K132027 · Ge Oec Medical Systems, Inc. · OWB · May 1, 2014 · Radiology
Device Facts
Record ID
K132027
Device Name
OEC 9800 PLUS
Applicant
Ge Oec Medical Systems, Inc.
Product Code
OWB · Radiology
Decision Date
May 1, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Series OEC® 9800 Plus is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Device Story
OEC 9800 Plus is a mobile fluoroscopic X-ray system used by surgeons to visualize anatomy and pathology during medical procedures. The system captures real-time X-ray images, providing pre-surgical, intra-operative, and post-surgical visualization. It consists of an X-ray tube, image intensifier (Radiological Imaging Unit), and a workstation with an LCD monitor. Surgeons operate the system via hand or wireless foot switches to acquire images, which are then displayed for clinical decision-making. The device facilitates localization of surgical regions of interest and records surgical activity. Modifications include an alternative X-ray tube, alternative image intensifier, replacement of CRT monitor with LCD, integration of two circuit boards into one, an alternative solid-state drive, and a different OEM injector system model. These updates maintain the fundamental scientific technology of the predicate device while improving component availability and performance.
Clinical Evidence
No clinical studies on human patients were conducted. Substantial equivalence was supported by bench testing, including component, subsystem, and system-level verification. Performance testing covered X-ray acquisition, image quality, image processing, and mechanical subsystems. Simulated use testing was performed by experienced clinicians to validate user needs, image storage/recall, and monitor performance. The device complies with mandatory and voluntary standards including IEC 60601-1, IEC 60601-1-2, and 21 CFR 1020.30-32.
Technological Characteristics
Mobile fluoroscopic X-ray system. Components: X-ray tube, image intensifier, LCD monitor, solid-state drive. Connectivity: DICOM 2008. Power: Electrical. Software: Embedded firmware for image acquisition, processing, and data management. Standards: 21 CFR 1020.30-32, UL 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, NFPA 99.
Indications for Use
Indicated for patients undergoing diagnostic, surgical, and interventional procedures requiring fluoroscopic and spot-film imaging, including cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
### 510(k) Summary
# MAY 0 1 2014
K132027
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: November 20, 2013
GE Healthcare Surgery Submitter: 384 Wright Brothers Drive Salt Lake City, UT 84116
| Primary Contact Person: Karen Russell | |
|------------------------------------------|-------------------------------------------|
| | Regulatory Affairs Leader |
| | GE Healthcare Surgery |
| | Phone: (801) 536-4930 Fax: (801) 517-6566 |
| Secondary Contact Person: Jeff Wagner | |
| | Regulatory Affairs Manager |
| | GE Healthcare Surgery |
| | Phone: (801) 517-6415 Fax: (801) 517-6566 |
OEC® 9800 Plus Device: (Trade Name): Mobile Fluoroscopic Imaging System Common/Usual Name: 21 CFR 892.1650 Image-intensified fluoroscopic x-ray system Classification Names:
> Product Code: 90OWB, 90OXO and 90JAA
K111551 OEC® 9800 Plus Predicate Device(s):
Device Description:
The OEC® 9800 Plus is a system used to assist trained The system is used to provide X-Ray images Surgeons. while the Surgeon performs a medical procedure. Images from the system help the Surgeon to visualize the patients' This visualization helps to localize surgical anatomy. regions of interest and pathology. The images provide realtime visualization and records of pre-surgical anatomy, in vivo-surgical activity and post surgical outcomes.
The proposed device will add an alternative supplier for the X-Ray Tube. The current tube and the proposed tube have the same specifications and will be interchangeable. Verification activities confirmed that the proposed tube meets the same specification as the current supplier's tube.
{1}------------------------------------------------
The proposed device will add an alternative supplier for the Radiological Imaging Unit (RIU) or Image Intensifier. The current RIU and the proposed RIU have the same specifications and will be interchangeable. Verification activities confirmed that the alternate RIU meets the same specification as the current supplier's RIU.
The proposed device will provide an optional wireless foot switch and hand switch. The new option is equivalent to the predicate in that functionality is identical. The wireless option provides placement flexibility and reduced cable clutter.
Two printed circuit board assemblies have been combined into a single board in the proposed device due to parts obsolescence. Both the proposed and predicate assemblies common specifications as confirmed through share verification testing.
The proposed device will include a LCD monitor which replaces the obsolete CRT monitor. Both the proposed and predicate assemblies share common specifications as confirmed through verification testing.
The proposed device includes an alternative solid state drive. This drive provides a faster write speed. Both drives meet their specifications.
The proposed device will include a different model of an OEM injector system. Both the proposed and predicate OEM iniector systems share common specifications as confirmed through verification testing.
