K132491 · Edge Biologicals, Inc. · JTZ · Jun 30, 2014 · Microbiology
Device Facts
Record ID
K132491
Device Name
MUELLER HINTON AGAR
Applicant
Edge Biologicals, Inc.
Product Code
JTZ · Microbiology
Decision Date
Jun 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1700
Device Class
Class 2
Indications for Use
Mueller Hinton Agar is a standard basal medium intended for in vitro antimicrobial disk diffusion susceptibility testing of isolated colonies of common, rapidly growing bacteria by the Bauer-Kirby method as standardized by the Clinical and Laboratory Standards Institute (CLSI). This product has not been evaluated for gradient diffusion testing.
Device Story
Mueller Hinton Agar is a basal culture medium used in clinical microbiology laboratories for antimicrobial susceptibility testing (AST). The device functions as a growth substrate for isolated bacterial colonies. In the Bauer-Kirby procedure, a standardized bacterial suspension is swabbed onto the agar surface; sterile paper disks impregnated with specific antimicrobial agents are applied. During 16-18 hours of incubation at 35±2°C, the bacteria grow as a solid lawn while the antimicrobial agent diffuses radially. The resulting circular zone of growth inhibition is measured by laboratory personnel. The zone diameter is compared against CLSI M100 interpretive criteria to classify the organism as susceptible, intermediate, or resistant. This process assists healthcare providers in selecting appropriate antibiotic therapy for patients. The medium is formulated to ensure appropriate levels of thymine, thymidine, calcium, and magnesium to prevent interference with sulfonamide, trimethoprim, and aminoglycoside activity.
Clinical Evidence
Bench testing only. Multi-site reproducibility study conducted at three sites using six CLSI-recommended ATCC quality control strains (S. aureus, E. faecalis, E. coli, P. aeruginosa). Performance evaluated against CLSI M100-S24/M02-A11 standards. Results showed ≥95% of test results fell within acceptable zone diameter ranges for all organism-antimicrobial combinations tested.
Technological Characteristics
Basal medium containing beef extract, acid hydrolysate of casein, starch, and agar. Controlled concentrations of calcium, magnesium, thymine, and thymidine. Form factor: agar plates (15x90mm and 15x150mm). Storage: 2-8°C. Principle: agar diffusion of antimicrobial agents from paper disks.
Indications for Use
Indicated for in vitro antimicrobial disk diffusion susceptibility testing of isolated colonies of common, rapidly growing (non-fastidious) bacteria.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K132491
B. Purpose for Submission:
To obtain substantial equivalence for Mueller-Hinton Agar Medium
C. Measurand:
Antimicrobial Susceptibility Test (Disk Diffusion)
D. Type of Test:
Culture Medium for Antimicrobial Susceptibility Test (Agar)
E. Applicant:
Edge Biologicals, Incorporated
F. Proprietary and Established Names:
Mueller Hinton Agar
G. Regulatory Information:
1. Regulation section:
866.1700
2. Classification:
II
3. Product code:
JTZ – Culture Media, Antimicrobial Susceptibility Test, Mueller Hinton Agar/Broth
4. Panel:
83 Microbiology
{1}
H. Intended Use:
1. Intended use(s):
Mueller Hinton Agar is a standard basal medium intended for *in vitro* antimicrobial disk diffusion susceptibility testing of isolated colonies of common, rapidly growing bacteria by the Bauer-Kirby method as standardized by the Clinical and Laboratory Standards Institute (CLSI). This product has not been evaluated for gradient diffusion testing.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
Prescription Use
4. Special instrument requirements:
Not Applicable
I. Device Description:
Mueller Hinton agar is a standard basal medium recommended by the CLSI for routine antimicrobial disk diffusion susceptibility testing of rapidly growing (non-fastidious) bacteria. Beef and casein extracts and soluble starch in an agar base make up the nutritive base of the medium. Starch protects the test organism from toxic materials that may be in the medium. Calcium, magnesium, thymine, and thymidine concentrations are controlled.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Mueller Hinton Agar with 5% Sheep Blood
2. Predicate 510(k) number(s):
K960420
{2}
3. Comparison with predicate:
Table 1 - Similarities and Differences of Mueller Hinton Agar and the Predicate
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Mueller Hinton Agar
(Edge Biologicals, Inc.) | Mueller Hinton Agar with 5% Sheep Blood – K960420
(Becton Dickinson) |
| Intended Use | Disk diffusion antimicrobial susceptibility testing | Same |
| Source of Microorganisms for Testing | Bacterial colonies isolate from culture | Same |
| Inoculum | 0.5 McFarland Suspension, streaked over entire plate | Same |
