K132220 · Limerick, Inc. · HGX · Feb 20, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K132220
Device Name
PJ'S SERENITY
Applicant
Limerick, Inc.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Feb 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Electric Breast Pump, model PJ2012 is used to express and collect milk from the breast to alleviate engorgement of the breast, maintain the ability of lactation and provide mother's milk for future feeding when separation of mother and baby occurs. The device is intended for multiple uses.
Device Story
Electric breast pump (model PJ2012) for hospital use; mechanically interfaces with breast via shield to express/collect milk. Uses electric motor-driven diaphragm to generate vacuum; air valve regulates pressure. Microcontroller (Microchip PIC18F26K22) controls motor speed via pulse width modulation and monitors vacuum sensor to maintain user-selected levels. User adjusts vacuum (15-270 mm Hg) and speed (15-275 cycles/min) via front panel knobs; LCD displays settings and elapsed time. Electronic memory stores rhythm/intensity. 1-micron filter prevents fluid/pathogen ingress. Benefits include engorgement relief and lactation maintenance. Operated by mother or clinician.
Clinical Evidence
Bench testing only. Testing included vacuum pressure/cycle rate verification, biocompatibility (cytotoxicity, irritation, sensitization), software validation, EMC, and electrical safety. Device met all acceptance criteria.
Indicated for breastfeeding women to express and collect milk to alleviate engorgement, maintain lactation, and provide milk for future feeding during mother-baby separation.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. In the center of the circle is an abstract symbol that resembles an eagle or other bird with outstretched wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
Limerick, Inc. Patricia Kelly 2150 N. Glenoaks Blvd. Burbank, CA 91504
K132220 Re: Trade/Device Name: PJ's Serenity Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Regulatory Class: II Product Code: HGX Dated: February 2, 2015 Received: February 6, 2015
Dear Patricia Kelly,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - Patricia Kelly
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132220
Device Name
PJ's Screnity
Indications for Use (Describe)
The Electric Breast Pump, model PJ2012 is used to express and collect milk from the breast to alleviate engorgement of the breast, maintain the ability of lactation and provide mother's milk for future feeding when separation of mother and baby occurs. The device is intended for multiple uses.
| Type of Use (Select one or both, as applicable) | <table style="border:none;"><tr><td><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
·
{3}------------------------------------------------
Submission Number K13220/S004
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
:
| Date: | 2014/06/13 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Limerick, Inc. |
| Primary Contact Person: | Patricia Kelly |
| | President |
| | Limerick, Inc. |
| | Tel: (818) 566-3060 |
| | Fax: (818) 566-1260 |
| Secondary Contact Person: | Joan Ortiz |
| | Vice President |
| | Limerick, Inc. |
| | Tel: (818) 566-3060 |
| | Fax: (818) 566-1260 |
| Device:Trade Name: | PJ's Serenity Model PJ2012 |
| Common/Usual Name: | Powered breast pump |
| Classification Names: | Powered breast pump |
| Regulation Number: | CFR884.5160 |
| Product Code: | HGX |
| Predicate Device(s) | K051926, PJ's Comfort breast pump<br>K012275, PJ's Comfort portable breast pump |
