VITEK 2 AST ST MEROPENEM

K132507 · bioMerieux, Inc. · LON · Nov 15, 2013 · Microbiology

Device Facts

Record IDK132507
Device NameVITEK 2 AST ST MEROPENEM
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateNov 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Antimicrobial Susceptibility Test (Quantitative)Discriminate analysis algorithmEssential Agreement (EA) >= 90%, Category Agreement (CA) >= 90%Combined Clinical and Challenge: 98.6% EA, 96.6% CAClinical study: 699 clinical isolates. Challenge study: 100 isolates.

Indications for Use

VITEK® 2 Streptococcus Meropenem is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Meropenem is a quantitative test. Meropenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Streptococcus pneumoniae (penicillin-susceptible strains) Streptococcus agalactiae Streptococcus pyogenes The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

Device Story

VITEK® 2 AST-ST Meropenem is an automated, miniaturized, quantitative antimicrobial susceptibility test (AST) card. Input: pure culture organism suspension standardized to McFarland 0.5 using DensiCHEK Plus™. Operation: card contains 64 microwells with premeasured meropenem concentrations (0.25–4 µg/mL) and culture medium; system automatically fills card via vacuum, seals, and incubates at 35.5°C. Detection: optical scanning measures light transmittance every 15 minutes to monitor growth. Output: Minimum Inhibitory Concentration (MIC) and category interpretation (SIR). Used in clinical laboratories; operated by technicians. Output assists clinicians in determining in vitro susceptibility to meropenem, guiding antibiotic therapy decisions.

Clinical Evidence

External evaluation conducted using fresh and stock clinical isolates and challenge strains. Performance compared to CLSI broth microdilution reference method. Results: 98.6% overall essential agreement and 96.6% overall category agreement. Reproducibility and quality control results were acceptable.

Technological Characteristics

Miniaturized microdilution card with 64 microwells. Materials: culture medium base, meropenem antibiotic. Sensing: optical transmittance scanning. Energy: electrical (VITEK® 2 instrument). Connectivity: integrated with VITEK® 2/Compact systems. Software: automated growth-based MIC calculation algorithm. Sterilization: not specified.

