K132164 · Merit Medical Systems, Inc. · DYB · Sep 17, 2013 · Cardiovascular
Device Facts
Record ID
K132164
Device Name
HEARTSPAN STEERABLE INTRODUCER KIT
Applicant
Merit Medical Systems, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Sep 17, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The HeartSpan™ Steerable Introducer Kit is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Device Story
HeartSpan™ Steerable Introducer Kit facilitates catheter introduction into cardiac anatomy. Components include steerable sheath, dilator, and guide wire. Sheath features hemostasis valve to minimize blood loss, sideport for aspiration/infusion/pressure monitoring, and handle with rotating knob for >180° tip deflection (10-56 mm radius). Radiopaque tip marker aids fluoroscopic visualization. Used in hospitals/healthcare facilities by clinicians. Device is sterile, single-use. Sheath wall encapsulates stainless steel braid reinforcement and deflection wires. Provides access to left heart via interatrial septum.
Clinical Evidence
Bench testing only. Verification testing included tip articulation, leak resistance, kink resistance, dimensional analysis, curve radius, marker band location, detachment force, and side hole performance. Biocompatibility testing performed per ISO 10993 standards (genotoxicity, cytotoxicity, irritation, sensitization, systemic toxicity). Sterilization validated per ISO 11135-1.
Technological Characteristics
Catheter introducer; 8.6F diameter; 74cm length. Materials: polymers with stainless steel braid reinforcement and deflection wires. Features: hemostasis valve, sideport with three-way stopcock, rotating knob for >180° bi-directional deflection. Radiopaque tip marker. Lubricious coating. Sterilized via ethylene oxide. Complies with ISO 10993, ISO 11070, and ISTA P2A standards.
Indications for Use
Indicated for patients requiring introduction of cardiovascular catheters into the heart, including left-sided access via the interatrial septum.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Merit Medical Systems, Inc. HeartSpan™ Steerable Introducer Kit Special 510(k)
Attachment A 510(k) Summary
## 510(k) Summary Correspondent Name: Merit Medical Systems, Inc. Address: 65 Great Valley Parkway Malvern, PA 19355 General Telephone Number: 610-651-5000 SEP 1 7 2013 Provisions Fax Number: 610-651-5003 Contact Person: Alina Stubbs Date of Preparation: July 08, 2013 Registration Number: 2529252 Trade Name: HeartSpan™ Steerable Introducer Kit Subject Common/Usual Name: Steerable Introducer Kit Device Classification Name: Catheter Introducer (21 CFR §870.1340) Trade Name: HeartSpan™ Steerable Introducer Kit Classification Name: Catheter Introducer (21 CFR §870.1340) Premarket Notification: K122431 Predicate Manufacturer: Merit Medical Systems, Inc. Device 65 Great Vallev Parkway Malvern, PA 19355 (formerly operating as Thomas Medical Products, Inc.) Class II 21 CFR §870.1340 Classification FDA Product Code: DYB Review Panel: Cardiovascular The HeartSpan™ Steerable Introducer Kit is indicated for introducing Intended Use various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
{1}------------------------------------------------
Merit Medical Systems, Inc. HeartSpan™ Steerable Introducer Kit Special 510(k)
Device
Description
Attachment A 510(k) Summary
The HeartSpan™ Steerable Introducer set consists of a dilator, guide wire, and steerable sheath, which are designed for catheter introduction into the cardiac anatomy. The device is provided sterile (ethylene oxide) and intended for single use only. It is for use in hospitals or healthcare facilities.
The steerable introducer contains a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The introducer handle includes a rotating knob to enable clockwise and counterclockwise tip deflection >180° with a radius of 10 - 56 mm (measured at 180° deflection) . The steerable introducer also includes distal holes to facilitate aspiration and minimize cavitation, a radiopaque tip marker to improve fluoroscopic visualization, an atraumatic soft tip, and a lubricious coating on the inner and outer surfaces. The dilator is designed to conform to the inner diameter of the sheath, and has a tapered tip.
The materials of construction are primarily polymers with the exception of stainless steel braid reinforcement and deflection wires in the introducer shaft that are completely encapsulated in the sheath wall and do not contact the patient or bodily fluids.
