The CT TAVI Planning application is intended to be used for patients with aortic valvular disease, severe symptomatic aortic stenosis or tricuspid aortic valve. The intended part of the body for this application is the human heart, specifically the ascending aorta, aortic root, coronary ostia and left ventricle in order to assess the aortic valve in pre-operational planning and transcatheter aortic valve replacement procedures. CT TAVI Planning is a non-invasive post-processing application providing 3D model-based segmentation of the aortic valve and aortic arch. The CT TAVI Planning application provides assessment and measurements of relevant heart structures for TAVI-device sizing, and allows the user to select a starting angle for C-arm position from the possible optimal positions from the CT TAVI and select a C-arm angle the user feels is appropriate to use in the catheterization laboratory by the Interventional team performing the procedure (to be used during the procedure itself). The physician retains the ultimate responsibility for making the determination of patient eligibility or which device is implanted based on their standard practices and additional imaging modalities such as echocardiography.
Device Story
Non-invasive post-processing software; inputs contrast-enhanced, ECG-gated (axial/helical) or un-gated helical CT images. Performs 3D model-based segmentation of aortic valve, ascending aorta, aortic root, left ventricle, and coronary ostia. Provides automated detection of anatomical landmarks (annulus, LVOT, sinotubular junction, sinus of valsalva, coronary ostia) with manual correction capability. Calculates diameters, areas, and distances for TAVI-device sizing. Determines optimal C-arm angles for procedural guidance. Used by radiologists/cardiologists in clinical settings for pre-procedural planning. Output includes 2D/3D visualizations, measurements, and reports. Assists interventional teams in catheterization labs by providing procedural C-arm positioning data; supports clinical decision-making regarding valve sizing and implantation strategy.
Clinical Evidence
No clinical trials; validation performed via internal testing by clinical experts in non-hospital environments and external validation at beta sites by local medical staff (radiologists/cardiologists). Results confirmed software meets user needs and intended use.
Technological Characteristics
Post-processing software for CT images. Supports contrast-enhanced, ECG-gated, and un-gated helical CT. Features model-based segmentation, landmark detection, and measurement tools. Connectivity via standard DICOM/PACS integration. Software-based algorithm for anatomical plane/landmark detection.
Indications for Use
Indicated for adult patients with aortic valvular disease, severe symptomatic aortic stenosis, or tricuspid aortic valve requiring pre-operational planning for transcatheter aortic valve replacement (TAVI).
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in a large, bold font. Above the word, there is some handwritten text that appears to be "K 132 524 Page 1 of 5". The text is written in black ink and is slightly slanted.
# Philips Medical Systems
JUN 1 3 2014
# 510(k) Summary
# CT TAVI Planning Application
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
- I: General Information
21 CFR 807.92 (a){l), (2)
| Company Name: | Philips Medical Systems (Cleveland), Inc. |
|-----------------------|-----------------------------------------------------------------------------|
| Address: | 595 Miner Rd<br>Cleveland, Ohio 44143<br>USA |
| Contact Person: | Susan Quick |
| Telephone Number: | 440-483-2291 |
| Prepared (date): | 2013 June 28 |
| Manufacturing Site: | Philips Medical Systems<br>9617978 Matam Building 34<br>Haifa, Israel 31004 |
| Trade Name of Device: | CT T Transcatheter Aortic Valve Implantation AV<br>Planning Application |
| Classification: | Class II |
| Regulatory Section: | Computed Tomography X-Ray System and<br>accessories 21 CFR 892.1750 |
| ProCode: | 90 JAK |
{1}------------------------------------------------
510(k) Summary CT TAVI Planning Application
# 21 CFR 807.92(a) (3): Legally marketed predicate device to which substantial equivalence is claimed:
The CT TAVI Planning Application is comparable in type and substantial equivalence to the legally marketed devices currently in commercial distribution, namely:
| 1. Predicate Device: | Philips HeartNavigatior |
|----------------------|----------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| Predicate Device k#: | K111245 |
| 2. Predicate Device: | Vitrea CT TAVR Planning |
| Manufacturer: | Vital Images |
| Predicate Device k#: | K122578 |
# 21 CFR 807.92(a) (4): Description of the device that is the subject of this premarket notification:
#### 1. Summary of functions of the device and its major components
The CT TAVI Planning application is intended to be used for patients with aortic valvular disease, severe symptomatic aortic stenosis or tricuspid aortic valve. The intended part of the body for this application is the human heart, specifically the ascending aorta, aortic root, coronary ostia and left ventricle in order to assess the aortic valve in pre-operational planning and transcatheter aortic valve replacement procedures.