| Read Method | Measure zones of inhibition in millimeters | Same |
| Storage | 2 - 8 degrees Celsius | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Formulation | Beef Extract, Acid Hydrolysate of Casein, Starch, Agar | Beef Extract, Acid Hydrolysate of Casein, Starch, Agar, 5% Sheep Blood Agar |
| Test Microorganisms | Common, rapidly growing, non-fastidious microorganisms | Fastidious microorganisms |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI M02-A11, Performance Standards for Antimicrobial Disk Susceptibility Tests; Approved Standard – Eleventh Edition
- CLSI M100-S24, Performance Standards for Antimicrobial Susceptibility Testing; Twenty Fourth Informational Supplement
L. Test Principle:
In the early 1960's Bauer, Kirby and others developed a standardized procedure in which Mueller Hinton Agar, a medium originally devised for the isolation of gonococci, was selected as the test medium for determining the susceptibility of bacteria to antibiotic and chemotherapeutic agents. A subsequent international collaborative study confirmed the value of Mueller Hinton Agar for this purpose because of the relatively good reproducibility of the medium, the simplicity of its formula, and the wealth of experimental data that had been accumulated using this medium.
The Bauer-Kirby procedure is based on the diffusion through agar medium (such as
{3}
Mueller Hinton agar) of antimicrobial substances which are impregnated on sterile paper disks. This method employs disks with a single concentration of antimicrobial agent and zone sizes are correlated with minimum inhibitory concentrations. In the test procedure, a standardized suspension of the organism is swabbed over the entire surface of the agar medium. Sterile paper disks impregnated with specified amounts of antibiotic or other antimicrobial agents are then placed on the surface of the inoculated agar medium. The agar medium is incubated at $35 \pm 2^{\circ} \mathrm{C}$ for 16-18 hours. The organism will grow as a solid "lawn". The antimicrobial will diffuse outward (in a circle). If the antimicrobial agent has activity against the organism, a circular zone of growth inhibition will result. The zone of inhibition around the paper disk is measured. A determination as to whether the organism is susceptible, intermediate or resistant to the antimicrobial agent is determined by comparing the size of the zone of inhibition to the zone diameter interpretive criteria in the CLSI M100 Standard. There are various factors which can influence the disk diffusion susceptibility testing such as the agar medium, excess surface moisture on the medium, agar depth, antimicrobial disk potency, inoculum concentration, and medium pH.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The performance of Mueller Hinton Agar was evaluated in a multi-site reproducibility study. Antimicrobial disk diffusion susceptibility testing was conducted at three sites. Two of the test sites were external and one test site was internal. Six organisms representing Gram negative and Gram positive rapidly growing (non-fastidious) CLSI recommended quality control strains from American Type and Culture Collection (ATCC) were included in the study, namely, *S. aureus* ATCC 25923, *E. faecalis* ATCC 29212, *E. faecalis* ATCC 51299, *E. coli* 25922, *E. coli* ATCC 35218 and *P. aeruginosa* ATCC 27853. Antimicrobials representing the major drug classes relevant to each organism were included in the study. Testing was conducted on three separate days, in duplicate, using two different lots of Mueller Hinton Agar, two different plate sizes (15 X 90 mm and 15 X 1 50 mm), and where available, antimicrobial disks from two different manufacturers resulting in a total of 72 results for each organism-antimicrobial combination. Antimicrobial disks from only one manufacturer were available at the time of testing for Cefepime, Ceftaroline, High Level Gentamicin, and High Level Streptomycin, resulting in a total of 36 results for each of these antimicrobials.
The performance of Mueller Hinton Agar was evaluated relative to the acceptable range as indicated in the antimicrobial specific FDA drug label, or CLSI Standards M100-S24 or M02-A11. Acceptable performance for each organism-antimicrobial combination was determined as test results falling within the acceptable range $\geq 95\%$ of the time.
4
{4}
Levels of thymine and thymidine in Mueller Hinton Agar were evaluated by testing Enterococcus faecalis ATCC 29212 with trimethoprim/sulfamethoxazole. Excess amounts of thymine and thymidine can reverse the inhibitory effect of sulfonamides and trimethoprim, resulting in smaller, less distinct zones of inhibition or no zones at all. The results of the testing of this organism-antimicrobial combination were within range, indicating adequate levels of thymine and thymidine.
