← Product Code [FLL](/productcode/FLL) · K132514

# KINSA SMART THERMOMETER (K132514)

_Kinsa, Inc. · FLL · Dec 2, 2013 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132514

## Device Facts

- **Applicant:** Kinsa, Inc.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Dec 2, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

The Kinsa Smart Thermometer is intended to measure the human body temperature orally, rectally, or under the arm, and the devices are reusable for clinical or home use on people of all ages.

## Device Story

Kinsa Smart Thermometer is a reusable, thermistor-based clinical electronic thermometer. It lacks internal batteries, processors, or displays, instead connecting to a smartphone or mobile device via the headphone/microphone jack. The device operates by sending an audio signal through the thermistor in the metal tip; the signal is altered by temperature and read back through the mobile device's microphone input. A dedicated mobile application processes this signal and displays the temperature on the smartphone screen. The device is used by patients or clinicians in home or clinical settings. By leveraging the smartphone's power and display, it provides a digital temperature reading, facilitating monitoring and clinical decision-making for users of all ages.

## Clinical Evidence

Bench testing only. The device was evaluated for compliance with ASTM E1112-00 (Reapproved 2011) for clinical electronic thermometer accuracy, as well as AAMI/IEC 60601-1:2005+A1:2012(E) for electrical safety and AAMI/ANSI/IEC 60601-1-2:2007 for electromagnetic compatibility.

## Technological Characteristics

Thermistor-based sensing; stainless steel tip; thermoplastic rubber tip housing; ABS body housing. Powered by connected mobile device battery via headphone/microphone jack. Dimensions: ~4.25" x 0.5" x 0.3"; Weight: 4.4g. Compliant with ASTM E1112-00, IEC 60601-1, and IEC 60601-1-2. Connectivity: Wired connection to smartphone/mobile device.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Fudakang Digital Thermometer ([K101387](/device/K101387.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SECTION 5: 510(K) PREMARKET NOTIFICATION

## Summary of Safety and Effectiveness Information

K132514 Kinsa Smart Thermometer August 6, 2013

Regulatory authority: Safe Medical Devices Act of 1990, 21 CFR 807.92

### 1) Device name

Trade name:

Kinsa Smart Thermometer

Common name:

DEC 02 2013

Thermometer

### Classification Number/ Classification name/Product code:

Clinical electronic thermometers are class II devices under 21 CFR § 880.2910 (product code FLL) and are classified by the General Hospital Panel.

## 2) Submitter

Kinsa, Inc. 603 Greenwich Street #101B New York, NY 10014

### 3) Company contact

Lael J. Pickett Kinsa, Inc. 603 Greenwich Street #101B ' New York, NY 10014 Telephone: 651-900-5460/Fax: 917-210-4288 Email: ljp1286@yahoo.com

## 4) Classification

Device class:

Class II

Classification panel:

General Hospital

Product code:

FLL

#### Special Controls:

Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers

1

{1}------------------------------------------------

## 5) Legally Marketed Device to which Equivalence is Claimed

Fudakang Digital Thermometer (K101387)

## 6) Comparison to Predicate Device

| Feature                                               | Fudakang Digital<br>Thermometer (K101387)                                                                                                                                                                                                                  | Kinsa Smart Thermometer<br>(KXXXXXX)                                                                                                                                                                          |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                          | Measurement and monitoring of<br>human body temperature                                                                                                                                                                                                    | Same                                                                                                                                                                                                          |
| Indications for Use                                   | Intended for the measurement and<br>monitoring of human body<br>temperature by doctor or<br>consumers in the hospital or<br>home.<br>BT-A11CN, BT-A21CN and BT-<br>A41CN can be used axillary<br>measurement, oral measurements<br>and rectal measurement. | The Kinsa Smart Thermometer is<br>intended to measure the human<br>body temperature orally, rectally,<br>or under the arm, and the devices<br>are reusable for clinical or home<br>use on people of all ages. |
| Tip                                                   | Stainless Steel                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                          |
| Tip Housing                                           | Thermoplastic rubber                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                          |
| Body Housing                                          | Acrylonitrile butadiene styrene                                                                                                                                                                                                                            | Same                                                                                                                                                                                                          |
| Size                                                  | 4.9" long x 0.74" wide x 0.35" thick                                                                                                                                                                                                                       | Approximately 4.25" long x 0.5"<br>wide x 0.3" thick                                                                                                                                                          |
| Weight                                                | 11 grams                                                                                                                                                                                                                                                   | 4.4 grams                                                                                                                                                                                                     |
| Flex Tip                                              | Yes                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                          |
| Power Source                                          | Battery                                                                                                                                                                                                                                                    | Mobile device battery                                                                                                                                                                                         |
| Principles of<br>Operation                            | Thermosensor/ASIC (Thermistor<br>used as themosensor)                                                                                                                                                                                                      | Same                                                                                                                                                                                                          |
| Compatible with<br>Mobile Device (e.g.<br>Smartphone) | No                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                           |
| Display °F or °C                                      | Yes                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                          |
| Reusable                                              | Yes                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                          |
| Measurement Range                                     | 32.0°C - 42.9°C                                                                                                                                                                                                                                            | 35.0 to 42.0ºC                                                                                                                                                                                                |
| Accuracy                                              | 95.0°F - 102.0°F/±0.2°F                                                                                                                                                                                                                                    | 95.0°F - 107.6°F/±0.2°F                                                                                                                                                                                       |
|                                                       | 35.0°C - 39.0°C/±0.1°C                                                                                                                                                                                                                                     | 35.0°C - 42.0°C/±0.1°C                                                                                                                                                                                        |
| Response Time                                         | 60 seconds                                                                                                                                                                                                                                                 | 15 seconds                                                                                                                                                                                                    |
| ASTM E1112-00                                         | Compliant with ASTM E1112-00<br>(2006)                                                                                                                                                                                                                     | Compliant with ASTM E1112-00<br>(Reapproved 2011)                                                                                                                                                             |
| AAMI/IEC 60601-<br>1:2005+A1:2012(E)                  | Compliant with IEC 60601-1                                                                                                                                                                                                                                 | Compliant with AAMI/IEC 60601-<br>1:2005+A1:2012(E)                                                                                                                                                           |
| AAMI/ANSI/IEC<br>60601-1-2:2007                       | Compliant with IEC 6061-1-2                                                                                                                                                                                                                                | Compliant with AAMI/ANSI/IEC<br>60601-1-2:2007                                                                                                                                                                |

