6981M LEAD EXTENDER KIT, 6984M LEAD EXTENDER KIT, 6986M LEAD EXTENDER KIT, 5866-24M LEAD ADAPTOR KIT, 5866-38M LEAD ADAP
Applicant
Medtronic, Inc.
Product Code
DTD · Cardiovascular
Decision Date
Jul 30, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 6981M Lead extender kit is designed to extend, by 37 cm (14.6 in.), the length of a pacing lead system with a unipolar connector that meets the (IS-1 UNI) standard. The 6984M Lead Extender Kit is designed to extend, by 37 cm (14.6 in.), the length of a pacing lead system with a bipolar connector that meets the (IS-1 BI) standard. The 5866-24M Lead adaptor kit is designed to connect a Medtronic pacing lead with a bipolar connector (5mm bifurcated) to a pulse generator featuring a bipolar connector block which meets the (IS-1 BI) standard. The 5866-38M Lead adaptor kit is designed to connect two pacing leads with unipolar connectors (IS-1 UNI) to a pulse generator featuring a bipolar connector block which meets the (IS-I BI) standard. The Model 5866-40M Lead Adaptor Kit is designed to connect a Medtronic pacing lead with a bipolar connector (3.2 mm low-profile) to a pulse generator featuring a bipolar connector block which meets the (1S-1 BI) standard and has a short-pin cavity and no sealing rings. The 6986M Lead adaptor/Extender kit is designed to extend, by 37 cm (14.6 in.), the length of a 3.2 mm low-profile connector pacing lead system and connect to a pulse generator featuring a bipolar connector block that meets the (IS-1 BI) standard.
Device Story
Lead extender and adaptor kits facilitate physical and electrical connection between existing cardiac pacing leads and pulse generators. Devices bridge connector mismatches (e.g., unipolar to bipolar, specific pin/cavity configurations) or provide additional lead length (37 cm). Used in clinical settings by physicians during lead revision or pulse generator replacement procedures. Operation involves mechanical coupling of lead connectors to the adaptor/extender, which then interfaces with the pulse generator connector block. Ensures electrical continuity for pacing therapy. Benefits include avoiding lead extraction or replacement when pulse generators are upgraded or replaced.
Clinical Evidence
Bench testing only. Verifications included dielectric withstanding voltage, saline soak/electrical impedance, fluid leakage, composite tensile integrity/strength, lead body reverse bend, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation).
Technological Characteristics
Passive mechanical/electrical adaptors and extenders. Materials assessed for biocompatibility. Features include IS-1 UNI and IS-1 BI standard connector compatibility, 3.2 mm low-profile connectors, and 37 cm length extensions. Sterilized components.
Indications for Use
Indicated for patients requiring extension of existing pacing lead systems or adaptation of pacing leads to pulse generators with incompatible connector blocks. Prescription use only.
Regulatory Classification
Identification
A pacemaker lead adaptor is a device used to adapt a pacemaker lead so that it can be connected to a pacemaker pulse generator produced by a different manufacturer.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for the Submission of Research and Marketing Applications for Permanent Pacemaker Leads and for Pacemaker Lead Adaptor 510(k) Submissions.”
{0}------------------------------------------------
## 2.0 510(k) Summary
| Date Prepared: | June 28, 2013 | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Cardiac Rhythm Disease Management<br>8200 Coral Sea Street N.E.<br>Mounds View, MN 55112<br>Establishment Registration Number: 2182208 | JUL 3 0 2013 |
| Contact Person: | Michele L Machacek MS<br>Senior Regulatory Affairs Specialist<br>Medtronic Cardiac Rhythm Disease Management<br>Phone: 763.526.2382<br>Fax: 651.367.0603<br>Email: michele.l.machacek@medtronic.com | |
| | Alternate Contact:<br>Wendy Saunders<br>Senior Regulatory Affairs Manager<br>Medtronic Cardiac Rhythm Disease Management<br>Phone: 763.526.8134<br>Fax: 651.367.0603<br>Email: wendy.a.saunders@medtronic.com | |
| Device Name and Classification | | |
| Trade Name: | 6981M Lead Extender Kit |
|-------------|---------------------------|
| | 6984M Lead Extender Kit |
| | 5866-24M Lead Adaptor Kit |
| | 5866-38M Lead Adaptor Kit |
| | 5866-40M Lead Adaptor Kit |
{1}------------------------------------------------
## K132008
| Common Name: | 6986M Lead Extender Kit<br>Pacemaker Lead Adaptor |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 870.3620 |
| Product Code: | DTD |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
| Special Controls: | "Guidance for the Submission of Research and Marketing Applications for<br>Permanent Pacemaker Leads and for Pacemaker Lead Adaptor 510 (k)<br>Submissions" |
| Predicate Devices | |
| K911827 | 6981M Lead Extender Kit |
| K915724 | 6984M Lead Extender Kit |
| K911302 | 5866-24M Lead Adaptor Kit<br>5866-38M Lead Adaptor Kit<br>5866-40M Lead Adaptor Kit |
6986M Lead Extender Kit
## Device Description/Indications for Use
| Model | Device Description/Indications for Use |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 6981M Lead<br>Extender Kit | The 6981M Lead extender kit is designed to extend, by 37 cm (14.6 in.),<br>the length of a pacing lead system with a unipolar connector that meets<br>the (IS-1 UNI) standard. |
| 6984M Lead<br>Extender Kit | The 6984M Lead Extender Kit is designed to extend, by 37 cm (14.6<br>in.), the length of a pacing lead system with a bipolar connector that<br>meets the (IS-1 BI) standard. |
