← Product Code [DQA](/productcode/DQA) · K132516

# OXIPROBE E (OXIPROBE) (K132516)

_Bio Medical Technologies Co., Ltd. · DQA · Jul 30, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132516

## Device Facts

- **Applicant:** Bio Medical Technologies Co., Ltd.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Jul 30, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The OxiprobE is indicated for non-invasive, continuous, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate, and pulse amplitude. Individual models are labeled for the intended use. It is for prescription use only.

## Device Story

OxiprobE is a disposable, patient-dedicated pulse oximetry sensor; measures blood oxygen saturation, pulse rate, and pulse amplitude via non-invasive optical sensing. Device consists of light source and photo detector secured to patient skin via adhesive; connects to compatible pulse oximeter (e.g., Nellcor N-395) via extension cable (BM-2070). Used in clinical settings to provide continuous monitoring; output displayed on host oximeter for clinician review. Enables real-time physiological assessment; aids clinical decision-making regarding patient oxygenation status; prevents cross-contamination through single-use design.

## Clinical Evidence

No clinical data was required. Substantial equivalence supported by non-clinical bench testing, including electrical safety per EN ISO 60601-1:2006, performance requirements per EN ISO 80601-2-61:2011, and EMC/EMI testing. Accuracy for SpO2 (70-100% ± 3%) and pulse rate (50-150 bpm ± 2 bpm) was verified via bench testing.

## Technological Characteristics

Disposable pulse oximetry sensor; utilizes optical sensing (light source/photo detector). Adhesive-based form factor; tan woven material. Connectivity via BM-2070 extension cable to host oximeter. Non-sterile. Complies with EN ISO 60601-1 and EN ISO 80601-2-61 standards.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- BMT Sensors ([K092549](/device/K092549.md))

## Reference Devices

- Nellcor N-395 Pulse Oximeter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(K) SUMMARY

## Date Summary Prepared: July 30. 2014

JUL 3 0 2014

| Sponsor Information: |                                                                                 |
|----------------------|---------------------------------------------------------------------------------|
| Company Name:        | BIO Medical Technologies Co., Ltd.(BMT)                                         |
| Address:             | Megu Bldg #431-17<br>Shindang-dong, Jung-gu<br>Seoul, Republic of Korea 100-452 |
| Contact:             | Bruce Lee                                                                       |
| Telephone:           | +(82) 222340782                                                                 |
| FAX:                 | +(82) 222340748                                                                 |

#### Correspondent Contact Information:

| Company Name: | Paladin Medical, Inc.        |  |
|---------------|------------------------------|--|
| Address:      | P.O. Box 560                 |  |
|               | Stillwater, MN 55082         |  |
| Contact:      | Elaine Duncan, M.S.M.E., RAC |  |
| Title:        | President                    |  |
| Telephone:    | +(82) 222340782              |  |
| FAX:          | +(82 222340748               |  |

#### DEVICE INFORMATION:

- . Proprietary Name: OxiprobE
- Common Name: Oximetry Sensor .
- Classification Name: Oximeter .
- Regulation Number: 870.2700 .
- Classification Product Code: DQA .
- . Classification/Review Advisory Committee: Anesthesiology

#### DESCRIPTION of the DEVICE:

The OxiprobE Pulse Oximetry Sensors are used for non-invasive measurement of blood oxygen saturation and pulse rate measurement. The disposable sensors are designed to enable the sensor's light source and photo detector to be securely and properly positioned on the patient. The adhesive stabilizes these important optical components and provides a comfortable fit. The adhesive sensors are patient-dedicated and can travel with the patients. The single-use, disposable sensors do not present the risk of cross-contamination caused by products that are reused from patient to patient. This 510(k) seeks premarket clearance for a line extension model BM-500, compatible with the BM-2070 extension cable and the Nellcor N-395 Pulse Oximeter (a legacy oximeter.). The BM-2070 extension cable was cleared for use with the predicate BMT sensors under K092549.

#### INDICATION FOR USE:

The OxiprobE is indicated for non-invasive, continuous, beat-by-beat monitoring of oxygen saturation of functional arterial hemoclobin, pulse rate, and pulse amplitude. Individual models are labeled for the intended use. It is for prescription use only.