## 7) Device description

The Kinsa Smart Thermometer product is a thermometer that connects to a Smartphone or another mobile device (e.g. an iPod Touch). The product will read body temperature the same way a clinical digital thermometer does by being placed under the tongue in the mouth, rectum or alternatively, or under the arm. Like other clinical digital thermometers, the Kinsa Smart Thermometer is a thermistor-based product; however, it has the advantage of

{2}------------------------------------------------

being read on a mobile device display. Unlike other clinical digital thermometers, the Kinsa Smart Thermometer product requires no batteries or LCD displays. The Kinsa Smart Thermometer is reusable for clinical or home use on people of all ages.

The Kinsa Smart Thermometer will connect to Smartphones or other mobile devices via a headphone jack that accepts a microphone input. In this document the terms Smartphone, smartohone and mobile device are used interchangeably and are defined to include the following products: Apple iPhones 5, 4S and the Apple iPod Touch 5.

The Kinsa Smart Thermometer consists of four components:

- A. Thermometer (probe).
- B. An adapter to setup each Smartphone for temperature reading (only needed once per Smartphone).
- C. An optional, flexible extension cord that can be used to lengthen the distance between the thermometer and Smartphone so users can see the Smartphone screen while taking a temperature.
- D. Software.

#### 8) Summary of technologies

The Kinsa Smart Thermometer works by sending out an audio signal. This signal crosses the thermistor in the metal tip of the thermometer and is altered based on the temperature. The device reads back the signal change through the mobile device's microphone input. The Kinsa Smart Thermometer app (application) has software to process the signal and displays the precise temperature on the screen. The Kinsa Smart Thermometer meets and exceeds ASTM standards for accuracy and meets ISO standards for accuracy.

The Kinsa Smart Thermometer has no batteries, processor or LCD display, instead leveraging the power, processor and display of the user's Smartphone.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three legs. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

### December 2, 2013

Kinsa, Incorporated C/O Lael J. Pickett 603 Greenwich Street, #101B New York, NY 10014

Re: K132514

Trade/Device Name: Kinsa Smart Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: September 2, 2013 Received: September 4, 2013

Dear Mr. Pickett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/6 description: The image contains the name Kwame Nkrumah. The text is in a bold, sans-serif font. The letters are arranged in a slightly curved line. There is a logo to the right of the name, which includes a circle and a triangle.

for

Erin I. Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K132514

#### Device Name

Kinsa Smart Thermometer

#### Indications for Use (Describe)

The Kinso Smart Thermometer is intended to measure orally, rectally, or under the am, and the devices are reusable for clinical or home use on people of all ages.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

[x] Over-The-Counter Use (21 CFR 801 Subpart C)

## Please do not write below this line -- continue on a separate page if needed.

1999 THE FOR FOR FOR CONNECT AFFOR FOR USE ONLY SOUTH STORES AND PORT THE STORES THE FOR THE STORE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/5/Picture/13 description: The image shows the FDA logo. The logo is made up of the letters F, D, and A. The letters are all capitalized and are in a bold, sans-serif font. The letters are arranged in a horizontal line, with the F on the left, the D in the middle, and the A on the right. The logo is black and white.

Digitally signed by Richard C. Chapman Date: 2013.11.29 12:42:36 -05'00'

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on fast page.

---

**Source:** [https://fda-staging.innolitics.com/device/K132514](https://fda-staging.innolitics.com/device/K132514)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).