K132456 · Becton, Dickinson and Company · GQG · Nov 7, 2013 · Microbiology
Device Facts
Record ID
K132456
Device Name
BD VERITOR (TM) SYSTEM FOR RAPID DETECTION OF RSV
Applicant
Becton, Dickinson and Company
Product Code
GQG · Microbiology
Decision Date
Nov 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3480
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from a direct nasopharyngeal swab from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 6 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader.
Device Story
The BD Veritor System for Rapid Detection of RSV is a point-of-care chromatographic immunoassay. It utilizes a nasopharyngeal swab sample processed and applied to a test strip where RSV fusion protein binds to antibody-conjugated detector particles. The complex migrates to a reaction area for capture. The BD Veritor System Reader, a portable electronic device, performs an instrumented read of the test strip using reflectance-based measurement. The reader evaluates line signal intensities at defined positions, applies a scoring algorithm to determine the presence or absence of the analyte, and displays the result (positive, negative, or invalid) on an LCD screen. Used by healthcare professionals in laboratory or clinical settings to aid in the diagnosis of RSV infections. The instrumented read reduces subjectivity compared to visual interpretation, providing objective results to support clinical decision-making and patient management.
Clinical Evidence
Prospective multi-center study (n=523) of pediatric patients <6 years. Compared BD Veritor RSV to viral cell culture and FDA-cleared PCR. Against culture: 91.8% sensitivity (95% CI: 85.9-95.4%), 93.3% specificity (95% CI: 90.4-95.4%). Against PCR: 81.6% PPA (95% CI: 75.2-86.6%), 99.1% NPA (95% CI: 97.5-99.7%). Invalid rate 0.2%.
Technological Characteristics
Lateral flow immunoassay; colloidal-gold particle detection; monoclonal antibodies. Instrumented read via optical reflectance. Dimensions: benchtop reader. Connectivity: standalone. Sterilization: not specified. Software: algorithm-based signal processing. Materials: test strip membrane, detergent-based extraction reagent.
Indications for Use
Indicated for the qualitative detection of RSV fusion protein in direct nasopharyngeal swabs from symptomatic infants and pediatric patients under 6 years of age suspected of viral respiratory infection. Not for use as the sole basis for treatment decisions; negative results are presumptive and require confirmation by culture or molecular assay.
Regulatory Classification
Identification
Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K132456
B. Purpose for Submission:
This is a traditional 510(k) submission requesting market clearance for the BD Veritor™ System for Rapid Detection of RSV (Respiratory Syncytial Virus) directly from nasopharyngeal swabs collected from children with symptoms of respiratory infection.
C. Measurand:
The BD Veritor™ System for Rapid Detection of RSV qualitatively detects the fusion (F) protein located in the envelope of the virus.
D. Type of Test:
The BD Veritor™ System for Rapid Detection of RSV (also referred to as BD Veritor System RSV and BD Veritor RSV) is a chromatographic lateral flow immunoassay utilizing an optical reflectance reader for qualitative detection of RSV antigen.
E. Applicant:
Becton, Dickinson and Company, BD Diagnostics
F. Proprietary and Established Names:
BD Veritor™ System for Rapid Detection of RSV
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3480 - Respiratory syncytial virus serological reagents
2. Classification:
Class I
{1}
3. Product code:
GQG
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from a direct nasopharyngeal swab from patients suspected of having a viral respiratory infection. This test is intended for *in vitro* diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 6 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
For prescription use.
4. Special instrument requirements:
The BD Veritor™ System Reader
I. Device Description:
The BD Veritor™ System RSV test is a chromatographic immunoassay with an instrumented read for the qualitative detection of Respiratory Syncytial Virus antigen (RSV fusion protein) directly from a nasopharyngeal swab. It utilizes lateral flow technology and a simple optical reflectance reader. The device utilizes monoclonal antibodies for detection and capture of the RSV antigen. Each test strip is designed with spatially-distinct zones containing a positive and negative internal control. The positive control zone ensures that the sample has flowed correctly, and the negative control zone serves to address non-specific signal generation. When specimens are processed and added to the test device, RSV antigen
{2}
in the specimen binds to anti-RSV antibodies conjugated to detector particles in the RSV test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by the line of RSV antibody on the membrane. A positive result for RSV is determined by the BD Veritor System Reader (purchased separately) when antigen-conjugate is deposited at the Test "T" position and the Control "C" position on the BD Veritor System RSV assay device. The BD Veritor™ System Reader determines the line intensity at each of the spatially-defined control zones and utilizes specific algorithms to determine the presence or absence of any target analyte. The total assay time is approximately 10 minutes with reactivity determined by the optical reflectance reader.
