The Roll wadding plastic Applicator Compact Size Unscented Tampons are recognized to not enhance the growth of Staphylococcus aureus. They do not increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1). They have no effect on culture pH. They do not alter the growth of normal vaginal microflora.
#### 10. Performance Testing - Bench
The following testing has been performed to support substantial equivalence:
- Absorptive Capacity (Syngina) ●
- Fiber loss ●
- Cord strength .
- Expulsion force of the applicator ●
- Microbiology batch testing (total viable count (TVC), presence of yeasts/moulds, . Candida albicans, gram-negative bacteria, Staphylococcus aureus, Pseudomonas aeruginusa)
The results of this testing have shown that the performance of the subject device is equivalent to the predicates. Indeed, as for the predicates, testing gives similar results which are within the specifications previously defined for the predicates devices.
Fiber loss is less than 1 mg. Cord strength is less than 50N (same as predicate Roll Digital Tampon in viscose), Expulsion force of the applicator is less than the specification of 6N.
As for the predicates, the total viable count is less than 10 cfu/g of tampon and meets the specification of less than 200 cfu/g of tampon. The total presence of yeasts and moulds is less than 10 cfu/g of tampon and meets the specification of less than 20 cfu/g of tampon. As for the predicates, Candida albicans, gram-negative bacteria, Staphylococcus aureus and Pseudomonas aeruginusa are absent in 1g of tampon and meet the specifications of negativity.
{4}------------------------------------------------
The modifications have no impact on the technological performance and microbiology safety of the device.
#### 11. Statement of Substantial Equivalence
As part of demonstrating safety and effectiveness of ONTEX Roll wadding plastic Applicator Compact size Unscented Tampons and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, ONTEX has completed a number of tests. The Roll wadding plastic Applicator Compact size Unscented Tampons meet all the requirements for biocompatibility, and microbiology and confirms that the output meets the design inputs and specifications.
It can be shown in this 510(k) submission that the difference between the Roll wadding plastic Applicator Compact size Unscented Tampons and the predicate devices do not raise any questions regarding its safety and effectiveness.
Design, principals of operation, performance characteristics and intended use between the Roll wadding plastic Applicator Compact size Unscented Tampons and the predicate devices are identical. Bench testing and standard microbiological controls demonstrate that the Roll wadding plastic Applicator Compact size Unscented Tampons is substantially equivalent to the relevant aspects of the predicate devices, in terms of technological performance and microbiological safety. The Roll wadding plastic Applicator Compact size Unscented Tampons, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with outstretched wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2014
ONTEX BVBA % Rachel Paul Senior Consultant OA/RA Emergo Europe Consulting Prinsessegracht 20 The Hague 2514 AP Netherlands
Re: K132595
> Trade/Device Name: ONTEX ROLL WADDING PLASTIC APPLICATORS COMPACT SIZE UNSCENTED TAMPONS Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: February 21, 2014 Received: February 18, 2014
Dear Rachel Paul,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Rachel Paul
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRHOfTces/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number : K132595
Device Name: ONTEX ROLL WADDING PLASTIC APPLICATORS COMPACT SIZE UNSCENTED TAMPONS
Indications for Use:
The Ontex Roll wadding plastic Applicator Compact size Unscented Tampons are inserted into the vagina to absorb menstrual discharge.
. Prescription Use
AND/OR
Over-The-Counter Use _ X (21 CFR 801 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner-S
2014.04.23 11:05:37 -04'00'
Page 4 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.