K132472 · Justright Surgical, LLC · GDW · Feb 11, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132472
Device Name
JUSTRIGHT 5MM STAPLER, JUSTRIGHT 5MM RELOAD
Applicant
Justright Surgical, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The JustRight™ 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
Device Story
JustRight 5mm Stapler System is a manual, reloadable surgical instrument designed for endoscopic use through 5mm cannula sleeves. Device features a 360° rotating shaft and integrated knife blade. Operation involves clamping tissue, placing two 25mm staggered rows of titanium staples, and simultaneously dividing tissue between rows. Used by surgeons in OR settings for abdominal, gynecologic, pediatric, and thoracic procedures. Provides atraumatic tissue clamping and safety lock mechanism. Benefits include smaller incision requirements (5mm vs 12mm) compared to larger predicate staplers, facilitating minimally invasive access while maintaining staple line integrity and performance equivalent to larger devices.
Clinical Evidence
Bench testing only. In vitro testing evaluated device durability, staple formation, and actuation/reload performance. In vivo testing compared the device to the predicate, assessing tissue trauma, knife cutting performance, staple formation, and staple line integrity (burst testing). Biocompatibility testing was conducted per ISO 10993.
Indicated for patients undergoing abdominal, gynecologic, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis creation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
## 510(K) SUMMARY
K132472
## Submitter Information
| | Submitter's Name:<br>Address:<br>Boulder, | JustRight Surgical LLC<br>6235 Gunpark Drive #G<br>CO 80301 | FEB 1 1 2014 | |
|-------------------------------------------|-------------------------------------------|-------------------------------------------------------------|--------------|--|
| | Telephone:<br>Fax: | 720-287-7130<br>720-287-7130 | | |
| | Contact Person:<br>Telephone: | Michele Lucey<br>603-748-1374 | | |
| Date Prepared: | | December 12, 2013 | | |
| | Device Trade Name: | JustRight™ 5mm Stapler System | | |
| | Classification: | Class II | | |
| Product Code(s): | | GDW, GAG | | |
| Common and Usual Name: Implantable Staple | | | | |
| Classification Name: | | Implantable Staple | | |
| Regulation Number(s): | | 878.4750 | | |
| Predicate Devices:<br>K900129, K061095, | | Covidien Endo GIATM 30mm 2.0 Size Staple, K892233, | | |
#### Intended Use:
The JustRight™ 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
## Device Description/Technological Characteristics:
The JustRight™ 5mm Stapler places two 25mm staggered rows of titanium staples while simultaneously dividing the tissue between the two staple rows. The unformed staple size is 2mm from staple backspan to staple leg tip. The system is provided with a stapler handle with one pre-loaded staple cartridge and reload cartridges are provided separately to allow for multiple staple line applications within one surgical procedure. The JustRight™ Stapler is compatible for introduction and use through a 5mm cannula sleeve and the stapler handle has a rotation knob to allow for 360° rotation of the staple cartridge.
{1}------------------------------------------------
## Non-Clinical Performance Data:
In vivo and in vitro testing of the JustRight ™ 5mm Stapler System was performed to evaluate device function and durability in order to demonstrate that the device is safe and effective and performs as intended. The testing conducted is summarized as follows:
- In vitro Testing .
- Device durability (multiple firings) ୍
- Staple formation o
- Device actuation and reload performance ୍
- In vivo Testing (includes comparative testing to predicate device to demonstrate . substantial equivalence)
- Tissue trauma evaluation o
- Knife cutting performance ಂ
- Device actuation and reload performance ಂ
- Staple formation o
- Staple line integrity ে
- Biocompatibilty testing was conducted in accordance with ISO 10993. .
The results of these tests demonstrate that the JustRight™ 5mm Stapler System is substantially equivalent to the predicate device.
