The intended use of the ClearMRI Helios-RG8 Coil Accessory Kit is to enable the GE Signa Horizon Cx MR System to perform MR scans of the head or body utilizing newer FDA-cleared coils (up to 8 channels), without hindering its current capability utilizing the legacy GE FDA-cleared 4-channel coils. The ClearMRI Helios-RG8 Coil Accessory Kit inherits the same limitations of the GE Signa Horizon Cx MR system. The ClearMRI Helios-RG8 Coil Accessory Kit has only been validated for and is for use strictly with the GE Signa Horizon Cx MR System, software build 9.1.0311b. The ClearMRI Helios-RG8 Coil Accessory Kit allows the GE Signa Horizon Cx MR System to continue to produce transverse, sagittal, coronal and oblique images of the internal structures of the head or body with the GE Signa Horizon Cx MR System. The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The Helios-RG8 Coil Accessory Kit will support coil anatomies from Resonance Innovations, LLC: Long Bone Array in large (model 168GE1501). Accessories to medical devices are regulated under the same classification regulation as the device(s) with which they are intended to be used. The ClearMRI Helios-RG8 Coil Accessory Kit is intended to be used strictly as an accessory to the GE Signa Horizon Cx MR System, software build 9.1.0311b, and does not change the existing indications for the GE Signa Horizon Cx MR System (K962061).
Device Story
Accessory kit for GE Signa Horizon Cx MR system; enables use of newer 8-channel RF receive coils; includes digital receiver, multi-coil driver, and computer system. Hardware operates in background parallel to GE system; does not alter operator workflow. Produces transverse, sagittal, coronal, and oblique MR images of head/body. Used in clinical settings by trained physicians to interpret proton density, T1, T2, and flow properties for diagnostic purposes. Benefits include expanded coil compatibility for legacy MR systems.
Clinical Evidence
Bench testing only. Evaluated to NEMA standards, IEC 60601 medical device safety standards, and IEC 60601-1-2 EMI/EMC standards. No clinical data presented.
Technological Characteristics
Includes digital receiver, multi-coil driver, and computer system. Supports up to 8-channel RF coils. Operates in parallel with GE Signa Horizon Cx MR system (software build 9.1.0311b). Complies with IEC 60601 and NEMA standards.
Indications for Use
Indicated for patients requiring MR imaging of the head or body using the GE Signa Horizon Cx MR System. No specific age or gender restrictions; contraindications are those inherent to the GE Signa Horizon Cx MR System.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
General Electric Signa Horizon Cx Magnetic Resonance System (K962061)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary
K132381
Page 1 of 2
APR 0 4 2014
# 510(k) Summary
| Sponsor: | ClearMRI Solutions, Inc.<br>W315 S3283 Harvest View Drive<br>Waukesha WI 53189 |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mary Ann Ferguson, Project Manager<br>ClearMRI Solutions, Inc.<br>W315 S3283 Harvest View Drive<br>Waukesha WI 53189<br>Telephone: (978)-578-1196 |
| Date Prepared: | October 15, 2013 |
| Subject Device: | Trade Name: ClearMRI Helios-RG8 Coil Accessory Kit<br>Common/Usual Name: MRI Coil Accessory Kit<br>Classification Name: Magnetic resonance diagnostic device.<br>Regulation: 21 CFR §892.1000; Classification: II; Product Code: LNH |
| Predicate Devices: | General Electric Signa Horizon Cx Magnetic Resonance System<br>(K962061) |
# Device Description:
The ClearMRI Solutions, Inc. Helios-RG8 Coil Accessory Kit is designed as an accessory to a previously cleared GE Signa Horizon Cx MR (K962061), software build 9.1.0311b, receiver subsystem with the main modification allowing the operator to use newer RF receive coils with up to 8 channels. The accessory includes a digital receiver, multi-coil driver, and a computer system containing the ClearMRI software and functions in a similar way to the predicate device's comparable subsystem. The ClearMRI hardware is seamlessly background-driven from the GE Signa Horizon Cx system'.
