The ISE module of the Roche/Hitachi systems is intended for the quantitative determination of sodium, potassium, and chloride in serum, plasma, or urine using ion-selective electrodes. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
Device Story
The cobas c 501 ISE module performs quantitative electrolyte analysis (Na, K, Cl) in serum, plasma, or urine. The system utilizes ion-selective electrode (ISE) potentiometry; electrodes develop an electrical potential (EMF) proportional to ion concentration, governed by the Nernst equation. The device requires specific auxiliary reagents: reference electrolyte, diluent, internal standard, and cleaning solution. The primary modification involves changing the ISE calibrator 3 (S3) from serum-based to aqueous-based. The system is operated by laboratory personnel in a clinical chemistry setting. The analyzer automatically processes samples, including auto-dilution for out-of-range urine samples. Results are provided to clinicians to assist in diagnosing and monitoring electrolyte imbalances and metabolic disorders. The device benefits patients by providing rapid, accurate electrolyte measurements necessary for managing critical conditions like dehydration, hypertension, and cystic fibrosis.
Clinical Evidence
Bench testing only. Precision studies (within-run and intermediate) performed over 21 days (n=84 per analyte) showed %CVs generally <2%. Linearity studies (CLSI EP6-A) confirmed reportable ranges for serum and urine. Method comparison studies against reference methods (Flame Photometer, Coulometry) and the predicate device showed high correlation (r ≥ 0.998). Interference studies confirmed robustness against common drugs and endogenous substances (bilirubin, hemoglobin, lipemia). No clinical studies were required.
Technological Characteristics
Ion-selective electrode (ISE) potentiometry; utilizes selective membranes to generate EMF based on Nernst equation. Integrated into cobas c 501 analyzer. Reagents stored in plastic bottles at 5-12 °C. Calibration uses aqueous ISE Standards (Low and High).
Indications for Use
Indicated for quantitative determination of sodium, potassium, and chloride in serum, plasma, or urine. Used for diagnosis/treatment of electrolyte imbalances, aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate ADH secretion, cystic fibrosis, and diabetic acidosis.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
cobas c 501 analyzer ISE (k053165)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k132418
B. Purpose for Submission:
Modification of a previously cleared device (k053165)
C. Measurand:
Sodium, Potassium, Chloride
D. Type of Test:
Ion Selective Electrode Potentiometry
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
cobas c 501 ISE Indirect Na, K, Cl for Gen. 2
G. Regulatory Information:
1. Regulation section:
21CFR 862.1665, Ion Specific Electrode, Sodium
21CFR 862.1600, Ion Specific Electrode, Potassium
21CFR 862.1170, Ion Specific Electrode, Chloride
2. Classification:
Class II
3. Product code:
JGS
CEM
CGZ
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indication(s) for use below
2. Indication(s) for use:
The ISE module of the Roche/Hitachi systems is intended for the quantitative determination of sodium, potassium, and chloride in serum, plasma, or urine using ion-selective electrodes.
Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
cobas c 501
I. Device Description:
The main modification of the device is the ISE calibrator 3 (S3) was changed from serum based to aqueous based and the concentrations of the Na, K, and Cl were made to be the same as ISE calibrator 2 (S2). The following ISE calibrators, auxiliary reagents, and electrodes are required to calibrate and calculate results for the ISE module and were used in the performance studies of the cobas c 501 ISE system.
ISE Calibrators S1, S2, and S3:
- S1: ISE Standard Low (120 mmol/L Na⁺, 3 mmol/L K⁺, 80 mmol/L Cl⁻)
- S2: ISE Standard High (160 mmol/L Na⁺, 7 mmol/L K⁺, 120 mmol/L Cl⁻)
2
{2}
S3: ISE Standard High (160 mmol/L Na⁺, 7 mmol/L K⁺, 120 mmol/L Cl⁻)
## Auxiliary Reagents:
- ISE Reference Electrolyte (1 mol/L potassium chloride) provides a strong stable ion reference potential in the reference electrode necessary for each ISE measurement.
- ISE Diluent (HEPES buffer: 10 mmol/L, Triethanolamine: 7 mmol/L, preservative) is used for sample dilution.
