The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.
Device Story
System provides hemodynamic measurements for coronary/peripheral artery disease diagnosis/treatment. Inputs: intravascular blood pressure signals via fiber-optic sensor at distal tip of 0.014" monorail catheter. Operation: catheter advanced over existing guidewire to coronary ostium; system converts optical pressure signals into physiological parameters like Fractional Flow Reserve (FFR). Used in cath labs by physicians. Output: real-time pressure data and FFR calculations on touchscreen interface. Benefits: allows use of preferred guidewire for PCI, potentially avoiding multiple guidewire exchanges. Clinical decision-making: assists in assessing lesion severity to determine need for intervention.
Clinical Evidence
Prospective, multi-center, observational study (n=58). Primary endpoint: FFR measurement agreement between Navvus catheter and Radi predicate. Results: 100% device success rate; Bland-Altman bias -0.01 (limits -0.13 to 0.10); Passing-Bablok regression slope 95% CI 0.76-1.17. AUC for FFR was 87.4% (Navvus) vs 88.1% (Radi). RXi demonstrated lower drift (0.02 vs 0.06 mmHg) and lower clinically relevant drift rate (13% vs 33%). No device-related serious adverse events reported.
Technological Characteristics
Fiber optic pressure sensor embedded in 0.014" rapid exchange catheter. System includes hardware console, touchscreen UI, and embedded software. Pressure range -30 to 300 mmHg; accuracy ±3% or ±3 mmHg. Biocompatible materials per ISO 10993-1. Sterilized via ethylene oxide per ISO 11135. Connectivity: standalone system for cath lab use.
Indications for Use
Indicated for patients with coronary or peripheral artery disease requiring intravascular pressure measurements (e.g., FFR) to guide diagnosis and treatment. Applicable to adults (ages 42-86 in clinical study) with single or multi-vessel disease and reference vessel diameters ≥ 2.5 mm.
Regulatory Classification
Identification
A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.
{0}------------------------------------------------
JAN 30 2014
## Section 6
## 510(k) Summary [21 CFR 807.92]
| Date the Summary was Prepared |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| January 28, 2014 |
| 510(k) Submitter's Name and Address |
| ACIST Medical Systems<br>7905 Fuller Road<br>Eden Prairie, MN 55344<br>952-253-4519 (telephone)<br>952-941-4648 (fax)<br>matt.stepanek@acistmedical.com |
| Owner/Operator |
| ACIST Medical Systems<br>7905 Fuller Road<br>Eden Prairie, MN 55344<br>952-995-9300 (telephone)<br>952-941-4648 (fax)<br>matt.stepanek@acistmedical.com |
| Primary Contact |
| Matthew D. Stepanek<br>Principal Regulatory Affairs Specialist<br>ACIST Medical Systems<br>7905 Fuller Road<br>Eden Prairie, MN 55344<br>952-253-4519 (telephone)<br>952-941-4648 (fax)<br>matt.stepanek@acistmedical.com |
| Alternative Contact |
| Dennis Crane<br>Director of Regulatory and Clinical Affairs<br>ACIST Medical Systems<br>7905 Fuller Road<br>Eden Prairie, MN 55344<br>(952) 253 4515 (telephone)<br>952-941-4648 (fax)<br>dennis.crane@acistmedical.com |
{1}------------------------------------------------
| Sterilization Facility |
|-------------------------------------------------------------------------------|
| Sterigenics US, Inc.<br>7775 S Quincy St.<br>Willowbrook, IL 60527 |
| Establishment Registration Number: 1450293 |
| Console Manufacturing Facility |
| ACIST Medical Systems, Inc.<br>7905 Fuller Road<br>Eden Prairie, MN 55344 |
| Establishment Registration Number: 2134243 |
| Catheter Manufacturing Facility |
| MedVenture<br>2301 Centennial Blvd.<br>Jeffersonville, IN 47130 |
| Establishment Registration Number: 1222140 |
| Device Classification Names |
| Transducer, Pressure, Catheter Tip and Catheter, Pressure Monitoring, Cardiac |
| Device Common/Usual Name |
| Catheter Tip Pressure Transducer and Diagnostic Intravascular Catheter |
| Device Trade/Proprietary Name |
| Rapid Exchange (RXi) System and Navvus Catheter |
| Product Codes |
| DXO and OBI |
| Classification of Device |
| Class II |
| 21 CFR 870.2870 and 870.1200 |
{2}------------------------------------------------
| Predicate Device(s) | | |
|-----------------------------------------------------------------------------------------------------|--------------------|------------------|
| Device Name | Submission Number. | Clearance Date |
| Radi Analyzer® and PressureWire® Sensor<br>(RADI Medical Systems, St. Jude Medical, Minnesota, USA) | K062769 | December 4, 2006 |
#### Device Description
The Rapid Exchange (RXi) System and Navvus Catheter are intended to provide hemodynamic measurement information for use in the diagnosis and treatment of coronary or peripheral artery disease. When used in catheterization and related cardiovascular specialty laboratories, the Rapid Exchange (RXi) System and Navvus Catheter will compute and display physiological parameters based on the output of an optically-based pressure measuring sensor.
