Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
Device Story
Device is a gas conditioning insufflator system; attaches to CO2 source; warms, humidifies, and filters gas prior to delivery into surgical cavity. System comprises sterile, single-use disposable (filter, heater/humidifier, tubing, path-of-entry access port) and reusable control module. Access ports (10mm/12mm) include duckbill/tool seals; XLR model features additional cannula seal to prevent endoscope fogging. Obturator includes transparent optical window for visualization during insertion. Used in OR by surgeons; provides conditioned gas to maintain surgical cavity; reduces fogging; improves visualization; facilitates instrument entry.
Clinical Evidence
Bench testing only. Performance testing included flow/pressure characterization, gas temperature and humidity output, insertion/removal force testing, and seal leak integrity. XLR model specifically tested to confirm reduction in endoscope fogging and improved visualization.
Technological Characteristics
Gas conditioning system with inline filter, heater/humidifier, and tubing. Access ports available in 10mm and 12mm configurations. XLR model includes endoscopic cannula seal. Obturator features transparent optical window. System utilizes reusable control module for safety/heating circuits. Single-lumen gas delivery. Sterilized, single-use disposable components.
Indications for Use
Indicated for patients undergoing thoracic, abdominal, or gynecologic minimally invasive endoscopic surgical procedures requiring CO2 insufflation and instrument access.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
# 5.0 510(k) Summary
Insuflow SynergyTM XL and XLR Port
| Date Prepared: | July 12, 2013; updated August 19, 2013 | |
|----------------|------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitter: | Lexion Medical, LLC<br>545 Atwater Circle<br>St. Paul, MN 55103<br>Telephone: 651-855-1447<br>Fax: 651-636-1671 | |
| Contact: | Mr. Bernard Horwath<br>Regulatory Affairs Consultant<br>4486 Timberline Ct<br>Vadnais Heights, MN 55127<br>Telephone: 651-231-1761 | AUG 2 2 2013 |
#### Proprietary Name
Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® SynergyTM Modified Device: XLR Port (10mm and 12mm with endoscopic cannula seal)
Common/Usual Name: Gas Conditioner Insufflator Device with integral path of entry device
Classification Name: Class II per regulations 884.1730, Product Code: HIF Class II per regulations 876.1500, Product Code: GCJ
### Establishment Registration Number: 2135348
#### Predicate Devices:
LEXION Medical believes that Insuflow Synergy™ XL and XLR Port devices are substantially equivalent to the following devices:
- Insuflow Synergy™ Port, KI20640 .
- EndoPath Xcel Trocar, K032676
### Description:
The Insuflow® Synergy™ XL and XLR Port are a gas conditioning device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow Synergy™ XL and XLR Port consist of a sterile, disposable single use device with a filter, heater/humidifier, tubing set, and a path of entry access port device. A reusable control module houses the control and safety circuits for the system.
The technology for gas conditioning and path of entry are the same as the predicate marketed device. Regulated CO2 gas flows into the Insuflow Synergy™ XL and XLR Port, through the inline filter, continues along the tubing to enter the path of entry access device that contains the
{1}------------------------------------------------
#### K132204
#### pg 2 of 3
heating element and humidification media, and through the path of entry access device lumen for delivery into the patient's surgical cavity.
The integral path of entry access device in the Insuflow Synergy™ XL and XLR Port is designed and constructed similarly to the Insuflow® Synergy™ Port but comes in larger sizes just as the predicate EndoPath trocar device. The Insuflow® Synergy™ XL Port and the Insuflow® Synergy™ XLR Port have access devices in 10 mm and 12 mm configurations with a single-lumen working channel configuration for conditioned gas delivery. The access device has a working channel with duckbill and tool seals for instrument entry into the surgical cavity and delivers conditioned insufflation gas through the working channel. The Insuflow® Synergy™ XLR Port has an endoscope compatible configuration that has an additional seal at the bottom of the main cannula, which serves as a means to seal off around the endoscope, forcing all gas flow out of the peripheral gas exhaust ports. This minor change is intended to reduce fogging and improve visualization.
The obturator for the Insuflow Synergy™ XL and XLR Port has a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue layers during insertion.
#### Indications for Use:
The Insuflow® Synergy™ XL Port and the Insuflow® Synergy™ XLR Port have the same intended use as the Insuflow® Synergy™ Port predicate device.
Insuflow Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
### Technological characteristics, comparison to predicate device.
The Insuflow Synergy™ XL and XLR Port utilize the same gas conditioning technology, materials and path of entry access port technology as the predicate Insuflow® Synergy™ Port cleared under 510(k) K120640. Both devices perform a similar gas conditioning function with minor design configuration changes due to the larger trocar. The Insuflow Synergy™ XL and XLR Port path of entry access devices are technically equivalent to the predicate Insuflow Synergy™ Port cleared under 510(k) K120640 but are configured for 10 mm and 12 mm surgical tools where the predicate has a 5 mm configuration. The Insuflow® Svnergy™ Port is configured in 5 mm single and dual lumen configurations, while the Insuflow® Synergy™ XL and XLR Port have only a single lumen configuration, similar to the EndoPath Trocar which also is available in sizes up to 12 mm. Just as with the predicate EndoPath trocar, the obturator for the Insuffow Synergy™ XL and XLR Port has a transparent optical window at the distal end, which when used with an endoscope, provides visibility of individual tissue lavers during insertion.
The Insuflow Synergy™ XLR Port devices are endoscope compatible, utilizing a seal at the bottom of the access device cannula that prevents conditioned gas from fogging the endoscope lens.
{2}------------------------------------------------
K132204
pg 3 of 3
#### Discussion of performance testing.
Extensive performance testing has been conducted to assure that the Insuflow® Synergy™ XL and XLR Port perform in accordance with its specifications and applicable standards. Flow/pressure performance, gas temperature and humidity characterization, insertion/removal testing, and seal leak integrity testing were successfully completed. In addition, for the Insuflow® Synergy™ XLR Port endoscope compatible models, a test was successfully conducted to confirm that the cannula seal configuration does reduce scope fogging and improve visualization.
#### Conclusion:
Based on a comparison to the predicate devices and information provided, it is the conclusion of LEXION Medical that Insuflow Synergy™ XL Port and the Insuflow® Synergy™ XLR Port devices are substantially equivalent to devices already on the market being used for these applications (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2013
Lexion Medical, LLC % Bernard Horwath Regulatory Consultant HRG 4486 Timberline Court Vadnais Heights, MN 55127
Re: K132204
Trade/Device Name: Insuflow® Synergy" XL and Insuflow® Synergy" XLR Port Regulation Number: 21 CFR 870.1730 Regulatory Class: Class II Product Code: HIF, GCJ Dated: July 12, 2013 Received: July 23, 2013
Dear Bernard Horwath,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
Page 2 - Bernard Horwath
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 510(k) Number (if known): K132204
## Insuflow® Synergy™ XL Port and Insuflow® Synergy™ XLR Port : .
#### Indications for Use:
Insuflow® Synergy™ XL Port (10mm and 12mm) and Insuflow® Synergy™ XLR Port (10mm and 12mm with endoscopic cannula seal) have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
Prescription Use x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Office of Device Evaluation (ODE)
# Herbert P. Lerner -S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K132204 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.