ELITECH CLINICAL SYSTEMS CREATININE PAP SL, ELICAL 2, ELITROL I AND ELITROL II, URINE CONTROL BI-LEVEL
Applicant
Elitechgroup
Product Code
JFY · Clinical Chemistry
Decision Date
Jan 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K132399 · Jan 3, 2014
ELITECH CLINICAL SYSTEMS CREATININE PAP SL, ELICAL 2, ELITROL I AND ELITROL II, URINE CONTROL BI-LEVEL
Elitechgroup
Patient serum samples; Patient urine samples
Patient samples were used in method comparison studies to evaluate the performance of the ELITech Clinical Systems CREATININE PAP SL reagent against a predicate device (Roche Diagnostics CREP2) on the Selectra Pro Series Analyzer.
ELITech Clinical Systems CREATININE PAP SL is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine on ELITech Clinical Systems Selectra Pro Series Analyzers. It is not intended for use in Point of Care settings. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers. ELITech Clinical Systems ELITROL I and ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers. ELITech Clinical Systems URINE CONTROL BI-LEVEL is a set of 2 levels of urine controls used for in vitro diagnostic in the quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers.
Device Story
Enzymatic colorimetric assay for quantitative creatinine determination in human serum, plasma, and urine; utilizes Selectra ProM analyzer. Principle: Creatininase hydrolyzes creatinine to creatine; creatine hydrolyzed to sarcosine and urea; sarcosine oxidized to H2O2; H2O2 reacts with 4-amino-antipyrine and EHSPT via peroxidase to form colored quinoneimine. Absorbance measured at 546 nm; proportional to creatinine concentration. Used in clinical laboratories; operated by trained laboratory personnel. Output provides creatinine concentration values for clinical decision-making regarding renal function and dialysis monitoring.
Clinical Evidence
Bench testing only. Precision study (n=80 per pool) showed total CV% 1.5-2.9%. Linearity confirmed 0.1-30 mg/dL (serum) and 5-450 mg/dL (urine). Method comparison (n=100 serum, n=54 urine) against Roche Creatinine plus v2 showed high correlation (r=1.000). Interference testing performed per CLSI EP-7A. No clinical data.
Technological Characteristics
Enzymatic colorimetric assay. Reagents: MOPS buffer, creatinase, sarcosine oxidase, ascorbate oxidase, 4-aminoantipyrine, peroxidase, EHSPT. Form factor: Kit for automated clinical chemistry analyzers (Selectra Pro Series). Calibration: Multi-parametric, traceable to ID-MS. Controls: Lyophilized human serum or liquid urine. Connectivity: Integrated with Selectra Pro Series Analyzers.
Indications for Use
Indicated for quantitative in vitro diagnostic determination of creatinine in human serum, plasma, and urine to aid in diagnosis/treatment of renal diseases and monitoring renal dialysis. Not for Point of Care settings. Prescription use only.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Roche Diagnostics Creatinine plus ver. 2 (k024098)
Roche Calibrator for Automated Systems (C.f.a.s.) (k033501)
Roche Precinorm and Precipath (k041227)
Biorad Liquichek Urine Chemistry Control Level 1 and Level 2 (k020817)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k132399
B. Purpose for Submission:
New device
C. Measurand:
Creatinine
D. Type of Test:
Enzymatic, colorimetric, quantitative
E. Applicant:
ELITechGroup, Inc.
F. Proprietary and Established Names:
ELITech Clinical Systems CREATININE PAP SL
ELITech Clinical Systems ELICAL 2
ELITech Clinical Systems ELITROL I and ELITROL II
ELITech Clinical Systems URINE CONTROL BI-LEVEL
G. Regulatory Information:
| Regulation | Classification | Product Code | Panel |
| --- | --- | --- | --- |
| 21 CFR 862.1225
Creatinine Test system | Class II | JFY | Chemistry (75) |
| 21 CFR 862.1150
Calibrator | Class II | JIX | Chemistry (75) |
| 21 CFR 862.1660 Quality Control material (assayed and unassayed) | Class I, reserved | JJY | Chemistry (75) |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
ELITech Clinical Systems CREATININE PAP SL is intended for the quantitative in vitro diagnostic determination of creatinine in human serum, plasma and urine on ELITech Clinical Systems Selectra Pro Series Analyzers. It is not intended for use in Point of Care settings.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers.
