K132592 · Cook, Inc. · DYB · Oct 9, 2014 · Cardiovascular
Device Facts
Record ID
K132592
Device Name
FLEXOR RADIAL HYDROPHILIC INTRODUCER ACCESS SET
Applicant
Cook, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Oct 9, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Flexor® Radial Hydrophilic Introducer Access Set is intended to introduce diagnostic and interventional devices in radial artery access procedures.
Device Story
The Flexor Radial Hydrophilic Introducer Access Set facilitates radial artery access for diagnostic and interventional procedures. The set includes an introducer sheath, dilator, wire guide, and access needle. The device is used by clinicians in a clinical setting to provide a conduit for other medical devices into the radial artery. The hydrophilic coating on the sheath enhances lubricity to aid insertion. The device is used under fluoroscopic guidance. By providing a stable access point, the device enables the delivery of diagnostic or therapeutic tools, potentially benefiting patients by allowing minimally invasive cardiovascular procedures.
Clinical Evidence
Bench testing only. Testing included tensile testing (BS EN ISO 11070:1999), needle tubing testing (EN ISO 9626:1995), valve leakage and dislodgement testing, fluoroscopic visibility, insertion/extraction force, coating durability/lubricity, particulate evaluation, and wire guide performance. Biocompatibility was confirmed per ISO 10993-1:2009 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogen, hemocompatibility, complement activation, PTT, and thromboresistance). Design validation was performed via simulated clinical use.
Technological Characteristics
The device consists of an introducer sheath, dilator, wire guide, and access needle. Features include a hydrophilic coating for lubricity and a Check-Flo valve. Materials and performance meet BS EN ISO 11070:1999 and EN ISO 9626:1995 standards. The device is designed for fluoroscopic visibility. Biocompatibility is compliant with ISO 10993-1:2009.
Indications for Use
Indicated for patients requiring radial artery access for the introduction of diagnostic and interventional devices.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Greatbatch Medical RadialSource Transradial Access Kit (K110051)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with three parallel lines above them, resembling a stylized flag or banner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 9, 2014
Cook, Inc. C/O David Lehr Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47402
Re: K132592
> Trade/Device Name: Flexor® Radial Hydrophilic Introducer Access Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 3, 2014 Received: September 4, 2014
Dear Mr. Lehr,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K132592
Device Name
Flexor® Radial Hydrophilic Introducer Access Set
Indications for Use (Describe)
The Flexor® Radial Hydrophilic Introducer Access Set is introduce diagnostic and interventional devices in radial arterv access procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Flexor® Radial Hydrophilic Introducer Access Set Traditional 510(k) 510(k) Summary 21 CFR §807.92
## Submitter Information:
| Applicant: | Cook Incorporated |
|-----------------------|--------------------------------------------------|
| Address: | 750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | David Lehr, RAC |
| Email: | david.lehr@cookmedical.com |
| Contact Phone Number: | 812-335-3575 ext. 102309 |
| Contact Fax Number: | 812-332-0281 |
| Date Prepared: | 07 October 2014 |
| Device Information: | |
| Trade Name: | Flexor® Radial Hydrophilic Introducer Access Set |
| Common Name: | Introducer Set |
| Classification Name: | Catheter Introducer<br>DYB (21 CFR §870.1340) |
# Predicate Devices:
The Flexor® Radial Hydrophilic Introducer Access Set is substantially equivalent to the following devices: the Terumo Glidesheath Slender (K122980), the Pinnacle ROII Introducer Sheath (K082847), and the Greatbatch Medical RadialSource Transradial Access Kit (K110051).
## Comparison to Predicate:
It has been demonstrated that the Flexor® Radial Hydrophilic Introducer Access Set is comparable to the predicate devices in terms of intended use, duration of use, principles of operation, technological characteristics, insertion method, and anatomical location.
{4}------------------------------------------------
Cook Incorporated Flexor® Radial Hydrophilic Introducer Access Set Traditional 510(k) 07 October 2014
# Device Description:
The Flexor Radial Hydrophilic Introducer Access Set is comprised of an introducer sheath, a dilator, a wire guide, and an access needle.
## Intended Use:
The Flexor® Radial Hydrophilic Introducer Access Set is intended to introduce diagnostic and interventional devices in radial artery access procedures.
# Test Data:
The proposed Flexor Radial Hydrophilic Introducer Access Set was subjected to applicable testing to ensure reliable design and performance under the testing parameters. The following tests were conducted to ensure reliable design and performance:
- . Dilator and introducer sheath tensile testing - Testing verified that, under proper clinical use of the dilator and introducer sheath, the peak load values shall be in accordance with the applicable values of BS EN ISO 11070:1999. The predetermined acceptance criteria were met.
- . Connecting tube to check-flo tensile testing - Testing verified that, under proper clinical use of the device, the connection between the valve body and the connecting tube will meet the predetermined acceptance criteria.
- Needle tensile testing - Testing verified that, under proper clinical use of the needle, the peak load values shall be in accordance with the applicable values of BS EN ISO 11070: 1999. The predetermined acceptance criteria were met.
- . Needle tubing testing - Testing verified that, under proper clinical use of the needle, the tubing will resist breakage in accordance with EN ISO 9626: 1995. The predetermined acceptance criteria were met.
- . Check-Flo valve dislodgement testing - Testing verified that the Check-Flo valve will not be dislodged during insertion and withdrawal when utilized according to the device's intended use. The predetermined acceptance criteria were met.
- Check-Flo valve liquid leakage testing Testing verified that the Check-Flo valve . will not experience excessive leakage when utilized according to the device's intended use. The predetermined acceptance criteria were met.
- Dimensional verification Testing verified that each component met the ● dimensions within tolerances of the predetermined acceptance criteria.
- Fluoroscopic visibility evaluation Testing verified that each component was ● visible under fluoroscopy. The predetermined acceptance criteria were met.
- Design validation Design was validated through simulated clinical use. The ● predetermined acceptance criteria were met.
- . Insertion and extraction force evaluation – The insertion and extraction forces of the dilator through the check-flo valve were evaluated and characterized. The predetermined acceptance criteria were met, where appropriate.
- Dilator and introducer sheath coating slough-off Testing verified that the ● hydrophilic coating met the requirements for durability. The predetermined acceptance criterion was met.
{5}------------------------------------------------
Cook Incorporated Flexor® Radial Hydrophilic Introducer Access Set Traditional 510(k) 07 October 2014
- Particulate evaluation Particulate counts were recorded in continuous flow loop. ●
- Introducer lubricity testing - Lubricity of the introducer shaft was evaluated. The predetermined acceptance criterion was met.
- . Coating integrity testing - Surface of device, including the hydrophilic coating, was visually examined at multiple time points.
- Wire guide testing Testing verified all predetermined acceptance criteria were . met for resistance to fracture, resistance to damage by flexing, tensile, tip deflection, torque response and torque strength as recommended in BS EN ISO 11070:1999 and in "Coronary and Cerebrovascular Guidewire Guidance," January 1995.
- . Biocompatibility testing - Testing (i.e., cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogen, hemocompatibility, complement activation, partial thromboplastin time, and thromboresistance) demonstrated the device as biocompatible. In conformance with the applicable sections of ISO 10993-1:2009, the predetermined acceptance criteria were met.
In conclusion, the results of these tests provide reasonable assurance to support a determination of substantial equivalence to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.