The levels of the cations calcium and magnesium in Mueller Hinton Agar were evaluated by testing Pseudomonas aeruginosa ATCC 27853 with the aminoglycosides gentamicin and tobramycin. Excess cation content results in reduced zones, whereas low cation content results in unacceptably large zone sizes. The results of the testing of these antimicrobials were within range, indicating adequate levels of calcium and magnesium.
The results of the study for each organism-antimicrobial combination were within acceptable range $\geq 95\%$ of the time.
Reproducibility study results for Mueller Hinton Agar are represented in Tables 2 through 7 below.
Table 2 – Reproducibility Study Results - E. coli ATCC 25922
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range/ Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | Ampicillin/Sulbactam | 71/72 | 98.6 |
| 2 | Ampicillin | 72/72 | 100 |
| 3 | Cefazolin | 72/72 | 100 |
| 4 | Cefepime | 35/36* | 97.2 |
| 5 | Cefotaxime | 72/72 | 100 |
| 6 | Cefoxitin | 71/72 | 98.6 |
| 7 | Cefuroxime | 72/72 | 100 |
| 8 | Gentamicin | 72/72 | 100 |
| 9 | Imipenem | 72/72 | 100 |
| 1 | Levofloxacin | 72/72 | 100 |
| 11 | Nitrofurantoin | 72/72 | 100 |
| 12 | Piperacillin | 72/72 | 100 |
| 13 | Tobramycin | 72/72 | 100 |
| 14 | Trimethoprim/ Sulfamethoxazole | 72/72 | 100 |
*Antibiotic disks from only one manufacturer were available at the time of testing.
{5}
Table 3 - Reproducibility Study Results - E. coli ATCC 35218
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range/ Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | Ampicillin/Sulbactam | 72/72 | 100 |
| 2 | Amoxicillin/Clavulanic Acid | 72/72 | 100 |
| 3 | Ampicillin | 72/72 | 100 |
| 4 | Piperacillin/Tazobactam | 72/72 | 100 |
| 5 | Ticarcillin/Clavulanic Acid | 72/72 | 100 |
Table 4 - Reproducibility Study Results - E. faecalis ATCC 51299
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range / Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | Vancomycin** | 72/72 | 100 |
** Test plates were incubated for a full 24 hours prior to reading and interpretation.
Table 5 - Reproducibility Study Results - E. faecalis ATCC 29212
* Antibiotic disks from only one manufacturer were available at the time of testing
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range / Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | High Level Gentamicin | 36/36® | 100 |
| 2 | High level Streptomycin | 36/36® | 100 |
| 3 | Trimethoprim/ Sulfamethoxazole | 72/72 | 100 |
{6}
Table 6 – Reproducibility Study Results - Staphylococcus aureus ATCC 25923
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range/ Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | Erythromycin | 72/72 | 100 |
| 2 | Cefepime | 36/36* | 100 |
| 3 | Cefoxitin | 69/72 | 95.8 |
| 4 | Ceftaroline | 36/36* | 100 |
| 5 | Clindamycin | 72/72 | 100 |
| 6 | Oxacillin | 72/72 | 100 |
| 7 | Tetracycline | 72/72 | 100 |
| 8 | Tobramycin | 72/72 | 100 |
| 9 | Trimethoprim/ Sulfamethoxazole | 71/72 | 98.6 |
| 10 | Vancomycin** | 72/72 | 100 |
* Antibiotic disks from only one manufacturer were available at the time of testing.