- ISE Internal Standard (HEPES buffer: 10 mmol/L, Triethanolamine: 7 mmol/L, Sodium Chloride: 3.06 mmol/L, Sodium acetate: 1.45 mmol/L, Potassium Chloride: 0.16 mmol/L, preservative) is used for a baseline calibration which is performed once every ISE cycle.
- ISE Cleaning Solution: Sodium hydroxide solution (12%) is intended to clean the ion-selective electrodes, vessel and tubing.
ISE Electrodes: Sodium Electrode, Potassium Electrode, Chloride Electrode, and Reference Electrode
## J. Substantial Equivalence Information:
1. Predicate device name(s): cobas c 501 analyzer ISE
2. Predicate 510(k) number(s): k053165
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate device cobas c 501 ISE Indirect Na, K, Cl for Gen. 2 | Predicate device cobas c 501 analyzer ISE (k053165) |
| Intended Use | The ISE module of the Roche/Hitachi cobas c system is intended for the quantitative determination of sodium, potassium and chloride in serum, plasma or urine using ion-selective electrodes. | Same |
| Measurement Principle | ISE Potentiometry | Same |
| Onboard storage temperature | 5-12°C | Same |
| Ion Selective Electrodes | Potentiometric chloride, potassium, sodium and reference electrodes. | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate device cobas c 501 ISE Indirect Na, K, Cl for Gen. 2 | Predicate device cobas c 501 analyzer ISE (k053165) |
| Sample Dilution | 1:31 | Same |
| Reportable Range for Serum and Plasma | Sodium: 80-180 mmol/L Potassium: 1.5-10 mmol/L Chloride: 60-140 mmol/L | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate device | Predicate device |
| Standard 3 (High) Calibrator | ISE standard high (aqueous) | S3 (serum based compensator) |
| Reportable Range for Urine | Sodium: 20-250 mmol/L Potassium: 3-100 mmol/L Chloride: 20-250 mmol/L | Sodium: 10-250 mmol/L Potassium: 1-100 mmol/L Chloride: 10-250 mmol/L |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision of Clinical Chemistry Devices, Approved Guideline- $2^{\mathrm{nd}}$ Edition
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition.
# L. Test Principle:
The cobas c 501 with and ISE module is an Ion-Selective Electrode (ISE) system for the determination of sodium, potassium, and chloride in serum, plasma, and urine. An ISE makes use of the unique properties of certain membrane materials to develop and electrical potential (electromotive force, EMF) for the measurements of ions in solution. The electrode has a selective membrane in contact with both the test solution and an intimal filling solution. The internal filling solution contains the test ion at fixed concentration. Because of the particular nature of the membrane, the test joins will closely associate with the membrane on each side. The membrane EMF is determined by the difference in concentration of the test ion in the test solution and the internal filling solution. The EMF develops according to the Nernst equation for a specific ion in solution.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies for serum and urine applications of the cobasc501 ISE electrodes for the measurement of Na, K, and Cl were performed using 1 analyzer, 1 site, 1 lot of reagent, and 6 samples. The 6 samples used for the serum precision testing were comprised of 4 human serum sample pools of varying concentrations (low, mid, mid-low, and high) for each analyte; and Precinorm U and Precipath U controls. The 6 samples used for the urine precision testing were comprised of 4 human urine sample pools of varying concentrations (low, mid, mid-low, and high) for each analyte; and Liquichek Level 1 and Liquichek Level 2 controls. The within run precision study was performed with 21 single determinations for each analyte. The intermediate precision was performed over a period of 21 days, 2 runs per day, and 2 aliquots per run for a total of 84 samples per analyte. The result summaries for the precision studies are presented in the tables below:
Within Run Precision
Sodium (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 125.6 | 147.5 | 84.4 | 121.3 | 131.7 | 177.2 |
| SD | 0.3 | 0.4 | 0.2 | 0.2 | 0.3 | 0.4 |
| %CV | 0.2 | 0.3 | 0.3 | 0.2 | 0.2 | 0.2 |
Sodium (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 80.2 | 169.2 | 29.9 | 131.3 | 22.7 | 236.1 |
| SD | 0.2 | 0.2 | 0.3 | 0.2 | 0.3 | 0.2 |
| %CV | 0.2 | 0.2 | 0.3 | 0.2 | 0.3 | 0.2 |
Potassium (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 3.58 | 6.78 | 1.62 | 4.99 | 2.64 | 9.49 |
| SD | 0.02 | 0.04 | 0.01 | 0.02 | 0.01 | 0.05 |
| %CV | 0.6 | 0.6 | 0.6 | 0.4 | 0.4 | 0.5 |
{5}
Potassium (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 31.7 | 70.6 | 5.25 | 51.90 | 15.3 | 89.80 |
| SD | 0.2 | 0.3 | 0.03 | 0.2 | 0.1 | 0.4 |
| %CV | 0.6 | 0.4 | 0.6 | 0.4 | 0.7 | 0.5 |
Chloride (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 85.2 | 117.4 | 67.6 | 128.6 | 91.6 | 139.0 |
| SD | 0.3 | 0.4 | 0.2 | 0.3 | 0.3 | 0.4 |
| %CV | 0.4 | 0.3 | 0.2 | 0.2 | 0.4 | 0.3 |
Chloride (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 21 | 21 | 21 | 21 | 21 | 21 |
| Mean | 95.2 | 195.4 | 25.0 | 130.6 | 23.4 | 242.9 |
| SD | 0.3 | 0.5 | 0.1 | 0.4 | 0.1 | 0.7 |
| %CV | 0.3 | 0.2 | 0.4 | 0.3 | 0.3 | 0.3 |
## Total (Intermediate) Precision
Sodium (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 126.0 | 148.2 | 84.8 | 121.4 | 131.6 | 176.7 |
| SD | 0.7 | 0.5 | 1.0 | 0.8 | 0.7 | 0.6 |
| %CV | 0.5 | 0.4 | 1.1 | 0.6 | 0.5 | 0.4 |
Sodium (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 81.6 | 172.3 | 30.6 | 131.7 | 23.3 | 236.7 |
| SD | 1.3 | 2.6 | 0.9 | 0.6 | 0.9 | 1.3 |
| %CV | 1.6 | 1.5 | 3.0 | 0.5 | 3.8 | 0.6 |
{6}
Potassium (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 3.57 | 6.59 | 1.62 | 4.97 | 2.63 | 9.46 |
| SD | 0.04 | 0.05 | 0.03 | 0.04 | 0.07 | 0.07 |
| %CV | 1.0 | 0.7 | 1.6 | 0.8 | 0.7 | 1.0 |
Potassium (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 31.5 | 70.6 | 5.2 | 52.1 | 15.4 | 90.3 |
| SD | 0.5 | 1.17 | 0.04 | 0.7 | 0.1 | 1.4 |
| %CV | 1.7 | 1.7 | 0.7 | 1.3 | 0.9 | 1.5 |
Chloride (Serum)
| Sample | PNU | PPU | HS low | HS med | HS MDL | HS high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 86.2 | 119.2 | 68.5 | 129.0 | 92.3 | 139.0 |
| SD | 0.5 | 0.5 | 0.6 | 0.6 | 0.5 | 0.6 |
| %CV | 0.6 | 0.4 | 0.8 | 0.5 | 0.6 | 0.4 |
Chloride (Urine)
| Sample | Liq 1 | Liq 2 | HU low | HU med | HU MDL | HU high |
| --- | --- | --- | --- | --- | --- | --- |
| Unit | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L | mmol/L |
| n | 84 | 84 | 84 | 84 | 84 | 84 |
| Mean | 97.5 | 198.2 | 25.8 | 131.4 | 24.3 | 243.4 |
| SD | 1.6 | 2.3 | 0.6 | 0.7 | 0.6 | 1.8 |
| %CV | 1.6 | 1.2 | 2.3 | 0.5 | 2.4 | 0.7 |
b. Linearity/assay reportable range:
Linearity studies were performed according to CLSI EP6-A. Dilution series of 11 concentrations were prepared using low and high human serum, plasma, and urine sample pools for each of the analytes and tested in triplicate. Linear regression analysis was done according to EP6-A. Linear regression summary results of the study are presented in the table below:
{7}
