K132435 · Smith & Nephew, Inc. · LZO · Oct 22, 2013 · Orthopedic
Device Facts
Record ID
K132435
Device Name
SMITH & NEPHEW GLOBAL FEMORAL HEAD TRIALS
Applicant
Smith & Nephew, Inc.
Product Code
LZO · Orthopedic
Decision Date
Oct 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Indications for Use
Smith & Nephew Global Femoral Head Trials are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Hip Systems in their cleared Indications for Use as provided below. Smith & Nephew Uncemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Uncemented Hip Systems and their cleared Indications for Use. Smith & Nephew Cemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cemented Hip Systems and their cleared Indications for Use. Smith & Nephew PLUS Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew PLUS Hip Systems and their cleared Indications for Use.
Device Story
Global Femoral Head Trials are surgical instruments used during hip arthroplasty to assist in the implantation of Smith & Nephew hip systems. These trials serve as temporary placeholders to verify fit and sizing before the final implant is placed. Used in the operating room by orthopedic surgeons; the trials are manually manipulated to assess joint stability and range of motion. By providing a trial fit, the device helps the surgeon select the appropriate final implant size, potentially reducing surgical time and improving patient outcomes by ensuring proper joint reconstruction and pain relief. No electronic or automated processing is involved.
Clinical Evidence
No clinical data provided; bench testing only. Substantial equivalence is based on material and technological similarities to previously cleared predicate devices.
Technological Characteristics
Manufactured from polyphenylsulfone (PPSU) with up to 10% barium sulfate (radiopacifier) and colorants for size differentiation. Manual surgical instrumentation; non-powered; non-sterile (intended for sterilization by the user).
Indications for Use
Indicated for patients undergoing primary or revision hip arthroplasty due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), noninflammatory degenerative joint disease (e.g., osteoarthritis), avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction. Specific stems indicated for advanced hip joint wear, failed previous arthroplasties, varus femur forms, or proximal femur trumpet shapes.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
#### K132435 page 1 of 4
### > We are smith&nephew
#### Submitted by:
Date of Summary:
Contact Person and Address:
Name of Device:
Common Name:
Device Classification Name and Reference:
Smith & Nephew, Inc. Advanced Surgical Devices Division 1450 East Brooks Road Memphis, Tennessee 38116
October 17, 2013
Megan Fessenden
Regulatory Affairs Specialist
T 901-399-1104 F 901-721-2583
**OCT** 22 2013
Smith & Nephew, Inc. Global Femoral Head Trials
Orthopaedic Surgical Instrumentation
21 CFR 888.3350 Hip joint metal/polymer semi-constrained cemented prosthesis
21 CFR 888.3353 Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis 21 CFR 888,3358 Hip joint metal/polymer metal semiconstrained porous-coated uncemented prosthesis 21 CFR 888.3360 Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis 21 CFR 888.3390 Hip joint femoral (hemi-hip) metal/polymer
cemented or uncemented prosthesis
| Device Class: | Class II |
|---------------|-----------------------------------|
| Panel Code: | Orthopaedics/87 |
| Product Code: | LZO, KWY, LWJ, LPH, JDI, MEH, MBL |
#### Device Description
Per U.S Food and Drug Administration (FDA) regulation, device-specific instruments are accessory devices and take on the classification of the device(s) with which they are used. Instruments that are designed to assist in the implantation of Class II Smith & Nephew Hip Systems are classified as Class II devices and are subject to premarket notifications and regulations.
The subjects of this 510|k) are the Smith & Nephew Global Femoral Head Trials. The femoral head trials are instruments used to assist in the implantation of cleared Smith & Nephew hip systems.
#### Intended Use
Smith & Nephew Global Femoral Head Trials are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Hip Systems in their cleared Indications for Use as provided below.
{1}------------------------------------------------
#### K132435 page 2 of 4
#### Smith & Nephew Uncemented Femoral Stems
Smith & Nephew Uncemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Uncemented Hip Systems and their cleared Indications for Use.
Uncemented Femoral Stems are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
#### Smith & Nephew Cemented Femoral Stems
Smith & Nephew Cemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cemented Hip Systems and their cleared Indications for Use.
Cemented Femoral Stems are indicated for cemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative ioint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis: diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
#### Smith & Nephew Inc. PLUS Femoral Stems
Smith & Nephew PLUS Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew PLUS Hip Systems and their cleared Indications for Use.
Smith & Nephew Inc PLUS Femoral Stems are indicated for:
- The SL-PLUS Stem Primary is intended for treating patients who are candidates for . total hip arthroplasty because the natural femoral head and neck has been subject to disease or trauma. The SL-PLUS Stem is intended for advanced hip joint wear due to
Confidential 3
#### Confidential
{2}------------------------------------------------
degenerative, posttraumatic or rheumatoid arthritis; fracture or avascular necrosis of the femoral head.
- The SLR-PLUS STEM, a revision component, is also available to replace previously . failed femoral him arthroplasties. Both components can be used with or without cement. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.