Indications for Use

Indicated for antimicrobial susceptibility testing of Streptococcus species (S. pneumoniae, S. agalactiae, S. pyogenes) using isolated colonies. Intended for use as a laboratory aid in determining in vitro susceptibility to Meropenem.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K132507 B. Purpose for Submission: The addition of Meropenem to the VITEK® 2 AST Streptococcus card C. Measurand: Meropenem ≤ 0.06 - ≥ 4 µg/mL D. Type of Test: Antimicrobial Susceptibility Test (Quantitative) E. Applicant: bioMerieux, Inc. F. Proprietary and Established Names: VITEK® 2 AST-ST Meropenem G. Regulatory Information: 1. Regulation section: 866.1645 Short-term Antimicrobial Susceptibility Test System 2. Classification: II 3. Product code: LON – System, Test, Automated, Antimicrobial Susceptibility, Short Incubation 4. Panel: 83 Microbiology {1} H. Intended Use: 1. Intended use(s): VITEK® 2 Streptococcus Meropenem is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Meropenem is a quantitative test. Meropenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Streptococcus pneumoniae (penicillin-susceptible strains) Streptococcus agalactiae Streptococcus pyogenes The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. 2. Indication(s) for use: VITEK® 2 Streptococcus Meropenem is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Meropenem is a quantitative test. Meropenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Streptococcus pneumoniae (penicillin-susceptible strains) Streptococcus agalactiae Streptococcus pyogenes The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. 2 {2} 3. Special conditions for use statement(s): - For prescription use only - The ability of the VITEK® 2 AST Streptococcus Meropenem to detect resistance to meropenem in S. pneumoniae, S. agalactiae and S. pyogenes is unknown because resistant strains were not available at the time of comparative testing. 4. Special instrument requirements: For use with the VITEK® 2 and VITEK® 2 Compact Systems I. Device Description: The VITEK® 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK® 2 test card contains 64 microwells, containing a premeasured amount of a specific antibiotic in a culture medium base. A control well, containing only culture medium, is included on all cards. A suspension of organism from a pure culture is prepared in a tube containing 0.45-0.5% sterile saline and standardized to a McFarland 0.5 using the DensiCHEK Plus™. The desired test card is placed in the cassette along with an empty transfer tube which is incorporated into the side of the susceptibility card and placed in the organism suspension. The cassette is placed into the VITEK® 2 instrument where a diluted susceptibility test suspension is automatically prepared by the Vitek® 2 System, using the original organism suspension. The cards are automatically filled with the susceptibility test suspension via vacuum, the tubes are cut, and the cards are sealed prior to proceeding to the Incubator Loading Station. Cards are transferred from the cassette into a carousel for incubation (35.5°C) and reading via optical scanning. Readings are performed every 15 minutes for the duration of the incubation cycle. The VITEK® 2 AST Streptococcus Meropenem has the following concentrations in the card: 0.25, 0.5, 1, 2, and 4 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result range for the VITEK® 2 AST Streptococcus Meropenem card is ≤ 0.06 - ≥ 4 µg/mL. The MIC interpretive criteria and equivalent concentrations are listed in Table 1 below. Table 1. MIC Interpretive Criteria and Equivalent Concentrations | Vitek® 2 AST-Streptococcus | Equivalent Standard Method Conc. By Efficacy in µg/mL | MIC Ranges and CLSI/FDA Categories (MIC in µg/mL) | | | | --- | --- | --- | --- | --- | | | | S | I | R | | Meropenem | 0.25, 0.5, 1, 2, 4 | Streptococcus agalactiae Streptococcus pyogenes | | | | | | ≤ 0.5* | - | - | | | | Streptococcus pneumoniae | | | | | | ≤ 0.25 | 0.5 | ≥ 1 | *The current absence of data on resistant isolates precludes defining any category other than "Susceptible". If isolates yield MIC results other than susceptible, they should be submitted to a reference laboratory for additional testing. {3} # J. Substantial Equivalence Information: 1. Predicate device name(s): VITEK® 2 Streptococcus Ceftriaxone 2. Predicate $510(\mathrm{k})$ number(s): K122359 3. Comparison with predicate: Table. 