K132291 · North American Science Assoc., Inc. · JOJ · May 20, 2014 · General Hospital
Device Facts
Record ID
K132291
Device Name
NAMSA CHEMICAL PROCESS
Applicant
North American Science Assoc., Inc.
Product Code
JOJ · General Hospital
Decision Date
May 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The NAMSA Chemical Process Indicator Strip for Steam, REF TST4-S is intended for use with individual materials (i.e. pouches, pack, tray) to demonstrate that the material has been exposed to a steam sterilization process to distinguish between processed and unprocessed goods. The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals as a process indicator: 121°C for 30 minutes (gravity cycle) . 134℃ for 3 minutes (pre-vacuum) .
Device Story
Chemical process indicator strip; used in healthcare facilities to verify steam sterilization exposure. Input: high-temperature steam during sterilization cycles. Operation: bismuth-based indicating ink (SSI-10-YBI) on paper substrate undergoes irreversible chemical color transition from yellow to dark brown/black upon exposure to specific time/temperature parameters (121°C/30min or 134°C/3min). Output: visual color change on strip. Healthcare providers place strip inside/on packs/trays; inspect color post-cycle to confirm processing. Benefits: provides immediate visual confirmation of sterilization exposure, preventing use of unprocessed goods.
Clinical Evidence
Bench testing only. Performance verified per ANSI/AAMI/ISO 11140-1:2005. Testing included exposure to various steam cycles (121°C/3min, 121°C/10min, 134°C/20sec, 134°C/30sec, 134°C/2min) and dry heat (140°C/30min). Results confirmed expected color transition and compliance with critical parameters (time, temperature, saturated steam). Biocompatibility and shelf-life testing also performed.
Technological Characteristics
Paper strip (0.75" x 4.5", 10 mil white tag stock) with bismuth-based indicating ink (SSI-10-YBI) and PET over-laminate. Complies with ANSI/AAMI/ISO 11140-1:2005 and ANSI/AAMI ST-60:1996. Single-use, non-electronic, passive chemical indicator.
Indications for Use
Indicated for use in healthcare facilities to distinguish between processed and unprocessed goods (pouches, packs, trays) following steam sterilization cycles (121°C gravity or 134°C pre-vacuum). No specific patient population restrictions.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
SteriTec Cross-Checks Indicator, REF CI 107 (K951113)
Submission Summary (Full Text)
{0}------------------------------------------------
# KI32291
# MAY 2 0 2014
#### 510(k) Summary
#### Submitter:
NAMSA 6750 Wales Road Northwood, Ohio 43619 P: 419.666.9455 F: 419,666,1715 E: productorders@namsa.com
Contacts:
Julie Wheeler General Manager, NAMSA Products 419.662.4488 jwheeler@namsa.com
Michelle Adamski Quality Assurance and Regulatory Affairs Manager, NAMSA Products 419.662.4829 madamski@namsa.com
Prepared on: July 18, 2013
- Device Name: NAMSA Chemical Process Indicator Strip for Steam
- Class II Medical Device, FDA Product Code JOJ, General Hospital Classification:
- Common Name: Indicator/Chemical Process Indicator
- NAMSA REF TST4-S Trade Name:
- Predicate Devices: SteriTec Cross-Checks Indicator, REF CI 107 K951113 SteriTec Products Mfg. Co., Inc.
### Statement of Intended Use:
The NAMSA Chemical Process Indicator Strip for Steam, REF TST4-S is intended for use with individual materials (i.e. pouches, pack, tray) to demonstrate that the material has been exposed to a steam sterilization process to distinguish between processed and unprocessed goods.
The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals as a process indicator:
- 121°C for 30 minutes (gravity cycle) .
- 134℃ for 3 minutes (pre-vacuum) .
Image /page/0/Picture/19 description: The image shows the word "NAMSA" in white text on a black background. The font is a sans-serif typeface, and the letters are bold. There is a trademark symbol to the right of the A.