** Test plates were incubated for a full 24 hours prior to reading and interpretation.
Table 7 – Reproducibility Study Results - Pseudomonas aeruginosa ATCC 27853
| Item No. | Name of Antimicrobial | Total number of Results within acceptable range/ Total number of test results | % of Results Within Acceptable Range |
| --- | --- | --- | --- |
| 1 | Cefepime | 36/36* | 100 |
| 2 | Ceftazidime | 72/72 | 100 |
| 3 | Gentamicin | 72/72 | 100 |
| 4 | Imipenem | 72/72 | 100 |
| 5 | Levofloxacin | 72/72 | 100 |
| 6 | Piperacillin | 72/72 | 100 |
| 7 | Tobramycin | 71/72 | 98.6 |
* Antibiotic disks from only one manufacturer were available at the time of testing
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Six organisms representing Gram negative and Gram positive rapidly growing (non-fastidious) CLSI recommended quality control strains from American Type and Culture Collection (ATCC) were included in the reproducibility study, namely, S. aureus ATCC 25923, E. faecalis ATCC 29212, E. faecalis ATCC 51299, E. coli 25922, E. coli ATCC 35218 and P. aeruginosa ATCC 27853. Antimicrobials
{7}
representing the major drug classes relevant to each quality control strain were tested. The results of the study for each organism-antimicrobial combination were within acceptable range ≥95% of the time.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Device performance was evaluated via the conduct of a multi-site reproducibility study. See Section M (1)(a) above for a detail description.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
{8}
# 5. Expected values/Reference range:
The acceptable zone diameter ranges for various antimicrobial agents are listed in Table 8 below for the six CLSI recommended quality control strains used in the Mueller Hinton Agar performance testing.
Table 8 – Acceptable Zone Diameter Ranges for each antimicrobial agent/QC strain tested
| Item No. | Name of Antimicrobial and disk potency in micrograms (μg) | Acceptable Zone Range in millimeters (mm) |
| --- | --- | --- |
| E. coli 25922 | | |
| 1 | Ampicillin/Sulbactam (10/10μg) | 19-24 mm |
| 2 | Ampicillin (10μg) | 16-22mm |
| 3 | Cefazolin (30μg) | 21-27mm |
| 4 | Cefepime (30μg) | 31-37mm |
| 5 | Cefotaxime (30μg) | 29-35mm |
| 6 | Cefoxitin (30μg) | 23-29mm |
| 7 | Cefuroxime (30μg) | 20-26mm |
| 8 | Gentamicin (10μg) | 19-26mm |
| 9 | Imipenem(10μg) | 26-32mm |
| 1 | Levofloxacin (5μg) | 29-37mm |
| 11 | Nitrofurantoin (300μg) | 20-25mm |
| 12 | Piperacillin(100μg) | 24-30mm |
| 13 | Tobramycin (10μg) | 18-26mm |
| 14 | Trimethoprim/Sulfamethoxazole (1.25/23.75μg) | 23-29mm |
| E. coli 35218 | | |
| 1 | Ampicillin/Sulbactam (10/10μg) | 13-19mm |
| 2 | Amoxicillin/Clavulanic Acid (20/10μg) | 17-22mm |
| 3 | Ampicillin (10μg) | 6 mm |
| 4 | Piperacillin/Tazobactam (100/10μg) | 24-30mm |
| | Ticarcillin/Clavulanic Acid (75/10μg) | 21-25mm |
| E. faecalis 51299 | | |
| 1 | Vancomycin (30μg) | ≤14mm |
| E. faecalis 29212 | | |
| 1 | High Level Gentamicin (120μg) | 16-23mm |
| 2 | High Level Streptomycin (300μg) | 14-20mm |
| 3 | Trimethoprim/Sulfamethoxazole (1.25/23.75μg) | >20mm |
{9}
| S. aureus 25923 | | |
| --- | --- | --- |
| 1 | Erythromycin (15μg) | 22-30mm |
| 2 | Cefepime (30μg) | 23-29mm |
| 3 | Cefoxitin (30μg) | 23-29mm |
| 4 | Ceftaroline (30μg) | 26-35mm |
| Item No. | Name of Antimicrobial | Acceptable Range |
| 5 | Clindamycin (2μg) | 24-30mm |
| 6 | Oxacillin (1μg) | 18-24mm |
| 7 | Tetracycline (30μg) | 24-30mm |
| 8 | Tobramycin (10μg) | 19-29mm |
| 9 | Trimethoprim/ Sulfamethoxazole (1.25/23.75μg) | 24-32mm |
| 10 | Vancomycin (30μg) | 17-21mm |
| P. aeruginosa 27853 | | |
| 1 | Cefepime (30μg) | 24-30mm |
| 2 | Ceftazidime (30μg) | 22-29mm |
| 3 | Gentamicin (10μg) | 17-23mm |
| 4 | Imipenem (10μg) | 20-28mm |
| 5 | Levofloxacin (5μg) | 19-26mm |
| 6 | Piperacillin (100μg) | 25-33mm |
| 7 | Tobramycin (10μg) | 19-25mm |
# N. Proposed Labeling:
The device labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The information submitted in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.