← Product Code [LFL](/productcode/LFL) · K132512

# HARMONIC ACE CURVED SHEARS WITH SCISSOR HANDLE, HARMONIC ACE CURVED SHEARS WITH ERGONOMIC HANDLE (K132512)

_Ethicon Endo-Surgery, LLC · LFL · Oct 24, 2013 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132512

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, LLC
- **Product Code:** [LFL](/productcode/LFL.md)
- **Decision Date:** Oct 24, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Harmonic Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open and endoscopic procedures.

## Device Story

Harmonic ACE Curved Shears are sterile, single-patient-use ultrasonic surgical instruments. They consist of a handle housing with hand control buttons, a rotating shaft, a curved ultrasonic blade, and a clamp arm. The device uses ultrasonic energy to coagulate and transect soft tissue and vessels up to 5mm in diameter. Operated by surgeons in open or laparoscopic procedures, the device provides audible/tactile feedback for full closure. The ergonomic handle design improves accessibility of activation buttons. The device serves as an alternative to electrosurgery, lasers, or steel scalpels, aiming to minimize thermal injury while maintaining hemostasis.

## Clinical Evidence

No clinical studies were required. Substantial equivalence supported by preclinical porcine vessel sealing validation studies, including transection times, hemostasis, thermal spread measurements, and blood pressure challenges. Functionality and reliability testing confirmed the modified devices meet design requirements.

## Technological Characteristics

Ultrasonic surgical instrument; hand-controlled; rotating shaft with curved blade and clamp arm. Features integrated audible/tactile mechanism for full closure indication. Ergonomic handle design. Energy source: ultrasonic. Sterilization: sterile, single-patient-use.

## Predicate Devices

- HARMONIC ACE Curved Shears with Scissor Handle ([K051036](/device/K051036.md), [K060245](/device/K060245.md))
- HARMONIC ACE Curved Shears with Pistol Handle ([K051036](/device/K051036.md), [K060245](/device/K060245.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ethicon Endo-Surgery, LLC

្រី

## Special 510(k) Summary

| Company<br>Information       | Ethicon Endo-Surgery, LLC<br>4545 Creek Road<br>Cincinnati, OH 45242                                                                | OCT 24 2013 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact<br>Information       | David M. Locke<br>Regulatory Affairs Associate, Energy<br>Ethicon Endo-Surgery<br>(419) 233 - 2611 ext. 1281<br>dlocke1@its.jnj.com |             |
| Date Prepared                | Aug 09, 2013                                                                                                                        |             |
| Device Name                  | Trade Names:<br>○ HARMONIC ACE Curved Shears with Scissor<br>○ HARMONIC ACE Curved Shears with Ergonomic Handle                     |             |
|                              | Classification Name:<br>Instrument, Ultrasonic Surgical                                                                             |             |
|                              | Common Name:<br>Ultrasonic scalpel                                                                                                  |             |
| Device Class                 | Unclassified                                                                                                                        |             |
| Panel                        | General & Plastic Surgery                                                                                                           |             |
| Product Code ·               | LFL                                                                                                                                 |             |
| Classification<br>Regulation | Unclassified                                                                                                                        |             |

{1}------------------------------------------------

Page 2 of 3 K132512 Ethicon Endo-Surgery, LLC 510(k) Premarket Notification (Special) for Harmonic ACE14S, ACE23E and ACE36E The Predicate devices are: Predicate O Harmonic ACE curved Shears with Scissor Handle and Hand Device Information Control cleared under K051036 & K060245 Harmonic ACE Curved Shears with Pistol Handle and Hand o Control cleared under K051036 & K060245 The Harmonic ACE Curved Shears with Scissor Handle (ACE-S) is Device Description a sterile, single patient use, ultrasonic surgical instrument consisting of a scissor handle housing assembly with hand control buttons, a rotating shaft with a curved, ultrasonic blade and a clamp arm on a 14 cm length shaft. The handle housing has an integrated audible/tactile mechanism for indicating full closure. The instrument is designed for use in open or laparoscopic procedures. The Harmonic ACE instruments are used for the coagulation of vessels up to and including 5mm in diameter. The Harmonic ACE Curved Shears with Ergonomic Handle (ACE-E), Hand Control are sterile, single patient use, ultrasonic surgical instruments consisting of an ergonomic housing assembly with hand control buttons, a rotating shaft with curved, ultrasonic blades and clamp arms. The handle housing has an integrated audible/tactile mechanism for indicating full closure. The instruments are designed for use in open or laparoscopic procedures and are available in various shaft lengths. The Harmonic ACE instruments are used for the coagulation of vessels up to and including 5mm in diameter. Description of There have been minor design and labeling changes to the ACE-S Changes to and the ACE-E Harmonic instruments since the clearance of K051036 and K060245 respectively. None of these changes the Device required notification to the agency per FDA guidance document: Deciding When to Submit a 510(k) for a Change to an Existing Device (K97-1). These changes are being incorporated in this submission to provide the Agency with information for the most current version of these devices. The Harmonic ACE-E instruments were changed from a pistol grip handle design to a handle with a more ergonomic design with activation buttons positioned in a more accessible location. The 14 cm length shaft continues to use the ACE-S scissors handle. Additional minor design changes for manufacturability and user ease of assembly were implemented. Changes were verified and validated in accordance with design control requirements. Labeling for the ACE-E and ACE-S instruments was updated to show the device with the ergonomic handle and to enhance warnings and precautions. There has been no change in intended use.