The Rapid Exchange (RXi) System and Navvus Catheter consist of a single patient use, catheter-based sensing device, a hardware system containing embedded software, a user interface touchscreen, and electronics for converting measured pressure signals into Fractional Flow Reserve (FFR) measurements. The intravascular blood pressure sensor is optically-based and adhered to the end of a fiber optic. The sensor and fiber optic are integrated into a monorail catheter, which connects to the system to deliver the measured pressure data.
#### Proposed Intended Use & Indication for Use
The Intended Use/Indication for Use of the Rapid Exchange (RXi) System and Navvus Catheter is as follows:
The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.
{3}------------------------------------------------
#### Summary of the Technological Characteristics to the Predicate Device(s)
The Rapid Exchange (RXi) System and Navvus Catheter employs many of the same technology characteristics as the predicate device to perform coronary pressure and fractional flow reserve (FFR) measurements. The Rapid Exchange (RXi) System and Navvus Catheter includes an intracoronary pressure measurement system that would allow the physician to use the best guidewire for the patient's anatomy and would facilitate access to coronary lesions. Using the Rapid Exchange (RXi) System and Navvus Catheter could avoid multiple guidewire exchanges if the patient went on to PCI after the FFR measurement, because the physician could use the same quidewire for the PCI.
The Rapid Exchange (RXi) System and Navvus Catheter and the predicate are intended for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Rapid Exchange (RXi) System. Navvus Catheter and predicate are designed to fit inside a percutaneous guiding catheter for the purpose of directing the catheter through a vessel. The signal output from the sensor is used for calculation and presentation of any physiological parameters, functions or indices based on pressure e.g. Fractional Flow Reserve (FFR).
The predicate devices are 0.014" guidewires with an integrated pressure sensor, together with a detachable cable for connection to a diagnostic computer.
The devices are for single use and are not to be re-sterilized.
The Rapid Exchange (RXi) System and Navvus Catheter device modifications are deemed substantially equivalent to the RAD! PressureWire Sensor, as there are no differences in the product performance, device indications for use/intended use and/or device functional scientific technology. The subject device uses the same surgical approach procedures as the predicate device.
ACIST Medical Systems considers the Rapid Exchange (RXi) System and Navvus Catheter product performance to be significantly equivalent to the predicate device, Radi Analyzer and PressureWire Sensor.
> ACIST Medical Systems Traditional 510(k) Notification Rapid Exchange (RXi) System and Navvus Catheter Page 4 of 9
{4}------------------------------------------------
| Characteristic | Radi Analyzer® and PressureWire®<br>(RADI Medical Systems, St. Jude<br>Medical, Minnesota, USA) (K062769) | ACIST Rapid Exchange (RXi) System and<br>Navvus Catheter |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | PressureWire is designed to fit inside a<br>percutaneous catheter for the purpose of<br>directing the catheter through a vessel. The<br>signal output from the sensor is used for<br>calculation and presentation of any<br>physiological parameters, functions or<br>indices based on pressure or temperature,<br>e.g. Fractional Flow Reserve (FFR). | The ACIST Rapid Exchange (RXi) System is<br>indicated for obtaining intravascular pressure<br>measurements for use in the diagnosis and<br>treatment of coronary and peripheral artery<br>disease. The Navvus Catheter is intended for<br>use with the ACIST RXi System. |
| Conditions of<br>Use | • Operating temperature 15°C to 30°C<br>(59°F to 86°F)<br>• Transport temperature -40°C to 70°C (-<br>40°F to 158°F)<br>• Storage temperature: Room temperature<br>• Relative humidity operating 30-75% | • Operating temperature 18°C to 30°C<br>• Transport temperature -29°C to 60°C<br>• Storage temperature: Room temperature<br>• Relative humidity operating 10-95% non-<br>condensing |
| Specifications<br>and<br>Properties | • Pressure Accuracy: +/-1mmHg plus +/-1%<br>of reading (over the pressure range -30 to<br>50mmHg); +/-3% of reading (over the<br>range 50 to 300mmHg)<br>• Pressure range: -30 to 300mmHg<br>• Frequency response: 0-25 Hz<br>• Zero thermal effect: 0.3 mmHg/degC<br>• Sensitivity thermal effect: 0.3%/degC<br>• Zero drift: < 7 mmHg/h | • Pressure Accuracy: +/- 3% of reading or +/- 3<br>mmHg, whichever is greater<br>• Pressure range: -30 to 300mmHg<br>• Frequency response: Not specified<br>• Zero thermal effect: <0.4 mmHg/degC<br>• Sensitivity thermal effect: Not specified<br>• Zero drift: < 7 mmHg/h |
