# SNAP WOUND CARE SYSTEM (K132080)

_Spiracur, Inc. · OKO · Mar 12, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K132080

## Device Facts

- **Applicant:** Spiracur, Inc.
- **Product Code:** [OKO](/productcode/OKO.md)
- **Decision Date:** Mar 12, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4683
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The SNaP Wound Care System is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, venous or pressure), surgically closed incisions, flaps and grafts.

## Device Story

SNaP Wound Care System is a non-powered, portable, single-use suction device for negative pressure wound therapy (NPWT). Device utilizes constant-force springs to mechanically generate negative pressure gradient; removes wound exudate, irrigation fluids, and infectious materials. Used in clinical or home settings; operated by clinicians or patients. System includes dressing kits (foam, bridge dressing, hydrocolloid SecurRing) to maintain seal over various anatomical contours. Output is continuous negative pressure applied to wound bed; facilitates wound healing by removing debris and exudate. Labeling update allows use in hyperbaric oxygen (HBO) environments and includes instructions for resetting the suction cartridge. No design modifications were made to the core suction mechanism.

## Clinical Evidence

Bench testing only. Testing included verification, simulated wound performance, biocompatibility, and packaging/shelf-life studies. Specific bench evaluations confirmed functionality of the cartridge reset feature and performance of the SNaP Bridge Dressing and SecurRing in delivering negative pressure. Oxygen compatibility testing characterized ignition probability and severity of dressing materials in hyperbaric oxygen environments.

## Technological Characteristics

Non-powered, single-use suction apparatus. Energy source: constant-force mechanical springs. Components: suction cartridge, foam dressings, hydrocolloid seals (SecurRing), bridge dressing. Connectivity: none. Sterilization: not specified. Software: none.

## Regulatory Identification

A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- SNaP® Wound Care System ([K081406](/device/K081406.md))
- SNaP® Wound Care System ([K111393](/device/K111393.md))
- SNaP® Wound Care System ([K112341](/device/K112341.md))
- SNaP® Wound Care System ([K113032](/device/K113032.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# MAR 1 2 2014

Spiracur Inc.

SNaP® Wound Care System TRADITIONAL 510(k) PREMARKET NOTIFICATION

### SECTION 6 510(k) SUMMARY

# 510(k) Notification K132080

### GENERAL INFORMATION

# Applicant:

Spiracur Inc. 1180 Bordeaux Drive Sunnyvale, CA 94089 U.S.A. Phone: 408-701-5300 Fax: 408-701-5301

#### Contact Person:

Sarah L. Canio Regulatory Consultant Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: 408-400-0856 ext. 109 Fax: 408-400-0865 Email: sarahi@experiengroup.com

# Date Prepared: March 10, 2014

#### DEVICE INFORMATION

Classification : 21CFR§878.4683, Class II

# Product Code: OKO, Negative Pressure Wound Therapy Non-Powered Suction Apparatus

Trade Name: SNaP® Wound Care System

#### Generic/Common Name:

Non-powered suction apparatus device intended for negative pressure wound therapy

#### Predicate Device·

SNaP® Wound Care System (K081406, K111393, K112341, K113032)

{1}------------------------------------------------

# SECTION 6 510(k) SUMMARY

#### Intended Use

The SNaP Wound Care System is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing through the removal of excess exudates, infectious material and tissue debris. The SNaP Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, venous or pressure), surgically closed incisions, flaps and grafts.

#### PRODUCT DESCRIPTION

The SNaP® Wound Care System ("SNaP System") is an existing (K081406, K111393, K112341, K 113032) non-powered, portable, single-use suction device intended for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. The SNaP System utilizes dedicated constant-force springs to mechanically generate the negative pressure gradient. The proposed SNaP Wound Care System will have the same indication for use as the predicate device, the SNaP Wound Care System (K113032). In addition, Spiracur has added the SNaP® Bridge Dressing Kit ("SNaP Bridge Dressing") and SNaP® SecurRing" Hydrocolloid ("SNaP SecurRing") to the SNaP Dressing Kit product family.

