The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician. It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient positioning and management of internal motion, and decision making in response to target position, size, shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures. XVI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers. Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cycle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion. The Elekta digital linear accelerator can be used for treatments that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepalic malignancies treated using SBRT, prostate, and bone metastases.
Device Story
XVI R5.0 is a kV imaging accessory for Elekta linear accelerators; utilizes gantry-mounted kV radiation source and detector to acquire 2D, 3D, and 4D images of patient anatomy at point of treatment. System enables visualization of internal structures, soft tissue, and implanted markers to support patient positioning and motion management. Symmetry™ software option acquires/displays volumetric images across breathing cycle phases for respiration-induced motion evaluation. Operated by clinicians in radiation therapy departments; output used for real-time image guidance, treatment decision-making, and dose placement accuracy. System facilitates intra-fraction imaging to monitor target displacement and organ deformation relative to critical structures.
Clinical Evidence
Bench testing only. Verification and validation performed at module, integration, and system levels under clinically representative conditions. Testing conducted on production-equivalent systems at Elekta and hospital sites to confirm conformance to design specifications and safety standards (IEC 60601 series, ISO 14971).
Technological Characteristics
System comprises gantry-mounted kV radiation source and image detector. Connectivity supports distributed workflow. Complies with IEC 60601-1, IEC 60601-2-1, IEC 60601-1-3, IEC 60601-2-44, IEC 60601-2-54, IEC 62304, IEC 62366, and ISO 14971. Software-based image acquisition and registration.
Indications for Use
Indicated for patients undergoing radiation therapy for malignant and benign tumors (e.g., brain, head/neck, lung, breast, pancreatic, hepatic, prostate, bone metastases) requiring image-guided positioning and motion management during static, dynamic, gated, or un-gated treatment fractions.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
K131965
Page 1 of 3
# 510(k) SUMMARY
Date of preparation of summary: June 2010 2013
| Submitted by: | Elekta Limited<br>Linac House, Fleming Way, Crawley, West Sussex<br>RH10 9RR, United Kingdom | | |
|--------------------------------|----------------------------------------------------------------------------------------------|--|--|
| Telephone: +44 (0) 1293 654201 | Fax: +44 (0) 1293 658445 | | |
| Contact name: | Mr Andrew Hedges | | |
| Trade Name: | XVI R5.0 | | |
| Common Name: | X-Ray Volume Imaging system, XVI | | |
| Classification Name: | Medical Linear Accelerator Accessory, 21 CFR 892.5050 | | |
| Product Code: | IYE | | |
| Predicate Device: | Elekta XVI (K100115), Varian TrueBeam (K123291),<br>Brainlab ExacTrac Vero (K122451) | | |
#### Product Description:
This Traditional 510/k) describes modifications made to the XVI kilo-voltage imaging accessory of the Elekta range of digital linear accelerators. The primary reason for these modifications is to further enhance the imaging tools to support the monitoring and management of motion, by allowing visualization of tumor and surrounding critical structures during treatment delivery.
Improvements have also been made to the operator interface connectivity with other systems to support distributed workflow and in the provision of licensable options to tailor individual features.
#### Intended Use and Indications for use statement:
The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician.
It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient oositioning and management of internal motion, and decision making in response to target position, size, shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures.
XVI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers.
Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cvcle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.
The Elekta digital linear accelerator can be used for treatments that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepalic malignancies treated using SBRT, prostate, and bone metastases.
{1}------------------------------------------------
#### Summary of Technological Characteristics:
The XVI system consists of a kV radiation source mounted onto the linac gantry drum and a kV radiation image detector. Incorporation of the kV imaging system onto the same structure as the treatment system allows high quality images of the patient anatomy to be acquired at the point of treatment and their content to be spatially related to the planned MV treatment, as previously cleared under Control Number (K100115)
There has been no change made to the underlying technological characteristics of the product.
#### Substantial Equivalence
The release of XVI R5.0 provides the ability to acquire KV images during treatment delivery, expanding the indications for use of its predicate device XVI R4.5 (K100115). The capability for monitoring and managing target motion and patient position during treatment is substantially equivalent to the predicate devices Varian TrueBeam (K123291) and ExacTrac Vero (K122451). Both devices provide localization to assist with patient positioning before, during and affer treatment and facilitate patient setup, through visualization of internal anatomy.
