← Product Code [GEX](/productcode/GEX) · K132286

# HYPERION LONG PULSED ND: YAG LASER (K132286)

_Scarlet Rf, Inc. · GEX · Jan 22, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132286

## Device Facts

- **Applicant:** Scarlet Rf, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jan 22, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Hyperion is a long pulsed Nd:YAG laser intended for treatment of wrinkles, Removal of unwanted hair for stable long term or *permanent hair reduction, treatment of pseudofolliculitis barbae (PFB), surgical incision, excision, vaporization, ablation and coagulation of soft tissue, and photocoagulation and hemostasis of pigmented and vascular lesions in dermatology. *Permanent hair reduction is defined as the long-term stable reduction in the number of hairs re-grown when measured at 6, 9, and 12 months after the completion of a treatment regimen

## Device Story

Hyperion is a long-pulsed Nd:YAG laser system (1064 nm) for dermatological and surgical applications. System components include a main cabinet (power supply, cooling system, microprocessor, laser generator), optical fiber with handpieces, and a footswitch. Operated by clinicians in a medical setting, the device delivers laser energy to target tissue for ablation, coagulation, or hair reduction. The microprocessor controls laser emission parameters. Clinical benefit is achieved through precise thermal interaction with soft tissue, pigmented lesions, or vascular structures. The device is used for surgical incision, excision, and various dermatological treatments.

## Clinical Evidence

Bench testing only. Non-clinical testing included visual and mechanical inspection, electrical and mechanical safety testing (IEC 60601-1 and 60601-1-2), and software testing. No clinical data provided.

## Technological Characteristics

Long-pulsed Nd:YAG laser; 1064 nm wavelength. System includes main cabinet, optical fiber, handpieces, and footswitch. Electrical safety per IEC 60601-1 and IEC 60601-1-2. Microprocessor-controlled emission.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Candela, Inc. - Gentle YAG Laser System ([K022951](/device/K022951.md))
- Fotona d.d. - XP Nd:YAG Laser System ([K090126](/device/K090126.md))

## Reference Devices

- Cynosure, Inc. Apogee Elite Laser ([K034030](/device/K034030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### Section 4 510(k) Summary Statement Hyperion Long Pulsed Nd:YAG Laser

.

______________________________________________________________________________________________________________________________________________________________________________

# Section 4 - This 510(k) Summary of safety and effectiveness is submitted in accordance with the requirements of 21 CFR §807.92

| Submitter's Information             |                                                                                                                      |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Applicant:                          | Laseroptek Co. Ltd.                                                                                                  |
| Address:                            | 204 Hyundai I Valley<br>223-12 Sangdaiwon, Jungwon<br>Sungnam-Si, Gyeonggido, 462-714<br>Rep. of Korea (South Korea) |
| Contact Person:                     | Hong Chu, Ph.D. (President & CEO)                                                                                    |
| Telephone / Fax / E-mail:           | Tel) +82.31.737.9885<br>E-mail) hchu@laseroptek.com                                                                  |
| Submitter:                          | Scarlet RF, Inc.                                                                                                     |
| Address:                            | 17145 Von Karman Ave, Suite 106<br>Irvine, CA 92614, USA                                                             |
| Contact Person:                     | Kevin J. Choi                                                                                                        |
| Telephone / Fax / E-Mail:           | Tel) 310-634-4480<br>E-mail) kjychoi@gmail.com                                                                       |
| Preparation Date:                   | 04/22/2013                                                                                                           |
| Device Trade Name:                  | Hyperion                                                                                                             |
| Common Name:                        | Long Pulsed Nd:YAG Dermatological Laser                                                                              |
| Classification Name:                | Instruments, Surgical, Powered, Laser 79-GEX<br>(21 CFR §878-4810)                                                   |
| Legally Marketed Predicate Devices: | Candela, Inc. - Gentle YAG Laser System<br>(K022951)<br>Fotona d.d. - XP Nd:YAG Laser System<br>(K090126)            |

- Cynosure, Inc. Apogee Elite Laser (K034030) .

