CMOS VIDEO NASOPHARYNGOSCOPE SYSTEM

K132009 · Schoelly Fiberoptic GmbH · EOB · Apr 9, 2014 · Ear, Nose, Throat

Device Facts

Record IDK132009
Device NameCMOS VIDEO NASOPHARYNGOSCOPE SYSTEM
ApplicantSchoelly Fiberoptic GmbH
Product CodeEOB · Ear, Nose, Throat
Decision DateApr 9, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4760
Device ClassClass 2

Indications for Use

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

Device Story

System comprises flexible, steerable endoscope and camera control unit (CCU). Endoscope features plastic exterior, control lever for distal tip angulation, integrated LED light source, and CMOS imaging sensor at shaft tip. Light transmitted via fiberoptic bundles; video signal captured by CMOS sensor and transferred to CCU. CCU regulates light intensity and connects to external monitors, PCs, or recorders for visualization/documentation. Includes ventilation system for shaft protection and leakage testing capability. Used in clinical settings by physicians/specialists for airway/nasal visualization. Output displayed on video monitor to assist clinical diagnosis. Benefits include enhanced visualization of anatomy for diagnostic procedures.

Clinical Evidence

Bench testing only. Performance data demonstrated compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, and ISO 10993. Reprocessing validation performed per AAMI TIR12:2010 and AAMI TIR30:2003, including manual cleaning, high-level disinfection, and STERRAD 100S sterilization.

Technological Characteristics

Flexible steerable endoscope with plastic housing; CMOS imaging sensor; LED light source with fiberoptic transmission; ventilation system for leakage testing. Connectivity to external monitors/PCs. Reprocessing via manual cleaning, high-level disinfection, or STERRAD 100S. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, ISO 10993, ISO 14937.

Indications for Use

Indicated for endoscopic diagnosis within nasal lumens and airway anatomy in patients requiring visualization via video monitor. Restricted to use by medically qualified professionals trained in rhino/laryngoscopic techniques.

Regulatory Classification

Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Schoelly Fiberontic GmbH 510(k) Premarket Notification (Traditional) 06/21/2013 CMOS Video Nusopharyngoscope System K132009 ## Section 5-510(k) Summary | General Information | | |-------------------------|--------------------------------------------------------------------------| | Preparation date: | 06/21/2013 | | Owner's Name: | Schoelly Fiberoptic GmbH (Registration: 8043903) | | Address: | Robert-Bosch-Str. 1-3<br>79211 Denzlingen<br>Germany | | Telephone Number: | +49-7666-980-0 | | Fax Number: | +49-7666-908-380 | | Contact Person: | Dr. Sandra Baumann | | Subject Device Name: | | | Trade Name: | Schoelly CMOS Video Nasopharyngoscope System | | Trade Name: | Schoelly CMOS Video Nasopharyngoscope System | | Common/Usual Name: | Video Nasopharyngoscope System | | Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II | | Predicate Device Name: | | | Trade Name: | Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM | | Trade Name: | Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM | | Common/Usual Name: | Video Rhino-Laryngoscope System | | Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II | | Premarket Notification: | K103467 (Karl Storz Endoscopy-America, Inc.), SE date June 28, 2012 | | Predicate Device Name: | | | Trade Name: | Schoelly Naso-Laryngo-Pharyngoscope | | Trade Name: | Schoelly Naso-Laryngo-Pharyngoscope | | Common/Usual Name: | Nasopharyngoscope | | Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II | | Premarket Notification: | K083553 (Schoelly Imaging, Inc.), SE date January 16, 2009 | #### Device Description The Schoelly CMOS Video Nasopharyngoscope System consists of a flexible and steerable endoscope and a camera control unit (CCU) for regulation of light intensity and connection to a monitor, PC, medical video recorder or printer for image display or image documentation. The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and an integrated LED light source. Light is transmitted through fiberoptic bundles illuminating the anatomy under investigation. The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope shaft and transferred to the CCU. The endoscope further incorporates a ventilation system to protect the shaft. The exhaust valve at the endoscope handle can further be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories. The Schoelly CMOS Video Nasopharyngoscope System is delivered in a non-sterile condition and is already CE marked. {1}------------------------------------------------ 06/21/2013 CMOS Video Nasopharyngoscope System #### Predicate Devices The Karl Storz predicate device is based on the same technology has the same essential design and dimension characteristics, the same essential optical parameters and the same principle of operations. It is made out of the same primary materials and the same standards were applied and met for design verification and validation as compared to the proposed Schoelly CMOS Video Nasopharyngoscope System. The Schoelly predicate device differs with respect to the image transfer technology; however this device is built out of identical components/systems for light transmission, angulation of the endoscope's distal tip or ventilation. Further, the manufacturing process for major parts of the endoscope is identical as compared to the endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System. #### Indications for Use The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor. The indications for use of the proposed devices are identical to the indications for use of the Karl Storz predicate device. #### Non-clinical Performance Testing Performance data demonstrated that the Schoelly CMOS Video Nasopharyngoscope System has met pre-determined acceptance criteria and is substantially equivalent to the predicate devices. The device is as safe, as effective, and performs as well as or better than the predicate devices. The risks associated with use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits of predicate devices are the same as compared to the proposed one. Testing has been conducted as per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, and ISO 10993 to address the safety and performance aspects of the device. #### Reprocessing The endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System is the subject of completed reprocessing validations including manual cleaning, high-level disinfection and STERRAD® 100S sterilization. Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilties: A Guide for Device Manufacturers) and AAMI TIR30:2003 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices). Devices that have been used for testing had been exposed to extensive clinical use including tip angulation, soiling, and reprocessing to address the FDA Draft Guidance Processing Medical Devices in Healthcare Settings (dated: May 2, 2011). Sterilization studies have been performed in accordance with ISO 14937. #### Conclusion The Schoelly CMOS Video Nasopharyngoscope System meets all the pre-determined acceptance criteria of the testing performed to confirm substantial equivalence to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 # April 9, 2014 Schoelly Fiberoptic GmbH c/o Ms. Pamela Papineau Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, MA 01432 Re: K132009 Trade/Device Name: Schoelly CMOS Video Nasopharyngoscope System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: March 3, 2014 Received: March 5, 2014 Dear Ms. Papineau: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Ms. Pamela Papineau Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman. M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K132009 Device Name: Schoelly CMOS Video Nasopharyngoscope System Indications for Use: The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhinoflaryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of Center for Devices and Radiological Health (CDRH) Sunny Park 2014.04.08 14:30:52 -04'00' Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...