#### Intended Use:
The intended use of the ClearMRI Helios-RG8 Coil Accessory Kit is to enable the GE Signa Horizon Cx MR System to perform MR scans of the head or body utilizing newer FDA-cleared coils (up to 8 channels), without hindering its current capability utilizing the legacy GE FDAcleared 4-channel coils. The ClearMRI Helios-RG8 Coil Accessory Kit inherits the same limitations of the GE Signa Horizon Cx MR system. The ClearMRI Helios-RG8 Coil Accessory Kit has only been validated for and is for use strictly with the GE Signa Horizon Cx MR System, software build 9.1.0311b. The ClearMRI Helios-RG8 Coil Accessory Kit allows the GE Signa Horizon Cx MR System to continue to produce transverse, sagittal, coronal and oblique images of the internal structures of the head or body with the GE Signa Horizon Cx MR System. The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
### ClearMRI Helios-RG8 Coil Accessory Kit
Page | 1
{1}------------------------------------------------
KT32381
Page 2 of 2
The Helios-RG8 Coil Accessory Kit will support coil anatomies from Resonance Innovations, LLC: Long Bone Array in large (model 168GE1501).
Accessories to medical devices are regulated under the same classification regulation as the device(s) with which they are intended to be used. The ClearMRI Helios-RG8 Coil Accessory Kit is intended to be used strictly as an accessory to the GE Signa Horizon Cx MR System, software build 9.1.0311b, and does not change the existing indications for the GE Signa Horizon Cx MR System (K962061).
## Technological Characteristics:
The ClearMRI Solutions Helios-RG8 Coil Accessory Kit includes a digital receiver, multi-coil driver and a computer system containing the ClearMRI software. The accessory kit is available as an add-on to a GE Signa Horizon Cx MR, software build 9.1.0311b, receiver subsystem that allows the operator to scan using a digital imaging system resulting in the benefits of newer multi-channel RF coils. The ClearMRI hardware is controlled by software running independently and in parallel with the GE Signa Horizon Cx System. It does not alter the existing operator workflow. The Helios-RG8 has no new technological characteristics that raise new safety or effectiveness questions.
### Performance Data:
The ClearMRI Helios-RG8 device will support up to eight channel coils. The accessory kit was evaluated to the appropriate NEMA standards and IEC 60601 medical device safety standards as well as IEC 60601-1-2 EMI/EMC standards.
## Substantial Equivalence:
The ClearMRI Solutions, Inc. (CMS) Helios-RG8 Coil Accessory Kit is designed as an add-on accessory to a GE Signa Horizon Cx MR, software build 9.1.0311b. receiver subsystem that allows the operator the use of newer RF coils (up to 8-channels). Accessories to medical devices are regulated under the same classification regulation as the device(s) with which they are intended to be used. The accessory kit does not significantly affect the performance of the legally marketed GE Signa Horizon Cx System, software build 9.1.0311b, and is therefore substantially equivalent. Nonclinical test data demonstrate that the device is safe and is substantially equivalent to the legally marketed predicate device.
Page | 2
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a symbol consisting of three curved lines that resemble a stylized human figure. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Flealth Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2014
ClearMRI Solutions, Inc. % Mr. Mark Job Responsible Third Party Official 1394 25" Street NW BUFFALO MN 55313
Rc: K132381
Trade/Device Name: ClearMRI Helios-RG8 Coil Accessory Kit Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 20, 2014 Received: March 21. 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I'DA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2-Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources.forYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara For
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### Indications for Use
510(k) Number: K132381
## Device Name:
ClearMRI Helios-RG8 Coil Accessory Kit
#### Indications for Use:
The intended use of the ClearMRI Helios-RG8 Coil Accessory Kit is to enable the GE Signa Horizon Cx MR System to perform MR scans of the head or body utilizing newer FDA-cleared coils (up to 8 channels), without hindering its current capability utilizing the legacy GE FDA-cleared 4-channel coils. The ClearMRI Helios-RG8 Coil Accessory Kit inherits the same limitations of the GE Signa Horizon Cx MR system. The ClearMRI Helios-RG8 Coil Accessory Kit has only been validated for and is for use strictly with the GE Signa Horizon Cx MR System, software build 9.1.0311b. The ClearMRI Helios-RG8 Coil Accessory Kit allows the GE Signa Horizon Cx MR System to continue to produce transverse, sagittal, coronal and oblique images of the internal structures of the head or body with the GE Signa Horizon Cx MR System. The images produced reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (T ), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
The Helios-RG8 Coll Accessory Kit will support coil anatomies from Resonance Innovations, LLC: Long Bone Array in large (model 168GE1501).
Accessories to medical devices are regulated under the same classification regulation as the device(s) with which they are intended to be used. The ClearMRI Helios-RG8 Coil Accessory Kit is intended to be used strictly as an accessory to the GE Signa Horizon Cx MR System, software build 9.1.0311b, and does not change the existing indications for the GE Signa Horizon Cx MR System (K962061).
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
> K132381 510(k)
ClearMRI Helios-RG8 Coil Accessory Kit
Page | I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.