The BD VeritorTM System for Detection of RSV kit consists of (a) 30 pouched devices, each containing one test strip, (b) 30 tubes, each containing 400 μL of RV Reagent D (a detergent solution), 30 flexible mini tip flocked swabs and one each of positive and negative external controls.
The test procedure instructs the user on proper collection of a nasopharyngeal swab. The patient specimen (a nasopharyngeal swab) is inserted and mixed in a prefilled unitized tube containing RV Reagent D (an extraction reagent) and added to the test device. RV Reagent D contains mucolytic agents that function to break down mucus in a patient specimen thereby exposing viral antigens and enhancing detection in the assay device. The swab is then removed and 3 drops of the tube contents are dispensed into the sample well of the device.

The test is allowed to stand at room temperature for 10 minutes after which time it is inserted into the reader.

The System Reader provides results on the screen. The following 3 results may be obtained.
{3}
4
| Reader Display | Interpretation |
| --- | --- |
| RSV: + | Positive Test for RSV
(RSV antigen present) |
| RSV: - | Negative Test for RSV
(no antigen detected) |
| CONTROL INVALID | Control line error |
An invalid test must be repeated.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Quidel QuickVue® RSV 10
2. Predicate 510(k) number(s):
K101918
3. Comparison with predicate:
| Item | BD Veritor™ System for RSV | Quidel QuickVue® RSV 10 |
| --- | --- | --- |
| | Similarities | |
| Intended Use | The BD Veritor™ System for Rapid Detection of Respiratory Syncytial Virus (RSV) is a chromatographic immunoassay with an instrumented read for the direct and qualitative detection of RSV fusion protein from direct nasopharyngeal swabs from patients suspected of having a viral respiratory infection. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in infants and pediatric patients under the age of 6 years. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative | The QuickVue® RSV 10 test is an immunoassay that allows for the rapid, qualitative detection of respiratory syncytial virus (RSV) antigen directly from nasopharyngeal swab and nasopharyngeal aspirate/wash specimens for symptomatic pediatric patients (less than six years old). The test is intended for use as an aid in the rapid diagnosis of acute RSV infection. Negative results do not preclude RSV infection and should not be used as the sole basis for treatment or for other management decisions. A negative test is presumptive. It is recommended that negative |
{4}
| Item | BD Veritor™ System for RSV | Quidel QuickVue® RSV 10 |
| --- | --- | --- |
| | Similarities | |
| | management decisions. A negative test is presumptive. It is recommended that negative test results be confirmed by viral cell culture or an alternative method, such as a FDA-cleared molecular assay. The test is intended for professional and laboratory use. It is to be used in conjunction with the BD Veritor™ System Reader. | test results be confirmed by cell culture. The test is intended for professional and laboratory use. |
| Specimen Types | Nasopharyngeal swabs | Nasopharyngeal swabs |
| Technology | Immunochromatographic (EIA) | Immunochromatographic (EIA) |
| Type of Test | Qualitative | Qualitative |
| Total Assay Time | 10 minutes | 10 minutes |
| | Differences | |
| Instrument | BD Veritor System Reader | None |
| Detection | Electronic readout | Visual readout |
| | | |
# K. Standard/Guidance Document Referenced (if applicable):
Not Applicable.