| Device Description: | The Electric Breast Pump, model PJ2012<br>is designed and manufactured by Limerick, Inc. It is<br>intended to express and collect milk from a lactating<br>woman's breast. This action helps to alleviate<br>engorgement of the breast, maintain the woman's<br>ability to lactate, and provide a mother's milk for<br>future feedings when separation of the mater and<br>baby occur.<br><br>The product uses a Single-Chip Microcontroller to<br>imitate a baby's suckling action. The device is<br>ergonomically designed to create comfortable milk<br>stimulation, massage, and suction from the breast.<br>There are 80 vacuum levels and 30 speeds available<br>to imitate the rhythm and action of a baby's<br>suckling. Selection of the vacuum and speed is<br>made by adjusting the control knobs on the front<br>panel of the pump. The control panel is soft and<br>viewing is provided by a LCD screen. Once<br>programmed, the pump's electronic memory stores<br>the selected rhythm and intensity of the device. |
| Indications for Use: | The Electric Breast Pump, model PJ2012 is<br>used to express and collect milk from the breast to<br>alleviate engorgement of the breast, maintain the<br>ability of lactation and provide mother's milk for<br>future feeding when separation of mother and baby<br>occurs. The device is intended for multiple users. |
| Technology: | The Electric Breast Pump, model PJ2012<br>is designed to mechanically interface with a<br>mother's breast via breast shield and withdraw, ther<br>collect, the breast milk. The device incorporates a<br>microcontroller with embedded firmware. The<br>microcontroller is a Microchip PIC18F26K22. The<br>firmware was developed using MicroEngineering<br>Labs PICBasic Pro. The device incorporates a<br>pump driven by an electric motor to produce a<br>vacuum. The device incorporates an air valve to<br>selectively allow air into the vacuum system. The<br>device incorporates an LCD display to provide<br>information to the user. The firmware provides a<br>means for the user to start and stop pump operation. |
{4}------------------------------------------------
:
{5}------------------------------------------------
While in operation, it provides a means for the user to adjust the speed of the pump motor and adjust the maximum vacuum level reached. The motor speed is controlled using pulse width modulation of the power applied to the motor. The maximum vacuum level is controlled by monitoring the output of a vacuum sensor and comparing the vacuum level with the user setting. When the level is reached, the air valve is opened. The LCD display shows the settings for the motor speed and vacuum levels using an arbitrary numeric scale, such as 1 to 30. The LCD also shows an elapsed time in minutes from the start of pump cycling.
| Specification | Predicate Device | Predicate Device | Proposed Device | Discussion of<br>Differences |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | PJ's Comfort | PJ's Comfort portable | PJ's Serenity | |
| K Number | K051926 | K012275 | K132220 | |
| Indications for<br>Use | The intended use of the<br>electrically powered<br>(diaphragm type)<br>suction device is to<br>express milk from the<br>breast of lactating<br>women. | The intended use of the<br>electrically powered<br>(diaphragm type)<br>suction device is to<br>express milk from the<br>breast of lactating<br>women. | The Electric Breast<br>Pump, model PJ2012 is<br>used to express and<br>collect milk from the<br>breast to alleviate<br>engorgement of the<br>breast, maintain the<br>ability of lactation and<br>provide mother's milk<br>for future feeding when<br>separation of mother<br>and baby occurs. The<br>device is intended for<br>multiple users. | The<br>Indications<br>for Use<br>statements<br>between the<br>subject and<br>predicate<br>devices are<br>not identical,<br>but the<br>intended use<br>of the<br>devices-to<br>express milk<br>from the<br>breast of<br>lactating<br>women-is<br>the same. |
| Patient<br>Population | Breastfeeding<br>Women | Breastfeeding<br>Women | Breastfeeding<br>Women | Same |
| Pump Type | Diaphragm | Diaphragm | Diaphragm | Same |
| Vacuum Range | 40-270mm Hg | 150-220 mm Hg | 15-270<br>mm Hg. | Similar |
| Cycle Levels | 16-70 cycles/min. | 30-45 cycles/min. | 15- 275 cycles/min | The<br>difference in<br>cycle rate<br>between the |
| | | | | subject and<br>predicate<br>devices does<br>not represent<br>new<br>technology,<br>and raises no<br>new types of<br>safety and<br>effectiveness<br>questions. |
