THE GYNOCULAR

K132423 · Gynius AB · HEX · Apr 8, 2014 · Obstetrics/Gynecology

Device Facts

Record IDK132423
Device NameTHE GYNOCULAR
ApplicantGynius AB
Product CodeHEX · Obstetrics/Gynecology
Decision DateApr 8, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1630
Device ClassClass 2

Indications for Use

The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy.

Device Story

Gynocular is a handheld, battery-powered colposcope providing magnified visualization (5x, 8x, 12x) of cervical, vaginal, and anogenital tissues. Device uses high-intensity LEDs for illumination and includes a green filter. Operated by physicians or trained colposcopists in clinical settings; can be used handheld or mounted on a standard tripod for hands-free operation. Device does not contact the patient. Output is direct optical visualization of tissue, enabling the clinician to evaluate lesions and identify biopsy sites. Benefits include portability and performance comparable to stationary colposcopes, facilitating diagnostic procedures in various clinical environments.

Clinical Evidence

Clinical evidence includes a randomized, crossover pilot trial of 69 women testing positive for visual inspection with acetic acid. The study compared Swede scores obtained via the Gynocular versus a standard stationary colposcope, correlated with histological diagnosis after biopsy. Results showed 70% overall agreement between the two techniques. The study concluded the Gynocular has a significant level of agreement with stationary colposcopes in assessing cervical lesions.

Technological Characteristics

Handheld colposcope with 300 mm focal distance and 5x, 8x, 12x magnification. Features high-intensity LED illumination, green filter, and rechargeable lithium-ion battery. Includes tripod-mounting capability. Meets IEC 60601-1 and EN 60601-1 safety standards. Includes automatic thermal shut-off mechanism.

Indications for Use

Indicated for magnified visualization of vaginal, cervical, and external genitalia tissues to evaluate and select areas for biopsy in patients requiring colposcopy.

Regulatory Classification

Identification

A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.