ELITech Clinical Systems ELITROL I & ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers.
ELITech Clinical Systems URINE CONTROL BI- LEVEL is a set of 2 levels of urine controls used for in vitro diagnostic in the quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Pro Series Analyzers.
3. Special conditions for use statement(s):
In vitro diagnostic use only
The test system is not for use in Point of Care settings
4. Special instrument requirements:
ELITech Clinical Systems Selectra ProM Analyzer
I. Device Description:
ELITech Clinical Systems CREATININE PAP SL is available as kit only. It consists of a bi-reagent reagent 1 (R1) and reagent 2 (R2). R1 consists of; MOPS buffer (pH 7.50), EHSPT (N-Ethyl-N-(2-Hydroxy-3-Sulfopropyl)-m-Toluidine), creatinase, sarcosine oxidase, and ascorbate oxidase. R2 consists of: MOPS buffer (pH 7.50), 4-Aminoantipyrine, creatininase, peroxidase, and sodium azide.
2
{2}
ELITech Clinical Systems ELICAL2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration. ELICAL 2 is prepared from the blood of donors tested individually and found to be negative for HbsAg and to the antibodies to HCV and HIV according to FDA-approved methods.
ELITech Clinical Systems ELITROL I and ELITROL II are two levels of quality control products consisting of a lyophilized human serum containing constituents at desired levels. ELITROL I and ELITROL II are prepared from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods.
ELITech Clinical Systems URINE CONTROL BI-LEVEL is a liquid solution prepared from human urine supplemented with constituents of human and animal origin, chemicals, preservatives and stabilizers. Human sera corresponding to the URINE CONTROL BI-LEVEL were tested for each urine donor and found to be negative for HbsAg and antibodies to HCV and HIV-1/HIV-2 according to FDA-approved methods.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
k024098 – Roche Diagnostics Creatinine plus ver. 2
k033501 – Roche Calibrator for Automated Systems (C.f.a.s.)
k041227 – Roche Precinorm and Precipath
k020817 – Biorad Liquichek Urine Chemistry Control Level 1 and Level 2
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Items | Candidate Device | Predicate Device k024098 |
| | ELITECH Clinical Systems CREATININE PAP SL | Roche Diagnostics Creatinine plus ver. 2 |
| Intended use/Indication for use | Same | For the quantitative determination of the creatinine concentration in serum, plasma, and urine |
| Test method | Same | Enzymatic colorimetric |
| Sample type | Serum, plasma (lithium heparin), and urine | Serum, plasma (lithium heparin, Potassium EDTA), and urine |
| Measuring range | Serum/plasma: 0.10 – 30 mg/dL
Urine: 5 – 450 mg/dL | Serum/plasma: 0.06 – 30.5 mg/dL
Urine: 1.1 – 452 mg/dL |
{3}
| Similarities and Differences | | |
| --- | --- | --- |
| Items | Candidate Device | Predicate Device k033501 |
| | ELITECH Clinical Systems ELICAL 2 | Roche Calibrator for Automated Systems (C.f.a.s.) |
| Intended use/Indication for use | Same | For use in the calibration of quantitative methods on clinical chemistry analyzers |
| Matrix | Same | Lyophilized serum |
| Level(s) | Same | Single level |
| Number of analytes | 21 | 30 |
| Stability | Same | Lyophilized: To store at 2-8°C and protected from light until the expiry date After reconstitution, the stabilities are : - 8 hours between 15-25 °C. - 2 days between 2-8 °C. - 4 weeks between -25 and -15 °C (when frozen once) |
| Similarities and Differences | | |
| --- | --- | --- |
| Items | Candidate Device | Predicate Device k041227 |
| | ELITECH Clinical Systems ELITROL I and ELITROL II | Roche Diagnostics Precinorm U and Precipath U |
| Intended use/Indication for use | Same | For use in quality control by monitoring accuracy and precision for quantitative methods |
| Matrix | Same | Lyophilized serum |
| Levels | Same | 2 |
| Stability | Same | Lyophilized: To store at 2-8°C and protected from light until the expiry date After reconstitution, the stabilities are : - 12 hours between 15-25 °C. - 5 days between 2-8 °C. - 4 weeks between -25 and -15 °C (when frozen once) |
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Items | Candidate Device | Predicate Device k020817 |
| | ELITECH Clinical Systems URINE CONTROL BI - LEVEL | Biorad Liquichek Urine Chemistry Control Level 1 and Level 2 |
| Intended use/Indication for use | Same | For quality control in quantitative methods on clinical chemistry analyzers |
| Matrix | Same | Urine |
| Levels | Same | 2 |
## K. Standard/Guidance Document Referenced (if applicable):
Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition. CLSI document EP05-A2, Vol 24, No. 25, August 2004.