## Substantial Equivalence:
Comparison to predicate device:
| Device Feature | JustRight™ 5mm<br>Stapler System | Covidien Endo GIA™ | Comments on<br>Differences |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GDW, GAG | Same | |
| Regulation<br>Number | 878.4750 | Same | |
| Regulation Name | Implantable Staple | Same | |
| Implantable<br>Staple Material | Titanium (ASTM F67-06) | Same | |
| Intended Use | Intended for use in<br>abdominal, gynecologic,<br>pediatric and thoracic<br>surgery for resection,<br>transection and creation | Same | |
| Device Feature | JustRight™ 5mm<br>Stapler System<br>of anastomosis | Covidien Endo GIA™ | Comments on<br>Differences |
| Anatomical Sites | Abdominal and thoracic<br>procedures | Same | |
| Unformed staple<br>leg length | 2mm | Same | |
| Staple wire<br>diameter | .2mm | Same | |
| Staple shape | B shaped | B shaped | |
| Staple Image | Image: JustRight Staple | Image: Covidien Endo GIA Staple | |
| Indicated tissue<br>thickness | .75-1.0mm | Same | |
| Cartridge length | 30mm | 50mm | Overall size does not<br>affect function |
| Staple line<br>length | 25mm | 30mm | The difference in length<br>limits the flattened<br>length for transection |
| Number of<br>staples per line | 6 | 8 | The difference in length<br>limits the flattened<br>length for transection.<br>The number of staples<br>per mm is the same. |
| Number of staple<br>lines per<br>cartridge | 4 | 6 | Staple line integrity<br>(burst) testing was<br>conducted to compare<br>the JustRight Stapler to<br>the predicate. The<br>results showed<br>equivalence in staple<br>line integrity |
| Device Feature | JustRight™ 5mm<br>Stapler System | Covidien Endo GIATM | Comments on<br>Differences |
| Staple line<br>configuration | Staggered | Same | |
| Operation | Manual | Same | |
| Integrated knife<br>blade to cut<br>between staple<br>line | Yes | Same | |
| Staples beyond<br>knife cut | Yes | Same | |
| Endoscopic use | Yes | Same | |
| Minimum<br>endoscopic<br>cannula size | 5.5mm | 12mm | The cannula size is for<br>compatibility and does<br>not affect function |
| Shaft length | 20 cm | 6,16, and 26 cm | The shaft length of the<br>JustRight Stapler is<br>within the range offered<br>for the predicate |
| Shaft Diameter | 5mm | 12mm | Device performance is<br>not dependent on the<br>shaft diameter. The<br>diameter only means that<br>different sized trocar<br>cannulas are used for<br>insertion. |
| Shaft rotation | 360° | Same | |
| Cartridge<br>articulation | No | Yes | |
| Reloadable | Yes | Same | |
| Reload limits | Up to 15 staple firings per<br>handle | Up to 25 staple firings<br>per handle | The number of staple<br>firings for the JustRight<br>device is reasonable for<br>a single use device and<br>the number of staple |
| Device Feature | JustRight™ 5mm<br>Stapler System | Covidien Endo GIA™ | Comments on<br>Differences |
| | | | firings performed within<br>a single surgical<br>procedure. |
| Safety lock | Yes | Same | |
| Atraumatic tissue<br>clamp | Yes | Same | |
| How Supplied | Sterile single use only | Same | |
| Biocompatibility | Tissue contact materials<br>are biocompatible per<br>ISO 10993 | Same | |
| Sterilization | Ethylene Oxide | Same | |
| Sterility<br>Assurance Level | 10-6 | Same | |
| Endotoxin Limit | 20 EU per Product | Same | |
| Package | Tyvek/Blister Tray | Same | |
| Labeling | Conforms to 21 CFR Part<br>801 | Same | |
{2}------------------------------------------------
## K132472
{3}------------------------------------------------
## K132472
{4}------------------------------------------------
K132472
The JustRight™ 5mm Stapler System and the predicate device, the Endo Gia™ 30 mm 2.0 Stapler, have the same intended use, similar technological characteristics, identical staple size and material, and similar principals of operation. Where there are differences those have been explained in the table above. These differences do not affect substantial equivalence. The JustRight™ 5mm Stapler System is substantially equivalent to the predicate device.
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 11, 2014
JustRight Surgical LLC Ms. Michele Lucey Lakeshore Medical Device Consulting, LLC 128 Blye Hill Landing Newbury, NH 03255
Re: K132472
Trade/Device Name: JustRight 5mm Stapler System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: December 12, 2013 Received: December 26, 2013
Dear Ms. Lucey:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced aboye and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Michele Lucey
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause, - S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K132472
Device Name JustRight 5mm Stapler System
Indications for Use (Describe)
The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatic and thoracic surgery for resection and creation of anastomosis.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE OR FOR FOR FOR EQUISE ONLY AND AND A POLICE OF CONSECTION OF CONSECTION OF THE And Andrew Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/12 description: The image contains the text "Jiyoung Dang -S". The text is written in a bold, sans-serif font. The words "Jiyoung" and "Dang" are separated by a space. The "-S" is separated from "Dang" by a space. The text is black and the background is white.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.