The Customized Mandible Recon Plate Kit is intended to be used for rigid internal fixation of primary and secondary mandibular reconstructions.
Device Story
Customized Mandible Recon Plate Kit provides patient-specific titanium implants for mandibular reconstruction. Surgeon orders via 'eRequest Lifecycle' online platform using patient CT-scan data. Plate Design Software (PDS) creates virtual plate design (dimensions, screw holes, bar strengthening) based on anatomical data and surgeon input. Surgeon reviews and approves Design Proposal (PDF) online. Device manufactured to specific design requirements; shipped non-sterile. Used in clinical settings for rigid internal fixation. Benefits include patient-specific fit for complex mandibular defects. Compatible with Stryker Universal Mandibular System; may include anatomical Mandible Model and customized cutting/drill guides.
Clinical Evidence
No clinical testing was performed to support this submission. Bench testing included biocompatibility, cleaning, sterilization, reprocessing, MRI conditional, transportation, mechanical strength of plate, mechanical strength of locking mechanism, and end-user handling tests.
Technological Characteristics
Material: Commercially pure titanium (CP Ti Grade 2). Design: Patient-specific, derived from CT data. Profile heights: 2.0mm and 2.8mm. Energy source: N/A (mechanical). Connectivity: Internet-based ordering platform (eRequest Lifecycle). Sterilization: Delivered non-sterile. Software: Plate Design Software (PDS) for virtual plate creation.
Indications for Use
Indicated for primary mandibular reconstruction with bone graft, temporary bridging until delayed secondary reconstruction, and secondary mandibular reconstruction.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Stryker NewGen/Universal Mandibular System (K014263)
{0}------------------------------------------------
K 132519 DEC - 6 2013
র বাবা পারে আর মাধ্যমে স
| 510(k) Summary | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date prepared: | August 5, 2013 |
| Submitter: | Stryker Craniomaxillofacial<br>750 Trade Center Way<br>Portage, MI 49002<br>USA |
| Contact: | Rob Yamashita<br>Phone: (269) 389-4258 / Fax: (269) 329-2235<br>rob.yamashita@stryker.com |
| Proprietary Name: | Stryker Customized Mandible Recon Plate Kit |
| Common Name: | Bone Plate |
| Proposed Regulatory Class: | Class II |
| Product Codes: | JEY |
| Predicate Device: | Predicate devices for intended use, indications for use, and<br>technical characteristics are: |
| | 1. Stryker NewGen/Universal Mandibular System - K014263<br>2. Synthes Patient Specific Plates - K122647 |
| | The predicate device for the patient-specific ordering,<br>designing, and logistics process is: |
| | 3. Stryker PEEK Customized Cranial Implant Kit - K121153 |
| Intended Use: | The Customized Mandible Recon Plate Kit is intended to be<br>used for rigid internal fixation of primary and secondary<br>mandibular reconstructions. |
| Indication for Use: | The Customized Mandible Recon Plate Kit is indicated for<br>use in primary mandibular reconstruction with bone graft,<br>temporary bridging until delayed secondary reconstruction<br>and secondary mandibular reconstruction. |
. . . . .
{1}------------------------------------------------
. * :
## Device Description:
The Customized Mandible Recon Plate Kit includes customized, patient specific implants, the Customized Mandible Recon Plates. Additionally, the kit includes the Instruction for Use (IFU) and a printed version of the Design Proposal approved by the surgeon prior to plate manufacture. It may include an anatomical model, named Mandible Model.
The Customized Mandible Recon Plates (CMRP) and the corresponding Mandible Model are designed and manufactured for one specific patient. The products are ordered by a surgeon on a patient-by-patient basis over an internet platform referred to as the "eRequest Lifecycle online ordering system" which was already cleared with K111065 for ordering patient specific polymer implants to correct trauma and/or defects in mandibular, maxillofacial, or craniofacial bone. Based on patient specific anatomical data (CT-scan) and input from the surgeon, a virtual mandible recon plate with its specific dimensions (profile heights, length and 3D run of the plate, number and position of screw holes and bar strengthening) is created using the Customized Mandible Recon Plate Design Process (CMRP-DP) including the Plate Design Software (PDS).
Screenshots of the virtual plate are then presented to the surgeon in the Design Proposal (pdf-file). After the surgeon approves the Design Proposal via the above mentioned eRequest Lifecycle online ordering system, the Customized Mandible Recon Plate, as well as the Mandible Model (if ordered by the surgeon), are manufactured according to the specifically set design requirements. Once the manufacturing process is finalized and the devices are cleaned, they are packed, labeled and shipped to the location specified during the ordering process.
The CMRP implant is compatible with the Stryker Universal Mandibular System. Additionally. Stryker offers customized cutting and drill guides for use with the Customized Mandible Recon Plate.
# Technological and Operational Characteristics:
The Stryker Customized Mandible Recon Plate Kit is similar to its Predicate Devices in the following technological and operational characteristics:
- Material: ●
●
The Subject Device is made of biocompatible commercially pure titanium (CP Ti Grade 2); Predicates 1 and 2 are made of pure titanium.
Desian:
The transfer of CT patient data to a manufacturing process resulting in an individualized implant fulfilling the specific requirements of the patient is identical to the subject device and Predicate 3 and similar to Predicate 2.