# L. Test Principle:
The BD Veritor™ System RSV test is a chromatographic immunoassay where RSV viral antigens in the patient sample bind to anti-RSV antibodies conjugated to detector particles on the test strip. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by an antibody line imbedded on the membrane. The assay utilizes a proprietary enhanced colloidal-gold particle at the test lines as the means for identifying the presence of RSV viral antigens. The test devices are designed with four spatially-distinct zones including positive and negative control lines, a test line for the target analyte, and a background zone. The test line for the target analyte is labeled on the test device as 'T' for test position. The onboard positive control ensures the sample has flowed correctly and is indicated on the test device as 'C'. Two of the four distinct zones on the test device are not labeled. These two zones are an onboard negative control line and an assay background zone. The onboard negative control zone (not labeled on the test device) addresses non-specific signal generation. The remaining zone is used to measure the assay background and is also not labeled. The BD Veritor™ System RSV assay incorporates an active negative control
{5}
feature in each test to identify and compensate for sample-related, nonspecific signal generation. The BD Veritor™ System Reader uses a proprietary algorithm which subtracts nonspecific signal at the negative control line from the signal present at the test line. If the resultant test line signal is above a pre-selected assay cutoff, the specimen is scored as positive. If the resultant test line signal is below the cutoff, the specimen is scored as negative. The BD Veritor™ System Reader measures the amount of light reflected from various zones along the assay strip. The measurement of the assay background zone is also evaluated as the reflectance is compared to that of the control and test zones. The instrument analyzes the reflectance data to provide the proper interpretation.
The schematic of the test strip is shown below:

## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
An analytical study was conducted over a period of 12 days to evaluate repeatability of the BD Veritor RSV test when used by two different operators. The test samples were prepared in a stabilizing liquid buffer at target concentrations of RSV to generate positive observation rates of approximately 5% (high negative), 95% (low positive) and 100% (moderate positive). The reproducibility samples were prepared by drying the liquid viral samples onto swabs. A negative swab sample was also included. Each day, each operator tested a group of 8 swab samples that were blinded and randomized and included negatives and positives (~5% positive, ~95% positive and 100% positive). The following table summarizes the results.
| Sample | Operator 1 | | | | Operator 2 | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Positive/total tested swabs | % of positivity | Lower bound | Upper bound | Positive/total tested swabs | % of positivity | Lower bound | Upper bound |
| Negative Swab | 0/24 | 0% | 0.0% | 14.2% | 0/24 | 0% | 0% | 14.2% |
| 5% Positive Swab | 2/24 | 8.3% | 1.0% | 27.0% | 4/24 | 16.7% | 4.7% | 37.4% |
| 95% Positive Swab | 20/24 | 83.3% | 62.6% | 95.3% | 20/24 | 83.3% | 62.6% | 95.3% |
| Moderate Positive Swab | 24/24 | 100% | 85.8% | 100% | 24/24 | 100% | 85.8% | 100% |
{6}
Another study was conducted over a period of 5 days to evaluate reproducibility of the BD Veritor RSV test across 3 testing sites; one site was a clinical laboratory while the other two were point-of-care sites. The samples were prepared in the same manner as above for the precision study. Two operators at each site tested a panel of 12 samples each day. The panel consisted of 3 replicates of each nominal concentration. The panels were masked and randomized. The percent of positive samples with the 95% confidence intervals (CI) for each panel member are shown below.
| Sample | Site 1 | Site 2 | Site 3 | Total |
| --- | --- | --- | --- | --- |
| High negative RSV | 6.7% (2/30)
(1.8%, 21.3%) | 6.7% (2/30)
(1.8%, 21.3%) | 13.3% (4/30)
(5.3%, 29.7%) | 8.9% (8/90)
(4.6%, 16.6%) |
| Low positive RSV | 90.0% (27/30)
(74.4%, 96.5%) | 76.7% (23/30)
(59.1%, 88.2%) | 80.0% (24/30)
(62.7%, 90.5%) | 82.2% (74/90)
(73.1%, 88.8%) |
| Moderate positive RSV | 100% (30/30)
(88.6%, 100%) | 100% (30/30)
(88.6%, 100%) | 100% (30/30)
(88.6%, 100%) | 100% (90/90)
(95.9%, 100%) |
| Negative | 0% (0/30)
(0%, 11.3%) | 0% (0/30)
(0%, 11.3%) | 0% (0/30)
(0%, 11.3%) | 0% (0/90)
(0%, 4.1%) |
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable.
d. Detection limit:
The limit of detection for the BD Veritor™ RSV test was determined for five RSV strains. Initially 10-fold serial dilutions of RSV in saline were prepared and each was tested in triplicate to find the lowest level detectable by three readers. Further characterization was performed using closely spaced dilutions. The final LoD, a concentration producing 95% positivity, was confirmed by testing 60 replicates.