| 5866-24M Lead<br>Adaptor Kit | The 5866-24M Lead adaptor kit is designed to connect a Medtronic<br>pacing lead with a bipolar connector (5mm bifurcated) to a pulse<br>generator featuring a bipolar connector block which meets the (IS-1 BI)<br>standard. |
| 5866-38M Lead<br>Adaptor Kit | The 5866-38M Lead adaptor kit is designed to connect two pacing leads<br>with unipolar connectors (IS-1 UNI) to a pulse generator featuring a<br>bipolar connector block which meets the (IS-1 BI) standard. |
| 5866-40M Lead<br>Adaptor Kit | The Model 5866-40M Lead Adaptor Kit is designed to connect a<br>Medtronic pacing lead with a bipolar connector (3.2 mm low-profile) to<br>a pulse generator featuring a bipolar connector block which meets the<br>(1S-1 BI) standard and has a short-pin cavity and no sealing rings. |
:
{2}------------------------------------------------
| 6986M Lead<br>Extender Kit | The 6986M Lead adaptor/Extender kit is designed to extend, by 37 cm<br>(14.6 in.), the length of a 3.2 mm low-profile connector pacing lead<br>system and connect to a pulse generator featuring a bipolar connector<br>block that meets the (IS-1 BI) standard. |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Comparison to Predicate Devices
When compared to the predicate devices (K911827, K915724 & K911302), the Medtronic Lead Adaptor Kit and Lead Extender Kit models presented in this submission have the following similarities:
- . Intended use
- Technological characteristics .
- Operating principle ●
- Design features ●
- . Performance
- Biological safety .
- Packaging materials .
- Sterilization and sterility assurance level �
- . Shelf life
#### Performance Testing
The assessment of the material change being made to the Lead Adaptor Kit and Lead Extender Kit products was carried out using Design Controls compliant to 21 CFR 820.30. The safety and effectiveness of the product has been ensured through performance testing. Verifications included:
Performance Testing
- Dielectric Withstanding Voltage Test .
- Saline Soak / Electrical Impedance .
- Fluid Leakage .
- . Composite Tensile Integrity
- Composite Tensile Strength .
- Lead Body Reverse Bend .
Biocompatibility Assessment
- . Cytotoxicity
- Sensitization .
- Irritation or Intracutaneous Reactivity ●
{3}------------------------------------------------
## K132008
- Systemic Toxicity ●
- Genotoxicity ●
- . Implantation
Sterilization Testing
- Residual Testing .
#### Conclusion
Medtronic has demonstrated that the modifications made to the Lead Adaptor Kit and Lead Extender Kit products described in this submission result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with three arms or lines extending upwards, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 30, 2013
Medtronic, Inc. Michele Machacek 8200 Coral Sea Street NE Mounds View, MN 55112 US
Re: K132008
> Trade/Device Name: 6981M, 6984M, 6986M Lead Extender Kits 5866-24M, 5866-38M, 5866-40M Lead Adaptors Regulation Number: 21 CFR 870.3620 Regulation Name: Pacemaker Lead Adaptor Regulatory Class: Class II Product Code: DTD Dated: June 28, 2013 Received: July 1. 2013
Dear Michele Machacek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Owen P. Faris -S
- for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
#### 6981M Lead Extender Kit
Indications for Use:
The 6981M Lead extender kit is designed to extend, by 37 cm (14.6 in.), the length of a pacing lead system with a unipolar connector that meets the (IS-1 UNI) standard.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
.
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Digitally signed by
Owen P. Faris -S
Date: 2013.07.30
10:24:22 -04'00'
{7}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
### 6984M Lead Extender Kit
Indications for Use: The 6984M Lead Extender Kit is designed to extend, by 37 cm (14.6 in.), the length of a pacing lead system with a bipolar connector that meets the (IS-1 BI) standard.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
#### 5866-24M Lead Adaptor Kit
Indications for Use: The 5866-24M Lead adaptor kit is designed to connect a Medtronic pacing lead with a bipolar connector (5mm bifurcated) to a pulse generator featuring a bipolar connector block which meets the (IS-1 BI) standard.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 807 Subpart C) |
|-------------------------------------------------|--------|------------------------------------------------|
| X | | |
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{9}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
#### 5866-38M Lead Adaptor Kit
Indications for Use: The 5866-38M Lead adaptor kit is designed to connect two pacing leads with unipolar connectors (IS-1 UNI) to a pulse generator featuring a bipolar connector block which meets the (IS-I BI) standard.
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
#### 5866-40M Lead Adaptor Kit
Indications for Use: The Model 5866-40M Lead Adaptor Kit is designed to connect a Medtronic pacing lead with a bipolar connector (3.2 mm low-profile) to a pulse generator featuring a bipolar connector block which meets the (IS-1 BI) standard and has a short-pin cavity and no sealing rings.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{11}------------------------------------------------
510(k) Number (if known): K132008
Device Name:
#### 6986M Lead Extender Kit
Indications for Use:
The 6986M Lead adaptor/Extender kit is designed to extend, by 37 cm (14.6 in.), the length of a 3.2 mm low-profile connector pacing lead system and connect to a pulse generator featuring a bipolar connector block that meets the (IS-1 BI) standard.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.