The Melys Atrial Fibrillation Monitor is indicated for self-testing by patients who have been diagnosed with, or are susceptible to developing, atrial fibrillation and who would like to monitor and record their heart rhythms on an intermittent basis.
Device Story
Portable monitor for heart rhythm self-testing; uses finger sensor (K101692) to capture physiological data via light transmission; processes data to calculate Arrhythmia Index; displays real-time waveform, heart rate, and arrhythmia status via green/amber/red light indicator; stores history of 8 previous analyses; battery-powered; intended for home use by patients; provides indicative results to assist in monitoring heart irregularity; does not transmit data.
Clinical Evidence
Two clinical trials conducted to verify identification of atrial fibrillation compared to EKG diagnosis. Human factor study with 20 volunteers showed 100% agreement on ease of use and label comprehension. Comparative bench testing against predicate showed matching heart rate measurements and arrhythmia indications.
Technological Characteristics
Portable monitor with finger sensor; Lithium-ion Polymer battery; real-time waveform display; light-based arrhythmia indicator; IEC 60601-1 safety compliant; dimensions 4.52" x 4.13" x 1.37"; weight 200g.
Indications for Use
Indicated for adult patients diagnosed with or susceptible to atrial fibrillation for intermittent self-monitoring and recording of heart rhythms.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
# AAR 1 9 2014
# 510(k) Summary 807.92(c)
| SPONSOR | | 807.92(a)(1) |
|-----------------------------------------|------------------------------------------------------------|--------------|
| Company Name: | Advanced Fluidics, LLC | |
| Company Address | 8860 Columbia 100 Parkway, Suite 204<br>Columbia, MD 21045 | |
| Telephone: | (443) 864-5295 | |
| Contact Person: | Surya Raghu, Ph.D. | |
| Summary Preparation Date: June 28, 2013 | | |
Summary Preparation Date: June 28, 2013
## DEVICE NAME
Trade Name: Common/Usual N Classification Nar Regulation Numb Product Code: Device Class:
# 807.92(a)(2)
| | Melys Atrial Fibrillation Monitor |
|-------|-------------------------------------------------------|
| Name: | Atrial Fibrillation Monitor |
| | Telephone electrocardiograph transmitter and receiver |
| | 21 CFR 870.2920 |
| | DXH |
| | Class II |
# PREDICATE DEVICE
# 807.92(a)(3)
Legally Marketed Equivalent Device
| K Number | Device | Manufacturer |
|----------|----------------------------------------|----------------------------|
| K052767 | AfibAlert Atrial Fibrillation Detector | Technologies Research, LLC |
## DEVICE DESCRIPTION
# 807.92(a)(4)
The Melys Atrial Fibrillation Monitor is a lightweight portable device for measuring and displaying heart irregularity and pulse rate. The device consists of the following - 1) Atrial Fibrillation Monitor, 2) Finger Sensor (K101692), 3) Power Supply and 4) Instruction Manual.
The Melys Atrial Fibrillation Monitor device receives physiological data relating to heart rhythm from a finger sensor (K101692) attached to the patient's finger. The device captures, processes and records this data. Once sufficient data has been captured the device will run a mathematical analysis on the data and presents an indicative result of the patient's likelihood of suffering from atrial fibrillation or other heart irregularity.
{1}------------------------------------------------
The monitor contains a rechargeable battery, and a power supply is provided with the monitor for charging the monitor's battery. The finger sensor is connected to the monitor and with the pressing of the start button is ready to deliver trace measurements within 10 seconds. Each measurement is shown in a large display with a light and a number (the Arrhythmia Index Result). A history results bar at the bottom of the monitor will show the results of the previous 8 analyses with each light indicating each separate result by a green, amber, or red light. In the upper right corner of the display is an indication of the input signal strength and the level of battery charge.
#### DEVICE INDICATIONS FOR USE
#### 807.92(a){5)
The Melys Atrial Fibrillation Monitor is indicated for self-testing by patients who have been diagnosed with, or are susceptible to developing, atrial fibrillation and who would like to monitor and record their heart rhythms on an intermittent basis.
#### COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a){6)
{2}------------------------------------------------
| | Subject | Predicate | |
|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Parameters | Melys AFS Ltd. | Lechnologies Research, Inc. | Similarities and Differences |
| 510(k) Number | | K052767 | |
| Indications for Use | The Melys Atrial Fibrillation Monitor is<br>indicated for self-testing by patients, who<br>have been diagnosed with, or are<br>susceptible to developing, atrial fibrillation<br>and who would like to monitor and record<br>their heart rhythms on an intermittent<br>basis. | The AfibAlert™ is indicated for self-testing<br>by patients who have been diagnosed with,<br>or are susceptible to developing atrial<br>fibrillation and who would like to monitor<br>and record their heart rhythms on an<br>intermittent basis. | Same |
| Regulation Name | Telephone electrocardiograph transmitter<br>and receiver | Telephone electrocardiograph transmitter<br>and receiver | Same |
| Product Code | DXH | DXH | Same |
| intended User | Adult | Adult | Same |
| Prescription device for<br>home use | yes | yes | Same |
| Single Patient Use | yes | yes | Same |
| Monitors regularity of<br>heartbeat | yes | yes | Same |
| Alerts user to an<br>irregularity in their<br>pulse through a light<br>indicator. | yes | yes | Same |
| Principle of Operation | The monitor passes light through the<br>fingertip sensor and receives data from<br>which the waveform is directly created.<br>Displays regularity or irregularity with a<br>light indicator. | The monitor creates an electrocardiograph<br>reading from the data received from the<br>thumb, wrist or chest sensor from which a<br>waveform is derived.<br>Displays regularity or irregularity with a<br>light indicator. | The Melys monitor receives its<br>information from the finger rather<br>than the thumb, wrist or chest, using<br>a familiar, pulse oximeter style finger<br>sensor. |
| Device Design | Monitor obtains waveform via finger<br>sensor and displays waveform in real time<br>on the monitor with an arrhythmia index. | Cardiac event recorder with thumb, wrist,<br>or chest sensor that displays results by light<br>indicator (green, red), but no pulse | Different acquisition of waveform<br>but same result and light indicator<br>display of heart regularity or |
{3}------------------------------------------------
# page 4 of 6
| Functional Features | Light display (green, amber, or red) indicates heart regularity or irregularity.<br>Last 8 tests are stored on the display by light indicator not Arrhythmia Index number. | waveform.<br>Records, stores and transmits results over the telephone wire. | irregularity. |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User programmable | No | No | Same |
| Recording length | 10 seconds<br>(Recommended four (4) consecutive measurement should show same indication (red, amber or green) before a conclusion is drawn) | 45 seconds | Melys Atrial Fibrillation Monitor<br>Instructions recommends four (4) consecutive measurement should show same indication (red, amber or green) before a conclusion is drawn |
| Low battery detection | Yes | Yes | Same |
| Heart rate accuracy | $\pm$ 1 BPM | $\pm$ 1 BPM (average for data collected, updated every three (3) seconds) | Same |
| Physical and Environmental | | | |
| Weight (w. Batteries) | 200g 7.05 oz. | 4.6 oz. | Slight difference does not affect safety & efficacy |
| Size | 4.52" x 4.13" x 1.37" | 5.88" x 2.75" x 1.11" | Slight difference does not affect safety & efficacy |
| Operational<br>Temperature | 20°C to 40° C | 0°C to +45° C (+32° F to 113°F) | Slight difference does not affect safety & efficacy |
| Storage & Transport | 0-20°C | -120°C to +60°C (-4°F to 140°F) | Slight difference does not affect safety & efficacy |
| Power | | | |
| Battery type | Lithium-ion Polymer Battery (Minamoto)<br>3.7 V | Two (2) AAA 1.5 V Eveready x -92 | Both use battery as power source |
| Battery Life | 40 days continuous | 90 days nominal used 1 x per day | Difference does not affect safety & efficacy |
| Differences | | | |
| Memory Capability | No | Yes | The subject device displays the last eight Indicator lights on the monitor face but not the arrhythmia Index. |
| Transmission of Data | No | Yes | The subject devices does not transmit data |
{4}------------------------------------------------
- -
-
page 5 of 6