{7}
| Viral Strain | Calculated LOD (TCID50/mL) | No. Positive/Total | % Positive |
| --- | --- | --- | --- |
| VR-26 (Long Subgroup A) | 1.43X105 | 57/60 | 95.0 |
| VR-955 (9320 subgroup B) | 3.98X104 | 57/60 | 95.0 |
| VR-1540 (A-2) | 1.94X103 | 59/60 | 98.3 |
| VR-1580 (Washington subgroup B) | 1.08X104 | 58/60 | 96.7 |
| VR-1400 (Wild Type subgroup A) | 2.96X103 | 76/80 | 95.0 |
To demonstrate that the LoD findings were not influenced by matrix effects, the sponsor conducted a matrix equivalence study by testing two RSV strains in dilutions near the established LoD, using (1) saline, (2) BD Universal Transport Media (UTM), and (3) negative clinical wash/aspirate sample matrix. There were 20 replicates of each dilution tested with the BD Veritor RSV test.
| | Saline | | | Clinical Matrix | | | UTM | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Viral Strain | LOD (TCID50/mL) | No. Pos/Total | Rate of Positivity | LOD (TCID50/mL) | No. Pos/Total | Rate of Positivity | LOD (TCID50/mL) | No. Pos/Total | Rate of Positivity |
| VR-26 (Long Subgroup A) | 1.43E+05 | 57/60 | 95% | 1.52E+05 | 20/20 | 100% | 1.43E+05 | 20/20 | 100% |
| | | | | 1.43E+05 | 18/20 | 90% | 1.39E+05 | 18/20 | 90% |
| | | | | 1.32E+05 | 17/20 | 85% | 1.35E+05 | 17/20 | 85% |
| VR-955 (9320 subgroup B) | 3.98E+04 | 57/60 | 95% | 3.71E+04 | 20/20 | 100% | 4.41E+04 | 20/20 | 100% |
| | | | | 3.46E+04 | 17/20 | 85% | 3.09E+04 | 19/20 | 95% |
| | | | | 3.28E+04 | 18/20 | 90% | 2.63E+04 | 14/20 | 70% |
The observed rate of positivity for VR-26 (Long Subgroup A) RSV strain at concentrations close to the established LoD (95% positivity rate), based on 20 replicates, ranged from 85% to 100% in both, the clinical matrix and UTM. Similarly, the observed rate of positivity for VR-955 (9320 Subgroup B) RSV strain at concentrations close to the established LoD ranged from 85% to 100% in clinical matrix and from 70% to 100% in UTM. These results suggest that there is no significant matrix effect on the assay performance at concentrations near LoD.
# e. Analytical specificity:
The BD Veritor for RSV test was evaluated for potential cross-reactivity with other microorganisms. Various bacteria, viruses and yeast were tested at clinically relevant concentrations. All determinations were performed in triplicate with no cross reactivity observed for any of the tested organisms. The results are shown below.