2- Similarities and Differences of the VITEK ${}^{ \circ }2$ Streptococcus Meropenem and the Predicate | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Determine quantitative antimicrobial susceptibility to antimicrobial agents | Same | | Inoculum | Saline suspension of organism | Same | | Instrument | VITEK® 2 and VITEK® 2 Compact Systems | Same | | Test Card | VITEK® 2 Test Card format | Same | | Differences | | | | Item | Device | Predicate | | Antimicrobial Agent | Meropenem | Ceftriaxone | | Test Isolates | S. pneumoniae S. agalactiae S. pyogenes | S. pneumoniae S. agalactiae S. pyogenes Viridans group Streptococcus | | Reading Algorithm | Unique to meropenem | Unique to ceftriaxone | # K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; August 28, 2009. - Clinical and Laboratory Standards Institute (CLSI). Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard - Eighth Edition. CLSI document M07-A8, 2010. - CLSI. Performance Standards for Antimicrobial Susceptibility Testing; Twenty-First Informational Supplement. CLSI document M100-S21, 2011. {4} # L. Test Principle: The VITEK® 2 System optics use visible light to directly measure organism growth. The transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. An interpretive call is made between 4 and 16 hours for a "rapid" read but may be extended to 18 hours in some instances. The VITEK® 2 susceptibility card test is based on the microdilution minimum inhibitory concentration (MIC) technique with concentrations equivalent to standard method concentrations. Several parameters based on the growth characteristics observed are used to provide appropriate input for the MIC calculations. Discriminate analysis is used to develop the algorithm that determines the susceptibility result for all antimicrobials on the VITEK® 2 System. The MIC result must be linked to organism identification in order to determine a category interpretation. A category interpretation (SIR) will be reported along with each MIC result. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Reproducibility studies were performed using 10 isolates of Streptococcus pneumoniae at 3 sites on 3 separate days, in triplicate. The studies included both the auto- and manual dilution methods with the VITEK® 2 instrument system and the manual dilution method with the VITEK® 2 Compact instrument system. Greater than $95\%$ reproducibility was demonstrated with both the VITEK® 2 and VITEK® 2 Compact Systems. All MIC results were on-scale for both inoculation methods and all reading methods; therefore, best case and worst case scenarios were identical. A summary of the reproducibility study performance is illustrated in Table 3 below. Table 3. - Summary of Reproducibility Studies | VITEK® 2 Instrument Platform | Inoculation Method | Best Case | Worst Case | | --- | --- | --- | --- | | VITEK® 2 | Auto-dilution | 100% | 100% | | | Manual | 100% | 100% | | VITEK® 2 Compact | Manual | 100% | 100% | b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality control testing was conducted during device performance testing at each study site. The recommended quality control (QC) isolate, namely S. pneumoniae ATCC 49619, was tested on every test occasion with the reference method and the {5} VITEK®2 instrument platform, using both auto- and manual dilution methods. QC results for the VITEK®2 AST Streptococcus Meropenem were within the expected range > 95% of the time for both dilution methods. Testing was conducted a sufficient number of times to demonstrate that the device can produce QC results in the recommended range. A summary of the QC testing is illustrated in Table 4 below. Table 4. QC Testing - VITEK®2 Streptococcus Meropenem (Auto- and Manual Dilution Methods) | ORGANISM | Conc. (μg/mL) | Vitek®2 Auto-dilution | | Vitek®2 Manual dilution | | | --- | --- | --- | --- | --- | --- | | | | Test | Ref. | Test | Ref. | | S. pneumoniae ATCC 49619 Expected Range : 0.06 – 0.25 μg/mL | 0.03125 | | 3 | | 1 | | | 0.0625* | 70 | 196 | 32 | 100 | | | 0.125 | 133 | 3 | 72 | 3 | | | 0.25 | | 1 | | | | | 0.5 | | | | | *This value is indicated a ≤ 0.06 for the VITEK®2 System In those instances where the test result was out-of-range for all replicates of the reference method, all data from that day’s testing was considered invalid and the testing for that day was repeated. Inoculum density control was monitored using the DensiCHEK Plus™ instrument. The DensiCHEK Plus™ was standardized weekly with all results recorded and in expected range. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Clinical study testing was conducted at three external sites using the VITEK®2 AST Streptococcus Meropenem card on the VITEK®2 instrument platform and the broth microdilution reference method using Mueller-Hinton broth with lysed horse blood prepared as recommended by the CLSI. Performance testing included a total of 699 {6} clinical isolates, both fresh and stock (28.3%; 198/699). Testing of the clinical isolates was performed using the automated method of dilution of the AST card inoculum. Performance data comparing the VITEK®2 AST Streptococcus Meropenem card and the reference method are illustrated in Table A below. Challenge study testing was conducted, at one external site using the VITEK® 2 AST Streptococcus Meropenem card on the VITEK® 2 instrument platform and the broth microdilution reference method using Mueller-Hinton broth with lysed horse blood prepared as recommended by the CLSI. A challenge set consisting of a total of 100 isolates, including 50 isolates of S. agalactiae and 50 isolates of S. pneumoniae, was used for the performance testing. Testing of the challenge isolates was performed using both the auto- and manual method of dilution of the AST card inoculum. As illustrated in Tables A and B below, there is very little difference in the challenge testing device performance between the two methods of dilution of the AST card inoculum. Challenge testing of the VITEK®2 AST Streptococcus Meropenem card on the VITEK®2 Compact instrument platform was also conducted as a secondary procedural option against the broth microdilution reference method. Testing was conducted at one external site, using the same challenge set of 100 isolates. Testing was performed using the manual method of dilution of the AST card inoculum. The performance data is illustrated in Table C below. S. agalactiae, S. pyogenes isolates were analyzed using the FDA breakpoint (≤ 0.5 = S). S. pneumoniae isolates were analyzed using the FDA breakpoints (≤ 0.25 = S; 0.5 = I; ≥ 1 = R). | Table A. VITEK2/ Streptococcus Meropenem Auto-dilution | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | | | | | | | | | | | | | | S. agalactiae | 226 | 225 | 99.6 | 0 | 0 | 0 | 226 | 100 | 0 | 0 | 0 | 0 | | S. pneumoniae | 241 | 235 | 97.5 | 30 | 27 | 90 | 228 | 94.6 | 2 | 12 | 1 | 0 | | S. pyogenes | 232 | 230 | 99.1 | 0 | 0 | 0 | 232 | 100 | 0 | 0 | 0 | 0 | | Total Clinical | 699 | 690 | 98.7 | 30 | 27 | 90 | 686 | 98.1 | 2 | 12 | 1 | 0 | | Challenge | | | | | | | | | | | | | | S. agalactiae | 50 | 50 | 100 | 0 | 0 | 0 | 50 | 100 | 0 | 0 | 0 | 0 | | S. pneumoniae | 50 | 48 | 96.0 | 38 | 36 | 94.7 | 36 | 72.0 | 2 | 14 | 0 | 0 | | Total Challenge | 100 | 98 | 98.0 | 38 | 36 | 94.7 | 86 | 86.0 | 2 | 14 | 0 | 0 | | Combined Clinical and Challenge | 799 | 788 | 98.6 | 68 | 63 | 92.6 | 772 | 96.6 | 4 | 26 | 1 | 0 | {7} 8 | Table B. VITEK2/ Streptococcus Meropenem Manual Dilution | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Challenge | | | | | | | | | | | | | | S. agalactiae | 50 | 50 | 100 | 0 | 0 | 0 | 50 | 100 | 0 | 0 | 0 | 0 | | S. pneumoniae | 50 | 49 | 98.0 | 37 | 36 | 97.3 | 34 | 68.0 | 2 | 16 | 0 | 0 | | Total Challenge | 100 | 99 | 99.0 | 37 | 36 | 97.3 | 84 | 84.0 | 2 | 16 | 0 | 0 | | Table C. VITEK 2 Compact/ Streptococcus Meropenem Manual Dilution | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Challenge | | | | | | | | | | | | | | S. agalactiae | 50 | 50 | 100 | 0 | 0 | 0 | 50 | 100 | 0 | 0 | 0 | 0 | | S. pneumoniae | 50 | 49 | 98.0 | 37 | 36 | 97.3 | 35 | 70 | 2 | 15 | 0 | 0 | | Total Challenge | 100 | 99 | 99.0 | 37 | 36 | 97.3 | 85 | 85 | 2 | 15 | 0 | 0 | EA = Essential Agreement R = Resistant Isolates maj = major discrepancies CA = Category Agreement min = minor discrepancies vmj = very major discrepancies Evaluable results are those that fall within the test range of the reference method and could also be on-scale with the new device if within plus/minus one dilution. Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of VITEK®2 AST card within plus or minus one serial two-fold dilution of the antibiotic. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK®2 AST card result. Relative to all the clinical and challenge testing performed using the VITEK®2 AST Streptococcus Meropenem card, the overall %EA and %CA in most instances, met the acceptance criteria of greater than or equal to 90%. In each instance where the %CA fell below 90%, the percent essential agreement of evaluable results was very good and all the discrepancies were minor discrepancies. There were no categorical very major errors. Relative to all the clinical and challenge testing performed using the VITEK®2 AST Streptococcus Meropenem card, there were no instances of growth failure. b. Matrix comparison: Not Applicable {8} 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: The MIC interpretive criteria are illustrated in Table 5 below. Table 5. - MIC Interpretive Criteria | Antibiotic | Organism | Interpretive Criteria | | --- | --- | --- | | Meropenem | Streptococcus pneumoniae (penicillin-susceptible strains) | S ≤ 0.25 I = 0.5 R ≥ 1 | | | Streptococcus agalactiae Streptococcus pyogenes | S ≤ 0.5* | *The current absence of data on resistant isolates precludes defining any category other than “Susceptible”. If isolates yield MIC results other than susceptible, they should be submitted to a reference laboratory for additional testing. N. Proposed Labeling: The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...