| Design | Strain gauge type pressure sensor<br>embedded in a guidewire. | Fiber optic type pressure sensor embedded in a<br>0.014" compatible rapid exchange catheter. |
| Methods | Device is introduced as a guidewire and<br>placed in the vessel of interest. | Device is introduced over a guidewire placed in<br>the vessel of interest. |
| Principles of<br>Operation | Calculates FFR based on simultaneous<br>acquisition of pressure from two<br>transducers:<br>• Distal pressure from the sensor.<br>• Aortic pressure from a transducer reading<br>pressure in the ostium. | Calculates FFR based on simultaneous<br>acquisition of pressure from two transducers:<br>• The flexible distal tip of the catheter that<br>contains the optical pressure sensor is<br>advanced over the proximal end of the<br>0.014 in (0.36 mm) guidewire that is already<br>in place in the distal vasculature.<br>• The Navvus catheter is advanced over the wire<br>and positioned with the radiopaque marker<br>in the coronary ostium, 3 mm outside the tip<br>of the guide catheter. |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{5}------------------------------------------------
| • Equalization must take place with the Navvus Catheter distal tip just outside the tip of the guide catheter. Pressure is measured less than 3 mm proximal from the marker band visible under fluoroscopy. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • The Navvus catheter sensor is advanced past the lesion under study.. To measure FFRmin or make other measurements using the RXi system, refer to the RXi System User's Guide for further instruction regarding the use of the RXi System for recording FFR values using the Navvus Catheter. |
Systems Page 6 of 9
{6}------------------------------------------------
#### Summary of Non-Clinical Testing
The non-clinical testing included the assessment of the physical properties of the Rapid Exchange (RXi) System and Navvus Catheter and its ability to achieve its intended use and is substantially equivalent to the predicate device. Radi Analyzer and PressureWire Sensor. Bench testing of the device confirmed the suitability of the device for its intended use.
Animal testing provided evidence to confirm the same technology characteristics as the predicate device to perform coronary pressure and fractional flow reserve (FFR) measurements.
Biocompatibility assessment was to ensure that the biocompatibility has been established for the Rapid Exchange (RXi) System and Navvus Catheter; based on the similarity of the materials as the commercialized predicate device.
The following table is a summary, not limited to, the non-clinical testing conducted in accordance to industry standards/requlations:
| Test | Industry Standards |
|------------------------|------------------------------------------------------------------------------------------------------------------|
| Design<br>Verification | Quality System Requirements FDA 21 CFR Part 820.20 EN ISO 13485: 2003; Design<br>Controls FDA 21 CFR Part 820.30 |
| Device<br>Performance | Quality System Requirements FDA 21 CFR Part 820.20 EN ISO 13485: 2003; Design<br>Controls FDA 21 CFR Part 820.30 |
| Animal Model | FDA 21 CFR: Part 58 Good Laboratory Practice Regulations |
| Packaging | ISO 11607: 2006 Part 1 and Part 2: Packaging for Terminal Sterilized Medical Devices |
| Shelf Life | ISO 11607: 2006 Part 1, 6.4 Stability Testing |
| Biocompatibility | ISO 10993-1: 2009 Biological Evaluation of Medical Devices. |
| Sterilization | ISO 11135: Medical Devices - Validation and Routine Control of Ethylene Oxide<br>Sterilization and EN ISO 13485. |
The proposed Rapid Exchange (RXi) System and Navvus Catheter device performance and fundamental scientific technology remains unchanged to the predicate device. The proposed modifications do not change to the coronary pressure and fractional flow reserve (FFR) measurements in the legally marketed predicate device.
{7}------------------------------------------------
#### Summary of Clinical Testing
A prospective, observational, multi-center study was conducted. The primary objectives of the study were to assess the relationship in FFR measurements between the Navyus catheter and the Radi, including rate of device success defined as a valid FFR measurement by the Navvus Catheter where a valid FFR measurement was taken using the Radi. The study used paired measurements from the Navvus catheterand Radi within subjects as controls. Pressure measurements were assessed intra-operatively; subjects were evaluated for safety events through hospital discharge.
Three centers participated and enrolled 58 subjects. Patients with single or multi-vessel coronary artery disease indicated for FFR and a vessel reference diameter of ≥ 2.5 mm were eligible.
