← Product Code [JEY](/productcode/JEY) · K132519

# STRYKER CUSTOMIZED MANDIBLE RECON PLATE KIT (K132519)

_Stryker · JEY · Dec 6, 2013 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132519

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [JEY](/productcode/JEY.md)
- **Decision Date:** Dec 6, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Customized Mandible Recon Plate Kit is intended to be used for rigid internal fixation of primary and secondary mandibular reconstructions.

## Device Story

Customized Mandible Recon Plate Kit provides patient-specific titanium implants for mandibular reconstruction. Surgeon orders via 'eRequest Lifecycle' online platform using patient CT-scan data. Plate Design Software (PDS) creates virtual plate design (dimensions, screw holes, bar strengthening) based on anatomical data and surgeon input. Surgeon reviews and approves Design Proposal (PDF) online. Device manufactured to specific design requirements; shipped non-sterile. Used in clinical settings for rigid internal fixation. Benefits include patient-specific fit for complex mandibular defects. Compatible with Stryker Universal Mandibular System; may include anatomical Mandible Model and customized cutting/drill guides.

## Clinical Evidence

No clinical testing was performed to support this submission. Bench testing included biocompatibility, cleaning, sterilization, reprocessing, MRI conditional, transportation, mechanical strength of plate, mechanical strength of locking mechanism, and end-user handling tests.

## Technological Characteristics

Material: Commercially pure titanium (CP Ti Grade 2). Design: Patient-specific, derived from CT data. Profile heights: 2.0mm and 2.8mm. Energy source: N/A (mechanical). Connectivity: Internet-based ordering platform (eRequest Lifecycle). Sterilization: Delivered non-sterile. Software: Plate Design Software (PDS) for virtual plate creation.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Stryker NewGen/Universal Mandibular System ([K014263](/device/K014263.md))
- Synthes Patient Specific Plates ([K122647](/device/K122647.md))
- Stryker PEEK Customized Cranial Implant Kit ([K121153](/device/K121153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 132519 DEC - 6 2013

র বাবা পারে আর মাধ্যমে স

| 510(k) Summary             |                                                                                                                                                                                                                                |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date prepared:             | August 5, 2013                                                                                                                                                                                                                 |
| Submitter:                 | Stryker Craniomaxillofacial<br>750 Trade Center Way<br>Portage, MI 49002<br>USA                                                                                                                                                |
| Contact:                   | Rob Yamashita<br>Phone: (269) 389-4258 / Fax: (269) 329-2235<br>rob.yamashita@stryker.com                                                                                                                                      |
| Proprietary Name:          | Stryker Customized Mandible Recon Plate Kit                                                                                                                                                                                    |
| Common Name:               | Bone Plate                                                                                                                                                                                                                     |
| Proposed Regulatory Class: | Class II                                                                                                                                                                                                                       |
| Product Codes:             | JEY                                                                                                                                                                                                                            |
| Predicate Device:          | Predicate devices for intended use, indications for use, and<br>technical characteristics are:                                                                                                                                 |
|                            | 1. Stryker NewGen/Universal Mandibular System - K014263<br>2. Synthes Patient Specific Plates - K122647                                                                                                                        |
|                            | The predicate device for the patient-specific ordering,<br>designing, and logistics process is:                                                                                                                                |
|                            | 3. Stryker PEEK Customized Cranial Implant Kit - K121153                                                                                                                                                                       |
| Intended Use:              | The Customized Mandible Recon Plate Kit is intended to be<br>used for rigid internal fixation of primary and secondary<br>mandibular reconstructions.                                                                          |
| Indication for Use:        | The Customized Mandible Recon Plate Kit is indicated for<br>use in primary mandibular reconstruction with bone graft,<br>temporary bridging until delayed secondary reconstruction<br>and secondary mandibular reconstruction. |

. . . . .

{1}------------------------------------------------

. * :

## Device Description:

The Customized Mandible Recon Plate Kit includes customized, patient specific implants, the Customized Mandible Recon Plates. Additionally, the kit includes the Instruction for Use (IFU) and a printed version of the Design Proposal approved by the surgeon prior to plate manufacture. It may include an anatomical model, named Mandible Model.

The Customized Mandible Recon Plates (CMRP) and the corresponding Mandible Model are designed and manufactured for one specific patient. The products are ordered by a surgeon on a patient-by-patient basis over an internet platform referred to as the "eRequest Lifecycle online ordering system" which was already cleared with K111065 for ordering patient specific polymer implants to correct trauma and/or defects in mandibular, maxillofacial, or craniofacial bone. Based on patient specific anatomical data (CT-scan) and input from the surgeon, a virtual mandible recon plate with its specific dimensions (profile heights, length and 3D run of the plate, number and position of screw holes and bar strengthening) is created using the Customized Mandible Recon Plate Design Process (CMRP-DP) including the Plate Design Software (PDS).

Screenshots of the virtual plate are then presented to the surgeon in the Design Proposal (pdf-file). After the surgeon approves the Design Proposal via the above mentioned eRequest Lifecycle online ordering system, the Customized Mandible Recon Plate, as well as the Mandible Model (if ordered by the surgeon), are manufactured according to the specifically set design requirements. Once the manufacturing process is finalized and the devices are cleaned, they are packed, labeled and shipped to the location specified during the ordering process.

The CMRP implant is compatible with the Stryker Universal Mandibular System. Additionally. Stryker offers customized cutting and drill guides for use with the Customized Mandible Recon Plate.

# Technological and Operational Characteristics:

The Stryker Customized Mandible Recon Plate Kit is similar to its Predicate Devices in the following technological and operational characteristics:

- Material: ●
●

The Subject Device is made of biocompatible commercially pure titanium (CP Ti Grade 2); Predicates 1 and 2 are made of pure titanium.