Protocols for Determination of Limits of Detection and Limits of Quantification; Approved Guideline. CLSI document EP17-A, vol 24, No. 34, October 2004.
Method Comparison and Bias estimation Using Patient Samples; Approved Guideline—Second Edition. CLSI document EP09-A2-IR, Vol 30, No. 17, July 2010.
Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use: Guidance for Industry and FDA Staff, November 2004.
Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition. CLSI document EP07-A2, Vol 25, No. 27, November 2005.
Evaluation of the Linearity of the Measurement of Quantitative Procedures: a Statistical Approach; Approved Guideline. CLSI document EP06-A, Vol 23, No. 16, April 2003.
## L. Test Principle:
Creatininase hydrolyzes creatinine in sample to creatine. Creatine is hydrolyzed by creatinase to sarcosine and urea. Sarcosine is then oxidized by sarcosine oxidase to produce hydrogen peroxide $(\mathrm{H}_2\mathrm{O}_2)$. $\mathrm{H}_2\mathrm{O}_2$ reacts with 4-amino-antipyrine (4-AAP) and EHSPT (N-Ethyl-N-2(-Hydroxy-3-Sulfopropyl)-m-Toluidine) under the catalytic action of peroxidase to form a colored quinoneimine. The absorbance of the quinoneimine at $546~\mathrm{nm}$ is proportional to the concentration of creatinine in the sample.
{5}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A precision study was performed by assaying three serum pools at 3 concentrations over 20 operating days, on two Selectra PROM analyzers and with two lots of reagent. Assay runs were performed twice per day and each sample was tested in duplicate. Calibration was performed for each run. The results are summarized below:
Serum
| Samples | n | Mean (mg/dL) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV(%) | SD | CV(%) |
| pool 1 | 80 | 0.76 | 0.01 | 1.2 | 0.01 | 1.9 |
| pool 2 | 80 | 1.52 | 0.01 | 0.6 | 0.03 | 1.7 |
| pool 3 | 80 | 5.52 | 0.03 | 0.5 | 0.08 | 1.5 |
Urine
| Samples | n | Mean (mg/dL) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV(%) | SD | CV(%) |
| pool 1 | 80 | 83 | 1 | 0.8 | 2 | 2.2 |
| pool 2 | 80 | 159 | 1 | 0.7 | 4 | 2.3 |
| pool 3 | 80 | 308 | 6 | 1.9 | 9 | 2.9 |
b. Linearity/assay reportable range:
A linearity study across the claimed assay range was performed by preparing samples from patient serum pools at a low creatinine concentration and a high concentration. The low pool contained serum diluted with albumin/NaCl (6 g/dL/0.9%). Eleven samples in all were assayed in triplicate
The expected values were plotted against the observed values and a linear regression line was fitted with the following regression equation:
Serum
| Level | Expected value | Mean value |
| --- | --- | --- |
| 1 | 0.18 | 0.10 |
| 2 | 3.20 | 3.23 |
| 3 | 6.22 | 6.21 |
| 4 | 9.24 | 9.20 |
| 5 | 12.26 | 12.22 |
| 6 | 15.28 | 15.31 |
| 7 | 18.30 | 18.43 |
{6}
| 8 | 21.32 | 21.46 |
| --- | --- | --- |
| 9 | 24.34 | 24.41 |
| 10 | 27.36 | 27.27 |
| 11 | 30.38 | 30.25 |
$\mathrm{y} = 0.9893\mathrm{x} + 0.08,\mathrm{r} = 1.0000$
Based on these results, the sponsor claims that the assay is linear from $0.10 - 30\mathrm{mg / dL}$ .