The differences in technological characteristics between XVI and the predicate devices do not raise questions of safety and effectiveness.
| Functionality | XVI R5.0 | XVI R4.5 | TrueBeam<br>(Varian) | ExacTrac<br>Vero<br>(BrainLab) |
|-------------------------------------------|-------------|----------|----------------------|--------------------------------|
| Control Number | K - - - - - | K 100115 | K123291 | K122451 |
| Visualize bony<br>anatomy on image | No Change | Yes | Yes | Yes |
| Visualize soft tissue<br>on image | No Change | Yes | Yes | Yes |
| Visualize implanted<br>markers in patient | No Change | Yes | Yes | Yes |
| Image patient in<br>treatment position | No Change | Yes | Yes | Yes |
| 2D image acquisition | No Change | Yes | Yes | Yes |
| Stereoscopic Imaging | Enhanced | Yes | Yes | Yes |
| 3D image acquisition | No Change | Yes | Yes | Yes |
| 4D image acquisition | No Change | Yes | Yes | No |
| Radiation dose<br>required | Low (kV) | Low (kV) | Low (kV) | Low(kV) |
| 3D Volumetric<br>Registration | No Change | Yes | Yes | Yes |
| Dual Registration | No Change | Yes | No | No |
| 4D Volumetric<br>Registration | No Change | Yes | Unknown | No |
| Intra-fraction Imaging | Yes | No | Yes | Yes |
{2}------------------------------------------------
K131965
Page 3 of 3
### Summary of non clinical performance testing
Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against the requirement specification.
Regression testing has been performed successfully to verify the integrity of any changes.
Validation of the system under clinically representative conditions has been performed by competent and professionally qualified personnel. Results from verification and validation testing demonstrate that conformance to applicable technical design specification have been met and safety & effectiveness have been achieved.
Testing has been undertaken on both production equivalent systems at Elekta and at hospital sites.
The system is subject to compliance testing to voluntary consensus safety standards. Details of the standards employed in the design are specified in the Standard Data Report in section 9 which includes but not limited to IEC 60601-1, IEC 60601-2-1, IEC 60601-1-3, IEC 60601-2-44, IEC 60601-2-54, IEC 62304, IEC 62366 and ISO 14971.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a black and white seal of the U.S. Department of Health and Human Services. The seal features the department's symbol, which consists of a stylized caduceus-like design with three parallel lines curving upwards and ending in a flourish. The symbol is encircled by the words "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Citifie Stivet Spring, MD 20093-0002
September 25. 2013
Elekta Limited % Mr. Andrew Hedges Regulatory Affairs Engineer Linac House, Fleming Way Crawley, West Sussex RH10 9RR UNITED KINGDOM
Rc: K131965
Trade/Device Name: XVI R5.0 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: June 20, 2013 Received: June 27, 2013
Dear Mr. Hedges:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Mr. Hedges
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.lida.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K131962
Device Name: XVI R5.0
Indications for Use:
The Elekta X-Ray Volume Imaging system, XVI, is an electronic imaging device (EID), designed to be used with the Elekta range of medical linear accelerators and intended to be used as part of the radiation therapy treatment process for single or multiple fractions, of static and/or dynamic treatments, in gated or un-gated deliveries, in all areas of the body where such image guidance is determined by a licensed physician.
It provides real-time 2D, 3D & 4D image guidance before, during and after treatment delivery, and is intended to support confirmation of patient positioning, monitoring and management of internal motion, and decision making in response to target position, size. shape and displacement resulting from organ deformation and anatomical movement in relation to surrounding critical structures.
X VI facilitates precise and accurate dose placement, and patient set-up correction, through visualization of internal anatomy including target, critical structures and soft tissue with or without the use of implanted markers.
Symmetry™ is a software option within XVI that can be used to acquire and display volumetric images of sequential phases of the breathing cycle for the evaluation of respiration induced motion, to assist in identification of appropriate target locations within anatomical structures in motion.
The Elekta digital linear accelerator can be used for treatment that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michal D. O'Hara
an Sign-Off
Page 1 of 1
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.