Intended Use:
The Series OEC® 9800 Plus is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
The modified OEC® 9800 Plus device employs the same Technology: fundamental scientific technology as the predicate device.
{2}------------------------------------------------
## Determination of Substantial Equivalence:
## Summary of Non-Clinical Tests
The following quality assurance measures were applied to the development of the proposed device: Risk Analysis, Requirements Reviews, Design Reviews and testing which included component, subsystem, and system.
Design verification methodology was hierarchal. When component development was completed, component level tests were executed. Upon completion of component testing, subsystem and system verification were also performed. The Design Verification of the OEC 9800 Plus confirms that design output meets design input requirements.
testing was performed which included Functional mechanical, electrical, and overall system tests. X-ray acquisition, X-ray control and indication, patient data management, imaging modes (basic and advanced), image quality acquisition and assessment, image processing and presentation, screen measurements, motorized system, and subsystem mechanical tests were performed. Performance testing included image quality, and subsystem performance.
Product Simulated Use Testing was used to evaluate whether the device meets user needs and intended uses by simulating a use environment, and having testing performed by users with relevant clinical experience. The functional requirements of image storage and recall capability were validated using the proposed device. The proposed device was also used to produce images that were evaluated to confirm the new monitor met user needs and intended use.
Safety testing was performed in laboratory settings by qualified technicians to confirm that the product met the requirements of the standards listed in Table 1 below.
The OEC® 9800 Plus complies with the voluntary and mandatory standards listed in Table 1 below.
{3}------------------------------------------------
## Summary of Clinical Tests
The OEC 9800 Plus is based on modifications to a cleared predicate device. There were no changes to the intended use. Clinical studies on living human patients were not necessary to support substantial equivalence.
The verification and validation testing described above Conclusion: demonstrates that the OEC® 9800 Plus is safe, effective, and performs, for its intended use, in an equivalent manner to the predicate device and in accordance with its labeling. OEC® 9800 Plus is substantially equivalent in performance to the predicate OEC® 9800 Plus (K111551).
| Standards<br>No. | Standards<br>Organization | Standards Title | Version | Date |
|----------------------|---------------------------|---------------------------------------------------------------------------------------|-------------------------------|------------------------|
| 21 CFR<br>1020.30-32 | FDA | Federal Performance Standard for<br>Diagnostic X-ray Systems | 2012 | 2012 |
| 60601-1 | UL | Medical Electrical Equipment, Part 1:<br>General Requirements for Safety | 1st edition | 2003 |
| 60601-1 | IEC | Medical Electrical Equipment, Part 1:<br>General Requirements for Safety | A1 (1991)<br>and A2<br>(1995) | 1988,<br>1991,<br>1995 |
| 60601-1-2 | IEC | Collateral Std: Electromagnetic<br>Compatibility | 2.1 edition | 2004 |
| 60601-1-3 | IEC | Collateral Std: Radiation Protection in<br>Diagnostic X-ray Equipment | 1st edition | 1994 |
| 60601-1-4 | IEC | Collateral Std: Radiation Protection in<br>Diagnostic X-ray Equipment | 1.1 edition<br>(+A1) | 2000 |
| 60601-1-6 | IEC | Collateral Std: General Requirements<br>for Basic Safety and Essential<br>Performance | 2nd edition | 2006 |
| 60601-2-7 | IEC | Particular Std: Safety of H.V.<br>Diagnostic X-ray Generators | 2nd edition | 1998 |
| 60601-2-28 | IEC | Particular Std: Safety of X-ray Tube<br>and X-ray Source Assemblies | 1st edition | 1993 |
| 60601-2-32 | IEC | Particular Std: Safety of Associated<br>Equipment of X-ray Equipment | 1st edition | 1994 |
| NEPA 99 | NFPA | US National Electric Code Health<br>Care Facilities | 2005 | 2005 |
| DICOM | DICOM | US National Electric Code Health<br>Care Facilities | 2008 | 2008 |
Table 1 Product Standards Compliance
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
May 1. 2014
GE Healthcare Surgery Jeff Wagner Regulatory Affairs Manager 384 Wright Brothers Drive Salt Lake City, Utah 84116
K132027/S004 Re:
> Trade/Device Name: OEC 9800 Plus Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II -Product Code: OWB, OXO, JAA Dated: February 27, 2014 Received: April 2, 2014
Dear Mr. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Wagner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
### 510(k) Number (if known) K132027
#### Device Name GE OEC® 9800 Plus
#### Indications for Use (Describe)
Indications for Use: The OEC 9800 Plus is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, margical and interventional procedures. Clinical application may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden time for this collection of information is common to connections of the parties of complete time to review instruction. Send comments regarding this burden estimate or any other aspect afto te now the collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.