| Patient Interface | Finger Sensor | Thumb, or wrist sensor, and direct chest<br>contact | Most users will have encountered a<br>fingertip sensor such as the Melys<br>AFS Ltd fingertip sensor and<br>therefore require little instruction on<br>how to correctly use it. The finger<br>sensor passes light through the<br>fingertip and creates a waveform<br>from this data. The Lechnologies<br>thumb, wrist and chest sensors<br>require user education to use<br>correctly. The chest sensor requires<br>that the patient removes clothing<br>and is therefore less convenient than<br>a fingertip sensor. |
|-------------------|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Waveform display | Waveform shown in real time on the<br>monitor screen. | Waveform is not shown on the monitor,<br>but is available by a download-from the<br>Lechnologies Research, Inc's website. | While the waveform is available to<br>the user, the subject device is the<br>only device with the display in real<br>time.<br> |
{5}------------------------------------------------
## page 6 of 6
## NON-CLINICAL PERFORMANCE DATA
#### 807.92(b)(1)
#### Safety Testing
The Melys Atrial Fibrillation Monitor is compliant to the safety standards. IEC 60601-1 (1988), 200 Edition, Medical Electrical Equipment, Part 1: General Requirements for Safety +A1(91) + A2(95)
#### Comparative Test
A comparative test was conducted between the subject and predicate device. The Melys Atrial Fibrillation Monitor matched the predicate device in all tests, giving the same heart rate measurements and the same indications towards atrial fibrillation in all but Tests 9 and 10, where the Melys Atrial Fibrillation Monitor correctly indicated that no atrial fibrillation was present.
## Human Factor Study
A Human Factor Study was performed with 20 volunteers to evaluate ease of use and label comprehension. There was a 100% agreement that the device was easy to use and the instructions were clear and understandable.
## CLINICAL PERFORMANCE DATA
Two clinical trials were conducted for verification of Melys Atrial Fibrillation Monitor, safely and effectively identifying as having an arrhythmia those patients identified by EKG examination as having atrial fibrillation.
The Clinical Trial results show the EKG diagnosis compared with the Melys Atrial Fibrillation Monitor predictions.
#### CONCLUSION
## 807.92(b)(3)
807.92(b)(2)
Electrical, mechanical and software testing was conducted and data collected in accordance with applicable standards to ensure that the device performs in accordance with applicable standards to ensure that the device performs according to specification and to verify that the device is suitable for home use. A Human Factor Study was conducted to verify Ease of Use and Label Comprehension
Advanced Fluidics considers the Melys Atrial Fibrillation Monitor to be equivalent to the predicate device. This conclusion is based upon the devices' similarities in functional design, method of use for self-monitoring and indications for use.
{6}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Cantrol Center - WO66-G609 Silver Spring, MD 20993-0002
Image /page/6/Picture/3 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
March 19, 2014
Advanced Fluidics, LLC c/o E J Smith 1468 Harwell Ave Crofton, MD 21114 US
Re: K132206 Trade/Device Name: Melys Atrial Fibrillation Monitor Regulation Number: 21 CFR 870.2920 Regulation Name: Atrial Fibrillation Detector Regulatory Class: Class II Product Code: DXH Dated: February 12, 2014 Received: February 14, 2014
Dear E J Smith,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free
{7}------------------------------------------------
number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
Image /page/7/Picture/4 description: The image shows a logo with the letters FDA. The letters are stylized and overlapping, with a geometric design. The logo is black and white and appears to be a vectorized image.
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) KI32206
#### Device Name Melys Atrial Fibrillation Monitor
#### Indications for Use (Describe)
The Melys Atrial Fibrillation Montor is indicated for self-testing by patients who have been diagnosed with, or are susceptible to developing, atrial florillation and who would like to monitor and record their heart thythms on an internitient basis.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) |
|-----------------------------------------------------------------------------------|
| Concur |
| <div> <img alt="Signature" src="signature.png" style="max-width: 150px;"/> </div> |
| Date: 2014.03.19 |
| 15:24:12 -04'00' |
| for Bram Zuckerman |
| FORM FDA 3881 (1/14) | Page 1 | PRC Publishing Services (301) 443-6740 EF |
|----------------------|--------|-------------------------------------------|
|----------------------|--------|-------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.