#### SUBSTANTIAL EQUIVALENCE

The proposed labeling changes did not result in any design modifications to the cleared SNaP Wound Care System (K081406, K111393, K112341, K113032). The only modifications resulting from the labeling changes are to the SNaP System Instructions for Use provided to the physician, that now describes conditions for use of the SNaP System dressing components within enriched oxygen environments, such as hyperbaric oxygen (HBO) chambers, and also now includes instructions for resetting the SNaP Wound Care Cartridge. In addition, Spiracur has added the SNaP Bridge Dressing and SNaP SecurRing to the SNaP Dressing Kit product family. The SNaP Bridge Dressing is a design modification introduced to the SNaP Foam Dressing Kit (K112341) to facilitate dressing use on a variety of anatomical contours. The sterile, moldable hydrocolloid SNaP SecurRing serves the same function as the hydrocolloid layer of standard SNaP Dressing, to improve the seal of the dressing around the wound.

The additions of the SNaP Bridge Dressing Kit and the SNaP SecurRing to the SNaP Dressing Kit product family are to provide clinicians additional dressing options for use with the SNaP System, designed to help achieve a seal during the SNaP Dressing application. The additions to the SNaP Dressing Kit product family are substantially equivalent to existing items of the product family as they have the same intended use in the same wound types, and utilize similar performance specifications and comparable technological features to achieve the same mechanism of action: therefore, the additions do not raise any new issues of safety or effectiveness.

The SNaP System is substantially equivalent to the predicate device as both devices have the same intended use, have been historically cleared for use in the same wound types, and utilize similar performance specifications and comparable technological features to achieve the same

{2}------------------------------------------------

#### Spiracur Inc.

# SECTION 6 510(k) SUMMARY

mechanism of action: therefore, the modified labeling does not raise any new issues of safety or effectiveness.

# Testing in Support of Substantial Equivalence Determination

To support the additions of the SNaP Bridge Dressing Kit and the SNaP SecurRing to the SNaP Dressing Kit product family, Spiracur conducted necessary bench testing to ensure conformance to design specifications and to support a determination of substantial equivalence to existing items of the SNaP Dressing Kit product family. The testing performed includes:

- Verification testing ●
- . Simulated wound performance testing
- Biocompatibility testing .
- Packaging and shelf life testing .

The proposed labeling changes to the cleared SNaP Wound Care System did not result in any design modifications. All necessary bench testing was conducted on the SNaP Wound Care System and its components to ensure conformance to design specifications and to support a determination of substantial equivalence to the predicate device. The testing performed includes:

- Bench testing within a simulated hyperbaric oxygen environment to characterize . the probability of self-ignition and to characterize the potential severity of ignition of the SNaP Wound Care dressing materials, in order to evaluate the oxygen compatibility of the dressing materials.
- Bench testing conducted in conjunction with the modified device Instructions for . Use to evaluate the functionality of a latent design feature allowing for the resetting of the SNaP Wound Care System to assess the ability of the SNaP Wound Care System to deliver negative pressure wound therapy comparable to the predicate device (K081406, K111393, K112341, K113032).
- Simulated wound performance testing conducted in conjunction with the modified . device Instructions for Use to evaluate the functionality of a latent design feature allowing for the resetting of the SNaP Wound Care System to assess the ability of the SNaP Wound Care System to deliver negative pressure wound therapy comparable to the predicate device (K081406, K111393, K112341, K113032) when used in conjunction with the SNaP Bridge Dressing Kit and the SNaP SecurRing and all other dressing configurations after resetting.

# SUMMARY

The SNaP® Wound Care System is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, represented by three curved lines.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 12, 2014

Spiracur Incorporated % Ms. Sarah L. Canio Experien Group, LLC 755 North Mathilda Avenue Sunnyvale, California 94085

Re: K132080

Trade/Device Name: SNaP® Wound Care System Regulation Number: 21 CFR 878.4683 Regulation Name: Non-Powered suction apparatus device intended for negative pressure wound therapy Regulatory Class: Class II Product Code: OKO Dated: February 7, 2014 Received: February 10, 2014

Dear Ms. Canio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commence prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any i with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Ms. Sarah L. Canio

CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K132080

Devica Name SNaP® Wound Care System

#### Indications for Use (Describe)

miliations of to to total vir patients who would benefit from wound management via the application of negative pressure, particularly as the device may promote wound healing through the removal of excess exudate, infectious material and tissue debris. The SNaP® Wound Care System is indicated for removal of small amounts of exudate from chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, venous or pressure), surgically closed incisions, flaps and grafts.

#### Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

. .

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Jiyoung Dang

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete time to review the collection. Send comments regarding this burden estimate or any other aspect of this informalion collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

---

**Source:** [https://fda-staging.innolitics.com/device/K132080](https://fda-staging.innolitics.com/device/K132080)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).