Urine
| Level | Expected value | Mean value |
| --- | --- | --- |
| 1 | 5 | 5 |
| 2 | 50 | 50 |
| 3 | 95 | 96 |
| 4 | 140 | 138 |
| 5 | 185 | 185 |
| 6 | 230 | 228 |
| 7 | 276 | 281 |
| 8 | 321 | 322 |
| 9 | 366 | 364 |
| 10 | 411 | 407 |
| 11 | 456 | 457 |
$\mathrm{y} = 1.0195\mathrm{x} - 3,\mathrm{r} = 0.9999$
Based on these results, the sponsor claims that the assay is linear from $5 - 450\mathrm{mg / dL}$ .
The sponsor has also provided data to support that post auto dilution the extended range claim is $30 - 150.00\mathrm{mg / dL}$ for serum
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability
Real-time testing was conducted. The stability study protocol and the acceptance criteria have been reviewed and found to be acceptable. The study results support the following stability claims in the labeling:
| Item | Storage Conditions | | Claimed Stability |
| --- | --- | --- | --- |
| Reagent Packs | Closed-vial | 2 – 8 °C | 20 months |
| | Open-vial | On board | 28 days |
| Calibrator | Closed-vial | 2 – 8 °C | 11 months |
| | Open-vial | 15 – 25 °C | 8 hours |
| | | 2 – 8 °C | 2 days |
{7}
| | | -25 to -15 °C
(frozen once) | 4 weeks |
| --- | --- | --- | --- |
| Serum Controls | Closed-vial | 2 – 8 °C | 30 months |
| | Open-vial | 15 – 25 °C | 12 hours |
| | | 2 – 8 °C | 5 days |
| | | -25 to -15 °C
(frozen once) | 4 weeks |
| Urine Controls | Closed-vial | 2 – 8 °C | 24 months |
| | Open-vial | 2 – 8 °C | 30 days |
The creatinine in ELICAL 2 is traceable to ID-MS (Isotope dilution-Mass Spectrometry) and was previously cleared under k103376
The serum control materials ELITROL I and ELITROL II were previously cleared under k103376
URINE CONTROL BI – LEVEL
## Value Assignment
The target value is determined by the median of 80 results from 2 Selectra ProM analyzers. The median is acceptable to the sponsor when percentage deviation is less than 10%.
## Calibration frequency
Stability across a 14-day calibration interval was assessed by calculating the percentage bias of 2 urine control specimens on each day from the result obtained on day 0. The results support the calibration interval claims of 14 days. The sponsor also noted in the labeling that a recalibration is recommended when reagent lots change, and when quality control results fall outside the range established, and after a maintenance operation.
## d. Detection limit:
Limit of Blank (LoB) and Limit of Detection (LoD) were determined following CLSI guideline EP17-A. For LoB determination, albumin/NaCl (6 g/dL/0.9%) was used as the zero sample and assayed 60 times on 2 selectra ProM analyzers. For LoD determination, 4 diluted serum or urine samples with creatinine concentrations between LoB and 4 * LoB were assayed 15 times for each sample (N = 60).
For LoQ determination, 4 low concentration samples were prepared by saline dilution and were assayed 15 times each (N = 60). The CV% and the SD at each tested concentration were calculated. The sponsor’s acceptable total error for the determination Limit of Quantification is ≤ 0.02 and 4 mg/dL for serum and urine, respectively. If the confidence interval is within the acceptable total error limits, then the Limit of Quantification is acceptable.
{8}
# Summary:
Based on the limit of quantitation determined in this study (see result in the below table) and the result from the linearity study in M1.b., the sponsor's labeling claims a measuring range of 0.1 to $30\mathrm{mg / dL}$ for serum and 5 to $450\mathrm{mg / dL}$ for urine..
| | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Serum | 0.01 | 0.02 | 0.08 |
| Urine | 0.2 | 0.5 | 2.0 |
# e. Analytical specificity:
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of six concentrations for each interfering substance. Two different clinically relevant concentrations of creatinine (1.50 and $5.00\mathrm{mg / dL}$ ) were used for evaluation. Samples with increasing amounts of potential interferents were tested in triplicate and compared to a control sample without the interferent.