{1}------------------------------------------------
Description of Device: The NAMSA Chemical Process Indicator Strip for Steam utilizes NAMSA indicating ink SSI-10-YBI that when exposed to high temperature steam turns from an initial yellow color to a permanent dark brown/black signal color.
> The intended use for the NAMSA Chemical Process Indicator Strip for Steam processes is a Process Indicator intended for use by health care facilities to distinguish between processed and unprocessed goods.
The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a permanent dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals:
- 121℃ for 30 minutes (gravity cycle) ●
- 134℃ for 3 minutes (pre-vacuum) .
The Indicating Ink was developed and validated for compliance with ANSI/AAMI/ISO 11140-1:2005 quideline. As such, the critical parameters are time, temperature and presence of saturated steam.
### Table 1: Test and Performance Requirements for Class 1 Process Indicators for STEAM
| Test Condition | Test Time | Test Temperature | No change or a change this<br>is markedly different from<br>the visible change as<br>specified by the<br>manufacturer | Visible change as<br>specified by the<br>manufacturer |
|-----------------|-------------------|------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Saturated Steam | 3.0 min ± 5 sec | 121°C (+3/0°C) | Acceptable result | Unacceptable result |
| Saturated Steam | 10.0 min 0/-5 sec | 121°C (+3/0°C) | Unacceptable result | Acceptable result |
| Saturated Steam | 20sec | 134°C (+3/0°C) | Acceptable result | Unacceptable result |
| Saturated Steam | 0.5 min ± 5 sec | 134°C (+3/0°C) | Acceptable result | Unacceptable result |
| Saturated Steam | 2 min +5/0 sec | 134°C (+3/0°C) | Unacceptable result | Acceptable result |
| Dry Heat | 30 min ± 1 min | 140°C (+2/0°C) | Acceptable result | Unacceptable result |
Device compliance to ANSI/AAMI/ISO 11140-1:2005 & ANSI/AAMI ST-60:1996 is dependent upon substrate, thickness of ink application and presence of an over-laminate, reference material specification for NAMSA Chemical Process Indicator Strip for Steam.
#### Operational Principles:
The NAMSA Chemical Process Indicator Strip for Steam is intended for use with individual units, (e.g. packs, containers) to demonstrate that the goods have been exposed to a steam sterilization process and to distinquish between processed and unprocessed goods.
The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial vellow color to a dark brown/black signal color when exposed to high temperature steam at various time and temperature intervals.
Statement of Similarity to Legally Marketed. Predicate Device:
The NAMSA Chemical Process Indicator Strip for Steam has the following similarities to those legally marketed under 510(k) number K951113:
# NAMSA
PEOPLE > SCIENCE > SOLUTIONS
{2}------------------------------------------------
- . Same intended use
- Same technological characteristics through device design, indicator agent, . sterilization method and end point specifications and incorporation of similar materials
- . Have similar shelf lives, and
- . The same materials for packaging
A comparison of the characteristics of both the new device and predicate device and data resulting in equivalent performance has been generated.
### Comparison of the New Device to the Predicate Device
| | New Device | Predicate Device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Element | NAMSA Chemical Process Indicator<br>Strip for Steam | Cross-Checks Indicator |
| Intended Use | Process Indicator | Process Indicator |
| Device Design and Components | Paper strip, 0.75" W x 4.5"L, 10 mil white<br>tag stock<br><br>Chemical indicator on BOPP film, Yellow<br>to Dark, Indicator tamped onto paper<br>strip | Paper strip, 0.5625" W x 7.66" L, 80<br>pound cover<br><br>Chemical indicator, White to Dark,<br>Indicator printed directly on paper strip |
| | Over laminate of PET | Over laminate of PET |
| Indicator Agent | Indicating Ink<br>NAMSA formulation SSI-10-YBI, Bismuth<br>based agent to yield color transition | Indicating Ink<br>Steritec formulation, Lead based agent to<br>yield color change |
| Sterilization Method and Cycles | Steam 121°C 30 minutes<br>(gravity cycle)<br><br>Steam 134°C for 3 minutes (pre-vacuum<br>cycle) | Steam 132°C - 135°C 3 Minutes<br>standard hospital steam sterilizer |
| Endpoint Specifications | During steam sterilization, indicator goes<br>from an initial yellow color to a dark<br>brown/black signal color. | During steam sterilization, indicator goes<br>from an initial white color to a dark<br>brown/black signal color. |
| Shelf Life | 17 Months | 36 Months |
The following tests were performed in accordance with ANSI/AAM/ISO 11140-1.2005 for steam to verify the performance of NAMSA Chemical Process Indicator Strip for Steam was substantially equivalent to the predicate device,
- Steam Prevac at 121°C for 3 minutes .
