The Advance® 35LP Low Profile PTA Balloon Dilatation Catheter has been designed for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including illiac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Device Story
Over-the-wire PTA balloon dilatation catheter; used for percutaneous transluminal angioplasty of peripheral artery lesions and arteriovenous dialysis fistulae. Device consists of catheter shaft and inflatable balloon; compatible with 0.035-inch wire guide. Operated by physician in clinical setting. Balloon inflated to dilate stenotic lesions; restores vessel patency. Sterile, single-use device.
Clinical Evidence
Bench testing only. Testing included compliance, balloon profile, fatigue (10 cycles), burst pressure, inflation/deflation time (<60s), sheath compatibility, and tensile strength. All tests met predetermined acceptance criteria and conformed to ISO 10555-1 and ISO 10555-4 standards.
Technological Characteristics
Over-the-wire PTA balloon catheter; 5.2F or 5.7F; 80cm or 135cm lengths; balloon diameters 3-12mm; balloon lengths 1.5-20cm. Compatible with 0.035-inch wire guide. Sterile, single-use. Mechanical design; no software or electronics.
Indications for Use
Indicated for percutaneous transluminal angioplasty (PTA) of peripheral artery lesions (iliac, renal, popliteal, infrapopliteal, femoral, iliofemoral) and obstructive lesions in native or synthetic arteriovenous dialysis fistulae.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters. The logo is simple and professional, and it is likely used on Cook Medical's products and marketing materials.
COOK INCORPORATEC
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE 812.339.2235 τOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
K132020 og. 1 of 2
### 510(k) SUMMARY
Submitted By:
Steven Lawrie, MS, MA Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 Phone: (812) 335-3575 x 104518 Fax: (812) 332-0281 Date Prepared: October 22, 2013
OCT 2 5 2013
Device:
Trade Name:
Common Name: Classification Name: Advance® 35LP Low Profile PTA Balloon Dilatation Catheter PTA Balloon Catheter Catheter, Angioplasty, Peripheral, Transluminal LIT (21 CFR §870.1250)
### Indications for Use:
The Advance® 35LP Low Profile PTA Balloon Dilatation Catheter is indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
## Predicate Device:
The Advance® 35LP Low Profile PTA Balloon Dilatation Catheters are identical in terms of intended use, principles of operation, materials of construction, and technological characteristics to the predicate device. Additional combinations of balloon diameter and length have been included. The device, subject of this submission, is substantially equivalent to the Advance® 35LP Low Profile PTA Balloon Dilatation Catheters cleared under 510(k) number K091527.
#### Comparison to Predicate Device:
It has been demonstrated that the Advance® 35LP Low Profile PTA Balloon Dilatation Catheters are comparable to the predicate device. The predicate device and the device subject of this submission are indicated for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including iliac, renal, popliteal, femoral and iliofemoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The predicate device and the proposed device are completely identical in terms of design, intended use, and materials.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, sans-serif font on the top line. Below that, the word "MEDICAL" is in a smaller, sans-serif font. The background is black, and the text is white.
COOK
INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
MONE 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
K132020 pg. 2 of 2
# Device Description:
The Advance® 35LP PTA Balloon Dilatation Catheters are over-the-wire catheters that will be available with inflated balloon diameters of 3, 4. 5, 6, 7, 8, 9, 10, and 12 millimeters and balloon lengths of 1.5. 2. 2.5. 3. 4. 6. 8. 10. 12. 14. 17. and 20 centimeters. The catheters are 5.2 French or 5.7 French, dependent upon device specification, and will be available in lengths of 80 and 135 centimeters. The catheters are compatible with a 0.035 inch (0.89 millimeter) diameter wire guide. The catheters will be supplied sterile and are intended for one-time use.
## Test Data:
The following tests were performed to demonstrate that the Advance® 35LP Low Profile PTA Balloon Dilatation Catheters met applicable design and performance requirements and support a determination of substantial equivalence.
- . Compliance Testing - Testing showed that, under simulated body temperature conditions, each balloon met its labeled diameter at the nominal pressure. The predetermined acceptance criteria were met.
- 미 Balloon Profile Testing - Testing showed that diameters for each catheter were less than the maximum outside diameter appropriate for the intended sheath size. The predetermined acceptance criteria were met.
- 0 Balloon Fatigue Testing - Testing showed that the balloons were from leakage and damage on inflation, withstanding 10 cycles of inflation/deflation. In conformance with the applicable sections of ISO 10555-4, the predetermined acceptance criteria were met.
- . Balloon Burst Testing - Testing showed that the balloons will burst at or above the minimum rated burst pressure, with all failure modes being linear tears. The predetermined acceptance criteria were met.
- 0 Balloon Inflation/Deflation Testing - Testing showed that the balloons inflated to rated burst pressure within 60 seconds and fully deflated within 60 seconds. The predetermined acceptance criteria were met.
- l Sheath Compatibility Testing - Testing showed that the catheters were capable of being inserted and retracted from an appropriately sized sheath without experiencing excessive resistance. The predetermined acceptance criteria were met.
- D Tensile Strength Testing - Testing showed that under proper clinical use of the device, the peak load values shall be in accordance with the applicable values of ISO 10555-1. The predetermined acceptance criteria were met.
In conclusion, the results of these tests support a determination of substantial equivalence to the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 25, 2013 ·
Cook Inc. Mr. Steven Lawrie, MS, MA Regulatory Affairs Specialist 750 Daniels Way P.O. Box 489 Bloomington, IN 47402
Re: K132020
Trade/Device Name: Advance® 35LP Low Profile PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: II Product Code: LIT Dated: September 26, 2013 Received: September 27, 2013
Dear Mr. Lawrie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.
{3}------------------------------------------------
Page 2 - Mr. Steven Lawrie, MS, MA
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Special 510(k): Device Modification PTA Balloon Catheter: Advance 351.P Low Profile PTA Balloon Dilatation Catheter Cook Incorporated June 28 2013
### Indications for Use
510(k) Number (if known): K132020
Device Name: Advance® 35LP Low Profile PTA Balloon Dilatation Catheter
Indications for Use for the Advance 35LP Low Profile PTA Balloon Dilatation Catheter:
The Advance® 35LP Low Profile PTA Balloon Dilatation Catheter has been designed for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries including illiac, renal, popliteal, infrapopliteal, femoral and iliofemoral as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Prescription Use XX (Part 21 CFR 801 Subpart D)
OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kenneth J. Cavanaugh -S
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.