K132124 · Klarity Medical Products, LLC · IYE · Nov 1, 2013 · Radiology
Device Facts
Record ID
K132124
Device Name
KLARITY VACUUM BAG, KLARITY MOLD CUSHION
Applicant
Klarity Medical Products, LLC
Product Code
IYE · Radiology
Decision Date
Nov 1, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The Klarity vacuum cushions are designed for the positioning and repositioning of patients receiving external-beam radiation therapy. The Klarity Mold Cushions are designed for the positioning and repositioning of patients receiving external-beam radiation therapy.
Device Story
Klarity Vacuum Cushions and Mold Cushions are patient immobilization devices for external-beam radiation therapy. Vacuum Cushions: polyurethane-coated nylon bags filled with polystyrene spheres; air evacuation via vacuum pump creates rigid, patient-conforming shape. Mold Cushions: polystyrene spheres surrounded by 1/16" polycaprolactone thermoplastic layer and nylon stockinette; heated to 150°F in oven or water bath to become moldable; hardens upon cooling to provide stable support. Used in clinics/hospitals by radiation therapists to ensure reproducible patient positioning for tumor treatment. Benefits include enhanced accuracy of radiation delivery by allowing patients to duplicate previous positions.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, construction, and functional similarity to legally marketed predicate devices.
Technological Characteristics
Vacuum Cushions: Polyurethane-coated nylon, polystyrene spheres, self-closing valve. Mold Cushions: Polystyrene spheres, 1/16" polycaprolactone thermoplastic, nylon stockinette. Thermoplastic softens at 150°F. Both are non-powered, passive positioning accessories. No software or electronic components.
Indications for Use
Indicated for stable support and positioning of patients diagnosed with cancerous tumors undergoing external beam radiation therapy in clinical or hospital settings. Administered by trained medical professionals.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Klarity Patient Immobilization Devices
## 510(K) submission
.
| V<br>1 | 2<br>1<br>ﮧ ﮐﮯ ﻟﯿﮯ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﮐﮯ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﺍﯾﺮﺍﻥ ﻓﺮﺍﻧﺴﯿﺴﯽ ﺍﯾﺮﺍﻥ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﮯ ﺷﮩﺮ ﺍ | r | 1 | A | 211 |
|--------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|---|-----|
| ////3 | | | | | |
| Device 1: | Klarity Vacuum Cushions |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | July 1, 2013 |
| Submitter: | Klarity Medical Products, LLC<br>80 Westgate Drive<br>Newark, Ohio 43055<br>Tel: 740.788.8107 |
| | NOV 0 1 2013 |
| Est. Registration # | 1530561 |
| Contact Person: | Peter M Larson<br>President<br>Klarity Medical Products, LLC<br>Email: peter@klaritymedical.com |
| Trade Name: | Klarity Vacuum Cushions |
| Common Name: | Vacuum Cushion Immobilization System |
| Manufacturing Site: | Klarity Medical Products, LLC<br>80 Westgate Drive<br>Newark, Ohio 43055<br>Tel: 740.788.8107 |
| Classification Name: | Medical charged-particle radiation therapy system, |
| Classification: | Class 2 devices, 892.5050, IYE |
| Predicate Device: | The SecureVac Immobilization System manufactured and<br>marketed by Bionix Development Corporation of Toledo, Ohio.<br>This device was classified as Class II by FDA and was given<br>marketing clearance and assigned document control number<br>K040773. |
| Intended Use: | The Klarity vacuum cushions are designed for the positioning and<br>repositioning of patients receiving external-beam radiation therapy. |
| Claim of Substantial Equivalence: | This product is substantially equivalent to existing patient<br>positioning vacuum bags currently being marketed as<br>accessories to radiation therapy systems.<br>One equivalent device is the SecureVac Immobilization System<br>manufactured and marketed by Bionix Development Corporation<br>of Toledo Ohio. This device was classified as Class II by FDA |
{1}------------------------------------------------
and was given marketing clearance and assigned document control number K040773.
The Bionix SecureVac Immobilization System consists of vacuum bags constructed with polyurethane coated nylon material and filled with small polystyrene spheres. Each bag is sealed airtight and fitted with a self-closing valve for ease of use. When air is evacuated. the SecureVac cushion contracts to hold a rigid shape over the course of the radiation therapy treatment.
The Klarity Vacuum Cushions are similarly constructed with a polyurethane coated nylon material. The vacuum bags are filled with small polystyrene spheres. The Klarity Vacuum cushions are airtight and fitted with a self-closing valve for ease of use. When air is evacuated, the cushions contract to hold a rigid shape over the course of the radiation therapy treatment.
The Klarity Vacuum Cushions function similarly to the Bionix Secure Vac Systems. In clinical practice both devices are used in the same manner. The cushion is placed on a treatment table or baseboard, and the patient is then positioned on top of the vacuum cushion. A vacuum pump is connected to the check valve on the cushion and the air is evacuated. As the air is removed, the cushion compresses around the polystyrene spheres forming a rigid structure that conforms to the anatomy of the patient. When the air is completely evacuated the vacuum pump is disconnected. The cushion now forms a rigid indentation conforming to the anatomy of the patient. For each therapy session the patient can easily duplicate his/her previous position by slipping into the preformed depression made in the cushion, which enhances the accuracy of the radiation treatment.