CT TAVI Planning is a non-invasive post-processing application providing 3D model-based segmentation of the aortic valve and aortic arch. The CT TAVI Planning application provides assessment and measurements of relevant heart structures for TAVI-device sizing, and allows the user to select a starting angle for C-arm position from the possible optimal positions from the CT TAVI and select a C-arm angle the user feels is appropriate to use in the catheterization laboratory by the Interventional team performing the procedure (to be used during the procedure itself).
The physician retains the ultimate responsibility for making the determination of patient eligibility or which device is implanted based on their standard practices and additional imaging modalities such as echocardiography.
- . CT TAVI Planning allows users to select patient CT studies from various vendors.
- CT TAVI Planning requires contrast enhanced CT images. .
{2}------------------------------------------------
- CT TAVI Planning provides visualization of aortic valve calcification during the ● segmentation stage with tissues deselected.
- CT TAVI Planning provides segmentation of the ascending aorta, aortic root, left . ventricle and coronary ostia adjusted utilizing an algorithm which can be manually.
- . CT TAVI Planning provides detection with manual correction capability of: the annulus plane, Left Ventricular Outflow Tract (LVOT) plane, sinotubular junction plane, sinus of valsalva plane, ascending aorta plane, right coronary cusp landmark, left coronary cusp landmark, non-coronary cusp landmark, right coronary ostium and left coronary ostium.
- CT TAVI Planning provides contour detection with manual correction capability, . with diameter, area and distance measurements at each plane as well as distance to each coronary ostia from the annulus plane.
- CT TAVI Planning provides 2D and 3D views. .
- Planning provides C-arm angles which can be utilized to determine the best angle . for deployment during the procedure.
- CT TAVI Planning provides a 3D display of the ascending aorta and aortic root at . the C-arm angle chosen within the application.
- CT TAVI Planning provides tools to generate reports to distribute findings. .
### 21 CFR 807.92(a) (5): Intended Use
The CT TAVI Planning application is intended to be used for patients with aortic valvular disease, severe symptomatic aortic stenosis or tricuspid aortic valve. The intended part of the body for this application is the human heart, specifically the ascending aorta, aortic root, coronary ostia and left ventricle in order to assess the aortic valve in pre-operational planning of transcatheter aortic valve replacement procedures.
CT TAVI Planning is a non-invasive post-processing application providing 3D model-based segmentation of the aortic valve and aortic arch. The CT TAVI Planning application provides assessment and measurements of relevant heart structures for TAVI-device sizing, and allows the user to select a starting angle for C-arm position from the possible optimal positions from the CT TAVI and select a C-arm angle the user feels is appropriate to use in the catheterization laboratory by the Interventional team performing the procedure (to be used during the procedure itself).
The physician retains the ultimate responsibility for making the determination of patient eligibility or which device is implanted based on their standard practices and additional imaging modalities such as echocardiography.
{3}------------------------------------------------
### 21 CFR 807.92(a) (6): Technological Characteristics:
The CT TAVI Planning Application is intended for visualization, assessment and quantification of medical images, specifically quantitative distance and area measurements for CT images from adult patients with a diagnosis of aortic stenosis in which a non-surgical approach is recommended.
The application supports contrast-enhanced. prospectively ECG-gated axial or retrospectivelygated helical CT images as well as un-gated helical scans.
The application provides visualization and measurement tools for qualitative and quantitative visualization and assessment of the input data.
It provides tools to determine the size and shape of the aortic root anatomy, ascending aorta and left ventricular outflow tract.
The CT TAVI Planning Application provides model-based segmentation of the aortic valve, ascending aorta and left ventricle, detection of the coronary ostia, planes detection and dimensions measurements of the aortic annulus, left ventricular outflow tract, simotubular junction, sinus of valsalva, ascending aorta and distance to coronary ostia for TAVI-device sizing. While both predicate devices and CT TAVI Planning provide detection and manual correction of the annulus plane, Philips CT TAVI Planning, utilizing the same planes and landmarks detection algorithm as Philips HeartNavigator, has extended that functionality to also include additional planes relevant to the planes at which physicians currently measure for pre-planning: Left Ventricular Outflow Tract (LVOT). Sinotubular Junction, Sinus of Valsalva, and Ascending Aorta planes.