| Analyte | Matrix | Slope | Intercept | Range Tested (mmol/L) |
| --- | --- | --- | --- | --- |
| Sodium | Serum | 0.9804 | 2.7508 | 70-190 |
| Sodium | Plasma | 1.0076 | 0.8378 | 70-185 |
| Sodium | Urine | 1.0100 | 2.1287 | 15-250 |
| Potassium | Serum | 0.9978 | 0.0437 | 1.5-11.0 |
| Potassium | Urine | 0.9781 | 0.3260 | 3.0-110 |
| Chloride | Serum | 1.0690 | 1.1321 | 60-145 |
| Chloride | Plasma | 0.9788 | 3.4818 | 60-140 |
| Chloride | Urine | 0.9918 | 2.4216 | 15-250 |
The plasma potassium linearity study resulted in a polynomial regression in which the $3^{\mathrm{rd}}$ order polynomial regression was considered significant. Therefore, the deviation of the results from the $3^{\mathrm{rd}}$ order linear regression were compared against the best fit line. The plasma potassium polynomial regression results are summarized in the following table:
| Sample | Best line regression mmol/L | Predicted 3rdorder mmol/L | Absolute difference | Relative % difference |
| --- | --- | --- | --- | --- |
| 1 | 0.948 | 0.993 | 0.045 | 4.7418 |
| 2 | 2.1783 | 2.1786 | 0.0003 | 0.0158 |
| 3 | 3.4086 | 3.386 | -0.0226 | -0.6628 |
| 4 | 4.6388 | 4.6101 | -0.0288 | -0.62 |
| 5 | 5.8691 | 5.8461 | -0.0231 | -0.3928 |
| 6 | 7.0994 | 7.089 | -0.0104 | -0.1461 |
| 7 | 8.3297 | 8.334 | 0.0044 | 0.0526 |
| 8 | 9.5599 | 9.5763 | 0.0163 | 0.1706 |
| 9 | 10.7902 | 10.8107 | 0.0205 | 0.902 |
| 10 | 12.0205 | 12.0326 | 0.0121 | 0.1007 |
| 11 | 13.2508 | 13.2369 | -0.0138 | -.1044 |
{8}
Urine auto-rerun dilution validation study:
An auto-rerun validation study was performed to evaluate the accuracy of the analyzer’s auto dilution function to obtain results for urine samples with analyte concentrations above the measuring range. Three samples (low, middle, and high) of the extended range were used for the study. Each sample was manually diluted with 1:1.5 ratio using a sample volume of 9.7 µL. Each sample was measured in triplicate on one cobas c 501 analyzer which uses a decreased sample volume dilution method. The results from the manually diluted urine samples were compared to the results from the auto-rerun dilution samples. The study protocol and pre-determined acceptance criteria were reviewed and found to be adequate. The results of the auto-rerun study support the sponsor’s extended measuring range claims for urine sodium (250 to 375 mmol/L), urine potassium (100 to 150 mmol/L) and urine chloride (100 to 150 mmol/L) samples.
The results of the linearity studies and the auto-rerun validation for urine samples support the sponsor’s following claimed measuring ranges:
Serum Sodium: 80-180 mmol/L
Serum Potassium: 1.5-10.0 mmol/L
Serum Chloride: 60-140 mmol/L
Urine Sodium: 20-250 mmol/L (auto-dilution extended range: 251-375 mmol/L)
Urine Potassium: 3-100 mmol/L (auto-dilution extended range: 101-150 mmol/L)
Urine Chloride: 20-250 mmol/L (auto-dilution extended range: 251-375 mmol/L)
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The ISE Standard Low (S1) and ISE Standard High (S2 and S3) are standardized against primary calibrators prepared gravimetrically from purified salts. The Sodium Chloride, Potassium Dihydrogenphosphate and Sodium Hydrogencarbonate salts with a minimum of 99.5% purity and USP grade were used.
Stability:
Shelf life: The ISE Standard Low (S1) and ISE Standard High (S2 and S3) are stable at 15-25°C for 60 months or until the expiration date listed on the labeling. Open vial: The ISE standards are intended to be used immediately after opening and are intended to be used for one calibration only.