{8}
| Sample Description | Concentration Tested | Cross Reactivity with RSV (Yes or No) |
| --- | --- | --- |
| Adenovirus Type 1 | 1.58 X 10^{5} TCID_{50}/mL | No |
| Adenovirus Type 7 | 1.01 X 10^{5} TCID_{50}/mL | No |
| Bacteriodes fragilis | 5 X 10^{6} CFU/mL | No |
| Bordetella pertussis | 5 X 10^{6} CFU/mL | No |
| Candida albicans | 5 X 10^{6} CFU/mL | No |
| Chlamydia pneumoniae | 1.01 X 10^{5} TCID_{50}/mL | No |
| Corynebacterium diphtherium | 5 X 10^{6} CFU/mL | No |
| Cytomegalovirus | 1.58 X 10^{5} TCID_{50}/mL | No |
| Enterovirus VR-28 Human Coxsackievirus | 8.89 X 10^{5} TCID_{50}/mL | No |
| Escherichia coli | 5 X 10^{6} CFU/mL | No |
| Fusobacterium nucleatum | 1.5 X 10^{6} CFU/mL | No |
| Haemophilus influenzae | 5 X 10^{6} CFU/mL | No |
| Haemophilus parainfluenzae | 5 X 10^{6} CFU/mL | No |
| HSV Type 1 (HF) | 8.89 X 10^{5} TCID_{50}/mL | No |
| Human coronavirus | 1.01 X 10^{5} TCID_{50}/mL | No |
| Human metapneumovirus | 2.2 X 10^{5} TCID_{50}/mL | No |
| Human parainfluenza | 6.5 X 10^{6} TCID_{50}/mL | No |
| Influenza A/California/7/2009 H1N1 | 1.0 X 10^{6} TCID_{50}/mL | No |
| Influenza A/Brisbane/10/2007 H3N2 | 1.02 X 10^{5} TCID_{50}/mL | No |
| Influenza A/Victoria/3/75 H3N2 | 4.11 X 10^{5} TCID_{50}/mL | No |
| Influenza B/Brisbane/60/2008 | 6.31 X 10^{5} TCID_{50}/mL | No |
| Influenza B/Florida/4/2006 | 2.15 X 10^{5} TCID_{50}/mL | No |
| Influenza B/Lee/40 | 4.44 X 10^{5} TCID_{50}/mL | No |
{9}
| Kingella kingae | 5 X 10^6 CFU/mL | No |
| --- | --- | --- |
| Klebsiella pneumoniae | 5 X 10^6 CFU/mL | No |
| Lactobacillus casei | 5 X 10^6 CFU/mL | No |
| Legionella pneumophila | 5 X 10^6 CFU/mL | No |
| Measles | 1.58 X 10^4 TCID50/mL | No |
| Moraxella catarrhalis | 5 X 10^6 CFU/mL | No |
| Mumps virus | 1.05 X 10^5 TCID50/mL | No |
| Mycobacterium tuberculosis avirulent | 5 X 10^6 CFU/mL | No |
| Mycoplasma pneumoniae | 5 X 10^6 CFU/mL | No |
| Neisseria gonorrhoeae | 5 X 10^6 CFU/mL | No |
| Neisseria meningitidis | 5 X 10^6 CFU/mL | No |
| Neisseria mucosa | 5 X 10^6 CFU/mL | No |
| Neisseria perflaus | 5 X 10^6 CFU/mL | No |
| Neisseria subflava | 5 X 10^6 CFU/mL | No |
| Peptostreptococcus anaerobius | 5 X 10^6 CFU/mL | No |
| Porphyromonas asaccharolyticus | 5 X 10^6 CFU/mL | No |
| Prevotella oralis | 5 X 10^6 CFU/mL | No |
| Propionibacterium acnes | 5 X 10^6 CFU/mL | No |
| Proteus mirabilis | 5 X 10^6 CFU/mL | No |
| Pseudomonas aeruginosa | 5 X 10^6 CFU/mL | No |
| Rhinovirus | 1.58 X 10^4 TCID50/mL | No |
| Serratia marcescens | 5 X 10^6 CFU/mL | No |
| Staphylococcus aureus | 5 X 10^6 CFU/mL | No |
| Staphylococcus epidermidis | 5 X 10^6 CFU/mL | No |
| Streptococcus mutans | 5 X 10^6 CFU/mL | No |
| Streptococcus pneumoniae | 5 X 10^6 CFU/mL | No |
| Group A Streptococcus | 5 X 10^6 CFU/mL | No |
{10}
# f. Interference
Various substances were evaluated for potential interference with the BD Veritor System for RSV test at concentrations comparable to or greater than levels that may be present in patient respiratory samples. In this study RSV positive samples, prepared at an antigen concentration corresponding to a weak positive, were spiked with the substances listed below. No interference was observed at the concentrations shown.