The average subject was a sixty-five (65) year old male (80%), ranging in age from 42-86 years. The majority of subjects had a history of dyslipidemia (86%), hypertension (74%) and angina (64%),
Study procedures were performed through the femoral (32%) artery. Reference vessel diameters ranged between 2.5 and 5.5 mm (average 3.2 ± 0.6 mm) with percent diameter stenosis ranging from 40 - 90% (average 57.2 ± 10.0%).
| Results of Clinical Investigation: | In summary, the results of this study are as follows:<br>The RXI system was 100% successful in taking valid FFR measurements in all lesions in which the Radi was able to take valid FFR measurements. The overall bias, as determined from a Bland-Altman analysis, was -0.01. Using a Bland-Altman analysis, the observed 95% limits of agreement was -0.13 to 0.10. The Kolmogorov-Smirnov test for linearity showed a cumulative p-value of 0.72, meeting success criteria of being > 0.1. The intercept determined by a Passing-Bablok linear regression between FFR measurements made using RXi system and that made using Radi had a 95% confidence interval of -0.15 to 0.19, meeting success criteria of the interval containing 0. The slope determined by a Passing-Bablok linear regression between FFR measurements made using RXi system and that made using Radi showed a 95%. confidence interval of 0.76 to 1.17, meeting success criteria of the interval containing 1. Receiver Operator Curves, using the first measured Radi FFR measurement as the standard, created for both the Navvus catheter measured FFR and the last Radi FFR demonstrated very similar areas under the curve (87.4% and 88.1%, respectively). The RXi System demonstrated a statistically significant lower rate of drift compared to the Radi system, both for mean drift (0.02 ± 0.02 for RXi vs. 0.06 ± 0.12 for Radi) and clinically relevant drift (13% vs. 33% rate of occurrence). This clinical study also demonstrated a very good safety profile. Although serious adverse events occurred during five subject's hospitalization none of these events were related to the RXi system. In a similar fashion, none of the 17 reported adverse events were device related. Eleven of the 17 were related to the procedure, including: hematomas, mild hemorrhage, elevated cardiac enzymes, nausea, vasovagal reaction, headache, hypokalemia, temporary rhythm disturbances and wheeze. All of the adverse events were mild to moderate in nature. |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{8}------------------------------------------------
#### Substantial Equivalence
The proposed Rapid Exchange (RXi) System and Navvus Catheter device uses the same surgical approach procedures as the predicate device.
The proposed Rapid Exchange (RXi) System and Navvus Catheter device and the predicate device have similar indications for use/intended use, similar implant materials, similar sterilization methods; and similar delivery characteristics.
The device performance and fundamental scientific technology remains unchanged. The differences between the proposed device and the predicate device do not have any negative effect on the safety and effectiveness of the device.
The non-clinical test and clinical results conclude that the Rapid Exchange (RXi) System and Navyus Catheter to be substantially equivalent to the predicate device, Radi Analyzer and PressureWire Sensor.
#### Conclusion
Based on ACIST Medical Systems assessment of the Rapid Exchange (RXi) System, Navus Catheter and predicate device, we conclude that the Rapid Exchange (RXi) System and Navvus Catheter contains technological features that do not differ significantly from the currently marketed device. Bench testing completed according to the Verification and Validation Test Plans confirms that the design specifications for the Rapid Exchange (RXi) System and Navvus Catheter have been met and that the device meets the applicable requirements for the safety and performance standards cited. Therefore, ACIST Medical Systems concludes that the Rapid Exchange (RXi) System and Navvus Catheter are substantially equivalent to the legally marketed predicate device subject to this 510(k) based on design, performance, characteristics, and indication for use and does not introduce any new questions regarding safety or effectiveness of the product for its proposed intended use.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, with three wavy lines extending upwards.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 30, 2014
Acist Medical Systems Matthew Stepanek Principal Regulatory Affairs Specialist 7905 Fuller Road Eden Prairie, MN 55344 US
Re: K132474
> Trade/Device Name: Rapid Exchange (RXi) System and Navvus Catheter Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer, Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DXO, OBI Dated: December 19, 2013 Received: December 20, 2013
Dear Mr. Stepanek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{10}------------------------------------------------
#### Page 2 - Matthew Stepanek
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### Section 5
### Statement of Indications for Use
510(k) Number: K132474
Device Name: Rapid Exchange (RXi) System and Navvus Catheter
Indications for Use: The ACIST Rapid Exchange (RXi) System is indicated for obtaining intravascular pressure measurements for use in the diagnosis and treatment of coronary and peripheral artery disease. The Navvus Catheter is intended for use with the ACIST RXi System.
Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Owen P.
Faris-S
Date: 2014.01.30 15:14:14
-05:00
ACIST Medical Systems Traditional 510(k) Notification Rapid Exchange (RXI) System and Navvus Catheler Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.