Value assignment:
The ISE standard low is prepared by dissolving 1870.32 g of Sodium Chloride, 163.30 g of Potassium dihydrogenphosphate, and 1344.00 g of Sodium hydrogencarbonate in 400.00 kg of water.
9
{9}
The ISE standard high is prepared by dissolving 2805.50 g of Sodium Chloride, 381.05 g of Potassium dihydrogenphosphate, and 1344.00 g of Sodium hydrogencarbonate in 400.00 kg of water.
All standard solutions are checked against the reference methods and must meet the internal specifications before released. The expected values are as follows:
S1: ISE Standard Low (120 mmol/L Na⁺, 3 mmol/L K⁺, 80 mmol/L Cl⁻)
S2: ISE Standard High (160 mmol/L Na⁺, 7 mmol/L K⁺, 120 mmol/L Cl⁻)
S3: ISE Standard High (160 mmol/L Na⁺, 7 mmol/L K⁺, 120 mmol/L Cl⁻)
All 3 standards are used for the full calibration, S3 is considered as a compensator. The slope of the calibration curve is calculated from Standards 1 and 2; the ISE compensation will affect the intercept, not the slope.
d. Detection limit:
Studies were carried out in accordance with CLSI Guidance Document EP17-A for Sodium, Potassium and Chloride analytes. For determination of LoB one analyte free sample was measured in 5 replicates, 6 runs, 3 days, on 2 cobas c 501 analyzers. Total of 60 measurements were obtained per analyzer. For determination of LoD, five samples (one replicate) with low-analyte concentration were measured in 6 runs for 3 days on 2 cobas c 501 analyzer modules. In total 60 measurements were obtained per analyzer. For LoQ studies a low level sample set was prepared by diluting 3 human urine and serum samples with an analyte free diluent (ISE Diluent). The low level sample set was tested in single replicate for 3 days in 2 runs per day on two cobas c 501 analyzers. LoQ is defined as the concentration where total error is less than 20%. Results from the detection limit studies are summarized in the table below:
| | LoB (mmol/L) | LoD (mmol/L) | LOQ (mmol/L) | Claimed measuring range (mmol/L) |
| --- | --- | --- | --- | --- |
| Sodium (serum) | 3.9 | 5.7 | 9.7 | 80-180 |
| Potassium (serum) | 0.17 | 0.2 | 0.3 | 1.5-10 |
| Chloride (serum) | 3.4 | 4.7 | 6.7 | 60-140 |
| Sodium (urine) | 3.9 | 5.7 | 11.4 | 20-375 |
{10}
# e. Analytical specificity:
Drug interference: Serum and urine sample pools with low and high concentrations of each analyte were spiked with drugs and concentrations listed below. The analyte concentrations of the spiked aliquots were tested in triplicate and the mean of the triplicate determinations is compared to the analyte concentration of the reference sample which contains no drugs. The sponsor's definition of significant interference is $\geq \pm 10\%$ of the initial value.
Approximate Concentration of Analytes tested for drug interference
| Analyte | Matrix | Low (mmol/L) | High (mmol/L) |
| --- | --- | --- | --- |
| Sodium | Serum | 123 | 150 |
| Sodium | Urine | 60 | 215 |
| Potassium | Serum | 3.80 | 6.50 |
| Potassium | Urine | 20 | 95 |
| Chloride | Serum | 95 | 105 |
| Chloride | Urine | 105 | 225 |
There was no significant interference for serum sodium, potassium, and chloride when these analytes and interferents were tested in the concentrations indicated below.
Drug Interference for Serum Sodium, Potassium and Chloride
| Drug | Concentration at which no significant interference was observed (mg/dL) |
| --- | --- |
| Acetylcystein | 150 |
| Ampicillin-Na | 1000 |
| Ascorbic acid | 300 |
| Cyclosporin | 5 |
| Na-Cefoxitin | 2500 |
| Heparin | 5000 U |
| Intralipid | 10000 |
| Levodopa | 20 |
| Methyldopa | 20 |
| Metronidazole | 200 |
| Phenylbutazone | 400 |
{11}
There was no significant interference for urine sodium, potassium, and chloride when these analytes and interferents were tested in the concentrations indicated below.