| Substance | Concentration Tested | Interference with RSV Result |
| --- | --- | --- |
| Whole Blood | 2% | No |
| 4-Acetamidophenol | 10 mg/mL | No |
| Acetylsalicylic acid | 20 mg/mL | No |
| Albuterol | 0.083 mg/mL | No |
| Amantadine | 500 ng/mL | No |
| Ayr Saline Nasal Gel | 10 mg/mL | No |
| Beclomethasone | 500 ng/mL | No |
| Budesonide | 500 ng/mL | No |
| Chlorpheniramine maleate | 5 mg/mL | No |
| Dexamethasone | 10 mg/mL | No |
| Dextromethorphan | 10 mg/mL | No |
| Diphenhydramine HCl | 5 mg/mL | No |
| Fexofenadine | 500 ng/mL | No |
| FluMist® | 1% | No |
| Flunisolide | 500 ng/mL | No |
| Fluticasone | 500 ng/mL | No |
| Guaiacol Glyceryl Ether | 20 mg/mL | No |
| Ibuprofen | 10 mg/mL | No |
| Loratidine | 100 ng/mL | No |
| Menthol Throat Lozenges | 10 mg/mL | No |
| Mometasone | 500 ng/mL | No |
| Mupirocin | 500 ng/mL | No |
| Oseltamivir | 500 ng/mL | No |
| Oxymetazoline | 0.05 mg/mL | No |
| Phenylephrine | 1 mg/mL | No |
| Pseudoephedrine HCl | 20 mg/mL | No |
| Purified Mucin Protein | 1 mg/mL | No |
| Ribavirin | 500 ng/mL | No |
| Rubber | 100 ng/mL | No |
| Sulfadine | 100 ng/mL | No |
| Sulfadine | 100 ng/mL | No |
| Toluidine | 100 ng/mL | No |
| Toluidine | 100 ng/mL | No |
| Toluidine | 100 | No |
{11}
| Rimantadine | 500 ng/mL | No |
| --- | --- | --- |
| Synagis | 4 ug/mL | No |
| Tobramycin | 500 ng/mL | No |
| Triamcinolone | 500 ng/mL | No |
| Zanamivir | 1 mg/mL | No |
| Antiseptic Mouthwash (CVS) | 5% | No |
| Cool Mint Listerine Antiseptic | 5% | No |
| Homeopathic Allergy Medicine | 10 mg/mL | No |
| Ibuprofen Concentrated Drops | 25% | No |
| Infants Advil concentrated Drops | 25% | No |
| Nasal Spray | 10% | No |
| Nasal Spray | 10% | No |
| Nasal Spray | 10% | No |
| Pedia Care Drops for infants | 25% | No |
| Scope Outlast Mouthwash | 5% | No |
| Triaminic infants drops | 25% | No |
# g. Specimen stability:
The BD Veritor for RSV test is intended to be used with nasopharyngeal samples immediately after collection. However, since a delay in testing may occur in real life situations, the sponsor conducted a sample stability study to allow for such circumstances. In this study, RSV strains Long (A) and 9320 (B) were used to spike swabs at viral loads near the LoD. The swabs were held at room temperature and were sampled at $0.5\mathrm{hr}$ intervals from 0 hours up to 2 hours after being spiked. All swabs tested as positive at all-time points for both strains and no deterioration of signal was observed. These results support the decision to allow patient testing to be conducted for up to 1 hour after collection of the NP swab.
# h. Assay cut-off:
The output of the BD Veritor Reader is provided in arbitrary units (AU) measuring reflectance. A "zero" AU signal is assigned when reading a white surface and 100 AU is the maximum signal that the Reader can provide. The assay cutoff was initially evaluated using negative specimens (with determined zero concentration of analyte) to assess the inherent noise level of the system which was determined to be approximately 0.4-0.5 AU. Additional measurements of contrived samples with concentrations near the cutoff plus leftover clinical samples were used to estimate a preliminary assay cutoff to be 1.5 AU (3 X 0.5 AU). Following an ROC analysis of the results, the final assay cutoff was determined to achieve optimal balance between maximal sensitivity without compromising the specificity. The selected assay cutoff was ultimately validated with clinical positive and negative samples in the clinical study. The results provided to the user are qualitative for any valid result, either positive or negative.