## 21 CFR 807.92(b) (1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
Verification was conducted to provide objective evidence that the application met its design specifications. This was done by testing the function and implementation of the application as well as risk management file.
The narrative below provides a brief summary of the verification activities. The details of the testing including test plans and reports can be found in [TAVI-008] -Verification of the 510(k).
The verification plan for the TAVI Application (TAVI-007) outlines the verification activities conducted to verify the function of the application and product's stability. Testing included full functionality of TAVI application and risk management file (RMF) related to defined requirements. The requirements for TAVI planning are traced to the clinical DRS and detailed test scenarios were written and accordingly the verification was conducted [TAVI-0051. TAVI verification included sanity tests which gave an overview of the application maturity for each build, critical tests which included tests of the main features and workflow and RMF tests which in general check calculations, measurements, patient details and information and other scenarios which might lead to misdiagnosis. Full functionality test included all TAVI functionality cases and covered all of the detailed requirements which provided us with assurance that the tested features work as required. Verification also covered the defect fixes of TAVI.
All the planned activities for verification have been completed and there are no blocking
{4}------------------------------------------------
defects for beta phase.
K132524
page 5 of 5
### 21 CFR 807.92(b) (2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
The TAVI validation activities, as documented in the TAVI Validation Plan (TAVI-009), were performed in the following two ways to provide objective evidence that the TAVI Software Application meets the intended use and defined customer needs as documented in the Customer Requirement Specification (CRS) (ISP-600-P1,2-0002-01 ):
1) Internal validation in a non-hospital environment by clinical experts Clinical experts can be non-Philips employees that can simulate the work of a typical clinical user as long as they did not participate in the development process and verification process Validation will be done in Philips facilities using nominal systems or outside Philips on a laptop which meets the HW requirements (evidence in ISP-600-P5-0109-01, Using Portal Demo laptops for IntelliSpace Portal validation; and clinical expert report as part of the Verification Report ISP-600-P5-0093-01
#### 2) External validation at beta sites
System is installed in a hospital environment and used by the local medical staff (radiologists and/or cardiologists).
The results of the two validation activities found in the Validation Report for Transcatheter Aortic Valve Implantation (TAVI-0010) confirms that the TAVI software application meets user needs and intended uses.
21 CFR 807.92(b) (3): The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a) (3) of this section:
The nonclinical and clinical tests have demonstrated that the device is safe and works according its intended use.
The CT TAVI Planning Application does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers CT TAVI Planning Application to be substantially equivalent to the Philips HeartNavigator (K111245) and the Vital Images Vitrea CT TAVR Planning (K133578).
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three lines representing the snakes and a central staff-like element.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 13, 2014
Philips Medical Systems % Ms. Susan Quick Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
Re: K132524
Trade/Device Name: CT Transcatheter Aortic Valve Implantation (TAVI) Planning Application Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 13, 2014 Received: May 15, 2014
Dear Ms. Quick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Ms. Quick
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K132524
Device Name CT TAVI Planning Application
## Indications for Use (Describe)
The CT TAVI Planning application is intended to be used for patients with aortic valvular disease, severe symptomatic aortic stenosis or tricuspid aortic valve. The body for this application is the human heart, specifically the ascending aorta, aortic root, coronary ostia and left ventricle in order to assess the aortic valve in pre-operational planning of transcatheter aortic valve replacement procedures.
CT TAVI Planning is a non-invasive post-processing application providing 3D model-based segmentation of the aortic valve and aortic arch. The CT TAVI Planning application provides assessment and measurements of relevant heart structures for TAVI-device sizing, and allows the user to select a starting angle for C-arm position from the possible optimal positions from the CT TAVI and select a C-arm angle the user feels is appropriate to use in the catheterization laboratory by the Interventional team performing the procedure (to be used during the procedure itself).
The physician retains the ultimate responsibility for making the determination of patient eligibility or which device is implanted based on their standard practices and additional imaging modalities such as echocardiography.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Robert A Ochs
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.