| Filter Between<br>kit and pump | Yes | Yes | Yes | Same |
| Adjustable<br>Suction Levels | Yes | Yes | Yes | Same |
| Software | Yes | Yes | Yes | Same |
| Anatomical<br>Sites | Breast | Breast | Breast | Same |
| Energy Used<br>And/or delivered | AC<br>Battery<br>Car adapter | AC<br>Battery<br>Car adapter | AC | No battery<br>No Car<br>adapter for<br>PJ's Serenity<br>as it is only<br>used in<br>hospital |
| Designed and<br>Materials | All food or human<br>contact components<br>are manufactured<br>from materials that<br>meet FDA food<br>additive criteria as<br>set forth in 21 Code<br>of Federal<br>Regulations Part 176,<br>177 and 178. | All food or human<br>contact components<br>are manufactured<br>from materials that<br>meet FDA food<br>additive criteria as set<br>forth in 21 Code of<br>Federal Regulations<br>Part 176, 177 and<br>178. | All food or human<br>contact components<br>are manufactured<br>from materials that<br>meet FDA food<br>additive criteria as set<br>forth in 21 Code of<br>Federal Regulations<br>Part 176, 177 and<br>178. | Same |
| Performance | Stimulation, suction<br>and collect | Stimulation, suction<br>and collect | Stimulation, suction<br>and collect | Same |
| Standards Met | IEC60601-1 2005,<br>3rd Edition<br>IEC60601-2: 2007,<br>3rd Edition<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | IEC60601-1 2005, 3rd<br>Edition<br>IEC60601-2: 2007,<br>3rd Edition<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | IEC60601-1 2005, 3rd<br>Edition<br>IEC60601-2: 2007,<br>3rd Edition<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Biocompatibility | Not cytotoxic<br>irritating or dermal<br>sensitizer | Not cytotoxic<br>irritating or dermal<br>sensitizer | Not cytotoxic<br>irritating or dermal<br>sensitizer | Same |
| Mechanical<br>Safety | Electromechanical<br>cycling suction<br>regulator | Electromechanical<br>cycling suction<br>regulator | Electromechanical<br>cycling suction<br>regulator | Same |
| Operating<br>Temperature | 5-40 degree C | 5-40 degree C | 5-40 degree C | Same |
| Electrical Safety | | | | |
| Bench Test | Performs within<br>specifications | Performs within<br>specifications | Performs within<br>specifications | Same |
| 1-Micron filter | Provides a barrier.<br>against bacteria, fluid<br>and virus from<br>entering the pump | Provides a barrier<br>against bacteria, fluid<br>and virus from<br>entering the pump | Provides a barrier<br>against bacteria, fluid<br>and virus from<br>entering the pump | Same |
| Accessory Kit | 2 silicone breast cups<br>2 braces<br>2 bottle caps<br>2 silicone gaskets<br>2 storage containers<br>2 tubes with "Y"<br>adapter<br>2 tube connectors<br>1 filter<br>1clamp | 2 silicone breast cups<br>2 braces<br>2 bottle caps<br>2 silicone gaskets<br>2 storage containers<br>2 tubes with "Y"<br>adapter<br>2 tube connectors<br>1 filter<br>1clamp | 2 silicone breast cups<br>2 braces<br>2 bottle caps<br>2 silicone gaskets<br>2 storage containers<br>2 tubes with "Y"<br>adapter<br>2 tube connectors<br>1 filter<br>1clamp | Same |
| Packaging | Corrugated | Corrugated | Corrugated | Same |
## Determination of Substantial Equivalence:
{6}------------------------------------------------
{7}------------------------------------------------
Submission Number K13220/S004
## Summary of non-clinical tests:
The sponsor has performed bench testing to demonstrate the electric breast pump performs within the specifications:
PJ's Serenity Model number PJ2012 Vacuum levels 15-270 mm Hg. Cycles/min
PJ's Serenity Model number PJ2012 has met acceptance criteria of performance testing including: biocompatibility (in vivo cytotoxicity, irritation, and sensitization testing), software validation, EMC, electrical safety, and vacuum pressure / cycle rate testing.
#### Conclusion:
Limerick, Inc considers the PJ's Serenity electric breast pump to be substantially equivalent to the predicated devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.