Based on the similarity of design and construction of the Klarity Vacuum Cushions and the Bionix SecureVac Immobilization System, it is reasonable to expect that these devices will have similar propertics and attenuation factors, and should function in substantially equivalent fashion during the radiation therapy process. Both devices are intended for use in positioning and repositioning patients during radiation therapy procedures and are employed in clinically identical fashions. Therefore, it is reasonable to conclude that the Klarity Vacuum Cushions and the Bionix SecureVac Immobilization System are substantially equivalent with respect to use, safety and effectiveness.
CONFIDENTIAL: Not to be reproduced outside of Klarity Medical Products. LLC without written permission
{2}------------------------------------------------
| | | | Device 2: Klarity Mold Cushions |
|--|--|--|---------------------------------|
|--|--|--|---------------------------------|
| Date Prepared: | July 1, 2013 |
|----------------------|-------------------------------------------------------------------------------------------------|
| Submitter: | Klarity Medical Products, LLC<br>80 Westgate Drive<br>Newark, Ohio 43055<br>Tel: 740.788.8107 |
| Est. Registration # | 1530561 |
| Contact Person: | Peter M Larson<br>President<br>Klarity Medical Products, LLC<br>Email: peter@klaritymedical.com |
| Trade Name: | Klarity Mold Cushion |
| Common Name: | Moldable Head and Neck cushions |
| Manufacturing Site: | Klarity Medical Products, LLC<br>80 Westgate Drive<br>Newark, Ohio 43055<br>Tel: 740.788.8107 |
| Classification Name: | Medical charged-particle radiation therapy system |
Class 2 devices, 892.5050, IYE Classification:
- The MoldCare head and neck cushion marketed by Civco of Predicate Device: Orange City, Iowa. This device was classified as Class II by FDA and was given marketing clearance and assigned document control number K982624.
- The Klarity Mold Cushions are designed for the positioning and Intended Use: repositioning of patients receiving external-beam radiation therapy.
## Claim of Substantial Equivalence:
This product is substantially equivalent to existing patient positioning cushions currently being marketed as accessories to radiation therapy systems.
One equivalent device is MoldCare head and neck cushion marketed by Civco (formerly Med-Tec) of Orange City, Iowa. This device was classified as Class II by FDA and was given marketing clearance and assigned document control number K982624. The
{3}------------------------------------------------
device is currently marketed by Civco as the AccuForm™ head and neck cushion.
The Civco MoldCare cushion has an inner body of polystyrene spheres mixed with a water activated resin. This mixture is surrounded by a soft nylon fabric cover. Before use, and when sold, the cushions are packaged in an air-tight aluminum foil case, to prevent moisture from reaching the cushions.
To use the cushion, a therapist will remove the packaging and apply water to the fabric of the cushion. The cushion is then placed and shaped under the patient's head or other body part. The inner resin reacts with the water and hardens in about 10 minutes, creating a solid formed support, stabilizing the patient.
The Klarity Mold Cushions are similarly constructed but with a different method of becoming secure. The cushion has an inner component of small polystyrene spheres. These are surrounded by a 1/16" layer of polycaprolactone thermoplastic that softens and is moldable at 150° F. The technology of this thermoplastic is commonly used in other radiation therapy devices, including our Klarity thermoplastic masks (510(k)#022708). The thermoplastic layer has a nylon stockinette material surrounding it, for patient comfort.
To use the Klarity Mold cushion, a therapist must first place the cushion in an oven, or in a hot water bath to heat it to about 150° F. At this temperature the cushion becomes moldable by hand and will remain moldable for about 5 to 8 minutes as it cools to room temperature. The warm cushion is extremely comfortable.
Like the MoldCare cushion. the Klarity Mold cushion hardens to create a firm support cushion, conforming to and stabilizing the patient. The cushions also conform to the structure underneath the cushion, providing a means to specifically locate a patient on a treatment table or other support device.
Based on the similarity of design and construction of the Klarity Mold Cushions and the Civco Moldcare cushion, it is reasonable to expect that these devices will have similar properties and should function in a substantially equivalent fashion. Both devices are intended for use in positioning and re-positioning patients during radiation therapy procedures and are employed in clinically similar fashions. Therefore, it is reasonable to conclude that the Klarity Mold Cushion and the Civco Moldcare cushion, are substantially equivalent with respect to use, safety and effectiveness.
CONFIDENTIAL: Not to be reproduced outside of Klarity Medical Products. LLC without written permission
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is oriented to follow the curve of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2013
Klarity Medical Products LLC % Mr. Peter Larson President 80 Westgate Drive NEWARK OH 43055
Re: K132124
Trade/Device Name: Klarity Vacuum Cushions and Klarity Mold Cushions Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 27, 2013 Received: October 4, 2013
Dear Mr. Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Larson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mary Pastel
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
7/1/13
## Indications for Use
510(k) Number: K132124
Device Names:
- 1. Klarity Vacuum Cushions
2. Klarity Mold Cushions
Indications for Use:
To be used by trained medical professionals for the stable support and positioning of patients undergoing external beam radiation therapy treatment in a clinic or hospital setting.
In particular, these items provide the stable positioning of patients in various positions, chosen to facilitate the most accurate treatment of tumors.
Patients are those that have been diagnosed and are undergoing treatment for cancerous turnors. Treatment is supervised and administered by licensed doctors and therapists trained in the application of radiation therapy treatments.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S Pastel
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
K132124 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.