The sponsor defined non-significant interference as the highest level tested that does not cause $\geq \pm 10\%$ change between the tested samples and the control sample.
Serum
| Analyte | Highest concentration tested with no significant interference (mg/dL) (≤10%) |
| --- | --- |
| Triglyceride | 3000 |
| Unconjugated Bilirubin | 30.0 |
| Conjugated Bilirubin | 14.8 |
| Hemoglobin | 500 |
| Glucose | 500 |
| Creatinine | 5.0 |
| Uric acid | 20 |
| Ascorbic Acid | 20.0 |
Calcium dobesilate, Methyl-dopa and L-dopa cause falsely low results at therapeutic concentrations.
Urine
| Analyte | Highest concentration tested with no significant interference (mg/dL) (≤10%) |
| --- | --- |
| Conjugated Bilirubin | 29.5 |
| Hemoglobin | 500 |
| Ascorbic Acid | 20.0 |
| Methyl dopa | 10 |
| Calcium dobesilate | 50.0 |
{9}
| Glucose | 5000 |
| --- | --- |
| pH | 2.5 – 12 |
A summary of these results and interferences is found in the labeling.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Serum
The performance of the Creatinine PAP SL reagent on the Selectra ProM analyzer was compared with the Roche Creatinine plus v2 reagent on the cobas c111 analyzer. Study performed according to CLSI protocol EP9-A2 using 100 serum samples covering the measuring range with values from 0.10 to 30.1 mg/dL. The linear regression results are summarized below:
| | Regression parameters | Confidence Interval, 95% | |
| --- | --- | --- | --- |
| | | Lower 95% | Upper 95% |
| slope | 0.979 | 0.977 | 0.981 |
| Intercept | 0.05 | 0.03 | 0.07 |
| r | 1.000 | | |
Urine
The performance of the Creatinine PAP SL reagent on the Selectra ProM analyzer was compared with the Roche Creatinine plus v2 reagent on the cobas c111 analyzer. Study performed according to CLSI protocol EP9-A2 using 54 urine samples covering the measuring range with values from 5 – 413 mg/dL. The linear regression results are summarized below:
| | Regression parameters | Confidence Interval, 95% | |
| --- | --- | --- | --- |
| | | Lower 95% | Upper 95% |
| slope | 1.063 | 1.055 | 1.070 |
| Intercept | 2 | 1 | 3 |
| r | 1.000 | | |
b. Matrix comparison:
To demonstrate comparable performance between serum, and lithium heparin, 40 matched samples were compared with the Creatinine PAP SL reagent on the Selectra ProM analyzer. Study performed according to CLSI protocol EP9-A2 covering the measuring
{10}
range with values from 0.11 to 29.49 mg/dL. The linear regression results are summarized below:
| | Regression parameters | Confidence Interval, 95% | |
| --- | --- | --- | --- |
| | | Lower 95% | Upper 95% |
| slope | 0.994 | 0.990 | 0.997 |
| Intercept | 0.00 | -0.03 | 0.03 |
| r | 1.000 | | |
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
None
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
| Matrix | Men | Women |
| --- | --- | --- |
| Serum, mg/dL | 0.72 - 1.18 | 0.55 – 1.02 |
| Serum, μmol /L | 64 – 104 | 49 – 90 |
| | | |
| Urine, mg/kg/24h | 14 – 26 | 11 – 20 |
| Urine, μmol/kg/24h | 124 – 230 | 97 – 177 |
Newman, D.J., Price C.P., Tietz Fundamentals of Clinical Chemistry, 5th Ed., Burtis, C.A. & Ashwood, E.R. (W.B. Saunders eds. Philadelphia USA), (2001), 414.
Ceriotti, F., Reference Intervals for Serum Creatinine Concentrations: Assessment of Available Data for Global Application. Clin. Chem., (2008), 54, 559.
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{11}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.