- Steam Prevac and Gravity at 121°C for 10 minutes .
- Steam Prevac at 134°C for 20 seconds .
- . Steam Prevac at 134°C for 30 seconds
- . Steam Prevac at 134°C for 2 minutes
- . Dry heat at 140°C for 30 minutes
In summary, the data generated demonstrates NAMSA Chemical Process Indicator Strip for Steam is substantially equivalent to the predicate device.
# NAMSA
PEOPLE > SCIENCE > SOLUTIONS
{3}------------------------------------------------
# Description of Testing:
Per the FDA recognized guidance document, testing was performed for steam sterilization processes using multiple lots of NAMSA Chemical Process Indicator Strip for Steam.
- Exposure at Various Times & Temperatures following ANSI/AAMI/ISO . 11140-1:2005 for Steam
- Transference Testing per ANSI/AAMI/ISO 11140-1:2005
- . Biocompatibility
- . Shelf Life
Image /page/3/Picture/6 description: The image shows the word "NAMSA" in white letters against a black background. The font is a bold, sans-serif typeface. The letters are evenly spaced and the word is centered in the image.
PEOPLE > SCIENCE > SOLUTIONS
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes, positioned above two curved shapes that resemble legs or feet. The bird is enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
May 20, 2014
North American Science Association Incorporated Ms. Julie Wheeler General Manager 6750 Wales Road Northwood, OH 43619
Re: K132291
> Trade/Device Name: NAMSA Chemical Process Indicator Strip for Steam Regulation Number: 21 CFR 880.2800 Regulation Name: Indicator, Physical/Chemical Sterilization Process Regulatory Class: Class II Product Code: JOJ Dated: April 17, 2014 Received: April 21. 2014
Dear Ms. Wheeler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Wheeler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Shoth, M.D. Clinical Deputy Director A CONTRACTOR DAGRID/ODE/CDRH FFOR
Erin I. Keith, M.S. Acting Director · Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use Statement
| 510(k) Number: | K132291 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | NAMSA Chemical Process Indicator Strip for Steam |
| Indications for Use: | The NAMSA Chemical Process Indicator Strip for Steam, REF TST4-S is intended for use with individual materials (i.e. pouches, pack, tray) to demonstrate that the material has been exposed to a steam sterilization process to distinguish between processed and unprocessed goods. The NAMSA Chemical Process Indicator Strip for Steam will transition from an initial yellow color to a dark brown/black signal color when exposed to high temperature steam at the following time and temperature intervals as a process indicator: |
- 121°C for 30 minutes (gravity cycle) ●
- 134℃ for 3 minutes (pre-vacuum cycle) .
② SINGLE USE ONLY
AND/OR
× Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabetl Clave
Prescription Use
(Part 21 CFR 801 SubpartD)
Claverie -S
Claverie -S
DN: c=US, o=U.S. Government,
0.9.2342.19200300.100.1.1=13000558 64, cn=Elizabeth F. Claverie -5 Date: 2014.05.19 16:36:26 -04'00'
Digitally signed by Elizabeth F.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.