K132188 · Fujifilm Medical System U.S.A., Inc. · LLZ · Oct 25, 2013 · Radiology
Device Facts
Record ID
K132188
Device Name
MAMMOGRAPHY VIEWER SMV658
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 25, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ASPIRE Bellus is intended to receive digital mammography images and to display these images on monitors for radiologists' review for diagnostic and screening purposes. To assist radiologists, ASPIRE Bellus provides functions such as image review, measurement, post-processing, film printing, displaying mammography CAD results, and image manipulation. ASPIRE Bellus does not accept lossy compressed mammographic images, which should not be used for primary diagnostic interpretation. Display monitors connected to ASPIRE Bellus for diagnostic interpretation of mammographic images must be cleared for use in digital mammography. All images sent to or imported into the ASPIRE Bellus must conform to regulatory requirements. Image quality must conform to applicable quality guidelines.
Device Story
ASPIRE Bellus (SMV658) is a medical image processing and analysis software workstation; receives digital mammography images via DICOM protocol from acquisition systems and PACS; displays images on cleared diagnostic monitors for radiologist review. Provides tools for image review, measurement, post-processing, film printing, and display of mammography CAD results. Used in clinical settings by radiologists to assist in screening and diagnostic decision-making. Software runs on commercial Windows-compatible computers; facilitates image manipulation and analysis to support clinical workflow.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level functionality, segmentation accuracy, measurement accuracy, interfacing, usability, and serviceability testing. Bench performance testing conducted using actual clinical images to confirm accuracy requirements. All tests passed based on pre-defined requirements.
Technological Characteristics
Medical application software for Windows OS; runs on commercial general-purpose computers. Connectivity via DICOM standard for network image retrieval. Features include image visualization, enhancement, manipulation, and measurement tools. Classified as moderate concern software per FDA guidance.
Indications for Use
Indicated for radiologists reviewing digital mammography images for diagnostic and screening purposes. Contraindicated for use with lossy compressed mammographic images for primary diagnostic interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
FUJIFILM Medical Systems U.S.A., Inc. SMV658 V3.0 510(k)
# 5. 510(k) Summary
## Date Prepared:
September 10, 2013
#### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902
Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin
OCT 2 5 2013
## Device Trade Name:
ASPIRE Bellus / Mammography Viewer SMV658
#### Device Common Name:
Medical Image Processing and Analysis Software
### Regulation Number:
21 CFR 892.2050
#### Device Classification:
Class II
## Device Classification Name :
Picture Archiving and Communications System (PACS)
#### Pane I:
Radiology
#### Product Code :
LLZ
#### Date Received:
TBD
#### Decision Date:
TBD
#### Decision:
TBD
{1}------------------------------------------------
## FUJIFILM Medical Systems U.S.A., Inc. SMV658 V3.0 510(K)
## Predicate Device:
- . MammoWorkstation (K081630), Image Diagnostic International GmbH
## Description of the Device
FUJIFILM's Mammography Viewer SMV658 (V3.0) system receives mammography images directly from FUJIFILM's digital mammography acquisition systems using DICOM protocol and PACS via network. These images are displayed on mammography diagnostic monitors for doctors' review. The SMV658 system works as mammography workstation depending on its configuration (monitors) and license information (software key). The system provides visualization and image enhancement tools (such as image review, measurement, postprocessing, film printing, displaying mammography CAD results, and image manipulation) to assist the radiologists' review of mammography images for diagnostic and screening purposes.
## Indication for Use
ASPIRE Bellus is intended to receive digital mammography images and to display these images on monitors for radiologists' review for diagnostic and screening purposes.
· To assist radiologists, ASPIRE Bellus provides functions such as image review, measurement, post-processing, film printing, displaying mammography CAD results, and image manipulation.
ASPIRE Bellus does not accept lossy compressed mammographic images, which should not be used for primary diagnostic interpretation. Display monitors connected to ASPIRE Bellus for diagnostic interpretation of mammographic images must be cleared for use in digital mammography. All images sent to or imported into the ASPIRE Bellus must conform to regulatory requirements. Image quality must conform to applicable quality guidelines.
## Technological Characteristics
The proposed Mammography Viewer SMV658 and the predicate device, MammoWorkstation (K081630), are medical application software running on Windows operating system installed on commercial general-purpose Windows-compatible computers. These devices are connected to mammography systems with DICOM standard and retrieve image data via network communications. These devices provide image visualization, enhancement, and manipulation tools for mammography images with various user interfaces and measurement tools for analysis of rendered images. Both the Mammography Viewer SMV658 and the predicate devices support the workflows, UI, and reporting functions for supporting the radiologists' review of mammography images for screening and diagnosis purposes.
Mammography Viewer SMV658 introduces no new safety or efficacy issues other than those already indentified with the predicate device. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices."
{2}------------------------------------------------
FUJIFILM Medical Systems U.S.A., Inc. SMV658 V3.0 510(k)
## Testing
Mammography Viewer SMV658 is tested successfully with reference to its Software Requirements Specification, as well as design verification and validation documents and Traceability Matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Mammography Viewer SMV658 software, which is found to be safe and effective and substantially equivalent to the currently cleared predicate devices.
Testing involved system level functionality test, segmentation accuracy test, measurement accuracy test, interfacing test, usability test, serviceability test, labeling test, as well as the test for risk mitigation method analyzed and implemented in the risk management process. In addition, we conducted the bench performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.
Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.
## Conclusion
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol with three lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the bird symbol.
Public Health Service
Foud and Drug Administration 10903 New Hampshire Avenue Document Control Center - WObo6-Gods Silver Spring, MD 20993-0002
October 25. 2013
FUJIFILM Medical System U.S.A., Inc. % Jyh-Shyan Lin, Ph.D. Senior Manager, Regulatory, Quality and Clinical Affairs 419 West Avenue STAMFORD CT 06902
Re: K132188
Trade/Device Name: Mammography viewer SMV658/ASPIRE Bellus Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ. Dated: October 2, 2013 Received: October 4, 2013
Dear Dr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. "The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRFI does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Dr. Lin
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safey/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
ரொ
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132188
Device Name: ASPIRE Bellus/ Mammography Viewer SMV658
Indications for Use:
ASPIRE Bellus is intended to receive digital mammography images and to display these images on monitors for radiologists' review for diagnostic and screening purposes. To assist radiologists, ASPIRE Bellus provides functions such as image review, measurement, post-processing, film printing, displaying mammography CAD results, and image manipulation.
ASPIRE Bellus does not accept lossy compressed mammographic images, which should not be used for primary diagnostic interpretation. Display monitors connected to ASPIRE Bellus for diagnostic interpretation of mammographic images must be cleared for use in digital mammography. All images sent to or imported into the ASPIRE Bellus must conform to regulatory requirements. Image quality must conform to applicable quality guidelines.
Prescription Use メ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$\sin(x)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 510(k) == K132188
Page 1 of _I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.