Drug Interference for Urine Sodium, Potassium and Chloride
| Drug | Concentration at which no significant interference was observed (mg/dL) |
| --- | --- |
| Acetaminophen | 3000 |
| N-Acetyl cystein | 10 |
| Salicyurlic acid | 6000 |
| Ascorbic acid | 4000 |
| Na-Cefoxitin | 120000 |
| Gentamycin Sulfate | 400 |
| Ibuprofen | 4000 |
| Levodopa | 1000 |
| Methyldopa | 2000 |
| Metronidazole | 200 |
| Ofloxacin | 900 |
| Phenylbutazone | 300 |
| Tetracycline | 300 |
| Rifampicin | 60 |
# Endogenous Interference
Serum sample pools with low and high concentrations of each analyte were spiked with conjugated and unconjugated bilirubin, hemoglobin, and intralipids in the concentrations listed below. Urine sample pools with low and high concentrations of each analyte were spiked with conjugated bilirubin and hemoglobin. The analyte concentrations of the spiked aliquots were tested in triplicate and the mean of the triplicate determinations is compared to the analyte concentration of the reference sample which contains no endogenous substances. The sponsor's definition of significant interference is $\geq \pm 10\%$ of the initial value.
{12}
Serum and Urine Sodium, Potassium and Chloride Concentrations Tested
| | Analyte | Analyte Concentration Low (mmol/L) | Analyte Concentration High (mmol/L) |
| --- | --- | --- | --- |
| Interferent | Serum | | |
| Unconjugated bilirubin | Sodium | 126 | 147 |
| | Potassium | 2.83 | 5.99 |
| | Chloride | 92 | 111 |
| Conjugated bilirubin | Sodium | 125 | 148 |
| | Potassium | 2.98 | 6.29 |
| | Chloride | 95 | 115 |
| Hemoglobin | Sodium | 120 | 141 |
| | Potassium | 2.78 | 5.89 |
| | Chloride | 93 | 111 |
| Lipemia | Sodium | 113 | 133 |
| | Potassium | 2.65 | 6.62 |
| | Chloride | 84 | 101 |
| Interferent | Urine | | |
| Conjugated bilirubin | Sodium | 39 | 210 |
| | Potassium | 18.8 | 65.6 |
| | Chloride | 46 | 96 |
| Hemoglobin | Sodium | 35 | 198 |
| | Potassium | 18 | 63 |
| | Chloride | 47 | 190 |
Results Summary:
Serum: Concentration of interferent at which no significant interference was observed.
| Interferent | Sodium | Potassium | Chloride |
| --- | --- | --- | --- |
| Hemoglobin | 1000 mg/dL | Avoid hemolyzed specimens* | 1000 mg/dL |
| Conjugated Bilirubin | 60 mg/dL | 60 mg/dL | 60 mg/dL |
| Unconjugated Bilirubin | 60 mg/dL | 60 mg/dL | 60 mg/dL |
| Lipemia | 2000 mg/dL | 2000 mg/dL | 2000 mg/dL |
* Hemolysis interferes with potassium determinations.
Sponsor has the following limitation in the labeling:
"Hemoglobin levels higher than 90 mg/dL increase the apparent potassium concentrations significantly (appropriately H index 90). Potassium concentration in
{13}
erythrocytes is 25 times higher than in normal plasma. The level of interference may be variable depending on the exact content of erythrocytes. Avoid hemolyzed specimens"
Urine: Concentration of interferent at which no significant interference was observed.
