K132409 · Merit Medical Systems, Inc. · DQY · Aug 29, 2013 · Cardiovascular
Device Facts
Record ID
K132409
Device Name
CONCIERGE GUIDING CATHETER
Applicant
Merit Medical Systems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 29, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Concierge Guiding Catheter is intended for the intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Device Story
Concierge Guiding Catheter is a single-lumen catheter used by clinicians to facilitate the delivery of interventional or diagnostic devices into coronary or peripheral vasculature. Device consists of Pebax body, stainless steel wire braid reinforcement, PTFE lubricious inner lumen, and soft radiopaque tip. Available in 5F, 6F, 7F, and 8F sizes, 100cm length, and various shapes. Operates as a conduit for other devices; no active components or electronic processing. Used in clinical settings; provides a pathway for diagnostic/interventional tools to reach target vascular sites.
Clinical Evidence
Bench testing only. No clinical data. Testing included dimensions, air/liquid leak, tip support/attachment, tensile strength, shaft kink/stiffness, simulated use, packaging integrity (dye penetration, burst, seal peel), and biocompatibility (cytotoxicity, hemocompatibility, chemical characterization).
{0}------------------------------------------------
# K132409
# Section 5
; (
4
## 510(k) Summary ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| General<br>Provisions | Submitter Name: Merit Medical Systems, Inc.<br>Address: 1600 West Merit Parkway<br>South Jordan, UT 84095<br>Telephone Number: (801) 208-4160<br>Fax Number: (801) 826-4171<br>Contact Person: Vicki Godwin<br>Date of Preparation: August 1,2013<br>Registration Number: 1721504<br>AUG 2 9 2013 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | Trade Name: Concierge® Guiding Catheter<br>Common/Usual Name: Guiding Catheter<br>Classification Name: Percutaneous Catheter |
| Predicate<br>Device | Trade Name: Concierge® Guiding Catheter<br>Classification Name: Percutaneous Catheter<br>Premarket Notification: K121051<br>Manufacturer: Merit Medical Systems, Inc. |
| Classification | Class II<br>21 CFR § 870.1250<br>FDA Product Code: DQY<br>Review Panel: Cardiovascular |
| Intended Use | The Concierge Guiding Catheter is intended for the intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems. |
| Device<br>Description | The Concierge Guiding Catheter is a single lumen catheter that incorporates a Pebax body reinforced with stainless steel wire braid, a PTFE lubricious inner lumen, and a soft radiopaque tip. It is available in 5F, 6F, 7F and 8F sizes, 100cm long and is produced in a variety of shapes. |
| Comparison to<br>Predicate<br>Device | The technological characteristics of the subject Concierge Guiding<br>Catheters are substantially equivalent to those of the predicate<br>device. The 5F, 7F and 8F Concierge Guiding Catheter sizes are<br>modifications of the predicate 6F Concierge Guiding Catheter. In<br>addition, changes were made to the device material and packaging |
| Safety &<br>Performance<br>Tests | No performance standards have been established under Section 514<br>of the Food, Drug and Cosmetic Act for these devices. Performance<br>testing of the subject Concierge Guiding Catheter was conducted<br>based on the risk analysis and based on the requirements of the<br>following international standards:<br>ISO 10555-1:1995, Sterile, single-use intravascular<br>catheters - Part 1: General requirements ISO 10555-2:1996, Sterile, single-use intravascular<br>catheters - Part 2: Angiographic catheters ISO 11135-1: 2007, Sterilization of health care products<br>- Ethylene oxide - Part 1: Requirements for<br>development, validation and routine control of a<br>sterilization process for medical devices ISO 10993-1: 2009, Biological Evaluation of Medical<br>Devices Part-1: Evaluation and Testing within a risk<br>management process ISO 10993-5: 2009, Biological Evaluation of Medical<br>Devices Part-5: Tests for In Vitro Cytotoxicity ISO 10993-7: 2008, Biological Evaluation of Medical<br>Devices Part-7 Ethylene Oxide Sterilization Residuals ISO 10993-17: 2002, Biological Evaluation of Medical<br>Devices - Part 17: Establishment of allowable limits for<br>leachable substances ISO 10993-18: 2005, Biological Evaluation of Medical<br>Devices - Part 18: Chemical Characterization of<br>Materials ISO 2233: 2000: Packaging - complete, filled transport<br>packages and unit loads - conditioning and testing BS EN ISO 11607: 2009, Packaging for terminally<br>sterilized medical devices Part 1: Requirements for<br>materials, sterile barrier systems and packaging systems<br>(ISO 11607-1:2006) ASTM D4169-09: 2009, Standard practice for<br>performance testing of shipping containers and systems ASTM F756-08, 2008: Standard Practice for<br>Assessment of Hemolytic Properties of Materials ASTM F1929-98(2004): 1998, Standard test method for<br>detecting seal leaks in porous medical packaging by dye<br>penetration |
{1}------------------------------------------------
。 『
r
23
{2}------------------------------------------------
- ASTM F2096-11: 2011, Standard test method for ● detecting gross leaks in medical packaging by internal pressurization (bubble test)
- . ASTM F88/F88M-09: 2009 Standard test method for seal strength of flexible barrier materials
- ASTM F1140-07, Standard test methods for internal . pressurization failure resistance of unrestrained packages
The following is a list of all significant testing that was successfully completed:
#### Device Testing
- Dimensions ●
- Air Leak
- Liquid Leak .
- Catheter Tip Support and Attachment ●
- Tensile .
- Shaft Kink .
- . Shaft Stiffness
- . Simulated use testing
#### Packaging testing
- Visual Inspection .
- . Dye Penetration
- . Underwater Leak Test
- . Seal Peel Tensile Test
- Burst Strength Test �
- Aseptic Peel Opening .
#### Biocompatibility testing
- Cytotoxicity .
- Hemocompatibility
- . Chemical Characterization
The results of the testing demonstrated that the subject Concierge Guiding Catheters met the pre-determined acceptance criteria applicable to the safety and efficacy of the device.
Summary of Substantial Equivalence
Based on the indications for use, design, safety and performance testing, the subject Concierge Guiding Catheters meet the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the Concierge Guiding Catheter, manufactured by Merit Medical Systems, Inc.
Safety & Performance Tests cont.
ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of an eagle or bird with three wing-like shapes. The emblem is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
Merit Medical Systems, Inc. c/o Vicki Godwin 1600 West Merit Parkway South Jordan, Utah 84095
Re: K132409
Trade/Device Name: Concierge Guiding Catheter Regulation Number: 21 CFR § 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 1, 2013 Received: August 2, 2013
Dear Ms. Godwin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Vicki Godwin
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. L. Hillen
for Bram D. Zuckerman, M.D. Director of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Section 4
### Indications for Use
K132409 510(k) Number (if known):
Device Name: Concierge® Guiding Catheter
Indications for Use:
The Concierge® Guiding Catheter is intended for the intravascular introduction of interventional/diagnostic devices into the coronary or peripheral vascular systems.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over - The - Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Z. Hilleman
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.