#### PREDICATE DEVICE USED FOR SUBSTANTIAL EQUIVALENCE:

| Device Feature     | Predicate BMT Sensors<br>(K092549)                                                                                                                                                                                                                                                     | Subject BMT Sensors (K132516)                                                                                                                                                                                                                                                    | Comparison* |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Indication for Use | The OxiprobE is indicated for<br>non-invasive, continuous, beat-<br>by-beat monitoring of oxygen<br>saturation of functional arterial<br>hemoglobin, pulse rate, and<br>pulse amplitude. Individual<br>models are labeled for the<br>intended use. It is for<br>prescription use only. | The OxiprobE is indicated for non-<br>invasive, continuous, beat-by-beat<br>monitoring of oxygen saturation of<br>functional arterial hemoglobin, pulse<br>rate, and pulse amplitude. Individual<br>models are labeled for the intended use.<br>It is for prescription use only. | Same        |
| Intended Use       | BM-200: Adult / Infant                                                                                                                                                                                                                                                                 | BM-500: Infant/Adult                                                                                                                                                                                                                                                             | Same        |

{1}------------------------------------------------

### 510(k) Summary-Continued

| Device Feature             | Predicate BMT Sensors<br>(K092549)           | Subject BMT Sensors (K132516) | Comparison* |
|----------------------------|----------------------------------------------|-------------------------------|-------------|
|                            | BM-400 Pediatric                             |                               |             |
| Usage Type                 | Single Use Disposable                        | Single Use Disposable         | Same        |
| Sterility                  | Non-Sterile                                  | Non-Sterile                   | Same        |
| Adhesive Tape Type         | BM-200: Tan Woven<br>BM-400 Clear Woven      | BM-500: Tan Woven             | SE          |
| Emitter - Detector Spacing | BM-200 = 26mm<br>BM-400 = 20mm               | BM-500 = 20mm                 | Same        |
| Tape Length                | BM-200 = 110mm<br>BM-400 = 81mm              | BM-500 = 80mm                 | Same        |
| SpO2 Accuracy<br>(%)       | BM-200: 70 - 100 ± 2<br>BM-400: 70 - 100 ± 3 | BM-500 : 70 - 100 ± 3         | Same        |
| Pulse Rate Accuracy (bpm)  | BM-200: 50 - 150 ± 2<br>BM-400: 50 - 150 ± 2 | BM-500 : 50 - 150 ± 2         | Same        |

#### PERFORMANCE TESTING:

Nonclinical Bench Testing (No new clinical data was required to support this 510(k)) was included to support substantial equivalence as listed below:

- EN ISO 60601-1:2006, 30 edition: including ●
	- Humidity preconditioning treatment (sub-clause 5.7) -
	- Accessible parts (sub-clause 5.9.2) -
	- Legibility of markings (sub-clause 7.1.2) -
	- -Durability of markings (sub-clause 7.1.3)
	- Leakage Current Tests (sub-clause 8.7) -
	- Temperature test (11.1) -
	- -Interruption of the power supply (sub-clause 11.8)
	- Power Limitation test (sub-clause 13.1.2) -
	- Push test (sub-clause 15.3.2) -
	- Drop Test (hand-held) (sub-clause 15.3.4.1 -
	- With the exception of: Usability EN 60601-1-6, or "accuracy" which was tested elsewhere by the sponsor, and those clauses not applicable to pulse oximetry sensors as determined by the certified testing laboratory
- EN ISO 80601-2-61:2011, Medical electrical equipment Part 2-61 Particular requirements . for basic safety and essential performance of pulse oximeter equipment for all units subject of this submission with the exception of those parts which the laboratory determined to be not applicable to sensors.
- . Electromagnetic Compatibility and Interference testing (EMC / EMI) was performed on all of the BMT Pulse Oximetry Sensors which are the subject of this 510(k) was conducted to determine compliance with the applicable requirements of the ISO 80601-2-61:2011 Standard as discussed on the FDA's Pulse Oximeter Guidance.
- Pulse and SpO2 accuracy. .

#### CONCLUSION

It can be concluded from the results that the model BM 500 for use with the Nellcor Model N395 Pulse Oximeter, compatible with the predicate BM-2070 extension cable is substantially equivalent to the predicate.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the United States. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 30, 2014

BIO Medical Technologies, Co. Ltd. c/o Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082

Re: K132516

Trade/Device Name: OxiprobE Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: July 18, 2014 Received: July 21, 2014

Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The read may merest of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 Ms. Elaine Duncan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum.

Sincerely yours.

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director

DAGRID/ODE/CDRI! FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K132516

Device Name: OxiprobE

Indications for Use:

The OxiprobE is indicated for non-invasive, continuous, beat-by-beat monitoring of oxygen saturation of functional arterial hemoglobin, pulse rate, and pulse amplitude. Individual models are labeled for the intended use. It is for prescription use only.

| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR                                          |   |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C)  |   |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF  
NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/4/Picture/7 description: The image shows the text 'Todd D. Courtney -S' at the top. Below that is the date and time '2014:07:30 13:19:01'. The timezone offset '-04'00'' is at the bottom left. The letters 'FDA' are in the center of the image.

---

**Source:** [https://fda-staging.innolitics.com/device/K132516](https://fda-staging.innolitics.com/device/K132516)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com