The final patient specific plate for mandibular reconstruction itself is similar to the stock plates of Predicate Device 1. Predicate Device 2 provides patient specific plates for mandibular
{2}------------------------------------------------
reconstruction, therewith being similar to the Subject Device.
The Subject Device is offered in 2.0mm and 2.8mm profile height, whereas the Predicate Devices 1 and 2 range from 1.5mm to 2.8mm profile height.
- The basic operational principle of the Subject Principal of . Device, as well as Predicate Devices 1 and 2, is Operation: the stabilization of mandibular reconstruction. The method of site preparation and fixation are identical for both the Subject Device and the referenced Predicate Devices.
- The Subject Device and all Predicate Devices are Packaging: . delivered non-sterile.
#### Clinical Testing:
:
:
No clinical testing was performed to support this submission.
#### Non-Clinical Testing:
The Stryker Customized Mandible Recon Plate Kit proved to meet acceptance criteria in evaluations conducted for biocompatibility, cleaning, sterilization, multiple all reprocessing, residual moisture after sterilization, MRI conditional, transportation, clinical transportation, mechanical strength of plate, mechanical strength of locking mechanism between plate and screw and handling of system (end user & end product test).
#### Substantial Equivalence to Predicate Devices:
The Stryker Customized Mandible Recon Plate Kit combines the princibles of individualized design configuration with the PEEK Customized implant as one of the predicate devices and the principles of operation with the other two Predicate Devices 1 and 2. Its intended use and its indications for use are nearly identical to these. Material and technological characteristics are similar to the legally marketed Predicate Devices 1 and 2. The ordering, design and logistical process of the Subject Device is nearly identical to the one of Predicate Device 3.
### Performance Testing:
Verification and Validation (V&V) evaluation has been performed on the Subject Device in the following categories: biocompatibility, cleaning validation validation, multiple reprocessing, residual moisture after sterilization, MRI conditional,
{3}------------------------------------------------
transportation validation, clinical transportation, mechanical strength of plate, mechanical strength of locking mechanism between plate and screw, and handling of system (end user & end product test). The Subject Device fulfilled all set acceptance criteria for each category in accordance with either ISO or ASTM specifications, or internally predetermined acceptance criteria if no standards were applicable. All applied ISO standards are listed below.
... _ .
| Reference | Title |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | |
| DIN EN ISO 10993 ff as<br>valid 2013 | Biological evaluation of medical devices. Part 1:<br>Evaluation and testing within a risk management<br>system |
| Sterility/Reprocessing | |
| ISO 11138-1:2006 | Sterilization of health care products - Biological<br>indicators - Part 1: General requirements |
| DIN EN ISO 11737-1:2009 | Sterilization of medical devices - Microbiological<br>methods - Part 1: Determination of a population of<br>microorganisms on products |
| DIN EN ISO 11737-2:2009 | Sterilization of medical devices - Microbiological<br>methods - Part 2: Tests of sterility performed in the<br>validation of a sterilization process |
| DIN EN ISO 14161:2011 | Sterilization of health care products - Biological<br>indicators - Guidance for the selection, use and<br>interpretation of results |
| DIN EN ISO 14937:2009 | Sterilization of health care products - general criteria<br>for characterization of a sterilizing agent and<br>development, validation and routine control of a<br>sterilization process |
| EN ISO 15883-1:2009 | Washer-disinfectors: General requirements, terms<br>and definitions and tests |
| DIN EN ISO 17664:2004 | Sterilization of medical devices - Information to be<br>provided by the manufacturer for the processing of<br>resterilizable medical devices |
| DIN EN ISO 17665-1:2006 | Sterilization of health care products - Moist heat -<br>Part 1: Requirements for the development, validation<br>and routine control of a sterilization process for<br>medical devices |
| DIN ISO/TS 17665-2:2009 | Sterilization of health care products - Moist heat -<br>Part 2: Guidance on the application of ISO 17665-1 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, with three flowing lines representing the department's mission to promote health and well-being.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2013
Stryker Craniomaxillofacial Mr. Rob Yamashita Associate Manager, Regulatory Affairs 750 Trade Center Way Ste 200 Portage, MI 49002
Re: K132519
Trade/Device Name: Stryker Customized Mandible Recon Plate Kit Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Il Product Code: JEY Dated: October 22, 2013 Received: November 4, 2013
Dear Mr. Yamashita:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Yamashita
Station of Contract
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Kwame O. Ulmer -S for
Erin I. Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Device and Radiological Health
Enclosure
{6}------------------------------------------------
Stryker Customized Mandible Recon Plate Kh
Craniomaxillofacial
510(k) Number (if known):
Device Name:
:
Indications For Use:
Stryker Customized Mandible Recon Plate Kit
K132519
The Customized Mandible Recon Plate Kit is intended to be used for rigid internal fixation of primary and secondary mandibular reconstructions.
The Customized Mandible Recon Plate Kit is indicated for use in primary mandibular reconstruction with bone graft, temporary bridging until delaved secondary reconstruction and secondary mandibular reconstruction.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Traditional 610(k)
Mary S. Runner -: 2013.12.04 14:22:00 -05'00'
Page 23 of 655
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.