{12}
13
2. Comparison studies:
a. Method comparison with predicate device:
See clinical studies below.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
Performance characteristics of the BD Veritor System for Rapid Detection of RSV test were evaluated in a multi-center prospective study across eight geographically diverse testing sites using nasopharyngeal (NP) swabs collected from patients with signs and symptoms consistent with respiratory tract infection. The study was conducted between November 2012 and June 2013. A total of 540 patients were enrolled in the clinical study of which 17 were excluded due to protocol deviations, leaving 523 specimens to be included in the calculations of estimates of assay performance. All specimens were collected from patients < 6 years of age. Of those, 58.3% (305) were from subjects < 2 years old and 41.7% (218) were from patients 2-5 years of age. There were 223 (42.6%) females and 300 (57.4%) males. Informed consent was obtained for all patients prior to testing. Two NP swabs were collected sequentially, one swab from each nostril, from each subject. One swab was placed into transport media for reference testing. The other swab was either placed in an empty tube until ready for testing, or tested immediately on the BD Veritor™ System. The order of swabs used for reference testing vs. the direct testing with the device was randomized. The direct testing with the BD Veritor device was intended to be completed immediately. If the sample could not be tested immediately, it was held at room temperature (20-25°C) for no longer than one hour prior to testing. The reference testing consisted of RSV viral culture and PCR testing (using the remaining specimen) with an FDA cleared molecular test for RSV.
.Clinical Performance
The performance of the BD Veritor System for Rapid Detection of RSV test with NP swabs was estimated based on the comparison with results from viral cell culture (sensitivity and specificity) as well as in comparison with the results obtained by a PCR comparator method (positive percent agreement (PPA) and negative percent agreement (NPA)).
{13}
BD Veritor RSV test performance with NP swabs versus viral culture, based on 523 specimens:
| BD VERITOR RSV | CULTURE | |
| --- | --- | --- |
| | POSITIVE | NEGATIVE |
| POSITIVE | 123 | 26* |
| NEGATIVE | 11 | 363 |
| TOTAL | 134 | 389 |
SENSITIVITY: $91.8\%$ (95% CI: 85.9%, 95.4%)
SPECIFICITY: $93.3\%$ (95% CI: 90.4%, 95.4%)
*Of the 26 BD Veritor RSV Positive, Viral Cell Culture negative specimens, 23 were positive by an FDA cleared molecular assay.
BD Veritor RSV test performance with NP swabs versus an FDA cleared PCR molecular assay, based on 523 specimens:
| BD VERITOR RSV | PCR | |
| --- | --- | --- |
| | POSITIVE | NEGATIVE |
| POSITIVE | 146 | 3 |
| NEGATIVE | 33 | 341 |
| TOTAL | 179 | 344 |
PPA: $81.6\%$ (95% CI: 75.2%, 86.6%)
NPA: $99.1\%$ (95% CI: 97.5%, 99.7%)
Among the 523 tests performed with the BD Veritor RSV test, there was one invalid result. The overall invalid rate was calculated to be $0.2\%$ (95% CI: $0.0\%$ , $1.1\%$ ).
# 4. Clinical cut-off:
Not applicable.
# 5. Expected values/Reference range:
The rate of positivity observed in RSV testing will vary depending on the method of specimen collection, handling/transport system employed, detection method utilized, time of year, age of the patient, and disease prevalence. The overall prevalence of RSV infection observed during the time the BD Veritor RSV trial was $25.6\%$ when determined by the results obtained with cell culture, or $34.2\%$ when determined by the results obtained with PCR. The prevalence observed at each testing site during the study, depending on the reference method, is shown below.
{14}
| Site | Prevalence based on Viral Cell Culture | Prevalence based on PCR |
| --- | --- | --- |
| P1 | 32.9% (49/149) | 45.6% (68/149) |
| P2 | 12.9% (12/93) | 17.2% (16/93) |
| P3 | 7.7% (4/52) | 15.4% (8/52) |
| P4 | 30.2% (13/43) | 37.2% (16/43) |
| P5 | 26.3% (26/99) | 34.3% (34/99) |
| P6 | 65.2% (15/23) | 69.6% (16/23) |
| P7 | 21.1% (12/57) | 31.6% (18/57) |
| P8 | 42.9% (3/7) | 42.9% (3/7) |
| Overall | 25.6% (134/523) | 34.2% (179/523) |
The positive and negative predictive values for the BD Veritor RSV assay, under hypothetical prevalence conditions, were calculated using the overall performance estimates (when compared to culture and when compared to PCR, separately).