{1}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Intended Use of Hyperion:                 | Hyperion is a long pulsed Nd:YAG laser intended for treatment<br>of wrinkles, Removal of unwanted hair for stable long term or<br>*permanent hair reduction, treatment of pseudofolliculitis barbae<br>(PFB), surgical incision, excision, vaporization, ablation and<br>coagulation of soft tissue, and photocoagulation and hemostasis<br>of pigmented and vascular lesions in dermatology. *Permanent<br>hair reduction is defined as the long-term stable reduction in the<br>number of hairs re-grown when measured at 6, 9, and 12 months<br>after the completion of a treatment regimen |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description:                       | Hyperion is a long pulse Nd:YAG laser system that produces<br>laser emission at the wavelength of 1064nm. The system consists<br>of three interconnected sections: a) the main cabinet, which<br>houses the power supply, cooling system, microprocessor, and<br>the laser generator, b) the optical fiber with hand pieces which<br>can be used selectively depending on the size of the laser beam<br>exposure, c) the footswitch.                                                                                                                                                           |
| Performance Data:                         | Non-clinical testing of Hyperion included visual and mechanical<br>inspection, electrical and mechanical safety testing, software<br>testing, etc., in bench. Electrical testing was performed in<br>accordance with IEC 60601-1 and 60601-1-2 standards.                                                                                                                                                                                                                                                                                                                                      |
| Substantial Equivalence<br>(Intended Use) | Hyperion's intended use is identical to the intended use of the<br>predicate devices: Candela's Gentle YAG Laser System, Fotona's<br>XP Nd:YAG Laser System, and Cynosure's Apogee Elite Laser                                                                                                                                                                                                                                                                                                                                                                                                 |
| Substantial Equivalence<br>(Technical)    | Hyperion is based on the Nd:YAG (1064 nm) laser technology as<br>its predicated devices listed above. Hyperion long pulsed<br>Nd:YAG laser operates in a similar fashion as its predicate<br>devices, with the same principle of operation, the same<br>wavelength (1064 nm) and essentially the same power range and<br>the same indications for uses.                                                                                                                                                                                                                                        |
|                                           | Based upon analysis of the overall performance characteristic for<br>the device, Laseroptek Co. Ltd. believes that no significant<br>differences exist between Hyperion and its predicate devices.<br>Therefore, Hyperion should not raise new safety or effectiveness<br>issues.                                                                                                                                                                                                                                                                                                              |

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure with three stylized wing or feather shapes.

Public Health Service

Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

Laseroptek Company, Ltd % Mr. Kevin J. Choi Scarlet RF Incorporated 28 Stafford Drive West Windsor, New Jersey 08550

January 22, 2014

Re: K132286

Trade/Device Name: Hyperion Long Pulse Nd: YAG Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: December 24, 2013 Received: December 27, 2013

Dear Mr. Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. Kevin J. Choi

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshua

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K132286

Device Name Hyperion Long Pulse Nd: YAG Laser

Indications for Use (Describe)

Treatment of wrinkles

Removal of unwanted hair for stable long term or *permanent hair reduction

Treatment of PFB (pseudofolliculitis barbae)

Surgical incision, excision vaporization, ablation and coagulation of soft tissue

Photocoagulation and hemostasis of pigmented and vascular lesions in dermatology

*Permanent hair reduction is defined as the long-term stable reduction in the number of hair re-grown when measured at 6, 9, and 12 months after the completion of a treatment regimen.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

The proportion of the FOR FDA USE ONLY CONTRACT AND A PARTY CHANGE OF SCHOOL

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/4/Picture/17 description: The image shows the name "Neil R Ogden" followed by "-S". Below that is the date "2014.01.21" and the time "13:56:37". To the right of the time is a logo that is difficult to make out. The logo appears to have the numbers "05:00" in it.

FORM FDA 3881 (1/14)

PSC Publishing Services (301) 443-6740

FI

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

---

**Source:** [https://fda-staging.innolitics.com/device/K132286](https://fda-staging.innolitics.com/device/K132286)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).