| Interferent | Sodium | Potassium | Chloride |
| --- | --- | --- | --- |
| Hemoglobin | 1000 mg/dL | 100 mg/dL | 1000 mg/dL |
| Conjugated Bilirubin | 60 mg/dL | 60 mg/dL | 60 mg/dL |
f. Assay cut-off: Not applicable
## 2. Comparison studies:
a. Method comparison with predicate device:
Sodium, potassium and chloride values for human plasma and urine samples were obtained on the Roche/Hitachi cobas c 501 analyzers and where compared to those determined on the corresponding reference method and with the Roche/Hitachi 912. The reference methods used were: Flame Photometer IL 943 for Sodium and Potassium and the Chloride Analyzer 926S for Chloride. No more than 20% of the samples were either spiked or diluted to cover the measuring range. In addition, the samples were tested in singlicate on one cobas 501 with the candidate device and one c 501 with the predicate device. The results were calculated using Passing/Bablok Linear regression. The method comparison result summary is shown in the tables below:
Serum Sodium Method Comparison: Flame Photometer vs candidate device
| n | Range mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 103 | 87-178 | y=1.000x+0.300 | 0.999 |
Serum Sodium Method Comparison: Predicate device vs candidate device
| n | Range mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 103 | 87-176 | y=1.014x-1.176 | 1.000 |
Urine Sodium Method Comparison: Flame Photometer vs candidate device
| n | Range mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 100 | 24-250 | y=0.964+4.032 | 1.000 |
{14}
Urine Sodium Method Comparison:
Predicate device vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 100 | 25-245 | y=0.995x+0.687 | 1.000 |
Serum Potassium Method Comparison:
Flame Photometer vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 106 | 1.59-9.56 | y=1.007x-0.190 | 1.00 |
Serum Potassium Method Comparison:
Predicate device vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 106 | 1.52-9.49 | y=1.006x+0.024 | 1.00 |
Urine Potassium Method Comparison:
Flame Photometer vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 4.00-97.2 | y=1.018x-0.397 | 1.00 |
Urine Potassium Method Comparison:
Predicate device vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 4.05-97.4 | y=0.997x+0.062 | 0.999 |
Serum Chloride Method Comparison:
Coulmetry vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 62-136 | y=1.033x-1.800 | 0.998 |
Serum Chloride Method Comparison:
Predicate device vs candidate device
| n | Range
mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 61-138 | y=1.000x +0.50 | 0.999 |
{15}
Urine Chloride Method Comparison:
Coulometry vs candidate device
| n | Range mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 22-248 | y=1.020x -1.70 | 0.999 |
Urine Chloride Method Comparison:
Predicate device vs candidate device
| n | Range mmol/L | Regression Passing-Bablok | r |
| --- | --- | --- | --- |
| 105 | 21-250 | y=0.989x+0.669 | 1.000 |
# b. Matrix comparison:
A matrix comparison study was performed to compare the recovery of the proposed anticoagulated sample type, lithium heparin, to the recovery of serum sample type. Thirty matched serum (SST) and lithium heparin samples were run in singlet on one cobas c 501 analyzer, at one site, with one lot of ISE diluent. Testing was performed during one day. Passing/Bablok linear regression analysis was used to evaluate the mean data and the result summary is shown below:
Serum vs. Lithium Heparin Plasma
| Analyte | n | Range mmol/L | Slope | Intercept | Correlation Coefficient (r) |
| --- | --- | --- | --- | --- | --- |
| Sodium | 30 | 80-174 | 1.003 | -0.659 | 1.000 |
| Potassium | 30 | 1.77-8.98 | 0.9960 | 0.023 | 0.995 |
| Chloride | 30 | 64-136 | 1.009 | -0.754 | 0.999 |
Sponsor concluded that lithium heparin is an acceptable anti-coagulant to be used with the candidate device.
# 3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
# 4. Clinical cut-off: Not applicable
{16}
17
5. Expected values/Reference range:
The expected values for adult based serum, plasma and 24-hour urine output are cited from the literature¹:
Serum (Adults)
Sodium: 136-145 mmol/L
Potassium: 3.5-5.1 mmol/L
Chloride: 98-107 mmol/L
Plasma (Adults)
Sodium: 136-145 mmol/L
Potassium: 3.4-4.5 mmol/L
Chloride: 98-107 mmol/L
Urine, 24 hour (Adults)
Sodium: 40-220 mmol/24h
Potassium: 25-125 mmol/24h
Chloride: 110-250 mmol/24h
References:
1. Tietz Fundamentals of Clinical Chemistry, Fifth Edition, Edited by Carl A. Burlis and Edward R. Ashwood, W.B. Saunders Company, 2001:970, 1004,1009 (ISBN 0-7216-8634-6.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.