<< This space is intentionally left blank>>
{2}------------------------------------------------
・
| Comparison to<br>Predicate | Technical<br>Characteristics | Predicate Device<br>(K122431) | Modified Device<br>(K132164) |
|--------------------------------------------------|------------------------------------------------------|------------------------------------------|--------------------------------------------|
| | Hemostasis valve provided | Yes | Yes |
| | Dilator to Guide wire<br>Compatibility | up to 0.038" | up to 0.038" |
| | Guide wire Design /<br>Dimensions | Standard<br>Non-Coated<br>0.038" x 135cm | Super-Stiff<br>with PTFE<br>0.032" x 180cm |
| | Compatibility with<br>Standard Transseptal<br>Needle | Yes | Yes |
| | Length | 74 cm | 74 cm |
| | French size | 8.6F | 8.6F |
| | Bi-Directional Curving | Yes<br>≥180° in both<br>directions | Yes<br>≥180° in both<br>directions |
| | One-Handed Operation of<br>Curving Mechanism | Yes | Yes |
| | Wire braid reinforcement<br>completely encapsulated | Yes | Yes |
| | Radiopaque tip or marker | Yes | Yes |
| | Soft Atraumatic Tip | Yes | Yes |
| Side port for infusion and<br>contrast injection | Yes | Yes | |
.
Summary of the technological characteristics of the device compared to the predicate devices:
{3}------------------------------------------------
The HeartSpan™ Steerable Introducer Kit has been thoroughly tested through verification of product specifications and user requirements. The following quality assurance measures were applied during the development of the HeartSpan™ Steerable Introducer Kit:
- Risk Analysis .
.
.
- Requirements/Specification Reviews
- . Design Reviews
- Performance Testing (Verification) including but not limited to: .
Safety & Performance Tests
- Tip articulation o
- Leak resistance o
- Kink resistance O
- Length 0
- Inner/Outer diameter O
- Curve radius 0
- Marker band location 0
- Tip/tube detachment force O
- Side hole diameter and location o
- Sterilization validation (ethylene oxide)
- Biocompatibility Testing (Verification) .
The results of the testing demonstrated that the modified HeartSpan™ Steerable Introducer Kit meets the predetermined acceptance criteria applicable to safety and efficacy of the device.
<<This space is intentionally left blank>>
{4}------------------------------------------------
Merit Medical Systems, Inc.
HeartSpan™ Steerable Introducer Kit
Special 510(k)
| International<br>Standards | No performance standards have been established under section 514<br>of the Food, Drug and Cosmetic Act for this device. Performance<br>testing of the modified HeartSpan™ Steerable Introducer Kit was<br>conducted based on the risk analysis and based on the requirements<br>of the following international standards:<br>• ANSI/AAMI/ISO 10993-3:2003 Biological Evaluation of Medical<br>Devices - Part 3: Tests of Genotoxicity, Carcinogenicity, and<br>Reproductive Toxicity<br>• ANSI/AAMI/ISO 10993-4:2002 Biological Evaluation of Medical<br>Devices - Part 4: Selection of Tests for Interactions with Blood<br>• ANSI/AAMI/ISO 10993-5:1999 Biological Evaluation of Medical<br>Devices - Part 5: Tests for In Vitro Cytotoxicity<br>• ANSI/AAMI/ISO 10993-7: 2008, Biological Evaluation of Medical<br>Devices - Part 7: Ethylene Oxide Sterilization Residuals<br>• ANSI/AAMI/ISO 10993-10: 2010, Biological Evaluation of Medical<br>Devices - Part 10: Tests for Irritation and Skin Sensitization<br>• ANSI/AAMI/ISO 10993-11: 2006, Biological Evaluation of Medical<br>Devices - Part 11: Tests for Systemic Toxicity<br>• ISO 11070:1998 Sterile Single-Use Intravascular Catheter<br>Introducers - Annex B<br>• ANSI/AAMI/ISO 11135-1:2007 Sterilization of Health Care<br>Products - Ethylene oxide - Part 1: Requirements for<br>Development, Validation and Routine Control of Sterilization<br>Process for Medical Devices<br>• ISTA P2A (2011) - Packaged-Products 150 lb (68kg) or Less |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence | Merit Medical Systems, Inc. considers the HeartSpan™ Steerable<br>Introducer Kits to be substantially equivalent to the currently<br>marketed predicate device (HeartSpan™ Steerable Introducer Kit<br>K122431). This assessment is based upon analysis of similar<br>technological characteristics, bench testing, and indications for use. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2013
Ms. Alina Stubbs Regulatory Affairs Specialist II Merit Medical Systems, Inc. 65 Great Valley Parkway Malvern, PA 19355
Re: K132164
Trade Name: HeartSpan™ Steerable Introducer Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: II Product Code: DYB Dated: August 16, 2013 Received: August 19, 2013
Dear Ms. Stubbs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{6}------------------------------------------------
Page 2 - Ms. Alina Stubbs
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
## Section 4 Indications for Use Statement
510(k) Number (if known): K132164
HeartSpan™ Steerable Introducer Kit Device Name:
Indications for Use:
The HeartSpan™ Steerable Introducer Kit is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/7/Picture/12 description: The image shows the text "Bram D. Zuckerman -S 2013.09.17 16:01:06 -04'00'". The text appears to be a timestamp or some other kind of identifier. The text is black and the background is white. The text is arranged in two lines.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.