← Product Code [OSH](/productcode/OSH) · K132303

# IMPLANET SPINE SYSTEM (K132303)

_Implanet, S.A. · OSH · Oct 10, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132303

## Device Facts

- **Applicant:** Implanet, S.A.
- **Product Code:** [OSH](/productcode/OSH.md)
- **Decision Date:** Oct 10, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, or revision of a failed fusion attempt. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

## Device Story

Posterior spinal instrumentation system; includes polyaxial/monoaxial pedicle screws, rods, transverse connectors, and hooks. Used by surgeons in clinical settings to immobilize spinal segments; adjunct to fusion. Components constructed from Ti6Al4V titanium alloy. System provides mechanical stabilization of spine; supports fusion process in degenerative, traumatic, or deformity cases. Pediatric application limited to posterior approach for adolescent idiopathic scoliosis. Surgeon selects appropriate screw diameters/lengths and rod configurations to achieve desired spinal alignment. Benefits include stabilization of spinal segments to facilitate bone fusion and correction of deformities.

## Clinical Evidence

Bench testing only. Testing incorporated by reference from Calypso System (K120564) and supplemented with additional biocompatibility testing. Bench tests included static axial gripping, flexion/extension bending, axial torque (ASTM F1798), compression bending (ASTM F1717), torsion (ASTM F1717), dynamic compression bending (ASTM F1717), shelf life (ASTM F1980), and sterilization validation (ISO 11137, ISO 17665). Biocompatibility confirmed per ISO 10993 (cytotoxicity, acute systemic toxicity, chemical characterization).

## Technological Characteristics

System components: monoaxial/polyaxial pedicle screws, union rods, transverse connectors, hooks. Material: Ti6Al4V titanium alloy (ISO 5832-3). Screw diameters: 5.0-8.0 mm; lengths: 35-60 mm. Mechanical fixation via polyaxial/monoaxial screw-rod construct. Sterilization: validated per ISO 11137/17665. Biocompatibility: ISO 10993 compliant.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Calypso System ([K120564](/device/K120564.md))
- CD HORIZON® Spinal System ([K091445](/device/K091445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) SUMMARY

## Implanet S.A.'s IMPLANET Spine System

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Implanet S.A. Technopole Bordeaux Montesquieue Allée François Magendie Phone +33 557 995 555 Facsimile: +33 557 995 700

Contact Person: Franck Rigal, Director of Quality and Regulatory Affairs

Date Prepared: September 15, 2013

#### Name of Device

IMPLANET Spine System

#### Common or Usual Name

Spinal fixation device

#### Classification Name

888.3070 - Spinal Pedicle Fixation Orthosis - NKB, OSH, MNI, MNH

888.3050 - Spinal Interlaminal Fixation Orthosis - KWP

#### Predicate Devices

Implanet S.A.'s Calypso System (K120564)

Medtronic Sofamor Danek's CD HORIZON® Spinal System (K091445)

#### Intended Use / Indications for Use

The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), turnor, pseudarthrosis, or revision of a failed fusion attempt.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

**OCT 10 2013**

{1}------------------------------------------------

## Device Description

The IMPLANET Spine System is a posterior instrumentation system. The polyaxial screws are made of Ti6Al4V titanium alloy compliant with ISO 5832-3. The polyaxial screw is comprised of three sections: the pedicle screw; the head; and a ring that connects the screw to the head. The polyaxial screws are offered in diameters ranging from 5.0 to 7.5 mm and in lengths ranging from 35 to 60 mm.

The monoaxial pedicle screws are made of Ti6A!4V titanium alloy compliant with ISO 583203 and are available in 5.0 mm, 6.0 mm, 7.0 mm, and 8.0 mm diameters. The screws range in length from 35 to 60 mm.

The system includes both straight and pre-bent rods made of Ti6Al4V titanium alloy.

The transverse connectors are composed of Ti6Al4V titanium alloy compliant with ISO 5832-3. These connectors are used to build a transverse connection between two union rods.

The hooks are made of Ti6AI4V titanium alloy and are provided in multiple configurations.

The IMPLANET Spine System has principles of operation substantially similar to other pedicle screw-based systems for the indications listed.

The IMPLANET Spine System components may be used for posterior pedicle screw fixation in pediatric cases: polyaxial and monoaxial screws, rods, transverse connectors and rods.

The purpose of the subject 510(k) notice is the addition of new indications and the inclusion of intermediate screw sizes to the company's cleared system.

## Technological Characteristics

The IMPLANET Spine System consists of: monoaxial pedicle screws, union rods, transverse connectors, and hooks.

# Performance Data

This 510(k) premarket notification incorporates by reference the bench testing performed in support of Implanet S.A.'s Calypso System (K120564), which is listed below. All bench testing confirmed that the product met the necessary specifications. Sterilization and shelf life validation testing conducted for the Calypso System in accordance with recognized industry standards are also incorporated by reference. In addition, the biocompatibility of the device was confirmed in accordance with ISO 10993. A list of the tests performed on the previous version of the device to support substantial equivalence is provided below:

- Static axial gripping capacity, static flexion/extension bending, static axial torque . gripping capacity - ASTM F1798;
- Static compression bending ASTM F1717; .
- Static torsion ASTM F1717; .
- Dynamic compression bending ASTM F1717; .
- Shelf life ASTM F 1980; .
- Implant sterilization validation ISO 11137; t
- Instrument cleaning and sterilization validation ISO 17665; .
- Cytotoxicity ISO 10993; .
- . Acute systemic toxicity - ISO 10993.

{2}------------------------------------------------

In addition, a list of the tests performed again on the current version of the device to support substantial equivalence is provided below:

- Cytotoxicity ISO 10993; ●
- Chemical characterization ISO 10993. .

## Substantial Equivalence

When used for pediatric use, the IMPLANET Spine System is substantially similar to Implanet S.A.'s Calypso System and Medtronic Sofamor Danek's CD HORIZON® Spinal System. The IMPLANET Spine System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the IMPLANET Spine System and its predicate devices raise no new issues of safety or effectiveness. Thus, the IMPLANET Spine System is substantially equivalent.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of three curved lines, resembling a person embracing another person. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the graphic. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 10, 2013

Implanet S.A. % Ms. Janice M. Hogan Hogan Lovells US LLP 1835 Market Street, 2016 Floor Philadelphia, Pennsylvania 19103

Re: K132303

Trade/Device Name: IMPLANET Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, KWP. MNI. MNH Dated: July 24, 2013 Received: July 24, 2013

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 ) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

# Page 2 - Ms. Janice M. Hogan

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Erin I. Keith

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known): _ K132303

Device Name: IMPLANET Spine System

Indications for Use:

The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature palients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, or revision of a failed fusion attempt.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. These devices are to be used with autografi and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

**Amy S. Graf -S**

(for RPJ)

(Division Sign-off) Division of Orthopedic Devices 510(k) Number: K132303

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K132303](https://fda-staging.innolitics.com/device/K132303)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