Hypothetical Positive and Negative Predictive Values for BD Veritor RSV Test,
Based on Performance against Viral Culture:
| Performance against Viral Culture | Hypothetical Prevalence | Positive Predictive Value | Negative Predictive Value |
| --- | --- | --- | --- |
| Sensitivity 91.8% | 5% | 42.0% | 99.5% |
| | 10% | 60.4% | 99.0% |
| | 20% | 77.4% | 97.8% |
| Specificity 93.3% | 30% | 85.5% | 96.4% |
| | 40.0% | 90.2% | 94.5% |
Hypothetical Positive and Negative Predictive Values for BD Veritor RSV Test, Based on Performance against PCR:
| Performance against PCR | Hypothetical Prevalence | Positive Predictive Value | Negative Predictive Value |
| --- | --- | --- | --- |
| PPA 81.6% | 5% | 83.1% | 99.0% |
| NPA 99.1% | 10% | 91.2% | 98.0% |
| | 20% | 95.9% | 95.6% |
| | 30% | 97.6% | 92.6% |
| | 40.0% | 98.4% | 89.0% |
# N. Instrument Name:
BD Veritor™ System Reader
{15}
O. System Descriptions:
1. Modes of Operation:
The Veritor™ System Reader is a small, battery powered, bench top instrument that is used to read the Veritor lateral flow test cassette. After the extracted patient sample has been added to the test cassette, the test is developed at room temperature for 10 minutes. The cassette is then placed into the reader where it is scanned. The cassette is divided into distinct zones where the analyzer reads the negative background, positive control, and the Influenza A and B specific zones. The reader applies an algorithm to determine the background of the test as well as the specific signal from the A or B test zones. The reader has a finite number of reads and will prompt the end-user as the total number of reads approaches the lifetime of the unit.
2. Software:
The Veritor™ System Reader is the identical instrument that has been reviewed and cleared with the BD influenza A and B assay K112277. FDA has conducted an additional review of the applicant’s instrument Hazard Analysis and software development processes associated with this specific device, the BD Veritor System for RSV.
Yes ☐ X ☑ or No ☐
3. Specimen Identification:
Not applicable
4. Specimen Sampling and Handling:
Not applicable
5. Calibration:
The Veritor Reader is not configurable by the end user and is designed to have a finite lifetime based on number of tests performed or shelf life from date of manufacture. Device calibration is not required, however, a verification cartridge is provided with the reader to perform a functional test of the reader.
6. Quality Control:
The Quality Control testing during the clinical trial was performed each day of testing at each site using the external Positive and Negative control swabs. A functional check of the Reader with the cartridge verification was also performed daily. The summary of QC results is shown below.
16
{16}
| | QC POS | | | | QC NEG | | | | VERIFY | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | PASS | | FAIL | | PASS | | FAIL | | PASS | | FAIL | |
| Site | N | % | N | % | N | % | N | % | N | % | N | % |
| P1 | 92 | 98.9 | 1 | 1.1 | 93 | 100 | 0 | 0.0 | 93 | 100 | 0 | 0.0 |
| P2 | 96 | 99.0 | 1 | 1.0 | 96 | 98.0 | 2 | 2.0 | 94 | 100 | 0 | 0.0 |
| P3 | 34 | 94.4 | 2 | 5.6 | 35 | 100 | 0 | 0.0 | 36 | 100 | 0 | 0.0 |
| P4 | 31 | 100 | 0 | 0.0 | 31 | 100 | 0 | 0.0 | 31 | 100 | 0 | 0.0 |
| P5 | 25 | 100 | 0 | 0.0 | 25 | 100 | 0 | 0.0 | 25 | 100 | 0 | 0.0 |
| P6 | 18 | 100 | 0 | 0.0 | 18 | 100 | 0 | 0.0 | 17 | 94.4 | 1 | 5.6 |
| P7 | 34 | 100 | 0 | 0.0 | 34 | 100 | 0 | 0.0 | 36 | 100 | 0 | 0.0 |
| P8 | 9 | 100 | 0 | 0.0 | 9 | 100 | 0 | 0.0 | 9 | 100 | 0 | 0.0 |
| P9 | 20 | 100 | 0 | 0.0 | 20 | 100 | 0 | 0.0 | 20 | 100 | 0 | 0.0 |
| Total | 359 | 98.9 | 4 | 1.1 | 361 | 99.4 | 2 | 0.6 | 361 | 99.7 | 1 | 0.3 |
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.