← Product Code [KNF](/productcode/KNF) · K132135

# LINA GOLD LOOP HC (K132135)

_Lina Medical Aps · KNF · Jan 31, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132135

## Device Facts

- **Applicant:** Lina Medical Aps
- **Product Code:** [KNF](/productcode/KNF.md)
- **Decision Date:** Jan 31, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The LiNA Gold Loop HC is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynaecological procedures involving endoscopic hysterectorny. It is used with standard electrosurgical generators.

## Device Story

5mm single-use laparoscopic instrument; monopolar electrosurgical device. Used in gynecological procedures for soft tissue cutting/removal during endoscopic hysterectomy. Operated by surgeons in clinical settings. Connects to standard electrosurgical generators. Features hand-activation switch on handle; gold-plated cutting wire; shrink tubing. Device enables tissue excision via electrical current; provides surgeon control via integrated switch. Benefits include precise tissue removal during minimally invasive surgery.

## Clinical Evidence

No clinical testing performed. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), and sterilization validation (ISO 11135-1).

## Technological Characteristics

5mm diameter laparoscopic instrument; monopolar electrosurgical energy. Materials include gold-plated cutting wire and shrink tubing. Single-use; ethylene oxide sterilized. Hand-activated switch integrated into handle. Compatible with standard electrosurgical generators. Complies with ISO 10993-1, IEC 60601-1, and ISO 11135-1.

## Regulatory Identification

A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.

## Predicate Devices

- LiNA Loop ([K070315](/device/K070315.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### LiNA Medical Premarket Notification Special 510(k): LiNA Gold Loop HC

K132135 Page 1 of 2

510(k) Summary

5

1. Submission Sponsor LINA Medical ApS Formervangen 5 2600 Glostrup Denmark Phone: +45 43 29 66 66 Fax: -45 43 29 66 99 Contact: Louisa Memborg, Regulatory Affairs Officer

### 2. Submission Correspondent

Christine E. Nichols RAC Boston Biomedical Associates 386 West Main Street. Suite 7 Northborough, MA 01532

- 3. Date Prepared July 10, 2013
## 4. Device Identification

Trade/Proprietary Name:

Common/Usual Name:

Classification Name:

Classification Regulation:

Product Code: Device Class: Classification Panel: LiNA Gold Loop HC

Hand Controlled Loop

coagulator-cutter, endoscopic. unipolar (and accessories) 21 CFR 884.4160

KNF Class II Obstetrics and Gynecology Panel

#### Predicate Devices to which S.E is claimed: క.

LiNA Loop K070315

#### 6. Device Description

The LiNA Gold Loop HC is a 5 mm single use laparoscopic instrument. It is available with three different loop dimensions; EL-160-H (160mm x 80mm), EL-200-H (200mm x 100mm) and EL-240-H (240mm x 120mm), with an uninsulated cutting area of 30mm. The device is single use ethylene oxide sterilized and is compatible with most standard electrosurgical generators that provide a monopolar outlet.

Page 18 of 42

JAN 3 1 2014

{1}------------------------------------------------

#### 7. Intended Use

The intended use has not been changed and is identical to the intended use of the LiNA Loop. The LiNA Gold Loop HC is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynaecological procedures involving endoscopic hysterectorny. It is used with standard electrosurgical generators.

#### 8. Comparison of Technological Characteristics

This special 510(k) is a modification to the LiNA Loop previously cleared by the FDA with the 510(k) number (K070315). The modifications include: the addition of the hand activation switch on the handle of the device; and a material change to the shrink tubing and gold plating on the cutting wire. No changes were made to the intended use, indications for use. energy type, performance specifications, sterilization method or fundamental scientific technology.

#### 9. Non-Clinical Performance Data

Testing shows that the modified device complies with the same testing requirements as the predicate device which includes compliance to 1SO10993-1:2009. IEC60601-1:2006 and ISO11135-1:2007.

The LiNA Gold Loop HC passed all the testing in accordance with national and international standards.

The LiNA Gold Loop HC testing supports the claim of substantial equivalence.

#### 10. Clinical Testing

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device LiNA Loop. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

#### 11. Statement of Substantial Equivalence

The differences between the LiNA Gold Loop HC and the LiNA Loop do not raise any new questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the LiNA Gold Loop HC is substantially equivalent to the relevant aspects of the predicate device LiNA Loop in terms of design. components, principals of operation, sterilization, biocompatibility, performance characteristics, and intended use. LiNA Gold Loop HC. as designed is determined to be substantially equivalent to the referenced predicate device.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes that resemble a caduceus or a stylized representation of human figures.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 31, 2014

LiNA Medical ApS % Christine Nichols RAC Regulatory Affairs Manager Boston Biomedical Associates 386 West Main Street, Suite 7 Northborough, MA 01532

Re: K132135

Trade/Device Name: LiNA Gold Loop HC Regulation Number: 21 CFR& 884.4160 Regulation Name: Unipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: KNF Dated: January 13, 2014 Received: January 14, 2014

Dear Christine Nichols,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Christine Nichols RAC

Sales and Comments of

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### LiNA Medical Premarket Notification Special 510(k): LiNA Gold Loop HC

#### 4 Indications for Use Statement

510(k) Number (if known): K132135

Device Name: LiNA Gold Loop HC

Indications for Use:

The LiNA Gold Loop HC is a 5mm monopolar electrosurgical device intended for cutting and removal of soft tissue in gynecological procedures involving endoscopic hysterectomy. It is used with standard electrosurgical generators.

Prescription Use_ X

AND/OR

Over-The-Counter Use _

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert 2014.01

Page 17 of 42

Confidential and Proprietary to LiNA Medical

---

**Source:** [https://fda-staging.innolitics.com/device/K132135](https://fda-staging.innolitics.com/device/K132135)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
