← Product Code [FIE](/productcode/FIE) · K131950

# ANGEL TIP SAFETY INTRAVASCULAR NEEDLE SET (K131950)

_Summit International Medical Technologies, Inc. · FIE · Jan 16, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K131950

## Device Facts

- **Applicant:** Summit International Medical Technologies, Inc.
- **Product Code:** [FIE](/productcode/FIE.md)
- **Decision Date:** Jan 16, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Angel Tip Safety Intravascular Needle Set is indicated to access the peripheral vascular system for hemodialysis. Additionally, after withdrawal of the needle from the patient's vein, the needle can be manually retracted within a safety enclosure to minimize the risk of an accidental needle stick.

## Device Story

Sterile, single-use intravascular needle set for hemodialysis access; consists of needle, winged hub, tubing, adapter, and safety enclosure. Operated by clinician; needle inserted into peripheral vein. Upon withdrawal, user manually slides safety enclosure over needle until it latches. Safety mechanism features audible click and visual confirmation (tethers fully extended, clear enclosure). Design protects user from accidental needle stick by keeping fingers behind needle point. Benefits include reduced risk of sharps injury to healthcare personnel.

## Clinical Evidence

Bench testing only. Performance testing included tensile, leakage, Luer, corrosion, force to activate, and force to override tests. Results demonstrate safety and effectiveness comparable to predicate devices.

## Technological Characteristics

Needle diameters 14-18 Ga; lengths 1.0-1.5 inches. Components: needle, winged hub, microbore/macrobore tubing, adapter, adapter cap, safety enclosure. Manual retraction mechanism with active engagement, audible click, and visual tether indicators. Sterile, single-use.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Angel Wing Safety Blood Collection and Infusion Set ([K912563](/device/K912563.md))
- A.V. Fistula Needle Set "WingEater" ([K111948](/device/K111948.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 16, 2015

Summit International Medical Technologies, Inc. Michael Merchant President 31 Hayward Street, Suite H-1 Franklin, MA 02038

Re: K131950

> Trade/Device Name: Angel Tip Safety Intravascular Needle Set Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: FIE Dated: November 16, 2014 Received: December 18, 2014

Dear Michael Merchant,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -A" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, and the overall impression is one of simplicity and clarity.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K131950

Device Name Angel Tip Safety Intravascular Needle Set

#### Indications for Use (Describe)

The Angel Tip Safety Intravascular Needle Set is indicated to access the peripheral vascular system for hemodialysis. Additionally, after withdrawal of the needle from the patient's vein, the needle can be manually retracted within a safety enclosure to minimize the risk of an accidental needle stick.

Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

Date Prepared: January 14, 2015

Contact Information:

Michael Merchant President Summit International Medical Technologies, Inc. 20 Addison Avenue Franklin, MA 02038 USA Email: mmerchant@summitmedtech.com Telephone: (508) 528-3065 Fax: (508) 570-4902

# Trade Name:

Angel Tip Safety Intravascular Needle Set

### Common Name:

Safety Intravascular Needle Set

### Classification Name:

Needle, Fistula: per 21 CFR: 876.5540 Product code: FIE

## Predicate Devices

- K912563: Angel Wing Safety Blood Collection and Infusion Set ●
- K111948: A.V. Fistula Needle Set "WingEater"" ●

## Device Description

The safety needle portion of the Angel Tip Safety Intravascular Needle Set will be available in diameters ranging from 14 to 18 Ga and lengths of 1.0 to 1.5 inches.

The Angel Tip Safety Intravascular Needle Sets are sterile, single use devices designed to access the peripheral vascular system for hemodialysis. Additionally, after withdrawal of the needle from the patient's vein, the needle can be manually retracted within a safety enclosure to minimize the risk of an accidental needle stick. The Angel Tip Safety Intravascular Needle Set consists of a needle attached to a winged hub, microbore or macrobore tubing, adapter and adapter cap and a safety enclosure that attaches to the winged hub.

As the needle is removed from the patient's vessel, the user's finger actively slides the safety enclosure until it latches onto the needle using a one- or two-handed technique. The user will know the needle is engaged based on the tethers being fully extended and 

{4}------------------------------------------------

hearing an audible click. The safety enclosure is designed to allow the user's fingers to remain behind the needle point so that the risk of needle stick injury is minimized. The enclosure is clear so the user has a visual means of knowing the needle is captured.

# Device Intended Use

The Angel Tip Safety Intravascular Needle Set is indicated to access the peripheral vascular system for hemodialysis. Additionally, after withdrawal of the needle from the patient's vein, the needle can be manually retracted within a safety enclosure to minimize the risk of an accidental needle stick.

# Technological Characteristics and Substantial Equivalence

The Angel Tip Safety Intravascular Needle Set is substantially equivalent to the safety enclosure Angel Wing Safety Blood Collection and Infusion Set, cleared under MBO Laboratories, Inc. 510(k) K912563, in terms of design, manufacturing process, functional performance, and materials of construction. The Angel Tip Safety Intravascular Needle Set includes improvements made to the safety mechanism including putting the activation clip on top of the device for easier access, enclosing the clip in the housing to protect it, implementing active engagement to ensure the needle will remain in the housing once it has been retracted and providing both visual (tethers fully extended) and auditory (audible click) indications that the needle is engaged in the housing.

The Angel Tip Safety Intravascular Needle Set is substantially equivalent to the A.V. Fistula Needle Set "WingEater", cleared under JMS North America Corporation 510(k) K111948, in terms of indications for use and functional performance. There is no change to the fundamental scientific technology of the previously cleared devices or its indications for use as compared to the A.V. Fistula Needle Set "WingEater"".

# Nonclinical Testing

The results of the performance testing (tensile, leakage, Luer, corrosion, force to activate and force to override) demonstrate that the Angel Tip Safety Intravascular Needle Set is as safe, as effective and performs comparably to the predicate Angel Wing Safety Blood Collection and Infusion Set and A.V. Fistula Needle Set "WingEater".

# Conclusions

The predicate and the subject devices have similar indications for use, intended use, fundamental scientific technology and similar design, materials, and manufacturing processes. The results of the testing performed have demonstrated that the subject device does not raise new issues of safety or effectiveness and therefore is considered substantially equivalent to the cited predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K131950](https://fda-staging.innolitics.com/device/K131950)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
