The SheathLess Eaucath Coronary Guide Catheter is intended to provide a pathway through which therapeutic and diagnostic devices are introduced. The guide catheter is intended to be used in the coronary vascular system.
Device Story
SheathLess Eaucath Coronary Guide Catheter provides pathway for introduction of therapeutic/diagnostic devices (e.g., balloon catheters, guide wires) into coronary vascular system. Device consists of catheter tube, proximal hub/connector, strain relief/protector, and dilator. Available in 6.5Fr and 7.5Fr sizes. Used by clinicians in interventional cardiology procedures. Operates as percutaneous guide catheter; facilitates access to coronary vasculature. Benefits include providing stable conduit for interventional instruments.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including corrosion resistance, force at break, leakage testing, kinking, coating integrity, radiopacity, dimensional verification, and comprehensive biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, thromboresistance, coagulation, complement activation, pyrogenicity).
Technological Characteristics
Percutaneous guide catheter system comprising catheter and dilator. Available in 6.5Fr and 7.5Fr. Materials and design consistent with standard coronary guide catheters. Sterilization via ethylene oxide. Standalone device.
Indications for Use
Indicated for use in the coronary vascular system to provide a pathway for the introduction of therapeutic and diagnostic devices. Not for use in the cerebral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K132556
OCT 2 4 2013
Premarket Notification SheathLess Eaucath Coronary Guide Catheter
# ASAHI INTECC CO.,LTD.
1703 Wakita-cho, Moriyama-ku, Nagoya, Aichi 463-0024 Japan Tel.+81-52-768-1211 Fax.+81-52-768-1221 Branch office : Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore,Beijing Research Facilities and Factories : Osaka, Seto. Thailand, Hanoi
#### 510(k) Summary [as required by 21 CFR 807.92(c)]
### SheathLess Eaucath Coronary Guide Catheter
510(k) _____
APPLICANT
Asahi Intecc Co., Ltd. 1703 Wakita-cho, Moriyama-ku Nagoya, Aichi 463-0024 Japan
Owner/Operator Number: 3003775027
OFFICIAL Correspondent
Yoshi Terai President, CEO Asahi Intecc USA, Inc. 2500 Red Hill Avenue, Suite 210 Santa Ana, CA 92705 Tel: (949) 756-8252 FAX (949) 756-8165 e-mail: yoshi.terai@asahi-intecc.com
SheathLess Eaucath Coronary Guide Catheter
Trade Name:
Common Name: Guide Catheter
Class 2 per 21 CFR §870.1250 DEVICE
Classification:
CLASSIFICATION Catheter, Percutaneous NAME:
PRODUCT CODE DQY
PREDICATE DEVICES:
K021256 6F Launcher Guide Catheter (Medtronic Vascular) K022764 7F Launcher Guide Catheter (Medtronic Vascular) K023402 8F Launcher Guide Catheter (Medtronic Vascular) K972484 ACS Viking Catheter (Guidant Corp.) K091188 ASAHI ZenyteEX Guiding Catheter (ASAHI Intecc Co., Ltd. K092372 Heartrail III Guiding Catheter (Terumo Medical Corp.)
- K051601 Pinnacle Destination (Terumo Medical Corp.)
Date Prepared: July 25, 2013
Asahi Intecc Co., Ltd
{1}------------------------------------------------
### INTENDED USE:
The SheathLess Eaucath Coronary Guide Catheter is intended to provide a pathway through which therapeutic and diagnostic devices are introduced. The guide catheter is intended to be used in the coronary vascular system.
#### DESCRIPTION:
The SheathLess Eaucath Coronary Guide Catheter is intended for use in coronary vascular applications and is designed to provide a pathway through which medical instruments, such as balloon catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.
The SheathLess Eaucath Coronary Guide Catheter consists of a "Catheter" and · "Dilator". The catheter consists of a tube, which is to be inserted into vasculature, a proximal hub/connector, and strain relief/protector. This product is being offered in two sizes: 6.5Fr and 7.5Fr.
### Comparison Table With Predicate Devices:
Comparisons of the SheathLess Eaucath Coronary Guide Catheter and predicate devices show that the technological characteristics of the SheathLess Eaucath such as intended use, components, design, materials, sterilization method, shelf life and operating principle are similar to the current marketed predicate devices.
# Non Clinical Testing / Performance Data:
Non clinical laboratory testing was performed on the SheathLess Eaucath Coronary Guide Catheter to determine substantial equivalence. The following testing/assessments were performed:
Performance test/evaluation summary:
- Corrosion resistance .
- . Force at break
- Liquid leakage under pressure test .
- . Air leakage into hub assembly during aspiration
- . Kinking short term test
- . Coating integrity / Particulate Evaluation
- . Radiopacity
- Dimensional Verification .
Biocompatibility:
- Cytotoxicity .
- Intracutaneous / Irritation .
- Sensitization .
- Systemic Toxicity .
{2}------------------------------------------------
Premarket Notification SheathLess Eaucath Coronary Guide Catheter
- . In Vitro Hemolysis
- . In Vivo Thromboresistance
- Plasma Recalcification Time Coagulation .
- . C3a Complement Activation
- . SC5b-9 Complement Activation
- Pyrogen .
Package integrity:
- . Visual inspection
- . Seal strength test
Sterilization:
- Ethylene oxide sterilization evaluation .
- EO residuals .
- Endotoxin .
### CONCLUSION:
The SheathLess Eaucath Coronary Guide Catheter has the same intended use and similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended, and is as safe and effective as its predicate.
Therefore, the SheathLess Eaucath Coronary Guide Catheter is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around a stylized caduceus symbol. The caduceus is depicted with a modern, abstract design, consisting of three curved lines. The text is written in a bold, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
October 24, 2013 -
Asahi Intecc Co., Ltd. C/O Semih Oktay CardioMed LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228
Re: K132556
> Trade/Device Name: SheathLess Eaucath Coronary Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 23, 2013 Received: August 26, 2013
Dear Semih Oktay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2- Semih Oktay
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. A. Zillicherm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification SheathLess Eaucath Coronary Guide Catheter
## Indications for Use
| 510(k) Number (if known): | K132556 |
|---------------------------|--------------------------------------------|
| Device Name: | SheathLess Eaucath Coronary Guide Catheter |
| Indications for Use: | |
The SheathLess Eaucath Coronary Guide Catheter is intended to provide a pathway through which therapeutic and diagnostic devices are introduced. The guide catheter is intended to be used in the coronary vascular system.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Lillehemann
Page 1 of 1
Asahi Intecc Co., Ltd
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.