← Product Code [OCC](/productcode/OCC) · K132159

# PROADENO+ ASSAY (K132159)

_Gen-Probe Prodesse, Inc. · OCC · Aug 14, 2013 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132159

## Device Facts

- **Applicant:** Gen-Probe Prodesse, Inc.
- **Product Code:** [OCC](/productcode/OCC.md)
- **Decision Date:** Aug 14, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.3980
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K132159 · Aug 14, 2013 | PROADENO+ ASSAY | Gen-Probe Prodesse, Inc. | Retrospective pre-selected archived nasopharyngeal swab (NPS) specimens | Retrospective archived clinical specimens were used to compare the performance of the modified Universal Internal Control (UIC-A) against the current internal control (UIC-P) and original laboratory results to ensure the device modification did not alter assay performance. | Retrospective specimens; Archived clinical samples; Assay validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective comparative study | Archived nasopharyngeal swab (NPS) specimens; Sample Size: 330 retrospective samples (plus 36 contrived samples) | Current Universal Internal Control (UIC-P) and original laboratory results (Luminex RVP) | Assay performance equivalency |

## Indications for Use

The Prodesse® ProAdeno®+ Assay is a multiplex Real Time PCR in vitro diagnostic test for the qualitative detection of human Adenovirus (HAdV) DNA isolated and purified from nasopharyngeal (NP) swab specimens obtained from individuals exhibiting signs and symptoms of acute respiratory infection. This test is intended for use to aid in the diagnosis of HAdV infections in humans in conjunction with other clinical and laboratory findings. The test detects, but does not differentiate, serotypes 1-51. Negative results do not preclude HAdV infection and should not be used as the sole basis for treatment or other patient management decisions.

## Device Story

Prodesse ProAdeno+ Assay is a multiplex nucleic acid amplification test for human adenovirus detection. Input: nasopharyngeal swab specimens. Process: nucleic acid extraction (via bioMérieux NucliSENS easyMAG or Roche MagNA Pure LC) followed by PCR amplification on Cepheid SmartCycler II. Output: qualitative detection of adenovirus DNA. Used in clinical laboratories by high-complexity personnel. Modifications include: outsourcing of Universal Internal Control (UIC-A) manufacturing, changing control vector to RNA IVT/DNA plasmid, increasing DNA control concentration, and eliminating end-user dilution of positive controls. Healthcare providers use results to aid in respiratory infection diagnosis. Benefits include standardized internal control across Pro+ series assays and simplified positive control handling.

## Clinical Evidence

Retrospective clinical comparison study demonstrated the modified assay meets performance claims of the current ProAdeno+ Assay. Analytical studies (sensitivity, IC interference, extractor equivalency, sample stability) and a Positive Control Effectiveness Study confirmed performance functionality.

## Technological Characteristics

Multiplex nucleic acid amplification assay. Components: Universal Internal Control (RNA IVT and DNA plasmid), Positive Control (DNA). Instrumentation: bioMérieux NucliSENS easyMAG or Roche MagNA Pure LC extractors; Cepheid SmartCycler II thermal cycler. Storage: ≤ -70°C. High-complexity user environment.

## Regulatory Identification

A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.

## Special Controls

*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.

## Predicate Devices

- Prodesse® ProAdeno™+ Assay ([K102952](/device/K102952.md))
- Prodesse ProFlu+ Assay ([K110968](/device/K110968.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification Review Memorandum

To: Hologic, Inc. (Gen-Probe Prodesse, Inc.)

RE: K132159

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II devices requiring 510(k). The following items are present and acceptable:

1. The name and 510(k) number of the SUBMITTER'S previously cleared device:

Prodesse® ProAdeno™+ Assay

510(k) number: K102952

2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use and package labeling.

3. A description of the device MODIFICATION(S) to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

The 510k submission contained modifications to the Internal Control and Positive Controls. The modifications are summarized as follows:

a. Outsourcing of the manufacturing of the Internal Control resulted in minor changes made to the current Internal Control (UIC-P);

- Modification made to the current Universal Internal Control (UIC-P); The current Universal Internal Control (UIC-P) in ProAdeno+ Assay contains a dsDNA PCR product. The modified Universal Internal Control (UIC-A) will contain a RNA in vitro transcript (IVT) and a DNA plasmid to allow users to perform one nucleic acid extraction and test with any combination of the Pro+ Series Assays including ProFlu+, ProhMPV+, ProParaflu+, ProFAST+, and ProAdeno+.
- Minor changes made to the UIC-A sequence;
Due to the different vector being used in the modified Universal Internal Control (UIC-A), minor changes were made to the 5' and 3' ends of the UIC-A sequence.
- The concentration of the dsDNA control is increased from $10^{4}$ copies/μL to $5 \times 10^{4}$ copies/μL. Handling of the modified Universal Internal Control is identical to that of the current Universal Internal Control (UIC-P) included in the ProAdeno+ Assay.

b. The handling of the Positive Controls for the ProAdeno+ Assay will be changed to eliminate the customer dilution that occurs immediately prior to PCR setup, effectively raising the testing concentration one log.