0

{2}------------------------------------------------

| Ethicon Endo-Surgery, LLC                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K132512 Page 3 of |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
|                                                  | 510(k) Premarket Notification (Special) for Harmonic<br>ACE14S, ACE23E and ACE36E                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                   |
| Intended Use                                     | The Harmonic Shears are indicated for soft tissue incisions when<br>bleeding control and minimal thermal injury are desired. The<br>instrument can be used as an adjunct to or substitute for<br>electrosurgery, lasers, and steel scalpels in general, plastic, pediatric,<br>gynecologic, urologic, exposure to orthopedic structures (such as<br>spine and joint space), and other open and endoscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                   |
| Completion of<br>Design<br>Control<br>Activities | The Ethicon Endo-Surgery design control procedures were followed<br>to develop and test the minor modifications. The devices continue to<br>meet predetermined acceptance criteria. Design verification and<br>validation testing revealed no new issues of safety and efficacy<br>related to the changes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                   |
| Performance<br>Data                              | Preclinical porcine vessel sealing validation studies demonstrated<br>equivalent performance following handle modifications. Preclinical<br>study criteria for success included the following parameters:<br>transection times, hemostasis, thermal spread measurements and<br>blood pressure challenges on sealed vessels. Additionally,<br>functionality and reliability testing was performed to demonstrate<br>that the subject devices meet design requirements following device<br>modifications. No clinical studies were required to support a finding<br>of substantial equivalence. In conclusion, the testing results<br>demonstrate that the ACE-E and ACE-S devices are as safe and<br>effective as the predicate devices. The results from testing support<br>the substantial equivalence of the ACE-E and the ACE-S devices to<br>the predicate devices. |                   |
| Substantial<br>Equivalence                       | The modified devices are substantially equivalent to the unmodified<br>predicate devices in that they have the same intended use as the<br>predicate devices and the same technological characteristics that do<br>not raise different types of questions of safety and effectiveness. The<br>modification to a more ergonomic handle does not adversely affect<br>the hemostatic capability of the devices. Testing shows the modified<br>Harmonic Instruments seal vessels up to and including 5 mm as did<br>the predicate Harmonic devices.                                                                                                                                                                                                                                                                                                                         |                   |

.

.

.

・

.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic element resembling a human figure embracing the world, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 24, 2013

Ethicon Endo-Surgery, LLC Mr. David Locke Regulatory Affairs Associate, Energy 4545 Creek Road Cincinnati. Ohio 45242

Re: K132512

Trade/Device Name: HARMONIC ACE Curved Shears Regulatory Class: Unclassified Product Code: LFL Dated: October 18, 2013 Received: October 21, 2013

Dear Mr. Locke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use

K132512 510(k) Number (if known):.

Device Names:

- o HARMONIC ACE Curved Shears with Scissor Handle
- HARMONIC ACE Curved Shears with Ergonomic Handle o

Indications for Use:

The Harmonic Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space). and other open and endoscopic procedures.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use ______________ AND/OR (21 CFR 801 Subpart C)

(21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Long H. Chen -A | <div> <img alt="Signature" src="signature.png"/> </div> |
|-----------------|---------------------------------------------------------|
|-----------------|---------------------------------------------------------|

for MXM

| (Division Sign-off)          |         |
|------------------------------|---------|
| Division of Surgical Devices |         |
| 510(k) Number:               | K132512 |

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K132512](https://fda-staging.innolitics.com/device/K132512)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).