Special Controls

*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ GYNIUS AB 510(k) submission Gynoculan Colposcope ## APR 0 8 2014 ### 5. 510(k) Summary April 4th, 2014 (1) Submitter Information Name: Gynius AB Address: Gynius AB Dobelnsgatan 23 11140, Stockholm, Sweden Telephone #: +46 700237670 Email: info@gynocular.com Contact Person: Matthew Volsky CEO - Gynius AB (2) Name of Device Trade Name: Gynocular Common Name: Colposcope Classification: Colposcope, 21 CFR 884.1630, Class II Product Code: HEX (3) Predicate devices | Applicant | Medgyn<br>Products,<br>Inc. | Intermed<br>Group Inc. | Medgyn<br>Products,<br>Inc. | Wallach<br>Surgical<br>Devices, Inc. | |-------------|----------------------------------------------------|------------------------------------|-----------------------------|--------------------------------------| | Device Name | Digital Colposcope &<br>Image Management<br>System | The Intermed<br>Zoom<br>Colposcope | Gynescope | Wallach Zoom<br>Colopscope | | Model | AL-106 | IZM-310 | AL 103 | ColpoStar IH | | K-Number | K122973 | K031629 | K880195 | K853389 | (4) Device Description The Gynocular is a hand-held, colposcope, which provides a magnified visualization of the tissues of the cervix, vagina, vulva and anogential area. It is used to evaluate these tissues and select areas for biopsy, without contacting the patient. The user is a medical physician, or a trained colposcopist. The Gynocular is a colposcope with similar specifications to traditional colposcopes. The Gynocular has a 300 mm focal distance, 3-magnifications: 5x, 8x, and 12x. It is a handheld device that comes with a tripod-mounting clip that screws into a standard tripod enabling the 510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670 1 {1}------------------------------------------------ medical professional to also perform colposcopy hands-free mode for ease of biopsy. The Gynocular uses high intensity LEDs for illumination, has a green filter, and is powered by a rechargeable lithium-ion battery. We also provide a charging base for the Gynocular. (5) Indications for use (colposcopy): The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy. (6) Comparison to Predicate Devices | Applicant | Medgyn<br>Products, Inc. | Intermed<br>Group Inc. | Medgyn<br>Products,<br>Inc. | Wallach<br>Surgical<br>Devices, Inc. | Gynius AB | |---------------------|---------------------------------------------------|---------------------------------------|--------------------------------|--------------------------------------|-------------------------------------| | Device Name | Digital<br>Colposcope<br>&<br>Image<br>Management | The<br>Intermed<br>Zoom<br>Colposcope | Gynescope | WALLACH<br>ZOOM<br>COLPOSCO<br>PE | The<br>Gynocular | | Model | AL-106 | IZM-310 | AL 103 | ColpoStar IH | VI | | K-Number | K122973 | K031629 | K880195 | K853389 | n/a | | Magnification | Optical<br>Display<br>;<br>36x | Standard 10x | 4.3x/7x/11.<br>3x | 12x | 5x/8x/12x | | Working<br>distance | 200-300 mm | 310 mm | 300 mm | 300 mm | 300 mm | | Lamp | LEDs | Halogen | Halogen | Halogen | LEDs | | Filter | Green filter | Green filter | Green filter | Green filter | Green filter | | Ocular type | Monocular | Binocular | Binocular | Binocular | Monocular | | Stand | Stationary | Stationary | Stationary | Stationary | Portable/stationary on a<br>tripod. | | Energy<br>source | Standard AC<br>Power Supply | Standard AC<br>Power<br>Supply | Standard AC<br>Power<br>Supply | Standard AC<br>Power<br>Supply | Lithium-ion<br>Battery | The Gynocular's performance specifications are substantially equivalent to the listed predicate devices. #### (7) Performance Data . The following testing was conducted to support a substantial equivalence determination. - . Distortion testing: - The analysis shows that the Gynocular exhibits a minimal Barrel type geometrical distortion: 0.47%, 0.75% and 0.81% for 5x, 8x and 12x magnification, respectively. This distortion is at levels of professional camera lenses considered and is considered negligible for a visual examination of tissue. The analysis shows that the Gynocular exhibits a 510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670 2 {2}------------------------------------------------ negligible chromatic distortion: 3.3 um, 5.8 um and 3.3 um for 5x, 8x and 12x magnification, respectively. This chromatic distortion is equal and below the levels of visual acuity of the human eye (5-6 um at the retina). - . Magnification testing: 4.81x ±0.15,7.74x ±0.15,12.40x ±0.15 - � Optical resolution: 5x = 28.6 cy/mm 8x = 40.5 cv/mm 12x = 56.8cy/mm Design validation: . Firmware and mechanical tests were conducted to demonstrate that the mechanical functions and optical properties achieve the desired results. - . Thermal safety: After 1hr 55 min of continuous use, the Gynocular's outer heat sink reached 39.1 degrees C in an ambient temperature setting of 24 degrees C. The Gynocular has an automatic shut off mechanism if temperatures become unsafe. - Battery Performance: . The typical charging time from empty to full charge is 2.5 hours or less. The battery gives approximately 2hrs of illuminator on time (full intensity) from a full charge. The use of life for the battery is 500 full charging cycles. #### Non-clinical tests Appropriate risk analysis-driven product testing was conducted to evaluate product safety. The following standards were met: - · IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) - · EN 60601-1 2006 #### Clinical tests Gynius conducted a randomized, crossover, pilot clinical trial for evaluating agreement of diagnosis of cervical lesions by colposcopy using a standard colposcope and the Gynocular. The study population was 69 women that tested positive for visual inspection with acetic acid. Swede scores were given at the time of colposcopy and compared with the final histological diagnosis after directed cervical biopsy. The Swede scores were compared across the two techniques and there was a 70% agreement overall. Conclusions: The study shows that in visual inspection with acetic acid positive women, a battery-driven, pocket-sized colposcope - the Gynocular, has a significant level of agreement with a stationary colposcope in assessing cervical lesions. The study was accepted and published by The Journal of Lower Genital Tract Disease in May 2013. #### (8) Conclusions The Gynocular is equivalent in safety and efficacy to the legally marketed predicate devices. 510(k) Submission Gynius AB P.O. Box 55503 10204, Stockholm Sweden info@gynocular.com T: 011 46 700237670 {3}------------------------------------------------ Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, with a double-stranded snake winding around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the emblem. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 April 8, 2014 Gynius AB Matthew Volsky CEO Dobelnsgatan 23 Stockholm 11140 Sweden Re: K132423 > Trade/Device Name: The Gynocular Regulation Number: 21 CFR§ 884.1630 Regulation Name: Colposcope Regulatory Class: II Product Code: HEX Dated: March 6, 2014 Received: March 10, 2014 Dear Matthew Volsky, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not mislcading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {4}------------------------------------------------ Page 2 - Matthew Volsky You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Sincerely yours, Herbert P.Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ . GYNIUS AB 510(k) submission Gynocular™ Colposcope ## 4. Indications for Use 510(k) Number: K132423 Device Name: The Gynocular Indications for Usc: The Gynocular is intended to provide magnified visualization of the tissues of the vagina, cervix and external genitalia. It is used to evaluate these tissues and select areas for biopsy. Prescription for Use: X (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Herbert P 2014.04.08 13:36: 510(k) Submission Gynius AB P.O. Box 55503 1 0204, Stockholm Sweden info@gynocular.com T: 011 46 700237670
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...