K132160 · Toshibamedical Systems Corporation · LNH · Oct 16, 2013 · Radiology
Device Facts
Record ID
K132160
Device Name
VANTAGE TITAN 3T
Applicant
Toshibamedical Systems Corporation
Product Code
LNH · Radiology
Decision Date
Oct 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of noncontrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density) - Spin-lattice relaxation time (T1) . - Spin-spin relaxation time (T2) . - Flow dynamics . - . Chemical Shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Titan 3T is a 3 Tesla MRI system. This update (v2.30) adds five software functionalities: updated SAR calculation for increased slice flexibility; 3D Arterial Spin Labeling (ASL) for volumetric imaging; Advanced Moving Bed (AMB) for station-specific scan settings; 3D MR Spectroscopy (MRS) for multi-voxel proton data acquisition; and distortion correction for 3D volumes. Operated by trained clinicians in clinical settings. System processes proton density, T1, T2, flow, and chemical shift data to generate diagnostic images. Output interpreted by physicians to aid diagnosis. Benefits include improved scan efficiency, volumetric imaging capabilities, and reduced image distortion.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance verified through adherence to recognized consensus standards (IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-33, 60825-1, 62304, 62366) and internal design control risk assessments.
Technological Characteristics
3T MRI system. Software-based enhancements: SAR calculation, 3D ASL, AMB, 3D MRS, and 3D volume distortion correction. Compliance with IEC 60601-2-33:2010. Connectivity: Standard MRI system architecture. Sterilization: N/A (non-invasive diagnostic imaging).
Indications for Use
Indicated for diagnostic imaging of head or body anatomic structures in patients requiring cross-sectional transaxial, coronal, sagittal, or oblique MRI, including non-contrast MRA. Contraindications include standard MRI safety exclusions (e.g., incompatible implants).
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K132160
Toshiba America Medical Systems, Pre-Market Notification 510(k) Vantage Titan 3T, MRT-3010/A5
#### 510(k) SUMMARY AND EFFECTIVENESS
### 1. CLASSIFICATION and DEVICE NAME:
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number: | 21 CFR 892.1000 |
| Product Code: | LNH |
| Trade Proprietary Name: | Vantage Titan 3T |
| Model Number: | MRT-3010/A5 |
#### 2. ESTABLISHMENT REGISTRATION: 2020563
#### 3. CONTACT PERSON, U.S AGENT and ADDRESS:
Contact Person Charlemagne Chua Manager, Regulatory Affairs (714) 669-7896
OCT 16 2013
U.S. Agent Name: Paul Biggins Director, Regulatory Affairs (714) 730-5000
#### Establishment Name and Address:
Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, Ca. 92780
#### 4. MANUFACTURING SITE:
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
#### 5. DATE OF SUBMISSION:
July 10, 2013
#### 6. DEVICE DESCRIPTION:
The Vantage Titan 3T (Model MRT-3010/A5) is a 3 Tesla Magnetic Resonance Imaging (MRI) System and was cleared under K120487. This submission will include the following
Page 1 of 4
{1}------------------------------------------------
five software functionalities: Changes in the SAR calculation method, 3D ASL (3 dimensional Arterial Spin Labeling), Advanced Moving Bed (AMB), 3D MRS (3 dimensional MR Spectroscopy) and Distortion correction for entire volume are added to Vantage Titan 3T.
# 7. SUMMARY OF HARDWARE CHANGES
There are no major hardware changes associated with the software change.
# 8. SUMMARY OF SOFTWARE CHANGES
Existing software packages are grouped by functions of software and pulse sequences. New software (V2.30) packages (four software functionalities) as follows:
# a) Changes in the SAR calculation method
The new calculation method provides more flexibility in setting scan parameters, which enables scan with increased number of slices per unit time etc.
# b) 3D ASL (3 dimensional Arterial Spin Labeling)
This function of software adds a new feature for current ASL imaging so that images can be acquired from 3D volume instead of 2D slice.
# c) Advanced Moving Bed (AMB)
AMB enables individual scan settings for each station when Moving Bed is used. It allows acquisition of images with the most appropriate settings for each station.