- The SL-PLUS Lateralized Stem is intended for varus femur forms and trumpet shape . of the proximal femur (champagne flute). These stems are for uncemented use only. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.
- The SL-PLUS MIA Stem is indicated for individuals undergoing primary and revision . surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
- . The POLARSTEM Standard and Lateral Femoral Stems are indicated for: advanced degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis, fracture or avascular necrosis of the femoral head, failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, all forms of osteoarthritis, patients with hips at risk of dislocation, and femoral neck fracture or proximal fracture to hip joint.
#### Technological Characteristics
The subject devices are similar to existing femoral head trials currently being used by surgeons when implanting femoral hip systems. The trials are manufactured from polyphenylsulfone (PPSU) with up to 10% of the radiopacifier barium sulfate. In addition, colorants are blended into the PPSU resin to allow for differentiation between the various sizes. The same materials are used in the manufacturing of existing hip system trials.
#### Substantial Equivalence Information
When compared to the previously cleared device-specific instruments listed in Table 2, the proposed instruments utilize the same raw materials, manufacturing processes, and sterilization methods (where applicable). It should also be noted that the nature of body contact is the same as found with the predicate devices that have been identified in this premarket notification.
| Manufacturer | Description | Submission<br>Number | Clearance Date |
|----------------|------------------------|----------------------|----------------|
| Smith & Nephew | Radiopaque Trial Necks | K113039 | 10/5/2012 |
#### Table 2: Substantially Equivalent Predicates to the Global Femoral Head Trials
{3}------------------------------------------------
## K132435 page 4 of 4
#### Conclusion
This Traditional 510(k) Premarket Notification is being submitted to request clearance for the Global Femoral Head Trials. Based upon the technological and material similarities, the Global Formeral Frontially equivalent to the predicate devices identified in this premarket notification.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is centered horizontally and vertically in the image. The background of the image is white.
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 22, 2013
Smith & Nephew, Incorporated Ms. Megan Fessenden Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016
Re: K132435
Trade/Device Name: Global Femoral Head Trials for use with Smith & Nephew, Inc. Uncemented, Cemented and PLUS Femoral Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, KWY, LWJ, LPH, JDI, MEH, MBL Dated: August 26, 2013 Received: August 28, 2013
Dear Ms. Fessenden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
#### Page 2 - Ms. Megan Fessenden
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin 原Keith
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K 132435 page 1 of 4
Device Name: Global Femoral Head Trials for use with Smith & Nephew Inc. Uncemented Femoral Stems
Indications for Use:
Smith & Nephew Uncemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Uncemented Hip Systems and their cleared Indications for Use.
Uncemented Femoral Stems are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth L. Frank -S
র্তা - Page Page -
{7}------------------------------------------------
#### Giobal Femoral Head Trials Indications for Use
510(k) Number (if known): K132435 page 2 of 4
Device Name: Global Femoral Head Trials for use with Smith & Nephew, Inc. Cemented Femoral Stems
Indications for Use:
Smith & Nephew Cemented Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Cemented Hip Systems and their cleared Indications for Use.
Cemented Femoral Stems are indicated for cemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.
x AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use . (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth L. Frank -S
Page _________________________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
#### Global Femoral Head Tria Indications for Us
510(k) Number (if known):
K132435 page 3 of 4
Device Name: Global Femoral Head Trials for use with Smith & Nephew Inc PLUS Femoral Stems
Indications for Use:
Smith & Nephew PLUS Femoral Stem Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew PLUS Hip Systems and their cleared Indications for Use.
Smith & Nephew Inc PLUS Femoral Stems are indicated for:
- The SL-PLUS Stem Primary is intended for treating patients who are candidates for total . hip arthroplasty because the natural femoral head and neck has been subject to disease or trauma. The SL-PLUS Stem is intended for advanced hip joint wear due to degenerative, posttraumatic or rheumatoid arthritis; fracture or avascular necrosis of the femoral head.
- The SLR-PLUS STEM, a revision component, is also available to replace previously failed . femoral him arthroplasties. Both components can be used with or without cement. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.
- The SL-PLUS Lateralized Stem is intended for varus femur forms and trumpet shape of the . proximal femur (champagne flute). These stems are for uncemented use only. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _ ol
Elizabeth L. Frank -S
{9}------------------------------------------------
#### Global Femoral Head Trials Indications for Use
510(k) Number (if known): K132435 page 4 of 4
Device Name: Global Femoral Head Trials for use with Smith & Nephew Inc PLUS Femoral Stems
Indications for Use (continued):
- The SL-PLUS MIA Stem is indicated for individuals undergoing primary and revision . surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
- The POLARSTEM Standard and Lateral Femoral Stems are indicated for: advanced . degeneration of the hip joint as a result of degenerative, post-traumatic or rheumatoid arthritis, fracture or avascular necrosis of the femoral head, failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, all forms of osteoarthritis, patients with hips at risk of dislocation, and femoral neck fracture or proximal fracture to hip joint.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth L. Frank -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.