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics.

|  Similarities  |   |   |
| --- | --- | --- |
|  Element | Modified Prodesse ProFlu+ Assay | Current Prodesse ProFlu+ Assay (K110968)  |
|  Organisms Detected | Same | Human adenovirus  |
|  Analyte | Same | DNA  |
|  Technological Principles | Same | Multiplex nucleic acid amplification  |

{1}

Page 2 of 4

|  Specimen | Same | Nasopharyngeal Swab  |
| --- | --- | --- |
|  Types |  |   |
|  User Complexity | Same | High  |
|  Sample Preparation | Same | Up front sample processing is required to extract nucleic acid.  |
|  Method |  |   |
|  Instrumentation | Same | bioMérieux NucliSENS easyMAG or Roche MagNA Pure and Cepheid SmartCycler II Instrument  |
|  Time to result | Same | Approximately 3 hours  |
|  Controls | Same | Internal control in each sample. External control processed with each batch of samples. (see below for differences)  |
|  Differences  |   |   |   |
| --- | --- | --- | --- |
|  Element |   | Modified ProFlu+ Assay | Current Prodesse ProFlu+ Assay  |
|  Controls | Internal | Control Stocks outsourced
- Change in manufacturer leading to change in control vectors and minor sequence change at the 5' and 3' ends of RNA IVT and change from PCR product to plasmid.
- The concentration of the DNA control was increased from 10^{4} copies/μL to 5 x 10^{4} copies/μL. | Control stocks manufactured in house  |
|   |  Positive | PC does not require dilution;
- PC is provided as “at use concentration” | End user must dilute PC 1:10 prior to use for RT-PCR  |

# 5. A Design Control Activities Summary:

a. Identification of Risk Analysis method used to assess the impact of the modification on the device.

FMEA was performed to determine whether the current design changes create new risks or failure modes or affect the risk priority number (RPN) value. No additional risk or change in RPN value was identified in the Risk Analysis.

b. To demonstrate that the modifications in Universal Internal Control (UIC-A) do not change the assay performance, Analytical Studies and a Comparison Study were conducted.

- Analytical Performances:

- Analytical Sensitivity Confirmation

LoD, which was established in K102952 in 2010, was confirmed for Adenovirus using three strains of adenovirus tested side by side with the modified Universal Internal Control (UIC-A) and with the current Universal Internal Control (UIC-P). The confirmed LoDs are as follows:

HAdV-3 1X 10<sup>0</sup> TCID<sub>50</sub>/mL

{2}

Page 3 of 4

HAdV-4 1X 10⁻¹ TCID₅₀/mL
HAdV-19 1X 10⁰ TCID₅₀/mL

- IC Interference Study
The IC Interference Study demonstrated that the new control, UIC-A, did not inhibit the detection of target organisms at levels close to the LoD.

- Sample Stability Study
The study demonstrated that the stability of the samples would not be affected by a change in the internal control.

- Extractor Equivalency Studies
The equivalency of nucleic acid extraction methods between the bioMérieux NucliSENS easyMAG automated extractor and Roche MagNA Pure LC extractor were evaluated by spiking the cultured and tittered strain of HAdV-3 into a negative nasopharyngeal swab (NPS) matrix pool at the confirmed LoD concentration. The study demonstrated equivalency between the two extraction methods.

- Comparison Study:
The comparison study was conducted for all Pro+ Series Assays including ProFlu+, ProhMPV+, ProParaflu+, ProFAST+, and ProAdeno+ using 366 positive samples and 66 negative samples. Among the 366 positive samples, 330 were retrospective pre-selected archived NPS specimens with 30 positive samples per target (11 targets total) and 36 were contrived samples, generated by spiking individual negative retrospective NPS samples with whole organism (Influenza A/Seasonal H1 or Parainfluenza 2). Each sample was split into 3 aliquots; one aliquot was tested using the current Internal RNA Control (RIC), one aliquot was tested using the new Universal Internal Control (UIC-A), and one aliquot was tested using the current Universal Internal Control (UIA-P) for ProAdeno+ Assay. All samples were then split into 72 panels with 6 samples per panel, extracted and tested by four different operators. Half of the panel samples were extracted using the bioMérieux NucliSENS easyMAG method and the other half using the Roche MagNA Pure LC method. The samples spiked with the current UIC-P and samples spiked with the modified UIC-A were also compared to the original source laboratory results when the specimens were tested with Luminex RVP. The study demonstrated that the modified UIC-A did not change the assay performance.

The ProAdeno+ Assay package insert has been updated to reflect the changes in the internal control.

c. To assess the stability of the modified Universal Internal Control (UIC-A), an accelerated stability study for the UIC-A stored at -70°C and a freeze-thaw stability study were conducted with one lot of each assay component. The studies demonstrated that the modified UIC-A can be stored at the same condition (≤ -70°C for 18 months with up to 2 freeze thaw cycles) as the current Universal Internal Control (UIC-P).

d. To evaluate whether eliminating customer dilution prior to PCR setup will affect the effectiveness of the Positive Control at detecting any errors occurred in the FAM (HAdV) Channel, several defective mixtures that either do not have the Taq polymerase, or do not have enough reverse primers, or contain a PCR inhibitor, were tested with Positive Control at supplied concentration without customer dilution. The study demonstrated that the Adenovirus DNA Control used at the supplied concentration can effectively detect any global errors in the FAM (HAdV) Channel. The Pro Adeno+ Assay package insert has been updated to reflect the changes in the Positive Control.

{3}

Page 4 of 4

e. A declaration of conformity with design controls was submitted for the manufacturing facility which includes:

i) A statement signed by the Senior Director of R &amp; D, Gen-Probe Prodesse, Inc., was submitted confirming that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and

ii) A “Declaration of Conformity” statement signed by the Associate Director of Quality and Regulatory, Gen-Probe Prodesse, Inc., was submitted stating that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.

6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure.

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

---

**Source:** [https://fda-staging.innolitics.com/device/K132159](https://fda-staging.innolitics.com/device/K132159)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