# d) 3D MRS (3 dimensional MR Spectroscopy)
This function of software enables the acquisition of proton spectroscopic data from multiple voxel in a 3 dimensional volume of an object by applying phase encode aradients in three orthogonal directions.
# e) Distortion correction for entire volume
This function of software performs correction of image distortion due to gradient magnetic field non-linearity for entire 3D volume including slice direction.
# 9. SAFETY PARAMETERS
| Item | Vantage Titan 3T with<br>new application software<br>package<br>(subject device) | Vantage Titan 3T ,<br>K120487<br>(Predicate Device) | Notes |
|-----------------------|----------------------------------------------------------------------------------|-----------------------------------------------------|-------|
| Static field strength | 3T | 3T | Same |
| Operational Modes | 1st Operating Mode | 1st Operating Mode | Same |
{2}------------------------------------------------
Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) Vantage Titan 3T, MRT-3010/A5
| i. Safety parameter<br>display | SAR dB/dt | SAR dB/dt | Same |
|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------|
| ii. Operating mode<br>access<br>requirements | Allows screen access to<br>1st level operating mode | Allows screen access to<br>1st level operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified<br>in<br>IEC 60601-2-33(2010)) | 4W/kg for whole body (1st<br>operating mode specified<br>in<br>IEC 60601-2-33(2002)) | Change* |
| Maximum dB/dt | <1st operating mode<br>specified in<br>IEC 60601-2-33 (2010) | <1st operating mode<br>specified in<br>IEC 60601-2-33 (2002) | Change* |
| Potential<br>emergency<br>condition and<br>means provided for<br>shutdown | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard for<br>ferromagnetic objects | Shut down by Emergency<br>Ramp Down Unit for<br>collision hazard for<br>ferromagnetic objects | Same |
*Note: The difference between predicate and subject device is due to the conformance of the subject device to IEC 60601-2-33 (2010)
#### 10. IMAGING PERFORMACE PARAMETERS
No change from the previous predicate submission (K120487).
#### 11. INTEDED USE
Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of noncontrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- . Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1) .
- Spin-spin relaxation time (T2) .
- Flow dynamics .
- . Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
No changes to the previously cleared indication (K120487).
Page 3 of 4
{3}------------------------------------------------
### 12. DESIGN CHANGE
Following software package addition to Vantage Titan 3T (K120487).
- a) Changes in the SAR calculation method
- b) 3D ASL (3 dimensional Arterial Spin Labeling)
- c) Advanced Moving Bed (AMB)
- d) 3D MRS (3 dimensional MR Spectroscopy)
- e) Distortion correction for entire volume
### 13. SUMMARY OF DESIGN CONTROL ACTIVITIES
PS Risk List for software of changing packages are attached. The test methods used are the same as those submitted in the previously cleared submissions (K120487).
### 14. TRUTHFUL AND ACCURACY CERTIFICATION
A certification of the truthfulness and accuracy of the Vantage Titan 3T described in this submission is provided in this submission.
### 15. SUBSTANTIAL EQUIVALENCE
Toshiba Medical Systems Corporation believes that the Vantage Titan 3T (model MRT-3010/A5) Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate devices referenced in this submission.
Testing was done in accordance with applicable recognized consensus standards as listed below.
#### List of Applicable Standards
- · IEC60601-1:2005
- · IEC60601-1-2:2007
- · IEC60601-1-8:2003,Amd.1:2006
- · IEC60601-2-33:2010
- · IEC60825-1: 2007
- · IEC62304:2006
- · IEC62366:2007
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 16, 2013
TOSHIBA MEDICAL SYSTEMS CORPORATION % Mr. PAUL BIGGINS DIRECTOR REGULATORY AFFAIRS 2441 MICHELLE DRIVE TUSTIN CA 92780
Re: K132160
Trade/Device Name: Vantage Titan 3T, MRT-3010/A5 v2.30 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: September 6, 2013 Received: September 9, 2013
Dear Mr. Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Biggins
If you desire specific advice for your device on our labeling regulation (2) CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Michael D. O'Hara
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
l'or
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K132160
Device Name:
Vantage Titan 3T, v2.30 (MRT-3010/A5)
Indications for Use:
Vantaqe Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display analomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear mağnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Michael D. O'Hara
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.