Desian:

The transfer of CT patient data to a manufacturing process resulting in an individualized implant fulfilling the specific requirements of the patient is identical to the subject device and Predicate 3 and similar to Predicate 2.

The final patient specific plate for mandibular reconstruction itself is similar to the stock plates of Predicate Device 1. Predicate Device 2 provides patient specific plates for mandibular

{2}------------------------------------------------

reconstruction, therewith being similar to the Subject Device.

The Subject Device is offered in 2.0mm and 2.8mm profile height, whereas the Predicate Devices 1 and 2 range from 1.5mm to 2.8mm profile height.

- The basic operational principle of the Subject Principal of . Device, as well as Predicate Devices 1 and 2, is Operation: the stabilization of mandibular reconstruction. The method of site preparation and fixation are identical for both the Subject Device and the referenced Predicate Devices.
- The Subject Device and all Predicate Devices are Packaging: . delivered non-sterile.

#### Clinical Testing:

:

:

No clinical testing was performed to support this submission.

#### Non-Clinical Testing:

The Stryker Customized Mandible Recon Plate Kit proved to meet acceptance criteria in evaluations conducted for biocompatibility, cleaning, sterilization, multiple all reprocessing, residual moisture after sterilization, MRI conditional, transportation, clinical transportation, mechanical strength of plate, mechanical strength of locking mechanism between plate and screw and handling of system (end user & end product test).

#### Substantial Equivalence to Predicate Devices:

The Stryker Customized Mandible Recon Plate Kit combines the princibles of individualized design configuration with the PEEK Customized implant as one of the predicate devices and the principles of operation with the other two Predicate Devices 1 and 2. Its intended use and its indications for use are nearly identical to these. Material and technological characteristics are similar to the legally marketed Predicate Devices 1 and 2. The ordering, design and logistical process of the Subject Device is nearly identical to the one of Predicate Device 3.

### Performance Testing:

Verification and Validation (V&V) evaluation has been performed on the Subject Device in the following categories: biocompatibility, cleaning validation validation, multiple reprocessing, residual moisture after sterilization, MRI conditional, 

{3}------------------------------------------------

transportation validation, clinical transportation, mechanical strength of plate, mechanical strength of locking mechanism between plate and screw, and handling of system (end user & end product test). The Subject Device fulfilled all set acceptance criteria for each category in accordance with either ISO or ASTM specifications, or internally predetermined acceptance criteria if no standards were applicable. All applied ISO standards are listed below.

... _ .

| Reference                            | Title                                                                                                                                                                                    |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility                     |                                                                                                                                                                                          |
| DIN EN ISO 10993 ff as<br>valid 2013 | Biological evaluation of medical devices. Part 1:<br>Evaluation and testing within a risk management<br>system                                                                           |
| Sterility/Reprocessing               |                                                                                                                                                                                          |
| ISO 11138-1:2006                     | Sterilization of health care products - Biological<br>indicators - Part 1: General requirements                                                                                          |
| DIN EN ISO 11737-1:2009              | Sterilization of medical devices - Microbiological<br>methods - Part 1: Determination of a population of<br>microorganisms on products                                                   |
| DIN EN ISO 11737-2:2009              | Sterilization of medical devices - Microbiological<br>methods - Part 2: Tests of sterility performed in the<br>validation of a sterilization process                                     |
| DIN EN ISO 14161:2011                | Sterilization of health care products - Biological<br>indicators - Guidance for the selection, use and<br>interpretation of results                                                      |
| DIN EN ISO 14937:2009                | Sterilization of health care products - general criteria<br>for characterization of a sterilizing agent and<br>development, validation and routine control of a<br>sterilization process |
| EN ISO 15883-1:2009                  | Washer-disinfectors: General requirements, terms<br>and definitions and tests                                                                                                            |
| DIN EN ISO 17664:2004                | Sterilization of medical devices - Information to be<br>provided by the manufacturer for the processing of<br>resterilizable medical devices                                             |
| DIN EN ISO 17665-1:2006              | Sterilization of health care products - Moist heat -<br>Part 1: Requirements for the development, validation<br>and routine control of a sterilization process for<br>medical devices    |
| DIN ISO/TS 17665-2:2009              | Sterilization of health care products - Moist heat -<br>Part 2: Guidance on the application of ISO 17665-1                                                                               |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, with three flowing lines representing the department's mission to promote health and well-being.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 6, 2013

Stryker Craniomaxillofacial Mr. Rob Yamashita Associate Manager, Regulatory Affairs 750 Trade Center Way Ste 200 Portage, MI 49002

Re: K132519

Trade/Device Name: Stryker Customized Mandible Recon Plate Kit Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Il Product Code: JEY Dated: October 22, 2013 Received: November 4, 2013

Dear Mr. Yamashita:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Yamashita

Station of Contract

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Kwame O. Ulmer -S for

Erin I. Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Device and Radiological Health

Enclosure

{6}------------------------------------------------

Stryker Customized Mandible Recon Plate Kh

Craniomaxillofacial

510(k) Number (if known):

Device Name:

:

Indications For Use:

Stryker Customized Mandible Recon Plate Kit

K132519

The Customized Mandible Recon Plate Kit is intended to be used for rigid internal fixation of primary and secondary mandibular reconstructions.

The Customized Mandible Recon Plate Kit is indicated for use in primary mandibular reconstruction with bone graft, temporary bridging until delaved secondary reconstruction and secondary mandibular reconstruction.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Traditional 610(k)

Mary S. Runner -: 2013.12.04 14:22:00 -05'00'

Page 23 of 655

---

**Source:** [https://fda-staging.innolitics.com/device/K132519](https://fda-staging.innolitics.com/device/K132519)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com